K062700 · B.Braun Medical, Inc. · FRN · Oct 26, 2006 · General Hospital
Device Facts
Record ID
K062700
Device Name
INFUSOMAT SPACE VOLUMETRIC INFUSION PUMP SYSTEM
Applicant
B.Braun Medical, Inc.
Product Code
FRN · General Hospital
Decision Date
Oct 26, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic, Pediatric, 3rd-Party Reviewed
Indications for Use
The Infusomat® Space Volumetric Infusion Pump System includes an external transportable electronic volumetric infusion pump, dedicated administration sets, and pump accessories. The system is intended for use on adults, pediatrics, and neonates for the intermittent or continuous delivery of parenteral and enteral fluids through clinically accepted routes of administration. These routes include, but are not limited to intravenous, intra-arterial, subcutaneous, epidural, irrigation/ablation, and enteral. The system is used for the delivery of medications indicated for infusion therapy including but not limited to drugs like anesthetics, sedatives, analgesics, catecholamines, anticoagulants etc., blood and blood components, Total Parenteral Nutrition (TPN), lipids, and enteral fluids. The Infusomat® Space Volumetric Infusion Pump System is intended to be used by trained healthcare professionals in healthcare facilities, home care, outpatient, and medical transport environments.
Device Story
Infusomat® Space is a transportable, 12V DC/battery-powered volumetric infusion pump using linear peristaltic mechanism. System includes pump, dedicated silicone-tubing administration sets, and accessories (SpaceStation docking, SpaceCover). Operated by trained healthcare professionals in hospitals, home care, outpatient, and transport. Pump interfaces with administration sets containing free-flow protection clamps. DrugListEditor software allows creation of drug libraries (up to 3500 entries) downloaded to pumps via service tool. SpaceStation provides central power and communication for up to 24 pumps. Clinicians use pump interface to set infusion parameters; system provides continuous/intermittent fluid delivery. Benefits include organized bedside management, flexible infusion therapy, and automated maintenance/calibration alerts. Safety features include free-flow protection upon door opening and alarm signaling via SpaceCover Comfort.
Clinical Evidence
Bench testing only. No clinical data presented.
Technological Characteristics
Linear peristaltic pumping mechanism; 12V DC or battery power; thermoplastic housing; silicone tubing for pump interface; free-flow protection clamp; modular docking system (SpaceStation) for power/communication; supports drug library software (DrugListEditor); single-use, sterile, non-pyrogenic administration sets; latex-free; DEHP-free (except blood filter tubing).
Indications for Use
Indicated for adults, pediatrics, and neonates requiring intermittent or continuous delivery of parenteral and enteral fluids, including medications, blood/blood products, TPN, and lipids, via intravenous, intra-arterial, subcutaneous, epidural, irrigation/ablation, or enteral routes. For use by trained healthcare professionals in clinical, home, outpatient, and medical transport (road ambulance) settings.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
ALARIS Medical Systems, Alaris Medley™ Patient Care System (K030459)
ALARIS Medical Systems, Alaris MEDLEY™ Pump Module (K950419)
B. Braun Medical Inc. Millennium CRT Infusion Pump and Millennium CRT Infusion Pump Administration Sets (K972678)
Submission Summary (Full Text)
{0}------------------------------------------------
K062700
(PIPE)
August i, 2006
:
::
:
:
| 5. 510(k) Summary | |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT: | B. Braun Medical Inc.<br>901 Marcon Boulevard<br>Allentown, PA 18109-9341<br>610-266-0500 |
| | Contact: Scott Pease<br>Manager, Regulatory Affairs<br>Phone: 610-596-2376 |
| SUBMITTER: | B. Braun Melsungen AG<br>Carl-Braun-Strasse 1<br>34212 Melsungen<br>GERMANY |
| | Contact: Thomas Möller<br>Manager, Regulatory Affairs<br>Phone: 49 5661 71-1349 |
| DEVICE NAME: | Infusomat® Space Volumetric Infusion Pump System |
| COMMON OR<br>USUAL NAME: | Infusion Pump and Intravascular Administration Sets |
| DEVICE<br>CLASSIFICATION: | Class II per Code of Federal Regulation Title 21<br>§880.5725: Infusion Pump, product code FRN and Enteral<br>Infusion Pump, product code LZH; 21 CFR §880.5440:<br>Intravascular Administration Set, product code FPA and<br>Blood Transfusion Set, product code BRZ.<br>Classification Panel: General Hospital. |
| PREDICATE<br>DEVICES: | Predicate device for infusion pump: B. Braun Medical Inc.<br>Vista Basic with <i>fm</i> System 510(k) K023189; ALARIS<br>Medical Systems, Alaris Medley™ Patient Care System<br>510(k) K030459, ALARIS Medical Systems, Alaris<br>MEDLEY <sup>TM</sup> Pump Module 510(k) K950419.<br>Predicate device for administration sets: B. Braun Medical<br>Inc. Millennium CRT Infusion Pump and Millennium CRT<br>Infusion Pump Administration Sets 510(k) K972678. |
| DESCRIPTION: | The Infusomat® Space Volumetric Infusion Pump System<br>includes an external transportable electronic volumetric<br>infusion pump, dedicated administration sets, and pump<br>accessories. |
OCT 26 2006
K-57
RECEIVED
2006 SEP 11 P 2:07
FDA/CDRH/ODE/PMO
10
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{1}------------------------------------------------
人*62799
(P.2 of 6)
# Infusomat® Space Pump
The Infusomat® Space is a 12V DC or battery powered external, transportable, volumetric infusion pump. The Infusomat® Space utilizes a linear peristaltic pumping mechanism and is intended to provide infusions of parenteral and enteral fluids. The Infusomat® Space is intended for but not limited to be used by trained healthcare professionals in healthcare facilities, home care, outpatient, and medical transport environments. A trained Biomedical Technician must perform a complete set-up of the pump prior to use in a clinical setting. The system created by using the administration sets with
the Infusomat® Space is intended to provide intermittent or continuous flow of parenteral and enteral fluids to the patient. Parenteral fluids may include all standard fluids and/or medications indicated for infusion as well as blood and blood products.
#### SpaceStation
The B. Braun Space Station is a 12V DC or battery powered flexible docking and communication system for the medical workplace, in particular the intensive medical care. It serves the perspicuous accommodation of the infusion and infusion-syringe pumps Infusomat® Space and Perfusor® Space. The pillar and mounting system synchronized system components enables the individualized workplace design. The SpaceStation module allows the set-up of a complete
pump system with up to 24 pumps. Up to four pumps can be installed in every SpaceStation. The pumps are supplied with power via the integrated power supply and the built-in connectors. Up to six SpaceStations can be set-up as a column with a total of 24 pumps. SpaceStation placed next to each other can be connected via special connection cables, if the maximum number of 24 pumps in maximum three columns is not exceeded.
### SpaceCover
SpaceCover Standard or SpaceCover Comfort forms the top of each column. Every SpaceStation or a column of SpaceStations must be closed with a cover. Alarms are
11
{2}------------------------------------------------
Kψ6279φ
(P. 3 of 6)
August 1, 2006
signalled by a row of LEDs and a loudspeaker in the SpaceCover Comfort.
The SpaceCover Standard does not contain any additional electronics. It protects the upper connectors from humidity and damage and allows a single SpaceStation to be used as a carrying unit. The SpaceCover standard housing mainly consists of the bottom part and the upper part. The design of the SpaceCover comfort corresponds to that of the SpaceCover standard. SpaceCover Comfort offers a
greater system functionality and operating facility. The cover is equipped on its front side with a large and very well readable status and alarm display. All status and alarm conditions of the pumps within the system as well as of the pumps themselves are displayed.
### DrugListEditor
The DrugListEditor Space is a programm to create a drug pool with drug names including the corresponding parameter. The DLE-Space only may be used together with the infusion pumps Infusomat® Space and Perfusor® Space. A download of a drug list into the respective pump is possible by using the HiBaSeD service tool. The DrugListEditor is able to hold up to 3500 drug names with corresponding parameters. From this pool, single drug libraries can be compiled and stored. The drug pool is saved as a binary file on the computer.
### Administration sets
The administration sets are single-use, sterile, nonpyrogenic, disposable devices for use with the B. Braun Infusomat® Space Volumetric Infusion Pump for pump and gravity administration of fluids. There are currently 13 administration set configurations available including basic sets, burette sets, additive sets, filtered sets, epidural sets, low adsorption sets, add-on sets and blood sets.
The Infusomat® Space Volumetric Infusion Pump administration sets will be available in lengths ranging from 88 in. (224 cm) to 136 in. (345 cm), drops/mL ranging from 10 to 60 and priming volumes ranging from approximately 13 to 39 mL. All sets are latex free and all but one of the sets is DEHP-Free. The only DEHP containing item is the tubing portion of the blood filter component of the blood administration set. All other tubing components are DEHP free. The epidural
12
{3}------------------------------------------------
August 1, 2006
administration set has a yellow stripe in the tubing. All other tubing segments are clear.
Each set contains a segment of tubing intended to interface with the linear peristaltic mechanism of the pump. This segment of tubing is composed of silicone. There are two connectors at each end of the pump tube segment to ensure proper placement of the set into the pump. The silicone tubing provides greater resistance to loss of shape from the peristaltic pumping action compared to standard PVC tubing.
Each set also contains a free flow protection clamp. This clamp is intended to prevent free flow of fluid when the pump door is opened and the set is removed. The free flow protection clamp is specifically designed to interface with a mating receptacle in the pump. The free flow protection clamp is forced in to the open position when the pump door is closed; however, fluid flow does not begin until the run button is pressed. When the pump door is opened, the free flow protection clamp is forced into the closed position; thus stopping fluid flow and protecting against free flow.
The Infusomat® Space Volumetric Infusion Pump System INTENDED USE: includes an external transportable electronic volumetric infusion pump, dedicated administration sets, and pump accessories. The system is intended for use on adults, pediatrics, and neonates for the intermittent or continuous delivery of parenteral and enteral fluids through clinically accepted routes of administration. These routes include. but are not limited to intravenous, intra-arterial. subcutaneous, epidural, irrigation/ablation, and enteral. The system is used for the delivery of medications indicated for infusion therapy including but not limited to drugs like anesthetics, sedatives, analgesics. catecholamines, anticoagulants etc., blood and blood components, Total Parenteral Nutrition (TPN), lipids, and enteral fluids. The Infusomat® Space Volumetric Infusion Pump System is intended to be used by trained healthcare professionals in healthcare facilities, home care, outpatient, and medical transport environments.
{4}------------------------------------------------
Kψ627φφ
(P. 5 of 6)
August 1, 2006
## SUBSTANTIAL EQUIVALENCE:
The Infusomat® Space Volumetric Infusion Pump System, including an external transportable electronic volumetric infusion pump, dedicated administration sets, and pump accessories has the same intended use, operation, and function as the Vista Basic with fin System distributed by B. Braun Medical Inc., 510(k) K023189, Alaris MedleyTM Patient Care System distributed by ALARIS Medical Systems, 510(k) K030459, and Millennium CRT Infusion Pump and Millennium CRT Infusion Pump Administration Sets (now referred to as the Vista Infusion Pump and Adminsitration Sets) distributed by B. Braun Medical Inc., 510(k) K972678. There are no differences between the subject device and the predicate devices that raise new issues of safety and effectiveness.
The subject and the predicate devices are software controlled, electronic, external, volumetric infusion pumps, composed of injection molded thermoplastic components. The components of the subject and predicate devices are non solution contact. The subject and predicate devices are similar in operation, as they are all linear peristaltic volumetric infusion pumps. The subject and predicate devices achieve fluid delivery in a similar manner by the cyclical crushing action of the pumping fingers of the peristaltic mechanism on a straight piece of tubing. The subject and predicate devices are also similar in that they are used in the hospital or home health care setting to control the infusion rate of parenteral fluids to the patient.
The included accessories of the Infusomat® Space Volumetric Infusion Pump System are similar to the predicate devices in that they all offer the clinician maximum flexibility in providing infusion therapy. The subject and predicate systems provide possibilities to clean up and organize the patient bedside by storage different pumps in a filing and communication system. This filing and communication system provides central power supply to the single pumps and tubing retainers to prevent IV line confusion. Also the subject and the predicate provide a means for easy transportation of the modules as well as of the system. Additionally the subject and predicate devices are used to customize the menu items and drug library parameters for each facility. The calibration preventive
{5}------------------------------------------------
KY62744
(p. 6 of 6)
August 1, 2006
maintenance of the pump can also be stored in each device and the users notified when such calibration or routine maintenance is due to be completed.
The subject administration sets and the predicate administration sets are similar in that they are both singleuse, sterile, non-pyrogenic, disposable devices for use with an external linear peristaltic volumetric infusion pump. Both the predicate and subject administration sets contain a free flow protection clamp intended to prevent free flow when the door is opened and the set is removed. The predicate device contains a straight piece of tubing to interface with the pump. Whereas the subject device contains a dedicated piece of silicone tubing to interface with the pump. Both the predicate and subject devices contain standard administration set components (spike, injections sites, roller clamps, slide clamps, check valves, filters, luers, etc.). Both the subject and predicate administration sets are intended for use with a specific infusion pump controlled volumetric infusion of parenteral and enteral fluids. The subject and predicate administration sets are also intended for use independent of their respective pumps for gravity controlled infusion.
15
{6}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a circular design with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract emblem resembling an eagle or bird in flight, composed of stylized lines.
OCT 26 2006
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
B. Braun Medical, Incorporated C/O Mr. Stefan Preiss Responsible Third Party Official TÜV America, Incorporated 1775 Old Hwy 8 NW New Brighton, Minnesota 55112-1891
- Re: K062700
Trade/Device Name: Infusomat® Space Volumetric Infusion Pump System Regulation Number: 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: October 4, 2006 Received: October 11, 2006
Dear Mr. Preiss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 - Mr. Preiss
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Suoer Raonon
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Page
510(k) Number (if known):
Device Name:
Infusomat® Space Volumetric Infusion Pump System
ФБӘЛФФ
Indications For Use:
The Infusomat® Space Volumetric Infusion Pump System includes an external transportable electronic I rie musonial "Space" voluments many of canp of Jers, and pump accessories. The system is intended volumently multips pediatrics, and neonates for the intermittent or continuous delivery of parenteral and to tuse on adults, pediatives, than neemed outes of administration. These routes include, but are not emotal thirds intravenous, irrigation, and enteral. The system is used for the delivery of million in intravenous, firigationsharapy including but not limited to colloids and cristalloids, blood medications morealed for mission moraly metably and enteral fluids, and enteral fluids. The Infusomate and blood continutions, Tour rarented to be used by trained healthcare professionals in Spice Voluncific mirusion Fullp Dylean is intensport environments (only road ambulances).
Prescription Use __ (Per 21 CFR 801.109) OR
Over-The-Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Asta V.r
i-igh-On)
ion of Anesthesiology, General Hospital,
ion Control, Dental Devices
ber. L-062794
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.