K062682 · Lake Consumer Products, Inc. · NUC · Apr 4, 2007 · Obstetrics/Gynecology
Device Facts
Record ID
K062682
Device Name
CVS PERSONAL LUBRICANT AND MOISTURIZER
Applicant
Lake Consumer Products, Inc.
Product Code
NUC · Obstetrics/Gynecology
Decision Date
Apr 4, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
CVS Personal Lubricant & Moisturizer is intended as a personal lubricant for vaginal dryness and personal lubrication of the vaginal area to enhance condom use and to facilitate ease and comfort during intimate sexual activity. The lubricous nature of the product helps supplement the body's natural lubricating fluids, thereby reducing friction. CVS Personal Lubricant & Moisturizer is compatible with latex condoms. This device is not a contraceptive or spermicide, nor does it contain any such component.
Device Story
CVS Personal Lubricant & Moisturizer is a non-sterile, aqueous, water-soluble gel. Intended for OTC use by consumers to alleviate vaginal dryness and facilitate sexual activity. Applied topically to the vaginal area; compatible with latex condoms. Formula consists of water, glycerin, propylene glycol, polyquaternium 15, methylparaben, and propylparaben. Packaged in PETE bottles with foil seals and polypropylene caps. Benefits include friction reduction and enhanced comfort during intimacy. No contraceptive or spermicidal properties.
Clinical Evidence
Bench testing only. Stability testing (90-day accelerated) and USP <1227> Antimicrobial Effectiveness Test passed. Condom compatibility testing confirmed no adverse effects on tensile strength, elongation, or burst pressure of latex condoms. Perceptual quality comparison against predicate rated 10/10.
Technological Characteristics
Aqueous gel formulation containing water, glycerin, propylene glycol, polyquaternium 15, methylparaben, and propylparaben. Non-sterile. Packaged in PETE plastic bottles. Water-soluble. Compatible with latex condoms.
Indications for Use
Indicated for OTC consumer use as a vaginal moisturizer and personal lubricant to reduce friction and enhance comfort during intimate sexual activity. Compatible with latex condoms. Not a contraceptive or spermicide.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
{0}------------------------------------------------
# APR 4 - 2007
## 510(k) SUMMARY
| <b>Submitter:</b> | Lake Consumer Products, Inc. |
|-------------------|------------------------------|
| | 1 Pharmacal Way |
| | Jackson, WI 53037 |
- Contact Person: Mary L. Wundrock Vice President- Laboratory Services mwundrock@pharmacalway.com Phone: (262) 677-4121 Ex 7110 Fax: (262) 677-9006
September 6, 2006 Date Submitted:
Proprietary Name: CVS Personal Lubricant & Moisturizer
- Personal Lubricant Common Name:
- Predicate Name: Astroglide 510(k) Number: K935299
### Classification Name:
21 CFR 880.6375 Patient Vaginal Lubricant Product Code MMS Class I
21 CFR880.5300 Lubricant, Patient, Vaginal, Latex Compatible Product Code NUC Class II
### Description of Device:
The CVS Personal Lubricant & Moisturizer formula is non-sterile. clear, non- staining, non-greasy, aqueous gel intended for use as a moisturizer for vaginal dryness and personal lubrication of the vaginal area to enhance condom use and to facilitate ease and comfort during intimate sexual activity. CVS Personal Lubricant & Moisturizer is compatible with latex condoms as demonstrated in condom compatibility testing. This device is not a contraceptive or spermicide, nor does it contain any such component. CVS Personal Lubricant & Moisturizer is packaged in PETE bottles with a foil seal and polypropylene cap.
న-I
{1}------------------------------------------------
## Intended Use:
CVS Personal Lubricant & Moisturizer is intended as a personal lubricant for vaginal dryness and personal lubrication of the vaginal area to enhance condom use and to facilitate ease and comfort during intimate sexual activity. The lubricous nature of the product helps supplement the body's natural lubricating fluids, thereby reducing friction. CVS Personal Lubricant & Moisturizer is compatible with latex condoms. This device is not a contraceptive or spermicide, nor does it contain any such component.
### Comparison of Technological Characteristics:
CVS Personal lubricant & Moisturizer contains water, glycerin, propylene glycol, polyquaterium 15, methylparaben and propylparaben. CVS Personal Lubricant & Moisturizer is substantially similar to the predicate device.
A summary comparison of the features of CVS Personal Lubricant & Moisturizer and the predicate device Astroglide is provided in Table 1.
| Features | CVS Personal<br>Lubricant &<br>Moisturizer | Predicate Device-<br>Astroglide |
|-------------------------------|--------------------------------------------|---------------------------------|
| Labeled Condom<br>Compatible | Yes | Yes |
| Personal Lubricant | Yes | Yes |
| Contains Water | Yes | Yes |
| Contains Glycerin | Yes | Yes |
| Contains<br>Propylene Glycol | Yes | Yes |
| Contains<br>Polyquaternium 15 | Yes | Yes |
| Contains<br>Methylparaben | Yes | Yes |
| Contains<br>Propylparaben | Yes | Yes |
| Sterile | No | No |
| Container type | Plastic | Plastic |
# TABLE 1
CVS Personal Lubricant & Moisturizer is substantially equivalent to the following device that is currently in commercial distribution:
Astroglide, BioFilm Inc., K935299
ર-2
{2}------------------------------------------------
#### Substantial Equivalence:
CVS Personal Lubricant & Moisturizer is substantially similar to the predicate device, Astroglide. The products have the same intended use and technological characteristics. No new safety or effectiveness issues have been raised through testing. Both products are intended for use as personal lubricants, safe to use with condoms, water-soluble and for over-the-counter (OTC) use.
### Summary of Performance Data:
### Non-Clinical Studies
Stability- CVS Personal Lubricant & Moisturizer has successfully passed 90 day accelerated stability.
Preservative Effectiveness-CVS Personal Lubricant & Moisturizer formulation with the preservative system has successfully passed the requirements of the USP <1227> Antimicrobial Effectiveness Test.
Comparison with Predicate Device- CVS Personal Lubricant & Moisturizer was compared to Astroglide on the basis of perceptual qualities, physical and chemical properties, ingredient list review, label claims and packaging. Test results rated CVS Personal Lubricant & Moisturizer (a 10 of 10 rating) as comparable to Astroglide.
Condom Compatibility Testing- CVS Personal Lubricant & Moisturizer, when applied to latex condoms does not adverselv affect tensile strength, elongation or burst pressure, when compared to untreated condoms.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and three human figures in profile facing the eagle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Ms. Mary L. Wundrock Vice President - Laboratory Services Lake Consumer Products, Inc. 1 Pharmacal Way, P.O. Box 198 JACKSON WI 53037
FEB 2 4 2014
Re: . K062682 Trade/Device Name: CVS Personal Lubricant & Moisturizer Regulation Number: 21 CFR §884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated (Date on orig SE ltr): March 21, 2007 Received (Date on orig SE ltr): March 22, 2007
Dear Ms. Wundrock:
This letter corrects our substantially equivalent letter of April 4, 2007.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{4}------------------------------------------------
Page 2 -
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
**Benjamin R. Fisher -S**
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use Statement:
KC62682 510(k) Number:
Device Name: CVS Personal Lubricant & Moisturizer
Indications for Use:
A non-sterile personal lubricant for OTC consumer use, as a moisturizer for vaginal dryness and personal lubrication of the vaginal area to enhance condom use and to facilitate ease and comfort during intimate sexual activity. CVS Personal Lubricant & Moisturizer is compatible with latex condoms. This device is not a contraceptive or spermicide nor does it contain any such component.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Bergman
Division Sien-Division of Reproductive, and Radiological Dev 510(k) Number
4-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.