← Product Code [GEX](/productcode/GEX) · K062619

# DENLASER 800 PLUS (K062619)

_CAO Group, Inc. · GEX · Nov 14, 2006 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K062619

## Device Facts

- **Applicant:** CAO Group, Inc.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Nov 14, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The DenLaser 800 Plus is indicated for the procedures of removal of lesions, excision, incision, vaporization, ablation, hemostasis, and photocoagulation on soft tissue in the fields of otolaryngology (ear, nose, and throat), dentistry and oral surgery, arthroscopy, gastroenterology, dermatology, podiatry, plastic surgery, urology, gynecology, and plastic surgery.

## Device Story

DenLaser 800 Plus is a surgical diode laser system. Input: electrical power; Output: 810nm laser energy (max 5W). Principle: solid-state diodes generate laser energy transmitted via proprietary optical fiber to a metal handpiece with disposable single-use tips. Operation: foot-actuated switch; user-definable settings include continuous or pulse delivery modes and a 630nm aiming beam. Used in clinical settings by healthcare professionals for soft tissue surgical procedures. Laser energy enables precise cutting, ablation, and hemostasis. Benefits include controlled tissue interaction and reduced collateral damage compared to traditional methods.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on technological characteristics and performance standards compliance.

## Technological Characteristics

Solid-state diode laser; 810 ± 20nm wavelength; max 5W output. Features 630nm aiming beam, fiber delivery system, and foot-actuated switch. Complies with 21 CFR 1040.10/1040.11, IEC 60601-1:1998+A1, IEC 60601-2-22:1995, IEC 60825-1:1993+A1+A2, 47 CFR 15/18, and ETSI 301-489-1. Handpiece uses disposable single-use tips.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Odyssey 2.4G Diode Laser ([K050453](/device/K050453.md))
- Ceralas D 810 Diode Laser ([K032864](/device/K032864.md))
- Aurora SL Diode Laser ([K993285](/device/K993285.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### 510(k) Summary of Safety and Effectiveness

CAO Group, Inc. 4628 West Skyhawk Drive West Jordan. UT 84084 Tel: 801-256-9282 Fax: 801-256-9287

Robert Larsen, Preparation Date: August 15, 2006

NUV 1 4 2006

KO62619

#### Device Name:

| Trade Name:             | DenLaser 800 Plus          |
|-------------------------|----------------------------|
| Common Name:            | 810nm Diode Laser          |
| Product Classification: | Laser Instrument, Surgical |

# Legally Marketed Predicate Devices for Substantial Equivalence:

Odyssey 2.4G Diode Laser, manufactured by Ivoclar Vivadent, Inc. (K050453)

Ceralas D 810 Diode Laser, manufactured by Biolitec, Inc. (K032864)

Aurora SL Diode Laser, manufactured by Premier Laser Systems, Inc. (K993285)

### Rationale for Substantial Equivalence:

The aforementioned laser devices and their accompanying delivery systems share similar indications for use on soft tissue with the submitted device for cutting soft tissue, affecting lesions, and photocoagulation. The predicate devices and submitted device share similar design features including wavelength, operating controls, and laser delivery method. The devices share similar methods of control systems, safety features, and performance monitoring. The devices share similar performance specifications including power output and energy type. The devices share similar applications to soft tissue.

#### Description of Submitted Device:

The DenLaser 800 Plus is a device for delivering laser energy to human soft tissue for a variety of surgical procedures. This energy is generated by solid-state diodes, which provide a consistent and reliable generation of laser energy at 810 ± 20nm for a maximum of 5 watts of energy output. The laser energy is delivered to surgical site by means of a proprietary optical fiber system, which allows for the safe transmission of laser energy to the site without creating undue risk to the patient or operatory staff by

{1}------------------------------------------------

errant or collateral laser emissions. The device features some user definable settings, including a switchable 630nm aiming beam, adjustable power output for both the working beam and aiming beam, and continuous delivery or pulse delivery option.

The working end of the delivery fiber is contained within metal handpiece with a disposable single-use tip. This handpiece system is provided with the device. The activation of the working beam diodes is completed by use of a foot-actuated switch.

# Intended Uses of the DenLaser 800 Plus diode laser system:

The device is intended to be used for a variety of surgical procedures on soft tissue. See Indications for Use on page 2-1.

# Technological Characteristics and Substantial Equivalence:

The Odyssey 2.4G Diode Laser uses solid state diodes to generate laser energy in the 810nm range. This system uses a fiber delivery system to transmit laser energy to the surgical site. The system also features a 630nm aiming beam and features controls that allow for adjusting the output of the working beam, and switching between a continuous or pulsed-mode laser emission. The device features a wireless foot switch for actuating the working beam. The maximum output of the working beam is 5 watts.

The Ceralas D 810 Diode Laser uses solid state diodes to generate laser energy in the 810nm range. This system uses a fiber delivery system to transmit laser energy to the surgical site. The system also features a 635nm aiming beam and features controls that allow for adjusting the output of the working beam, and switching between a continuous or pulsed-mode laser emission. The maximum output of the unit is 15 watts.

The Aurora SL Diode Laser uses solid state diodes to generate laser energy in the 810nm range. This system uses a fiber delivery system to transmit laser energy to the surgical site. The system also features a 630-680nm aiming beam and features controls that allow for adjusting the output of the working beam. The maximum output of the unit is 6 watts.

#### Performance Standards:

The DenLaser 800 Plus diode laser complies with the performance requirements of 21 CFR 1040.10 and 1040.11, with permissible deviations relative to Laser Notice 50, dated July 26, 2001. The device also complies with IEC 60601-1:1998+A1, IEC 60601-2-22:1995, IEC 60825-1:1993+A1+A2, 47 CFR 15 and 18, and ETSI 301-489-1.

### Clinical Performance Data

{2}------------------------------------------------

# Conclusion

The DenLaser 800 Plus is substantially equivalent to the listed laser surgical devices without raising any issues of safety or effectiveness. This device shares similar intended uses, and similar functional and performance characteristics. The device is designed to comply with relevant federal and international safety and performance standards.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird symbol with three curved lines representing its body and wings. The bird is enclosed within a circular border, and the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is written around the border of the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

CAO Group, Inc. % Mr. Robert K. Larsen Operations Director 4628 West Skyhawk Drive West Jordan, Utah 84084

NOV 1 4 2006

Re: K062619 Trade/Device Name: DenLaser 800 Plus Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: August 15, 2006 Received: September 6, 2006

Dear Mr. Larsen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been rcclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Mr. Robert K. Larsen

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Sincerely yours,

Mark N. Melkerson

Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

| 510(k) Number (if known): | K062619 |
|---------------------------|---------|
|---------------------------|---------|

Device Name: DenLaser 800 Plus

Indications For Use:

The DenLaser 800 Plus is indicated for the procedures of

removal of lesions, excision, incision, vaporization, ablation, hemostasis, and photocoagulation

on soft tissue in the fields of

otolaryngology (ear, nose, and throat), dentistry and oral surgery, arthroscopy, gastroenterology, dermatology, podiatry, plastic surgery, urology, gynecology, and plastic surgery.

| Prescription Use           | X | AND/OR Over-The-Counter Use |  |
|----------------------------|---|-----------------------------|--|
| (Per 21 CFR 801 Subpart D) |   | (21 CFR 801 Subpart C)      |  |

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

![Signature](signature.png)
(Division Sign-Off)

Division of General, Restorative, and Neurological Devices

| 510(k) Number | L06219 | Page 1 of 1 |
|---------------|--------|-------------|
|---------------|--------|-------------|

2-1

---

**Source:** [https://fda-staging.innolitics.com/device/K062619](https://fda-staging.innolitics.com/device/K062619)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
