The WellnessPro 2010 is an Electrotherapy Unit that is used for the following: 1. Relief of chronic, intractable pain. 2. Adjunctive treatment of post-surgical or post-traumatic acute pain
Device Story
Wellness Pro 2010 is a TENS device providing electrical stimulation for pain relief. Device delivers two waveform modes: Continuous (train of impulses) and Premodulated (amplitude-modulated impulses). Clinicians select from 250 pre-set frequencies or store custom regimens in 1000 memory slots. System includes 7.5V power adapter, four electrode wires, and four self-adhesive electrodes. Operated by clinicians in clinical settings to manage pain via electrode placement and regimen selection. Output is electrical stimulation applied to patient skin; intended to provide symptomatic relief for chronic or acute pain conditions. Device offers simplified user interface compared to predicates.
Clinical Evidence
Bench testing only. No clinical data provided. Device performance verified through risk analysis and compliance with safety standards including UL 60601-1-1, IEC 60601-2-10, and ISO 10993.
Technological Characteristics
TENS device; 7.5V power adapter; 50mm self-adhesive electrodes. Waveforms: Continuous and Premodulated. Memory: 1000 slots. Standards: UL 60601-1-1:2003, IEC 60601-2-10:1987, IEC 60601-1-1:1998, EN 60601-1-2:2001, ISO 10993. Connectivity: None specified.
Indications for Use
Indicated for patients requiring symptomatic relief and management of chronic, intractable pain, post-traumatic acute pain, or post-surgical acute pain.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
K062616
JUL - 6 2007
# 510(k) Summary
# Wellness Pro 2010
IntelSource Group, Inc. 3104 E. Camelback Road #528 Phoenix, AZ 85016 Phone: 888-880-7888 Fax: 480-452-1518
Contact: Mathew Wolfson TEL: 602-790-8034 FAX: 480-452-1518
Summary Prepared: Revised February 26, 2007
Trade Name: Wellness Pro 2010
Common Name: TENS device
Classification Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief. 21 CFR 882.5890, Class II, Product Code 84GZJ (prescription) or Stimulator, Nerve, Transcutaneous, for Pain Relief.
Predicate Device Identification:
21CFR:882.5890 Product Code: GZJ Device Class: II Legally Marketed Device: Medrelief ST-100, ST-150, ST-200, ST-300 Manufacturer: Healthonics, Inc. K#: K060669
21CFR:882.5890 Product Code: GZJ Device Class: II Legally Marketed Device: Fenzian Treatment System Manufacturer: Eumedic, Ltd K#: K041575
{1}------------------------------------------------
## Device Description
The Wellness Pro 2010 is designed to provide electrical stimulation, for the relief of chronic, intractable pain and as an adjunctive treatment of post-surgical or post-traumatic acute pain, in a single compact, lightweight user friendly package. Stimulation does not always cause sensation to relieve pain.
The WellnessPro 2010 has 2 modes of waveforms: Continuous and Premodulated. The device gives the clinician ability to store frequencies or to choose from a set of 250 frequencies that allow quick and easy selection, for prescription of stimulation regimens that can be later stored in any of the 1000 available memory slots.
The Wellness Pro 2010 package is comprised of the following items:
- 1. 7.5 Volt medical power adapter, conforming to industry standards.
- 2. Four industry standard wires for electrodes conforming to FDA standard,
- 3. Four standard commercially available and FDA approved, round 50(mm) self adhesive electrodes.
- 4. An instruction Manual.
#### Waveform Description
The Continuous mode produces a continuous train of impulses. The stimulation parameters are not automatically interrupted nor vay. The continuous mode is quite versatile because it may be applied with a variety of rate settings.
Image /page/1/Figure/10 description: The image shows a square wave pattern. The wave alternates between a high and low state, creating a series of rectangular pulses. The pattern appears to repeat consistently across the image, with each pulse having a similar width and amplitude. The image is black and white.
Variations of this mode look like this:
Image /page/1/Figure/12 description: The image shows a graph with a repeating pattern. The pattern consists of a series of rapid, high-frequency oscillations that last for 3 to 3.5 seconds, followed by a period of low activity that lasts for 0.5 to 1 second. This pattern repeats twice in the image.
It is still the same continuous mode with a continuous train of impulses (3-3.5 sec) with a small pause (0.5-1 sec) in between.
{2}------------------------------------------------
The Premodulated mode produces an amplitude-modulated impulse. The current intensity is modulated: The frequency remains the same while the intensity is modulated, (the Amplitude Modulation Frequency).
Image /page/2/Figure/1 description: The image shows two identical waveforms separated by a small gap. Each waveform is labeled as lasting for "4 sec". The waveforms appear to be square waves with varying amplitudes, increasing and then decreasing over the 4-second duration.
Variations of this mode looks like this:
Image /page/2/Figure/3 description: The image shows a graph with time on the x-axis. There are two sections labeled "3 sec" and two sections labeled "1 sec". The graph shows a series of square waves with varying amplitudes and frequencies. The square waves are clustered into two groups, each lasting 3 seconds, separated by a 1-second interval of relatively constant value.
The next variation is just a combination of the previously described modes
Image /page/2/Figure/5 description: The image shows a waveform with time markers. The waveform consists of a series of square waves with varying amplitudes and frequencies. The time markers indicate segments of 3 seconds and 1 second.
Note: In order to name these modes and variations, we simply label them as variations 1,2,3,4 and 5. All wave variations have same indications of use. See Below.
## Indications for use
There is no clinical basis or benefit for selecting one variation over the other. It is up to the physician to choose the correct placement, regimen and waveform variation to be used for symptomatic relief and management of chronic, intractable pain, post- traumatic acute pain, post-surgical acute pain.
#### Intended Use
The WellnessPro 2010 is an Electrotherapy Unit that is used for the following:
- 1. Relief of chronic, intractable pain.
- 2. Adjunctive treatment of post-surgical or post-traumatic acute pain
{3}------------------------------------------------
#### Declaration of Conformity
- ISO 10993 .
- UL 60601-1-1: 2003 ●
- IEC 60601-2-10: 1987 .
- IEC 60601-1-1: 1998 = A1 = A2:1995 ●
- EN 60601-1-2:2001 ●
#### Summary of Safety and Effectiveness
The WellnessPro 2010 is substantially equivalent to the Healthonics, Inc - MedRelief ST-100, ST-150, ST-200 and ST-300 devices as well as the Eumedic, Ltd - Fenzian Treatment System because of similar indication for use and unit characteristics.
The WellnessPro 2010 does not raise new issues of safety and effectiveness based on the above similarities.
#### Basis for substantial equivalence
The Wellness Pro 2010 is equivalent in function and intended use to the predicate devices. Both offer similar user options and functions. However the Wellness Pro 2010 offers simplified user interface, memory slot options and does not have a frequency range as great as the Medrelif ST series.
## Non-Clinical Data:
1. Risk Analysis results demonstrate acceptable and mitigated potential hazards.
2. The device meets the requirements for EMC, Radiated Emissions, Electrostatic Discharge, Radiated Immunity and device safety: UL 60601-1-1: 2003, IEC 60601-2-10: 1987, EC 60601-1-1: 1998 = A1 = A2:1995, EN 60601-1-2:2001, ISO 10993.
#### Conclusion:
The device is designed and labeled and verified for performance and safety. The performance is equivalent to legally marketed predicate devices. Risk Analysis does not demonstrate any design or performance potential hazards that are not adequately mitigated.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES · USA
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL - 6 2007
IntelSource Group, Inc. % Mr. Mathew Wolfson 3104 E. Camelback Road, #528 Phoenix, AZ 85016
RE: K062616
Trade/Device Name: Wellness Pro 2010 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: II Product Code: GZJ Dated: June 22, 2007 Received: July 2, 2007
Dear Mr. Wolfson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Mr. Mathew Wolfson
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
for Peter
Mark N. Melkerson
Director
Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
:
Device Name: Wellness Pro 2010
Indications for Use:
The Wellness Pro 2010 is an Electrotherapy Unit that is used for the following:
1. Relief of chronic, intractable pain.
2. Adjunctive treatment of post surgical or post-traumatic acute pain
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
| 510(k) Number | 1-062616 |
|---------------|----------|
|---------------|----------|
| Page | of |
|------|----|
|------|----|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.