BIOSPACE BODY COMPOSITION ANALYZERS, MODEL INBODY 230
Applicant
Biospace Corporation Limited
Product Code
MNW · Cardiovascular
Decision Date
Feb 15, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2770
Device Class
Class 2
Indications for Use
For use only in healthy subjects for Measurement of: Estimated : Skeletal muscle mass, Extra-Cellular Water (ECW), Intra-Cellular Water (ICW), Total Body Water, (TBW), Body Fat, Body Lean + Dry Lean, Metabolic Rates, Segmental Lean Mass. Actual : Weight, Body Mass Index (BMI),and Impedance Values
Device Story
InBody 230 is an impedance plethysmograph body composition analyzer. Device uses bioelectrical impedance analysis (BIA) to differentiate lean tissue, fat, and water. Input: electrical signals applied to body via 8 tactile electrodes placed on thumbs, palms, heels, and fore-feet. Device measures impedance at 20 and 100 kHz frequencies. Principle: lean tissue is highly conductive due to bound water/electrolytes; fat/bone are poor conductors. Device calculates resistance and capacitance components to derive body composition parameters. Used in upright position. Output: estimated skeletal muscle mass, body water, body fat, lean mass, metabolic rates, segmental lean mass, weight, BMI, and impedance values. Provides quantitative data to assist in health monitoring of healthy subjects.
Clinical Evidence
Device validated via human clinical trial. Bench and clinical testing data indicate the device is as safe and effective as the predicate devices.
Technological Characteristics
Impedance plethysmograph; 8-electrode tactile system (thumbs, palms, heels, fore-feet); operating frequencies 20 and 100 kHz; AC line power; upright patient positioning; BIA measurement principle analyzing resistance and capacitance.
Indications for Use
Indicated for healthy subjects for measurement of body composition parameters including skeletal muscle mass, body water compartments (ECW, ICW, TBW), body fat, lean mass, metabolic rates, segmental lean mass, weight, BMI, and impedance values.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
Biospace Body Composition Analyzer Model InBody 3.0 (K042528)
Biospace Body Composition Analyzer Model InBody 520 (K052646)
Tanita Ironman Innerscan Body Composition Monitor Model BC-558 (K062652)
Omron Body Composition Monitor and Scale Model HBF-500 (K062043)
Submission Summary (Full Text)
{0}------------------------------------------------
## 167603 OF 2
510(k) Summary Biospace Corporation Limited 363 Yangjae-dong, Seocho-qu, Seoul 137-898, Korea Tel : +82-2-501-3939, Fax : +82-2-501-3978 Homepage : http:// www.biospace.co.kr January 15, 2006
Contact: Kichul Cha, CEO
FEB ] 5 2007
l . Identification of the Device: Proprietary-Trade Name: Biospace Body Composition Analyzers, Model Inbody 230 Classification Names: 74 MNW ANALYZER, BODY COMPOSITION Common/Usual Name: Body fat meter
- 2. Equivalent legally marketed devices: Biospace Body Composition Analyzer Model InBody 3.0, K042528, Model InBody 520, K052646. Predicates for skeletal muscle mass are: Tanita Ironman Innerscan Body Composition Monitor Model BC-558 (K062652 and Omron Body Composition Monitor and Scale Model HBF-500 (K062043)
- 3. Indications for Use (intended use) For use only in healthy subjects for Measurement of: Estimated : Skeletal muscle mass, Extra-Cellular Water (ECW), Intra-Cellular Water (ICW), Total Body Water, (TBW), Body Fat, Body Lean + Dry Lean, Metabolic Rates, Segmental Lean Mass.
Actual : Weight, Body Mass Index (BMI),and Impedance Values
- 4. Description of the Device: Model Inbody 230 is an impedance plethysmograph body composition analyzer. This device determines body composition parameters based on bioelectrical impedance analysis (BIA). BIA relies on the differing behavior of biological tissues in response to an applied electrical current. Lean tissue is generally highly conductive because it contains large amounts of bound water and electrolytes, while fat tissue and bone are relatively poor conductors. By analyzing the response to electrical signals, BIA thereby permits differentiation of lean tissue, fat, and water and, in some instances, derivation of related body composition parameters. The total impedance resulting from BIA incorporates both resistance and capacitance components.
- 5. Safety and Effectiveness, comparison to predicate device. The results of bench and clinical testing indicates that the new device is as safe and effective as the predicate devices.
{1}------------------------------------------------
| | Biospace Body Composition<br>Analyzer Model InBody 3.0 | Biospace Body Composition<br>Analyzer Model InBody 520 | Biospace Body Composition<br>Analyzer Model InBody 230<br>(NEW) |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | K042528 | K052646 | |
| Intended Use | Body composition analyzer | Body composition analyzer | Body composition analyzer |
| | Measurement Of: | Measurement Of: | Measurement Of: |
| | Estimated : | Estimated : | Estimated : |
| | Extra-Cellular Water(ECW),<br>Intra-Cellular Water(ICW),<br>Total Body Water,<br>ECW/TBW | Extra-Cellular Water(ECW),<br>Intra-Cellular Water(ICW),<br>Total Body Water,<br>ECW/TBW | Extra-Cellular Water(ECW),<br>Intra-Cellular Water(ICW),<br>Total Body Water, |
| Indications for<br>Use | Body Fat,<br>Body Lean + Dry Lean,<br>Metabolic Rates,<br>Segmental Lean Mass<br>Actual:<br>Weight,<br>Body Mass Index (BMI),<br>and Impedance Values:<br>5, 50, 250, 500kHz | Body Fat,<br>Body Lean + Dry Lean,<br>Metabolic Rates,<br>Segmental Lean Mass<br>Actual:<br>Weight,<br>Body Mass Index (BMI),<br>and Impedance Values:<br>5, 50, 500kHz | Skeletal Muscle Mass<br>Body Fat,<br>Body Lean + Dry Lean,<br>Metabolic Rates,<br>Segmental Lean Mass<br>Actual:<br>Weight,<br>Body Mass Index (BMI),<br>and Impedance Values:<br>20, 100kHz |
| Analysis Method | Bioelectrical Impedance | Bioelectrical Impedance | Bioelectrical Impedance |
| Operating<br>parameters | Frequency :<br>5, 50, 250, 500kHz | Frequency :<br>5, 50, 500kHz | Frequency :<br>20, 100kHz |
| Electrode Type | tactile | tactile | Tactile |
| Number / | 8 electrodes | 8 electrodes | 8 electrodes |
| Placement of<br>Electrodes | placed on thumbs, palms,<br>heels, and fore-feet | placed on thumbs, paims,<br>heels, and fore-feet | placed on thumbs, palms,<br>heels, and fore-feet |
| Impedance<br>Measuring Site | Right Arm, Left Arm, Trunk,<br>Right Leg, Left Leg | Right Arm, Left Arm, Trunk,<br>Right Leg, Left Leg | Right Arm. Left Arm, Trunk,<br>Right Leg, Left Leg |
| Patient Position | Upright | Upright | Upright |
| Power Source | AC line | SAME | SAME |
# 6. Substantial Equivalence Chart, Model Inbody 230 BODY COMPOSITION ANALYZER
PAGE 2. OF 2
### 7. Conclusion
After analyzing both bench and clinical testing data, it is the conclusion of Biospace that the Model After analyzing both benefi and chilled tooling data, it is as safe and effective as the predicate devices, model 250, DODT COMI OUTFORT ANTESTERS to to substantially equivalent to the predicate allu has lew technological urrerences, thas been validated via human clinical trial.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
Biospace Corporation Ltd. c/o Daniel Kamm, P.E. Regulatory Engineer Kamm & Associates P.O. Box 7007 DEERFIELD IL 60015
Re: K062603
Trade/Device Name: Biospace Body Composition Analyzer, Model InBody 230 Regulation Number: 21 CFR §870.2770 Regulation Name: Impedance plethysmograph Regulatory Class: II Product Code: MNW Dated: January 28, 2007 Received: January 31, 2007
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding-and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/2/Picture/10 description: The image is a circular logo for the FDA Centennial, celebrating 1906-2006. The central design features the FDA acronym in bold, stylized letters above the word "Centennial". Three stars are aligned beneath the word "Centennial". The logo is encircled by text, which is partially obscured due to the image quality.
Protecting and Promoting Public Health
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
510(k) Number (if known): K062603
Device Name: Biospace Body Composition Analyzer, Model Inbody 230
Indications For Use:
For use only in healthy subjects for Measurement Of: Estimated : Estimatou :
Skeletal muscle mass, Extra-Cellular Water (ECW), Intra-Cellular Water (ICW), Total Body Okeiolar massio mass, -------------------------------------------------------------------------------------------------------------------------------------------------------
Actual : Weight, Body Mass Index (BMI),and Impedance Values
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use X (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Syverson
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices
510(k) Number K062603
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.