← Product Code [DQY](/productcode/DQY) · K062594

# MODIFICATION TO POLARCATH PERIPHERAL DILATATION SYSTEM (K062594)

_Boston Scientific Corporation · DQY · Sep 20, 2006 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K062594

## Device Facts

- **Applicant:** Boston Scientific Corporation
- **Product Code:** [DQY](/productcode/DQY.md)
- **Decision Date:** Sep 20, 2006
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The PolarCath Peripheral Dilatation System is indicated to dilate stenoses in the peripheral vasculature (iliac, femoral, popliteal, infrapopliteal, renal and subclavian arteries) and for the treatment of obstructive lesions of polytetrafluoroethylene (PTFE) access grafts or native arteriovenous dialysis fistulae. The PolarCath Peripheral Dilatation System is also indicated for post-deployed stent expansion of self-expanding peripheral vascular stents.

## Device Story

PolarCath Peripheral Dilatation System is a percutaneous transluminal angioplasty (PTA) catheter system. Components include a catheter, inflation unit, connecting cable, rechargeable battery pack, and disposable 14g liquid nitrous oxide cartridge. Device operates by delivering cryotherapy via the inflation medium to dilate stenoses in peripheral arteries, treat obstructive lesions in dialysis access grafts/fistulae, and expand self-expanding stents. Used in clinical settings by physicians. System provides controlled dilatation through cryo-inflation technology. Benefits include improved vessel patency and stent apposition.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

System comprises a catheter, inflation unit, connecting cable, and rechargeable battery. Inflation medium is liquid nitrous oxide (14g disposable cartridge). Operates as a PTA catheter for peripheral vascular dilation. No software or complex algorithms described.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- PolarCath™ Peripheral Dilatation System ([K062594](/device/K062594.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### 3. 510(k) Summary

| Sponsor:           | Boston Scientific Corporation<br>160 Knowles Drive<br>Los Gatos, CA 95032 |
|--------------------|---------------------------------------------------------------------------|
| Contact Person:    | Sarah Peterson                                                            |
| Phone Number:      | 763-494-2133                                                              |
| Fax Number:        | 763-494-2981                                                              |
| Prepared:          | August 30, 2006                                                           |
| Trade Name:        | PolarCath™ Peripheral Dilatation System                                   |
| Common Name:       | Percutaneous Transluminal Angioplasty Catheter                            |
| Classification:    | II                                                                        |
| Product Code:      | LIT/DQY<br>21 CFR 870.1250                                                |
| Predicate Devices: | PolarCath™ Peripheral Dilatation System                                   |

# Device Description

The PolarCath Peripheral Dilatation System consists of a Catheter, Inflation Unit, connecting cable and a rechargeable battery pack with recharging unit and battery receptacle. The inflation medium (liquid nitrous oxide) is provided in a disposable 14 gram cartridge.

# Indications for Use

The PolarCath Peripheral Dilatation System is indicated to dilate stenoses in the peripheral vasculature (iliac, femoral, popliteal, infrapopliteal, renal and subclavian arteries) and for the treatment of obstructive lesions of polytetrafluoroethylene (PTFE) access grafts or arteriovenous dialysis fistulae. The PolarCath Peripheral Dilatation System is also indicated for post-deployed stent expansion of self-expanding peripheral vascular stents.

# Substantial Equivalence

The PolarCath Peripheral Dilatation System design, materials, manufacturing process and intended use are substantially equivalent to the predicate device and other marketed PTA catheters.

# Performance Data

The safety and effectiveness of the modified PolarCath Peripheral Dilatation System is demonstrated with design control activities and bench testing on file at Boston Scientific.

{1}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/2 description: The image shows a logo with an abstract eagle design. The eagle is depicted with stylized lines forming its head, wings, and tail. The logo is encircled by text that reads "MAN SERVICES-USA DEPARTMENT". The text is arranged in a circular fashion around the eagle emblem.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 2 0 2008

Boston Scientific, Inc c/o Ms. Sarah Peterson Regulatory Affairs Specialist 160 Knowles Drive Los Gatos, CA 95032

Re: K062594

Trade Name: PolarCath™ Peripheral Dilation System Regulation Number: 21 CFR 870.1250 Regulation Name: Angioplasty Catheter Regulatory Class: II (two) Product Code: DQY, LIT Dated: August, 2006 Received: September 1, 2006

Dear Ms. Peterson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

#### Page 2 -- Ms. Sarah Peterson

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Maco Sacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

R. Vochner

(A Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

2. 510(k) Indications for Use

Page 1 of 1

K062594 510(k) Number (if known):

Device Name:

PolarCath™ Peripheral Dilatation System

Indications for use:

The PolarCath Peripheral Dilatation System is indicated to dilate stenoses in the peripheral vasculature (iliac, femoral, popliteal, infrapopliteal, renal and subclavian arteries) and for the treatment of obstructive lesions of polytetrafluoroethylene (PTFE) access grafts or native arteriovenous dialysis fistulae. The PolarCath Peripheral Dilatation System is also indicated for post-deployed stent expansion of self-expanding peripheral vascular stents.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

ourna 2- Victories

(Division Sign-Off) Division of Cardiovascular Devices

510(k) Number Ko6 2594

---

**Source:** [https://fda-staging.innolitics.com/device/K062594](https://fda-staging.innolitics.com/device/K062594)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
