CALAVIEW COLORS (ETAFILCON A) SOFT (HYDROPHILIC) CONTACT LENS
K062541 · Innova Vision, Inc. · LPL · Jun 7, 2007 · Ophthalmic
Device Facts
Record ID
K062541
Device Name
CALAVIEW COLORS (ETAFILCON A) SOFT (HYDROPHILIC) CONTACT LENS
Applicant
Innova Vision, Inc.
Product Code
LPL · Ophthalmic
Decision Date
Jun 7, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
Calaview Colors (Etafilcon A) Soft (hydrophilic) Contact Lens is indicated for daily wear the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopic or hyperopic and may exhibit refractive astigmatism up to 2.00 diopters that does not interfere with visual acuity. The eye care practitioner may prescribe the contact lens for frequent replacement wear, with cleaning, rinsing and disinfection and scheduled replacement as prescribed by the eyecare professional. The contact lens may be disinfected using chemical (not heat) disinfection system.
Device Story
Calaview Colors are soft, hydrophilic contact lenses manufactured via spin-casting. Composed of Etafilcon A (HEMA and methacrylic acid copolymer cross-linked with TMPTMA and EGDMA). Lenses feature pigments in the non-optic area to mask or enhance iris color using FDA-approved additives (iron oxides, titanium dioxide, copper phthalocyanine, etc.). Supplied sterile in blister packs with isotonic phosphate buffered saline. Used by patients for daily wear to correct refractive errors; prescribed by eye care practitioners. Benefits include visual correction and cosmetic iris enhancement. Safety established through biocompatibility, leachability, and sterilization validation.
Clinical Evidence
Clinical study conducted on 65 human eyes over 6 months. Results showed nearly 100% vision correction and high patient satisfaction. No severe complications or significant side effects reported. Data supports safety and effectiveness for daily wear.
Indicated for daily wear correction of visual acuity in non-aphakic, non-diseased eyes with myopia, hyperopia, or refractive astigmatism up to 2.00 diopters. For prescription use only.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
ACUVUE 2 colours Brand (Etafilcon A) Contact Lens (K010114)
Discon (Etafilcon A) Soft (hydrophilic) Contact Lens (K051129)
Submission Summary (Full Text)
{0}------------------------------------------------
K062541
# SUMMARY OF SAFETY AND EFFECTIVENESS FOR Calaview Colors (Etafilcon A) Soft (hydrophilic) Contact Lens for Daily Wear
| | Submitter Information: | |
|-----------------|-----------------------------------------------------------------------------------------------|--------------|
| Company: | INNOVA VISION INC.<br>No. 231-1, Wen-Te Road, Chiung-lin Village,<br>Hsin-Chu County, Taiwan. | JUN - 7 2006 |
| Contact Person: | Mr. Michael w. Hsu,<br>Executive Vice President | |
| Phone: | 886-3-592-7299 | |
| FAX: | 886-3-592-6213 | |
| E-Mail: | mikehsu@innovavision.com.tw | |
| Date Prepared | July. 25, 2006 | |
### ldentification of Device:
| Classification Name: | Soft hydrophilic contact lens, per 21 CFR. 886.5925 |
|-----------------------|---------------------------------------------------------------------------------|
| Trade Name: | Calaview Colors (Etafilcon A) Soft (hydrophilic)<br>Contact Lens for Daily Wear |
| Common or usual Name: | Soft (hydrophilic) Contact lens (daily wear) |
| FDA Classification: | Class II |
### Predicate Device:
| ACUVUE 2 colours Brand<br>(Etafilcon A) Contact Lens cleared<br>via K010114 Johnson & Johnson Co. USA. | Discon (Etafilcon A) Soft (hydrophilic)<br>Contact Lens for Daily Wear cleared<br>via K051129 from Innova Vision Inc., Taiwan. |
|--------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
#### Indications for Use
Calaview Colors (Etafilcon A) Soft (hydrophilic) Contact Lens is indicated for daily wear the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopic or hyperopic and may exhibit refractive astigmatism up to 2.00 diopters that does not interfere with visual acuity.
The eye care practitioner may prescribe the contact lens for frequent replacement wear, with cleaning, rinsing and disinfection and scheduled replacement as prescribed by the eyecare professional. The contact lens may be disinfected using chemical (not heat) disinfection system."
{1}------------------------------------------------
### Description of Device
Calaview Colors Soft (hydrophilic) Contact Lens are available as non-spherical lenses manufactured by spin-casting method. The model illuminated with high water content (58 ± 1%). The hydrogel lens' material is a random copolymer composed of 2-hydroxyethyl methacrylate (HEMA) and methacrylic acid (MAA), which was cross-linked with 1,1,1-trimethylolpropane trimethacrylate (TMPTMA) and Ethylene Glycol Dimethacrylate (EGDMA) via UV photo- polymerization. The Calaview Colors Contact lens contains pigments around the non-optic area that will mask or enhance the color of the natural iris without blocking the light transmittance. The lens is colored with one or more of the FDA-approval color additives: iron oxides, titanium dioxide, phthalocyaninato copper, phtalocyanine green, vat orange 1. The Calaview Colors (Etafilcon A) Soft (hydrophilic) Contact Lens is available in the following opaque colors: Blue, green, violet, gray, hazel, white, yellow, gold, orange red and black. Lenses are supplied sterile in sealed blister packers containing sterile isotonic phosphate buffered saline solution. The compatibility and package integrity of the blister pack packaging system has been demonstrated and successfully used for other marketed lens products:
### Summary of Clinical Study
The Calaview Colors (Etafilcon A) Soft (hydrophilic) Contact Lens were tested in 65 human eyes separately within 6 months. Nearly a hundred percent of the participants' vision was corrected and nearly all were satisfied with the lens wearing and care of lenses. In general, these products are good and safe for customers.
Calaview Colors lenses have been wide-used around the world, including Taiwan. China, Europe, etc. Among the users being daily worn the Calaview Colors (Etafilcon A) Soft (hydrophilic) Contact Lens, all the procedures were in generally stable condition without severe complication. There are no significant side effects and complaints to be observed.
#### Nonclinical Studies
A series of non-clinical performance tests were performed to demonstrate the safety and effectiveness of the Calaview Colors (Etafilcon A) Soft (hydrophilic) Contact Lens, and establish substantial equivalence to predicate lenses-Discon Lens (K051129); ACUVUE 2 colours Brand (Etafilcon A) Contact Lens (K010114). All testing was conducted in accordance with the May 1994 FDA guideline titled Premarket Notification 510(K) Guidance Document for Class IV Contact lenses, and in conformance to applicable device regulations. The evidence of substantial equivalent to the predicate lens described as follow:
{2}------------------------------------------------
### a) Technological characteristics studies
There characterizations of Calaview Colors (Etafilcon A) Soft (hydrophilic) Contact Lens are equivalent and comparable to those of predicate lenses.
| Characteristic | Calaview Colors | Acuvue<br>Colors | Discon |
|-------------------------------------------------|-----------------------------------------------------|-------------------------------------------------------|-----------------------------------------------------|
| %Water content | 57 to 59 | 58 | 57 to 59 |
| Refractive index | 1.407 | 1.40 | 1.407 |
| Specific Gravity | 1.05 | 1.05 | 1.05 |
| Oxygen permeability (edged<br>corrected) @ 35°C | $24\times10^{-11}$<br>[(cm2/sec)(ml<br>O2/ml-mmHg)] | $22.4\times10^{-11}$<br>[(cm2/sec)(ml<br>O2/ml-mmHg)] | $24\times10^{-11}$<br>[(cm2/sec)(ml<br>O2/ml-mmHg)] |
| %Light Transmission | >93 | >85 | >93 |
| Base Curve Radius, mm | 8.00~9.00 | 7.85~10.00 | 8.00~9.00 |
| Diameter, mm | 13.8~14.2 | 12.0~15.0 | 13.8~14.2 |
| Center Thickness, mm | 0.08~0.20 | 0.06~1.00 | 0.08~0.20 |
| Power, Diopters | +6.0D~-20.0D | +20.0D~-20.0D | +6.0D~-20.0D |
### b) Biocompatibility
In accordance with the May 1994 Guidance Document for daily wear contact lenses, toxicity studies have been conducted on the model: Calaview Colors (Efafilcon A) Soft (hydrophilic) Contact Lens. The Irritation test in the rabbit eye and Systemic toxicity studies indicate the extracts would be considered as non-toxic and nor The Cytotoxicity testing demonstrates the lens is not cytotoxic under the irritated. conditions of the study.
#### c) Microbiology
Steam sterilization process has been validated to deliver a minimum SAL of 10°, thereby complying with the requirement of FDA Group IV. There is shelf-life stability data supporting that the lens remains sterile through the expiration date claimed for the product.
#### d) Leachability
Studies were conducted to determine the leachable materials from the finished lens. The results show that, at the levels of the detection reported, there are no leachable monomers and addictive residues.
{3}------------------------------------------------
### Substantial equivalence Statement
Testing performed on the Calaview Colors Soft (hydrophilic) Contact Lens for Daily Wear indicated that it can support the efficiency and security use as well as the predicate devices-Discon (Etafilcon A) Soft (hydrophilic) Contact Lens for Daily Wear (K051129) & ACUVUE 2 colours Brand (Etafilcon A) Contact Lens (K010114), when used in accordance with the instructions for use. It is due to the facts that the risks and benefits of the subject device are the same as soft contact lenses for to the daily wear.
In conclusion, it is Innova's conviction that data submitted in this 510(k) to validate the claim of substantial equivalency, substantiates our ability to manufacture a soft contact lens, the Calaview Colors Soft (hydrophilic) Contact Lens, with the same established safety profile and effectiveness as the predicate device-- Discon (Etafilcon A) Soft (hydrophilic) Contact Lens for Daily Wear (K051129) & ACUVUE 2 colours Brand (Etafilcon A) Contact Lens (K010114).
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a series of human profiles, rendered in black.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Innova Vision, Inc. c/o Ms. Jennifer Reich Harvest Consulting Corp. 2904 N. Boldt Dr. Flagstaff AZ 86001
JUN - 7 2007
Re: K062541
Trade/Device Name: Calaview Colors (etafilcon A) Soft Hydrophilic Contact Lens for Daily Wear
Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL Dated: May 25, 2007 Received: May 30, 2007
Dear Ms. Reich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
M.B. Eichelman S.M.D
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
510(k) Number:
Device Name: Calaview Colors Soft (hydrophilic) Contact Lens for Daily Wear
Indications for Use:
Calaview Colors (Etafilcon A) Soft (hydrophilic) Contact Lens is indicated for daily wear the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopic or hyperopic and may exhibit refractive astigmatism up to 2.00 diopters that does not interfere with visual acuity.
The eye care practitioner may prescribe the contact lens for frequent replacement wear, with cleaning, rinsing and disinfection and scheduled replacement as prescribed by the eyecare professional. The contact lens may be disinfected using chemical (not heat) disinfection system."
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|---------|
| (Division Sign-Off) | |
| Division of Ophthalmic Ear, | |
| Nose and Throat Devices | |
| 510(k) Number | K062541 |
Prescription Use:
or Over the Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.