BLUE POWDERED LATEX EXAMINATION GLOVES, NON-STERILE
Applicant
Multisafe Sdn. Bhd.
Product Code
LYY · General Hospital
Decision Date
Apr 13, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
Blue Powdered Latex Examination Gloves, Non-sterile is a disposable medical device made from natural rubber latex compound. It is intended to be worn on the hands as barrier against potentially infectious materials and other contamination.
Device Story
Blue Powdered Latex Examination Gloves are disposable, non-sterile natural rubber latex gloves. Used by healthcare and similar personnel as a protective barrier against infectious materials, bodily fluids, waste, and environmental contaminants. Device functions as a physical barrier; worn on hands. No electronic, software, or algorithmic components. Performance verified via water leak testing, dimensional analysis, physical property testing (tensile strength/elongation), and biocompatibility assessments (dermal irritation/sensitization).
Clinical Evidence
Bench testing only. No clinical data. Testing included water leak test (AQL 1.5), dimensions, palm width, thickness, tensile strength (min 18 MPa before aging, 14 MPa after), ultimate elongation (min 700% before, 500% after), powder content (ASTM 6124, <10.0 mg/glove), protein content (ASTM 5712, <200 µg/dm²), and biocompatibility (primary dermal irritation and dermal sensitization studies).
Indicated for use by healthcare and similar personnel as a protective barrier on hands against infectious materials, fluids, waste, or environmental contamination. Intended for disposable use.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
K062526
## 17.0 SMDA 510(K) SUMMARY
APR 1 3 2007
| 1. | Applicant | : | MULTISAFE SDN. BHD<br>Lot 764, Bidor Industrial Estates,<br>35500 Bidor,<br>Perak Darul Ridzuan,<br>Malaysia. |
|----|--------------------------|---|---------------------------------------------------------------------------------------------------------------|
| | Tel No. | : | 605-4348269 |
| | Fax No | : | 605-4348266 |
| | Name of Contact Person | : | 1. Mr. Abd Hadi bin Husin<br>2. Ms. Rosnani Binti Hassan Besari |
| | Date of Summary Prepared | : | 28th February 2007 |
| 2. | Name of Device | | |
| | Trade Name | : | Blue Powdered Latex Examination<br>Glove, Non-Sterile |
| | Common Name | : | Examination Gloves |
| | Classification Name | : | Patient Examination Gloves (Class 1 in<br>US and European markets, Class IIa in<br>Canadian Market) |
## 3. Identification Of the Legally Marketed Devices
Latex Patient Examination Gloves LYY, Powdered that meets all the requirements of FDA, ASTM D 3578, ISO 11193, EN 455 : Part 1, Part 2 and Part 3, CMDR, MDR and CGSB.
.
{1}------------------------------------------------
#### Description Of the Devices 4.
Latex Patient Examination Gloves, LYY, Powdered
#### 5. The Intended Use of Glove
A medical glove is worn on the hands of healthcare and similar personnel as a barrier against potentially infectious materials and other contamination between healthcare and patient's body, fluid, waste or environment.
- The Quality performance of the glove are shown in the table above (. meets ASTM D 3578-01 Standard and FDA's requirement
- The Biocompatibility Test consists of Primary Dermal Irritation Study 7. And the Dermal Sensitization Study. The gloves pass the Biocompatibility Test.
{2}------------------------------------------------
## Summary of Quality Performance 8.
...
!
:
The Quality Performance of gloves based on the requirements stated in item 6.
| TEST | REQUIREMENT | IN-HOUSE QUALITY<br>PERFORMANCE |
|-------------------------------------|--------------|---------------------------------|
| 1. Water Leak Test<br>(1000ml) | GII, AQL 2.5 | Pass GI, AQL 1.5 |
| 2. Length (mm)<br>Size: XS | Min. 220 | Minimum 290 mm for all<br>sizes |
| S | Min. 230 | |
| M | Min 230 | |
| L | Min 230 | |
| XL | - | |
| 3 Palm Width (mm)<br>Size : XS | $70\pm10$ | 75-78 |
| S | $80\pm10$ | 82-86 |
| M | $90\pm10$ | 92-95 |
| L | $100\pm10$ | 103-109 |
| XL | - | 111-115 |
| 4. Thickness (mm)<br>(Single layer) | | |
| Finger | Min 0.08 | Min 0.10 |
| Palm | Min 0.08 | Min 0.10 |
| 5. Physical<br>Properties | | |
| Before Ageing | | |
| Tensile Strength (MPa) | Min. 14 | Min 18 |
| Ultimate Elongation (%) | Min 700 | Min 700 |
| After Ageing | | |
| Tensile Strength (MPa) | Min.14 | Min. 14 |
| Ultimate Elongation (%) | 500 | 500 |
| 6. Powder Content | ASTM 6124 | Below 10.0 mg/glove |
| 7. Protein Content | ASTM 5712 | Below 200 µg/dm² |
14
. . .
{3}------------------------------------------------
#### Conclusion 9.
We concluded that the Blue Powdered Latex Examination Gloves, Protein Content Labeling (200 µg/dm2 or less), Non-Sterile meets:
- ASTM D3578 Standard
- · FDA pin hole requirement
- FDA minimum powder residual content
- · CMDR requirement
- CGSB requirement
- · Label Claim of 200 µg/dm2 or less of total water extractable protein content
{4}------------------------------------------------
......
. Comparison table and discussion of the similarities and difference of Blue Powdered Latex :amination Gloves (Non-Sterile) compared to Multisafe Powdered Latex Examination Gloves, (Non-.. erile).
| No | Description | Requirement | 510K No | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|------------------------|------------------------|
| | | | K993811 | K062526 |
| 1 | Water Tight Test<br>(1000ml) | GII, AQL 2.5 | Pass GI, AQL 1.0 | Pass GI, AQL 1.5 |
| 2 | Dimension Test | Min. 290 (Long Glove)<br>Min. 230 (Short Glove) | Min 230 for all sizes | Min 290 for all sizes |
| 3 | Palm Width (mm) | XS : 70 +/- 10<br>S : 80 +/- 10<br>M : 90 +/- 10<br>L :100 +/- 10 | Meet the specification | Meet the specification |
| 4 | Thickness (mm)<br>i) Finger<br>ii) Palm | Min. 0.08<br>Min. 0.08 | Meet the specification | Meet the specification |
| 5 | Physical Properties (Before<br>Aging)<br>i) Tensile Strength<br>(Mpa) | Min 14 | Min 18 | Min. 18 |
| | ii) Ultimate<br>Elongation (%) | Min. 700 | Min. 700 | Min. 700 |
| | Physical Properties<br>(After Aging) | | | |
| | i) Tensile Strength<br>(Mpa) | Min. 14 | Min. 14 | Min. 14 |
| | ii) Ultimate<br>Elongation (%) | Min. 500 | Min. 500 | Min. 500 |
| 6 | Powder Content | ASTM 6124 | Meet the specification | Below 10.0mg/glove |
| 7 | Protein Content | ASTM 5712 | Meet the specification | Below 200 µg/dm² |
| 8 | Biocompatibility Test<br>i) Primary Skin<br>Irritation | No Animal Irritation | Yes | Yes |
| | ii) Dermal<br>Sensitization Assay<br>(Animal Study) | No Animal Irritation | Yes | Yes |
| 9 | Colour Extraction | Others | Normal Creamy<br>White | Blue Colour |
| 10 | Ingredient<br>a) Latex<br>b) Accelerators<br>c) Antioxidant<br>d) Curing Agent<br>e) Activator | Natural Rubber latex<br>ZDBC<br>ZDEC<br>Sulphur<br>Zinc Oxide | √ | √ |
| f) Dispersing Agent<br>g) Filler / Pigment<br>h) Donning Powder<br>i) Pigment | Tamol<br>Titanium Dioxide<br>Modified Corn Starch<br>Flexobrite Blue | | | |
| | | √<br>√<br>√<br>X | √<br>√<br>√<br>√ | |
1 ୧
{5}------------------------------------------------
:
:
Conclusion :
125
P
1
e Blue Powdered Latex Examination Gloves (Non-Sterile) is similar with Multisafe Powdered Latex Examination Gloves (Non-Sterile) except the dimension of length and colour extraction (
MULTISAFE SDN. BHDA
R ABD HADI HUSIN (Plant Mahager)
। ।
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a serpent entwined around it, and three parallel lines above it. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" is arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# APR 1 3 2007
Mr. Abd Hadi Bin Husin Quality Management Representative/Plant Manager Multisafe Sdn Bhd Lot 764, Bidor Industrial Estate 35500 Bidor Perak Darul Ridzuan MALAYSIA
Re: K062526
Trade/Device Name: Blue Powdered Latex Examination Gloves, Non-Sterile Regulation Number: 880.6250 Regulation Name: Patient Examination Gloves Regulatory Class: I Product Code: LYY Dated: February 14, 2007 Received: March 19, 2007
Dear Mr. Husin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
# Page 2 - Mr. Husin
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sylitte Y. Micham Om.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
510(k) Number (if known): K062526
Device Name: Blue powdered latex Examination Gloves, Non-sterile
Indications For Use:
Blue Powdered Latex Examination Gloves, Non-sterile is a disposable medical device made from natural rubber latex compound. It is intended to be worn on the hands as barrier against potentially infectious materials and other contamination.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Stales H.M. Murphy, bo
(K) Number K 062526
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.