← Product Code [MAI](/productcode/MAI) · K062466

# ARTHREX INTERFERENCE SCREW (K062466)

_Arthrex, Inc. · MAI · Nov 29, 2006 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K062466

## Device Facts

- **Applicant:** Arthrex, Inc.
- **Product Code:** [MAI](/productcode/MAI.md)
- **Decision Date:** Nov 29, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Arthrex Interference Screws are intended to be used for fixation of tissue including ligament or tendon to bone and bone tendon bone. Interference fixation with PEEK, Titanium, or Bioabsorbable Polymer Interference Screws is appropriate for surgeries of the knee, shoulder, elbow, ankle, foot, and hand/wrist where the sizes offered are patient appropriate. Specifically; - Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Shoulder: Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction - Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Foot/Ankle: Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair, Flexor Hallucis Longus for Achilles Tendon reconstruction, tendon transfers in the foot and ankle - Anterior Cruciate Ligament Repair, Medial Collateral Ligament Knee: Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Illiotibial Band Tenodesis - Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction - Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction, Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty), Carpal Ligament Reconstructions and repairs, tendon transfer in the hand/ wrist

## Device Story

Arthrex Interference Screws are orthopedic fixation devices used to secure soft tissue (ligament/tendon) to bone. Manufactured from PEEK, titanium, or bioabsorbable polymers; screws provide mechanical interference fixation within bone tunnels. Used by orthopedic surgeons in clinical/OR settings for reconstructive procedures across multiple joints (knee, shoulder, elbow, ankle, foot, hand/wrist). Device functions as a mechanical fastener; no electronic or software components. Benefits include stable tissue-to-bone fixation to facilitate healing and joint stabilization.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Materials: PEEK, titanium, or bioabsorbable polymers. Principle: Mechanical interference fixation. Form factor: Threaded screw for bone fixation. No software, electronics, or energy sources.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Arthrex Interference Screw ([K052607](/device/K052607.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# NOV 2 9 2006

# VIII. 510(k) Summary of Safety and Effectiveness

# Arthrex Interference Screw

| Manufacturer / Sponsor | Arthrex, Inc.<br>1370 Creekside Boulevard<br>Naples, Florida 34108-1945                                                 |
|------------------------|-------------------------------------------------------------------------------------------------------------------------|
| 510(k) Contact:        | Ann Waterhouse, RAC<br>Regulatory Affairs Project Manager<br>Telephone: (239) 643-5553 ext. 1179<br>FAX: (239) 598-5508 |
| Trade Name:            | Arthrex Interference Screw                                                                                              |
| Common Name:           | Fastener, Fixation, bone, degradable and<br>nondegradable                                                               |
| Product Code /         | HWC, 21 CFR 888.3040<br>MAI, 21 CFR 888.3030<br>MBI, 21 CFR 888.3040                                                    |
| Classification Name:   | Fixation, Fastener, bone                                                                                                |
| Predicate Device:      | Arthrex Interference Screw, K052607                                                                                     |

## Device Description and Intended Use:

The Arthrex Interference Screws are intended to be used for fixation of tissue including ligament or tendon to bone and bone tendon bone. Interference fixation with PEEK, Titanium, or Bioabsorbable Polymer Interference Screws is appropriate for surgeries of the knee, shoulder, elbow, ankle, foot, and hand/wrist where the sizes offered are patient appropriate. Specifically;

- Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Shoulder: Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
- Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Foot/Ankle: Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair, Flexor Hallucis Longus for Achilles Tendon reconstruction, tendon transfers in the foot and ankle

{1}------------------------------------------------

- Anterior Cruciate Ligament Repair, Medial Collateral Ligament Knee: Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Illiotibial Band Tenodesis
- Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction
- Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction, Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty), Carpal Ligament Reconstructions and repairs, tendon transfer in the hand/ wrist

#### Substantial Equivalence:

The Arthrex Interference Screw is substantially equivalent to the predicate Arthrex Interference Screw in which the basic features and intended uses are the Any differences between the Arthrex Interference Screw and the same. predicate are considered minor and do not raise questions concerning safety and effectiveness. Based on the information submitted, Arthrex, Inc. has determined that the PEEK, PLLA, and Titanium Interference Screw are substantially equivalent to the currently marketed predicate device.

{2}------------------------------------------------

Image /page/2/Picture/12 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or other bird-like figure.

NOV 2 9 2006

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Arthrex, Inc. % Ann Waterhouse, RAC Project Manager, Regulatory Affairs 1370 Creekside Boulevard Naples, Florida 34108-1945

Re: K062466

Trade/Device Name: Interference Screws Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Codes: MAI, MBI, HWC Dated: November 6, 2006 Received: November 8, 2006

Dear Ms. Waterhouse:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{3}------------------------------------------------

#### Page 2 - Ann Waterhouse, RAC

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Вашава Внешир

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### Indications for Use Form ==============================================================================================================================================================================

KOG 2466 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

## Device Name: Arthrex Interference Screw

## Indications for Use:

The Arthrex Interference Screws are intended to be used for fixation of tissue including ligament or tendon to bone and bone tendon bone. Interference fixation with PEEK, Titanium, or Bioabsorbable Polymer Interference Screws is appropriate for surgeries of the knee, shoulder, elbow, ankle, foot, and hand/wrist where the sizes offered are patient appropriate. Specifically;

- Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Shoulder. Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
- Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Foot/Ankle: Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair, Flexor Hallucis Longus for Achilles Tendon reconstruction, tendon transfers in the foot and ankle
- Anterior Cruciate Ligament Repair, Medial Collateral Ligament Knee: Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Illiotibial Band Tenodesis
- Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Elbow: Reconstruction
- Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Radial Collateral Ligament Reconstruction, Reconstruction, Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty), Carpal Ligament Reconstructions and repairs, tendon transfer in the hand/ wrist

AND/OR Over-The-Counter Use Prescription Use X (21 CFR 801 Subpart C) (Per 21 CFR 801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

============================================================================================================================================================================== Concurrence of CDRH, Office of Device Evaluation (Open vision

Division of General, Restoracya. and Neurological Devices

510(k) Number / < 0 (24) 246 (s

---

**Source:** [https://fda-staging.innolitics.com/device/K062466](https://fda-staging.innolitics.com/device/K062466)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).