To assist persons who have difficulty rising from a seated position to a standing position.
Device Story
Modular standing frame for indoor use; supports users in standing, weight-bearing position. Pivot points allow transition from seated to standing; minimizes shear. Standard version uses manual hydraulic pump; optional 'Pow'r Up' version uses low-voltage motor-driven linear actuator powered by two 12V batteries. Actuated via hand-held control pendant. Includes emergency release. Configurations include Basic (passive standing), Shadow Support Tray (workstation), and Glider (leg exercise). Fit determined by therapist or physician. Provides postural support; assists users with physical disabilities in achieving standing position; facilitates passive standing or exercise.
Clinical Evidence
Bench testing only. Device tested against BS EN 12182:1999, IEC/EN 60601-1:1998, and UL 60601-1:2003 for electrical and mechanical safety.
Technological Characteristics
Modular steel tubing, plastic, and vinyl construction. Actuation via manual hydraulic cylinder or 24V DC linear mechanical worm drive. Powered by two 12V batteries. Hand-held pendant control. Dartex (polyurethane transfer coating on weft knitted fabric) seat covering. Indoor use. Standards: BS EN 12182:1999, IEC/EN 60601-1, UL 60601-1.
Indications for Use
Indicated for individuals with physical disabilities or difficulty transitioning from seated to standing positions. Accommodates heights 5' to 6'2" (152-188 cm) and weights up to 280 lbs (127 kg).
Regulatory Classification
Identification
An electric positioning chair is a device with a motorized positioning control that is intended for medical purposes and that can be adjusted to various positions. The device is used to provide stability for patients with athetosis (involuntary spasms) and to alter postural positions.
Special Controls
(1) Appropriate analysis and non-clinical testing must demonstrate that the safety controls are adequate to ensure safe use of the device and prevent user falls from the device in the event of a device failure;
(2) Appropriate analysis and non-clinical testing must demonstrate the ability of the device to withstand the rated user weight load with an appropriate factor of safety;
(3) Appropriate analysis and non-clinical testing must demonstrate the longevity of the device to withstand external forces applied to the device and provide the user with an expected service life of the device;
(4) Appropriate analysis and non-clinical testing must demonstrate proper environments of use and storage of the device to maximize the longevity of the device;
(5) Appropriate analysis and non-clinical testing (such as that outlined in the currently FDA-recognized editions of ANSI/AAMI/ES60601-1, “Medical Electrical Equipment—Part 1: General Requirements for Basic Safety and Essential Performance,” and ANSI/AAMI/IEC 60601-1-2, “Medical Electrical Equipment—Part 1-2: General Requirements for Basic Safety and Essential Performance—Collateral Standard: Electromagnetic Disturbances—Requirements and Tests”) must validate electromagnetic compatibility and electrical safety;
(6) Appropriate analysis and non-clinical testing (such as that outlined in the currently FDA-recognized editions of ANSI/AAMI/ISO 10993-1, “Biological Evaluation of Medical Devices—Part 1: Evaluation and Testing Within a Risk Management Process,” ANSI/AAMI/ISO 10993-5, “Biological Evaluation of Medical Devices—Part 5: Tests for In Vitro Cytotoxicity,” and ANSI/AAMI/ISO 10993-10, “Biological Evaluation of Medical Devices—Part 10: Tests for Irritation and Skin Sensitization”) must validate that the skin-contacting components of the device are biocompatible;
(7) Appropriate analysis and non-clinical testing (such as that outlined in the currently FDA-recognized editions of IEC 62304, “Medical Device Software—Software Life Cycle Processes”) must validate the software life cycle and that all processes, activities, and tasks are implemented and documented;
(8) Appropriate analysis and non-clinical testing must validate that the device components are found to be non-flammable;
(9) Appropriate analysis and non-clinical testing must validate that the battery in the device (if applicable) performs as intended over the anticipated service life of the device; and
(10) Adequate patient labeling is provided to the user to document proper use and maintenance of the device to ensure safe use of the device by the patient in the intended use environment.
{0}------------------------------------------------
## 510(K) SUMMARY FOR ALTIMATE MEDICAL'S EASYSTAND EVOLV
This summary of 510(k) safety and effectiveness information is being supplied in accordance with the requirements of the SMDA of 1990 and 21 CFR 807.92
The assigned 510(k) number is
K062402
Date: August 15, 2006
Submitted by: Invacare Corporation Registration No. 1525712 One Invacare Way Elyria, Ohio 44035-4190
Manufacturer: Altimate Medical Registration No. 2183634 262 West First St. Morton, MN 56270
Telephone: 440-326-3115 Fax: 440-326-3607
Contact Person: Mr. Ronald Clines
Trade Name: EasyStand Evolv
Common Name: Electric lift chair
Classification Name: Chair, positioning, electric per 21 CFR 890.3110
Legally Marketed Predicate Device(s): Altimate Medical StandEX; K885343, May 1, 1989 Invacare Lift Chair; K002171, August 10, 2000
Device Description: The EasyStand Evolv is a modular standing frame for indoor use that allows users with various degrees of physical disability to be supported in a standing, weight-bearing position. The device is modular and is configured around a centralized seat. It incorporates pivot points that allow a user to rise from a fully seated to a fully standing position. Pivot action is such that shear the user, seat and seat back is minimized. Device is modular in that multiple uses of the device can be accommodated depending on the overall configuration of the device. Current modules allow for passive standing (Basic), workstation use at an elevated or seated position (Shadow Support Tray) and standing with leg exercise (Glider).
Elevation action is provided by one of two means. The standard device utilizes a user controlled hydraulic pump. The optional electrical version uses a low voltage motor-driven linear actuator that is powered by two 12-Volt batteries and actuated by a hand held control pendant. On the electrical
{1}------------------------------------------------
version, the patient or a care provider activates the hand held control pendant to change the position of the seat. The electric version also provides an emergency release. Recharge of batteries is accomplished by connection of an external battery charger to the control pendant.
The EasyStand Evolv is designed for indoor use and will accommodate individuals ranging in height from 5' to 6'2" (152cm - 188 cm) and up to 280lbs (127kg)
Models available EasyStand Evolv - Basic EasyStand Evolv - Glider EasyStand Evolv- Shadow Support Tray
Supporting these three basic models are additional seating, postural supports and positional attachments that allow proper fit of the device to accommodate a patients particular medical condition. The powered lift feature (Pow'r Up) can be added to any of the three basic models. Fit of the device is to be determined only by a qualified therapist or physician.
Intended Use: To assist persons who have difficulty rising from a seated position to a standing position.
Substantial Equivalence: Products that are substantially equivalent to the Altimate Medical EasyStand Evolv are the Altimate Medical StandEx (K885343, May 1, 1989) and the Invacare Lift Chair (K002171, August 10, 2000).
The EasyStand Evolv is comparable to both the Invacare Lift Chair and the StandEx. It is comparable to the StandEx device in its intended use, construction and functionality. The intended use of providing support for a person in a standing position, providing a means for a person to rise from a seated to a fully standing position and offering a method of exercising the body (the Glider model) remains the same between the two devices. The difference between the two devices is the addition of the power lift feature. This power lift feature is comparable to the lift feature in the Invacare Lift Chair in that they both use an electrically powered, low voltage DC, linear mechanical worm drive actuated via a hand held pendant to adjust the seat. Both devices also have the same intended use. The differences between the EasyStand Evolv and the Invacare Lift Chair are that the main electrical power input needed to operate the actuators comes from a battery for the EasyStand Evolv, but for the Invacare Lift Chair, it comes from AC power converted to low voltage DC power; the fact that the EasyStand Evoly seat is covered with Dartex (a polyurethane transfer coating on weft knitted fabric) and the Invacare Lift Chair has a fabric upholstered seat; that the Invacare Lift Chair offers no method of exercise; and that the EasyStand Evolv supports the user to allow for passive standing.
Performance Standards: Although no performance standards or special controls have been developed under Section 514 of the FDC Act for electric positioning chairs, Altimate Medical has chosen to test the EasyStand Evolv against the following standards:
BS EN 12182:1999 Part 1 Medical Electrical Equipment General Requirements for Safety IEC/EN 60601-1: 1998; A1:1991; A2:1995 Part 1 Medical Electrical Equipment General Requirements for Safety
{2}------------------------------------------------
UL 60601-1:2003 First Edition and CAN/CSA 22.2 No. 601.1-M90 Standard for Electrical Medical Equipment Part 1: General Requirements for Safety
| Features | EasyStand Evolv | Invacare Lift Chair | StandEx |
|-----------------------------|--------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Altimate Medical | Invacare Corporation | Altimate Medical |
| 510(k) Number | TBD | K002171 | K885343 |
| Classification/Product Code | 890.3110 / INO | 890.3110 / INO | 890.5370 / ION |
| Date Cleared | TBD | August 10, 2000 | May 1, 1989 |
| Class | Class II | Class II | Class I |
| Intended Use | To assist persons who have difficulty rising from a seated position to a standing position | To assist persons who have difficulty rising from a seated position to a standing position | To assist persons to redevelop muscles or restore motion to joints or for use as an adjunct treatment for obesity. |
| Basic Construction | Steel Tubing / Plastic / Vinyl | Steel Tubing/Wood/Fabric | Steel Tubing / Plastic / Vinyl |
| Energy Requirements | Standard: Manual pump action<br>Electrical option: 24 Volt Battery (2) | AC-Power converted to low voltage DC | Manual pump action |
| Weight Capacity | 280 lbs. | 350 lbs. | 300 lbs. |
| Mechanical Action | Standard: hydraulic oil cylinder<br>Electrical option: DC linear mechanical worm drive | AC linear mechanical worm drive | Hydraulic oil cylinder |
Substantial Equivalence Discussion
The EasyStand Evolv is comparable to the StandEx device in its intended use, construction and functionality. The intended use of providing support for a person in a standing position, providing a means for a person to rise from a seated to a fully standing position and offering a method of exercising the body (the Glider model) remains the same between the two devices. The difference between the two devices is the addition of the power lift feature. This power lift feature is comparable to the lift feature in the Invacare Lift Chair in that they both use an electrically powered, low voltage DC, linear mechanical worm drive actuated via a hand held pendant to adjust the seat. Both devices also have the same intended use. The differences between the EasyStand Evolv and the Invacare Lift Chair are that the main electrical power input needed to operate the actuators comes from a battery for the EasyStand Evolv, but for the Invacare Lift Chair, it comes from AC power converted to low voltage DC power; the fact that the EasyStand Evolv seat is covered with Dartex (a polyurethane transfer coating on weft knitted fabric) and the Invacare Lift Chair has a fabric upholstered seat; that the Invacare Lift Chair offers no method of exercise; and that the EasyStand Evolv supports the user to allow for passive standing.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo features a stylized eagle with three curved lines representing its body and wings. The eagle is positioned to the right of the text, which reads "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" in a circular arrangement around the logo.
## Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 2 1 2006
Ivacare Corporation % Mr. Ronald Clines One Invacare Way Elyria, Ohio 44035
Re: K062402
Trade/Device Name: EasyStand Evolv Regulation Number: 21 CFR 890.3110 Regulation Name: Electric positioning chair Regulatory Class: Class II Product Code: INO Dated: August 15, 2006 Received: August 17, 2006
Dear Mr. Clines:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Ronald Clines
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120 . Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers; International and Gonsumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Raibara Buemp
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: EasyStand Evolv
Indications for Use: To assist persons who have difficulty rising from a seated position to a standing position.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Barbara Pouend
(Division Si: Off) Division of Ceueral, Restorative, anu Neurological Devices
Page 1 of _
510(k) Number K062472
Altimate Easy Stand Evolv 510k Submission Page 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.