K062366 · Peters Surgical · GAM · Oct 2, 2006 · General, Plastic Surgery
Device Facts
Record ID
K062366
Device Name
OPTIME, SINUSORB PGA
Applicant
Peters Surgical
Product Code
GAM · General, Plastic Surgery
Decision Date
Oct 2, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4493
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The synthetic absorbable surgical sutures OPTIME® and SINUSORB® PGA are indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic surgery; but not for use in cardiovascular and neurological surgery.
Device Story
OPTIME® and SINUSORB® PGA are synthetic absorbable braided surgical sutures composed of polyglycolic acid homopolymer. Sutures are coated with epsilon-caprolactone and calcium stearate or a mixture of polycaprolactone, calcium stearate, and fatty acid ester. Available undyed or violet dyed (D&C violet nº2, CI 60725), with or without attached needles. Used by surgeons in clinical settings for soft tissue approximation and ligation. Device provides temporary wound support during healing before being absorbed by the body. Substantial equivalence established to Coated VICRYL™ (Polyglactin 910) Suture.
Clinical Evidence
Bench testing only. Device performance demonstrated through compliance with USP and EP monographs for absorbable surgical sutures. No clinical data provided.
Technological Characteristics
Braided synthetic absorbable suture; material: polyglycolic acid homopolymer; coating: epsilon-caprolactone/calcium stearate or polycaprolactone/calcium stearate/fatty acid ester; sterile; conforms to USP and EP standards for absorbable surgical sutures.
Indications for Use
Indicated for general soft tissue approximation and/or ligation, including ophthalmic surgery. Contraindicated for cardiovascular and neurological surgery.
Regulatory Classification
Identification
An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Peters Surgical. The logo features a stylized eye on the left, with a pattern of small circles forming the iris and pupil. To the right of the eye, the text "Péters" is written in a bold, slightly italicized font, with the accent mark over the "e". Below "Péters", the word "SURGICAL" is written in smaller, sans-serif capital letters.
K062366
Page 1 of 2
### SECTION 11
OCT = 2 2006
### 510(k) SUMMARY
# [As Required by 21 CFR 807.92(c)]
Information supporting claims of substantial equivalence, as defined under the Federal Food, drug and Cosmetic Act, respecting safety and effectiveness is summarized below.
510(k) Summary Date prepared August 8, 2006
510(k) Submitter
PETERS SURGICAL Z.I. Les vignes 42 rue Benoît Frachon Bobigny, FRANCE 93013
Registration Number: 3004060107 Owner/Operator Number: 9009304 Phone: 33-148-106262 Fax: 33-148-912299
Official Correspondent
Annie LASSERRE Research & Development Manager PETERS SURGICAL Bobigny, FRANCE, 93013 Phone: 33-148-106259
### New Device Name
| Trade name: | OPTIME® and SINUSORB® PGA |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual name: | Synthetic Absorbable Suture, PGA Suture |
| Classification name: | Suture, Absorbable, Synthetic, Polyglycolic Acid |
| New Device Classification | Class II in 21 CFR §878.4493 by the General and Plastic<br>Surgery Devices Panel, Absorbable poly(glycolide/l-<br>lactide) surgical suture (GAM). |
| Predicate Device Name | Coated VICRYL™ (POLYGLACTIN 910) Suture,<br>ETHICON INC., K022269 |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Peters Surgical. The logo consists of a stylized image of an eye on the left, and the text "Peters" above the word "Surgical" on the right. The eye is made up of many small dots, and the text is in a simple, sans-serif font.
K 062366
page 2 of 2
#### Statement of intended use
The synthetic absorbable surgical sutures OPTIME® and SINUSORB® PGA are indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic surgery; but not for use in cardiovascular and neurological surgery.
OPTIME® and SINUSORB® PGA have the same intended use as the predicate device Coated VICRYL 100 (Polyglactin 910) Suture.
#### New Device Description
OPTIME® and SINUSORB® PGA are synthetic absorbable braids composed of homopolymer of glycolic acid and coated with epsilon-caprolactone and calcium stearate or a mixture composed of polycaprolacton, calcium stearate and fatty acid ester. These surgical sutures are available undyed or violet dyed with an FDA approved color additive: D&C violet nº2, CI 60725. The suture may be provided with or without a standard needle attached.
## Summary of Technological Characteristics of New Device compared to Predicate Device
OPTIME® and SINUSORB® PGA sutures have similar technological characteristics as the predicate device Coated VICRYL™ (Polyglactin 910) Suture. Like the predicate device, OPTIME® and SINUSORB® PGA are sterile, braided synthetic absorbable surgical sutures that conform to the requirements of the United States Pharmacopoeia (USP) and the European Pharmacopoeia (EP) for Absorbable surgical sutures. The polyglycolic acid material used for both "new" and "predicate" medical devices is commonly used in surgical applications and has been proven to be biocompatible.
#### Performance data
Non-clinical laboratory testing was performed demonstrating that the device complied with the USP Monographs and with the EP Monographs for Absorbable surgical sutures.
#### Conclusions
Based on the 510(k) summary (21 CFR 807) and the information provided herein, we conclude that our New Medical Devices OPTIME® and SINUSORB® PGA are substantially equivalent to the Predicate device Coated VICRYL™ Suture under the Federal Food, Druing, and Cosmetic Act.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle with three lines forming its wings. The eagle is positioned above a wavy line, which may represent water. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular pattern around the central image.
OCT - 2 2006
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Peters Surgical % Ms. Annie Lasserre Research and Development Manager Z.I. Les vignes 42 Rue Benoît Frachon Bobigny, France 93013
Re: K062366
Trade/Device Name: OPTIME®, SINUSORB® PGA Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable poly (glycolide/L-lactide) surgical suture Regulatory Class: Class II Product Code: GAM Dated: August 8, 2006 Received: August 14, 2006
Dear Ms. Lasserre:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable. the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Annie Lasserre
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a logally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Melkerson
Dep. Director
Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Peters Surgical. The logo features a stylized image of an eye on the left, with the company name "Péters SURGICAL" on the right. The word "Péters" is in a larger, stylized font, while "SURGICAL" is in smaller, block letters below it.
## SECTION 10
# STATEMENT OF INDICATIONS FOR USE
510(k) Number K062366
Device Name OPTIME®, SINUSORB® PGA
### Indications for use
The synthetic absorbable surgical sutures OPTIME® and SINUSORB® PGA are indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmio surgery, but not for use in cardiovascular and neurological surgery.
Prescription Use _x__ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------------------------------------|
| (Division Sign-Off)<br>Division of General, Restorative,<br>and Neurological Devices |
| |
510(k
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.