K062336 · Optimedica Corporation · GEX · Oct 19, 2006 · General, Plastic Surgery
Device Facts
Record ID
K062336
Device Name
OPTIMEDICA LASER INDIRECT OPHTHALMOSCOPE
Applicant
Optimedica Corporation
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 19, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The OptiMedica LIO is indicated for the delivery of laser energy in eyes with retinal pathology. The OptiMedica LIO is indicated for use in the following ophthalmic treatments and conditions: - Diabetic retinopathy (panretinal photocoagulation) - Macular degeneration - Peripheral neovascularization - Retinal breaks - Retinal detachments - Retinal tears - Lattice degeneration - Pneumatic retinopexy reattachment procedures - Segmental peripheral photocoagulation - Segmental photocoagulation - Cloudy vitreous cavities - Pediatric retinal repairs (under general anesthesia) - Delivery of laser energy through small pupils or to eyes with focal lens opacities
Device Story
OptiMedica Laser Indirect Ophthalmoscope (LIO) is a binocular indirect ophthalmoscope used for ocular diagnosis and treatment. Device illuminates and magnifies fundus; when connected to Pascal Retinal Photocoagulator, it delivers laser energy to patient eye. Used by ophthalmologists in clinical settings; enables precise laser delivery for retinal conditions. Device utilizes Heine OMEGA 500 binocular indirect ophthalmoscope; transmits 532nm treatment and 635nm aiming beams via fiber optic cable. Includes photopically neutral eye safety filter (OD>5) for visualization and protection. Clinical benefit includes ability to perform retinal photocoagulation through small pupils or lens opacities. Physician views fundus through ophthalmoscope; laser delivery affects retinal tissue to treat pathology.
Clinical Evidence
Bench testing only. No clinical data presented. Performance testing characterized optical similarity to predicate (Zeiss LIO) by measuring optical transmissions, delivered spot size, and intensity profiles at treatment and aiming wavelengths.
Technological Characteristics
Binocular indirect ophthalmoscope (Heine OMEGA 500). Wavelengths: 532nm (treatment), 635nm (aiming). Optical safety filter: OD>5. Spot size: 960um. Working distance: 275mm. Fiber length: 5 meters. Connectivity: Fiber optic connection to external laser photocoagulator. Cooling/Electrical: Same as predicate indirect ophthalmoscopes.
Indications for Use
Indicated for delivery of laser energy to treat retinal pathology, including diabetic retinopathy, macular degeneration, peripheral neovascularization, retinal breaks/tears/detachments, lattice degeneration, and for pneumatic retinopexy, segmental photocoagulation, cloudy vitreous, and pediatric retinal repairs. Contraindicated when visualization of target tissue is inadequate due to corneal opacities, cataracts, or vitreous hemorrhage.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
### SECTION 5 510(K) SUMMARY OF SAFETY AND EFFECTIVENE
#### 5.0 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT | OptiMedica Corporation |
|----------------|--------------------------------------------------------|
| TRADE NAME: | OptiMedica Laser Indirect Ophthalmoscope (LIO) |
| COMMON NAME: | Laser Indirect Ophthalmoscope (LIO) |
| DEVICE | GEX |
| CLASSIFICATION | 21 CFR 878.4810<br>Laser instrument, surgical, powered |
#### 5.1 SUBSTANTIALLY EQUIVALENT TO:
The Laser Indirect Ophthalmoscope (LIO) is substantially equivalent in intended use and the same or similar technological characteristics (including treatment and aiming wavelengths, spot size, working distance and indirect ophthalmoscope used) as the Zeiss LIO (K924588). The OptiMedica LIO is also substantially equivalent to the delivery devices included in 510(k)s K022181 (Novus VARIA and Delivery Devices ) and K052526 (Novus 3000 and Delivery Devices ) by Lumenis and K991258 (Novus Verdi Delivery Systems) by Coherent.
### DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION 5.2
The OptiMedica Laser Indirect Ophthalmoscope (LIO) is intended for use in the diagnosis and treatment of ocular pathology. The LIO illuminates and magnifies the fundus image. When connected to the Pascal Retinal Photocoagulator, the laser aiming and treatment beams are precisely focused and delivered to the patient's eye through the LIO.
{1}------------------------------------------------
### 5.3 INDICATION FOR USE
The OptiMedica LIO is indicated for the delivery of laser energy in eyes with retinal pathology. The OptiMedica LIO is indicated for use in the following ophthalmic treatments and conditions:
- . Diabetic retinopathy (panretinal photocoagulation)
- . Macular degeneration
- Peripheral neovascularization .
- ◆ Retinal breaks
- Retinal detachments �
- . Retinal tears
- Lattice degeneration
- Pneumatic retinopexy reattachment procedures .
- . Segmental peripheral photocoagulation
- . Segmental photocoagulation
- . Cloudy vitreous cavities
- Pediatric retinal repairs (under general anesthesia) .
- . Delivery of laser energy through small pupils or to eyes with focal lens opacities
### 5.4 CONTRAINDICATIONS
Laser surgery with the OptiMedica LIO is contraindicated when an appropriate procedure cannot be performed safely. This occurs when tissue targets cannot be visualized properly. Under such circumstances, an important tissue structure adjacent to a target tissue might be photocoagulated inadvertently. Corneal opacities, cataract formation, and vitreous hemorrhage can all interfere with the laser surgeon's view of appropriate target structures. Treatment should be delayed until the visualization has adequately improved.
### TECHNICAL CHARACTERISTICS ર્સ્ડ
# Indirect Ophthalmoscope
The OptiMedica LIO uses a Heine OMEGA 500 binocular indirect ophthalmoscope. Each of the predicate devices, including the Zeiss Laser Indirect Ophthalmoscope (K924588), the Lumenis Novus 3000 Delivery Devices (K052526), Lumenis Novus VARIA Delivery Devices (K022181), and Coherent Novus Verdi Delivery Systems (K991258) use a Heine binocular indirect ophthalmoscope as well.
# Usable Wavelengths
The OptiMedica LIO is designed to be used with the Pascal Photocoagulator. This system produces 532nm light for treatment, and 635 nm light for aiming, identical to the Lumenis Novus 3000 (K052526), Lumenis Novus VARIA (K022181), Coherent Novus Verdi (K991258), and the Zeiss Visulas system used with the Zeiss Laser Indirect Ophthalmoscope (K924588).
{2}------------------------------------------------
# Eye Filter Optical Density
The OptiMedica LIO eye safety filter optical density at the treatment wavelength is the same as all of the predicate devices, OD>5. The OptiMedica LIO also uses a photopically neutral filter to allow the high transmission of the aiming beam and white balance the image, making visualization easier and more realistic.
# Aerial Spot Size
The OptiMedica LIO spot size for delivery through indirect surgical lens is 960um. This is almost identical to that of the Lumenis Novus 3000 (K052526), Lumenis Novus VARIA (K022181), Coherent Novus Verdi (K991258), which are 973um. The Zeiss Laser Indirect Ophthalmoscope (K924588) uses a 1060 um spot.
# Working Distance
The OptiMedica LIO working distance of 275mm lies right in the middle of the range specified by the predicate devices. The Lumenis Novus 3000 (K052526), Lumenis Novus VARIA (K022181), Coherent Novus Verdi (K991258) systems specify a 270mm working distance. The Zeiss Laser Indirect Ophthalmoscope (K924588) uses a 280mm working distance.
# Fiber Length
The OptiMedica LIO fiber length of 5 meters is equal to that of the Zeiss Laser Indirect Ophthalmoscope (K924588), and slightly longer than that of the Lumenis Novus 3000 (K052526), Lumenis Novus VARIA (K022181), Coherent Novus Verdi (K991258) systems which use 4.6 meter long fibers.
# Electrical Requirements
The OptiMedica LIO and all of the predicate devices use the same sort of illumination source, and therefore have the same electrical requirements.
# Cooling Methods
The OptiMedica LIO and all of the predicate devices use the same sort of illumination source, and therefore have the same cooling requirements and methods.
# Weight
The OptiMedica LIO and all of the predicate devices use the same sort of binocular indirect ophthalmoscope, and therefore have roughly the same weight.
# Dimensions
The OptiMedica LIO and all of the predicate devices use the same sort of binocular indirect ophthalmoscope, and therefore have roughly the same dimensions.
{3}------------------------------------------------
### 5.6 PERFORMANCE DATA
Performance testing has been conducted with the OptiMedica LIO to establish the optical similarity of the OptiMedica LIO with respect to a predicate device, the Zeiss LIO, as a means to assess the relative clinical performance of the devices. This was accomplished by characterizing the optical performance of both laser delivery systems at their treatment and aiming wavelengths. Optical transmissions, as well as the delivered spot size and intensity profiles were measured.
#### 5.7 BASIS FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE
The indications for use for the OptiMedica LIO are similar to the predicate LIOs cited in this application. The safety of the materials used for the manufacture of LIO has previously been demonstrated. Testing demonstrates that the OptiMedica LIO is functionally equivalent to the predicate devices.
{4}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 9 2006
OptiMedica Corporation % ClinReg Consulting Services, Inc. Judy F. Gordon, D.V.M. 733 Bolsana Drive Laguna Beach. California 92651
Re: K062336
Trade/Device Name: OptiMedica Laser Indirect Ophthalmoscope Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: August 7, 2006 Received: August 10, 2006
Dear Dr. Gordon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Judy F. Gordon, D.V.M.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
fih xh
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (if known): 1 06233 p
Device Name: OptiMedica Laser Indirect Ophthalmoscope
Indications for Use:
The OptiMedica LIO is indicated for the delivery of laser energy in eyes with retinal pathology. The OptiMedica LIO is indicated for use in the following ophthalmic treatments and conditions:
- Diabetic retinopathy (panretinal photocoagulation) .
- Macular degeneration
- � Peripheral neovascularization
- Retinal breaks ●
- Retinal detachments .
- � Retinal tears
- Lattice degeneration
- Pneumatic retinopexy reattachment procedures .
- Segmental peripheral photocoagulation
- Segmental photocoagulation
- Cloudy vitreous cavities .
- Pediatric retinal repairs (under general anesthesia) ●
- Delivery of laser energy through small pupils or to eyes with focal lens opacities
The CDRH Indications for Use Statement for the OptiMedica LIO can be found on the following page.
Prescription Use
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
in
Division Sig -Off Division of General, Restorative, Jarralogical Devices
Page of
**510(K)** L061-32
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.