The Deltec Cozmo® Insulin Pump is a syringe infusion pump designed for Continuous Subcutaneous Insulin Infusion (CSII) for the control of diabetes. The Deltec CoZmonitor™ Blood Glucose Monitor is available as an accessory to be used solely with the Deltec Cozmo® Insulin Pump. Because the monitor is powered by FreeStyle™, blood sampling can be performed using the finger, forearm, upper arm, thigh, calf or hand. All testing is performed outside of the body (in other words, in-vitro diagnostic use only with FreeStyle™ Test Strip and FreeStyle™ Control Solution, as other products can give inaccurate results. The Deltec CoZmonitor™ Blood Glucose Monitoring System is designed to be used only with the Deltec Cozmo® Insulin Pump to help you and your healthcare provider monitor your diabetes management program. When used properly, the monitor provides a quantitative measurement of glucose (or sugar) in whole blood. Because the monitor is powered by FreeStyle™, blood sampling can be performed using the finger, forearm, upper arm, thigh, calf or hand. All testing is performed outside of the body (in other words, in-vitro diagnostic use only). Use only with FreeStyle™ Test Strip and FreeStyle™ Control Solution, as other products can give inaccurate results.
Device Story
Deltec Cozmo® Insulin Pump is a syringe-based infusion pump for CSII; used for diabetes management. Features include quick-start programming, food database, user-defined BG targets/alerts, insulin-on-board tracking, pump disconnect tracking, and 'Hypo Manager' (recommends carbohydrate intake for low BG). CoZmonitor® module attaches to pump for in-vitro blood glucose measurement using FreeStyle™ test strips. CoZmanager® 2.0 PC software allows data download and pump programming via IR connection. Device assists patients in basal pattern scheduling and therapy reporting. Healthcare providers use generated reports to adjust management programs. Benefits include improved glycemic control and simplified insulin delivery/monitoring.
Clinical Evidence
No clinical data. Effectiveness established via functional verification and validation of software-controlled programming functions, demonstrating performance to specifications, and by comparison to legally marketed predicate devices.
Technological Characteristics
Syringe infusion pump; IR connectivity for PC communication; software-controlled programming; includes blood glucose monitoring module. Materials and specific standards not detailed. Software-based feature set includes food database, insulin-on-board calculation, and hypo-management logic.
Indications for Use
Indicated for patients with diabetes requiring Continuous Subcutaneous Insulin Infusion (CSII) and blood glucose monitoring. Used for diabetes management and monitoring by patients and healthcare providers.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
Deltec Cozmo® Insulin Pump Model 1700
Animas IR 1250
Submission Summary (Full Text)
{0}------------------------------------------------
Ka323
NOV - 1 2006
# 510(k) SUMMARY
# Deltec Cozmo™ Insulin Infusion Pump
#### I. GENERAL INFORMATION
| Applicant's Name and Address: | Smiths Medical MD, Inc.<br>1265 Grey Fox Road<br>St. Paul, MN 55112 |
|--------------------------------|-----------------------------------------------------------------------------------------|
| Contact Person: | David H. Short<br>Director Regulatory and Clinical Affairs |
| Common/Usual Name: | Insulin Infusion Pump and Glucose Monitor |
| Proprietary Name: | Deltec Cozmo® Insulin Infusion Pump Model 1800<br>with CoZmonitor® Blood Glucose Module |
| Equivalence Device Comparison: | Deltec Cozmo® Insulin Infusion Pump and Animas<br>IR 1250 |
### II. DEVICE DESCRIPTION
The Deltec Cozmo® Insulin Infusion Pump with CoZmonitor® Blood Glucose Module is lightweight and easy to use. The Deltec Cozmo® Insulin Pump incorporates a number of features to assist users in the management of their diabetes. Some of these features include a quick start program to allow users easy pump set-up and programming, enhancements to existing features, the addition of a food data base, user established blood glucose targets and alerts, insulin-on-board to alert the user of remaining active insulin following bolus administration, pump disconnect feature that allows the user to disconnect from the pump and allowing the pump to then keep track of insulin that would have been delivered during the disconnect period, Hypo Manager that alerts the patient of low BG levels and recommends carbohydrate units to ingest to elevate BG to patient specific specified levels, Therapy Summary Report to assist with diabetes management, weekly schedule to allow the user to set basal patterns to match activities of daily living. The Deltec Cozmo® Insulin Pump is used with the CoZmonitor® Blood Glucose Module
The CoZmanager® 2.0 is a PC based program that allows the user to download information from the pump and to program the pump via an IR connection between the pump and the PC. CoZmanager® 2.0 allows the user to customize names of features as well as establish meal selections based on individual needs. A number of reports are available from the program.
{1}------------------------------------------------
### III. INTENDED USE OF THE DEVICE
The Deltec Cozmo® Insulin Pump is a syringe infusion pump designed for Continuous Subcutaneous Insulin Infusion (CSII) for the control of diabetes.
### IV. DEVICE COMPARISON
The Deltec Cozmo® Insulin Pump Model 1800 is similar to the currently distributed Deltec Cozmo® Insulin Pump Model 1700 pump, with the exception of updated software that enhanced certain features and added a food data base to allow users to better manage dietary requirements in conjunction with the use of their insulin pump.
The Deltec Cozmo® Insulin Pump Model 1800 is substantially equivalent to other commercially available insulin pumps such as the Animas IR 1250.
### V. SUMMARY OF STUDIES
#### A. Functional Testing
Verification and validation of software controlled programming functions related to proper pump operation were conducted and results of this testing showed the Deltec Cozmo® Insulin Pump Model 1800 met and performed to established specifications.
#### B. Clinical Studies
Clinical studies were not deemed necessary to establish effectiveness of the Deltec Cozmo® Insulin Pump Model 1800 as the pump is similar to other currently commercially available insulin pumps on the market today.
### C. Conclusion Drawn from the Studies
Based on the information provided above, the Deltec Cozmo® Insulin Pump Model 1800 is substantially equivalent to other currently commercially available insulin infusion pumps on the market today.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV - 1 2006
Mr. David H. Short Director Regulatory Affairs and Design Assurance Smiths Medical MD, Incorporated 1265 Grey Fox Road St. Paul, Minnesota 55112
Re: K062323
Trade/Device Name: Deltec Cozmo® Insulin Pump with CoZmonitor™ Glucose Monitor Regulation Number: 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: LZG Dated: October 9, 2006 Received: October 10, 2006
Dear Mr. Short:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
# Page 2 - Mr. Short
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Suitti Y. Michin Omd
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
# 510(k) Number (if known): K062323
Deltec Cozmo® Insulin Pump with CoZmonitor™ Glucose Monitor Device Name:
### Indications for Use:
# Deltec Cozmo® Insulin Pump
The Deltec Cozmo® Insulin Pump is a syringe infusion pump designed for Continuous Subcutaneous Insulin Infusion (CSII) for the control of diabetes.
Over-The Counter Use Prescription Use OR X (Per 21 CFR 801.109)
### Indications for Use:
# CoZmonitor™ Glucose Monitor
The Deltec CoZmonitor™ Blood Glucose Monitor is available as an accessory to be used solely with the Deltec Cozmo® Insulin Pump. Because the monitor is powered by FreeStyle™, blood sampling can be performed using the finger, forearm, upper arm, thigh, calf or hand. All testing is performed outside of the body (in other words, in-vitro diagnostic use only with FreeStyle™ Test Strip and FreeStyle™ Control Solution, as other products can give inaccurate results.
Prescription Use (Per 21 CFR 801.109) Over-The Counter Use X
# Indications for Use:
# Deltec Cozmo® Insulin Pump with CoZmonitor™ Glucose Monitor
The Deltec CoZmonitor™ Blood Glucose Monitoring System is designed to be used only with the Deltec Cozmo® Insulin Pump to help you and your healthcare provider monitor your diabetes management program. When used properly, the monitor provides a quantitative measurement of glucose (or sugar) in whole blood. Because the monitor is powered by FreeStyle™, blood sampling can be performed using the finger, forearm, upper arm, thigh, calf or hand. All testing is performed outside of the body (in other words, in-vitro diagnostic use only). Use only with FreeStyle™ Test Strip and FreeStyle™ Control Solution, as other products can give inaccurate results.
Prescription Use X (Per 21 CFR 801.109)
OR
OR
Over-The Counter Use
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Cim
Anesthesiology, General Hospital
Lion Center, Dental Devices
上462323
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.