VITROS CHEMISTRY PRODUCTS BENZ REAGENT, CALIBRATOR KIT 26, FS CALIBRATOR 1 AND DAT PERFORMANCE VERIFIERS I, II, III, IV
K062285 · Ortho-Clinical Diagnostics, Inc. · JXM · Dec 22, 2006 · Clinical Toxicology
Device Facts
Record ID
K062285
Device Name
VITROS CHEMISTRY PRODUCTS BENZ REAGENT, CALIBRATOR KIT 26, FS CALIBRATOR 1 AND DAT PERFORMANCE VERIFIERS I, II, III, IV
Applicant
Ortho-Clinical Diagnostics, Inc.
Product Code
JXM · Clinical Toxicology
Decision Date
Dec 22, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3170
Device Class
Class 2
Indications for Use
VITROS Chemistry Products BENZ Reagent is used on VITROS 5.1 FS Chemistry Systems for the semi-quantitative or qualitative determination of benzodiazepines (BENZ) in human urine using a cutoff of either 200 ng/mL or 300 ng/mL. Measurements obtained with the VITROS BENZ method are used in the diagnosis and treatment of benzodiazepines use or overdose. The VITROS Chemistry Products BENZ assay is intended for use by professional laboratory personnel. It provides only a preliminary test result. A more specific alternative chemical method must be used to confirm a result obtained with this assay. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drugof-abuse test result, particularly when evaluating a preliminary positive result. VITROS Chemistry Products Calibrator Kit 26 is used to calibrate VITROS 5,1 FS Chemistry Systems for the qualitative or semi-quantitative measurement of drugs of abuse. VITROS Chemistry Products FS Calibrator 1 is used in conjunction with VITROS Chemistry Products Calibrator Kits to calibrate VITROS 5,1 FS Chemistry Systems. VITROS Chemistry Products DAT Performance Verifiers are assayed controls used to monitor performance of urine drugs of abuse screening assays on VITROS 5,1 FS Chemistry Systems.
Device Story
VITROS BENZ assay is a homogeneous enzyme immunoassay for detecting benzodiazepines in human urine. Performed on VITROS 5,1 FS Chemistry Systems by professional laboratory personnel. Input: urine sample; Reagent 1 (antibody to diazepam, G6P, NAD); Reagent 2 (diazepam labeled with G6P-DH). Principle: competition between benzodiazepines in sample and enzyme-labeled diazepam for antibody binding sites. Enzyme activity decreases upon binding; concentration of benzodiazepines is directly proportional to measured enzyme activity. Active enzyme converts NAD to NADH, causing absorbance change measured spectrophotometrically at 340 nm. Output: semi-quantitative or qualitative benzodiazepine concentration. Used for preliminary screening; results require GC/MS confirmation. Clinical decision-making relies on professional judgment and confirmatory testing. Benefits include rapid, automated screening for drug use or overdose.
Clinical Evidence
No clinical trials; performance validated via bench testing comparing VITROS BENZ assay on VITROS 5,1 FS Chemistry System against Syva® EMIT® II Plus Benzodiazepine assay on Syva® 30R Biochemical System. Results demonstrated good agreement between the two immunoassay methods.
Technological Characteristics
Homogeneous enzyme immunoassay; liquid ready-to-use reagents; sheep polyclonal antibodies reactive to diazepam; spectrophotometric detection at 340 nm. Calibrated against lormetazepam. Used on VITROS 5,1 FS Chemistry Systems (automated clinical chemistry analyzer).
Indications for Use
Indicated for professional laboratory use for the qualitative or semi-quantitative determination of benzodiazepines in human urine to aid in the diagnosis and treatment of benzodiazepine use or overdose. Provides preliminary results requiring confirmation by GC/MS.
Regulatory Classification
Identification
A benzodiazepine test system is a device intended to measure any of the benzodiazepine compounds, sedative and hypnotic drugs, in blood, plasma, and urine. The benzodiazepine compounds include chlordiazepoxide, diazepam, oxazepam, chlorzepate, flurazepam, and nitrazepam. Measurements obtained by this device are used in the diagnosis and treatment of benzodiazepine use or overdose and in monitoring levels of benzodiazepines to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). A benzodiazepine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
EMIT II Plus Benzodiazepine Assay (k993985)
Liquicheck Urine Toxicology Controls (k022707)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k062285
B. Purpose for Submission:
New product
C. Measurand:
Benzodiazepines
D. Type of Test:
Semi-quantitative and qualitative homogeneous enzyme immunoassay
E. Applicant:
Ortho-Clinical Diagnostic
F. Proprietary and Established Names:
VITROS Chemistry Products BENZ Reagent
VITROS Chemistry Products Calibrator Kit 26
VITROS Chemistry Products FS Calibrator 1
VITROS Chemistry Products DAT Performance Verifiers I. II, III, IV, and V
G. Regulatory Information:
1. Regulation section:
21 CFR §862.3170, Benzodiazepine Test System
21 CFR §862.3200, Clinical Toxicology Calibrator
21 CFR §862.3280, Clinical Toxicology Control Material
2. Classification:
Class II (Reagent, Calibrator)
Class I Reserved (Control)
3. Product code:
JXM; DKB; DIF
4. Panel:
Toxicology (91)
H. Intended Use:
1. Intended use(s):
See Indications for Use below.
2. Indication(s) for use:
VITROS Chemistry Products BENZ Reagent is used on VITROS 5,1 FS
Chemistry Systems for the semi-quantitative or qualitative determination of
benzodiazepines (BENZ) in human urine using a cutoff of either 200 ng/mL or
300 ng/mL. Measurements obtained with the VITROS BENZ method are used in
the diagnosis and treatment of benzodiazepines use or overdose.
{1}
The VITROS Chemistry Products BENZ assay is intended for use by professional laboratory personnel. It provides only a preliminary test result. A more specific alternative chemical method must be used to confirm a result obtained with this assay. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when evaluating a preliminary positive result.
VITROS Chemistry Products Calibrator Kit 26 is used to calibrate VITROS 5.1 FS Chemistry Systems for the qualitative or semi-quantitative measurement of the drugs of abuse.
VITROS Chemistry Products FS Calibrator 1: For in vitro diagnostic use only. VITROS Chemistry Products FS Calibrator 1 is used in conjunction with VITROS Chemistry Products Calibrator Kits to calibrate VITROS 5,1 FS Chemistry Systems
VITROS Chemistry Products DAT Performance Verifiers are assayed control used to monitor performance of urine drugs of abuse screening assays on VITROS 5,1 FS Chemistry Systems.
3. Special conditions for use statement(s):
This device is for use by professional laboratory personnel. For in vitro diagnostic use only.
4. Special instrument requirements:
Ortho-Clinical Diagnostics VITROS 5,1 FS Chemistry System
I. Device Description:
The VITROS BENZ Reagent is a dual-chambered package containing ready-to-use liquid reagents that are used to detect benzodiazepines in urine. Sample, calibrators, and controls are automatically treated with surfactant (DAT Diluent 2) prior to addition of reagents. Treated sample is added to Reagent 1 containing antibody reactive to diazepam, glucose-6-phosphate and nicotinamide adenine dinucleotide $(\mathrm{NAD}^{+})$, followed by Reagent 2 containing diazepam labeled with the enzyme glucose-6-phosphate dehydrogenase (G6P-DH).
VITROS Chemistry Products Calibrator Kit 26 is prepared from human urine to which drugs of abuse, metabolites of drugs of abuse, organic salts, surfactants and preservative have been added.
VITROS Chemistry Products FS Calibrator 1 is prepared from sodium chloride and processed water. These products are used to calibrate VITROS 5,1 FS Chemistry Systems for the qualitative and semi-quantitative measurement of cocaine.
VITROS DAT Performance Verifiers I, II, III, IV & V are prepared from a human urine pool to which analytes, surfactant and preservative have been added. These are assayed controls used to monitor performance of the VITROS COCM Reagent on VITROS 5,1 FS Chemistry Systems.
The product labeling for the Calibrator Kit 26 and Performance Verifiers contain warnings regarding the presence of human sourced materials and recommend the use
{2}
of Universal Precautions when handling these products.
## J. Substantial Equivalence Information:
1. Predicate device name(s):
EMIT II Plus Benzodiazepine Assay
Liquicheck Urine Toxicology Controls
2. Predicate 510(k) number(s):
k993985; k022707
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | For use in the qualitative and semi-quantitative analysis of benzodiazepines in human urine. | Same |
| Reagent | Liquid, ready to use | Same |
| Principle | Homogeneous enzyme immunoassay | Same |
| Matrix | Urine | Same |
| Antibody | Sheep polyclonal | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Instrumentation | VITROS 5,1 FS Chemistry Systems | Multiple automated clinical chemistry analyzers |
| Calibrators: Number of level | Six | Qualitative: Three at each cutoff value
Semi-quantitative: Five |
| Controls: Number of levels | Five | Two |
## K. Standard/Guidance Document Referenced (if applicable):
CSLI EP9-A2: Method Comparison and Bias Estimation Using Patient Samples
CLSI EP5-A: Evaluation of Precision Performance of Clinical Chemistry Devices
CLSI EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures, A Statistical Approach
CLSI EP7-P: Interference Testing in Clinical Chemistry
CLSI EP17-A: Protocols for Demonstration, Verification and Evaluation of Limits of Detection and Quantitation
CLSI EP12-A: User Protocols for Evaluation of Qualitative Test Performance
## L. Test Principle:
The VITROS BENZ assay is a homogeneous enzyme immunoassay that is performed using the VITROS Chemistry Products BENZ Reagent with the VITROS Chemistry
{3}
Products Calibrator Kit 26, VITROS Chemistry Products FS Calibrator 1, and VITROS Chemistry Products FS Diluent Pack 4 (DAT Diluent/DAT Diluent 2) on VITROS 5,1 FS Chemistry Systems.
The assay is based on competition between benzodiazepines in the treated urine sample and diazepam labeled with the enzyme glucose-6-phosphate dehydrogenase (G6P-DH) for antibody binding sites. Enzyme activity decreases upon binding to the antibody, therefore the concentration of benzodiazepines in the urine sample is directly proportional to measured enzyme activity. Active enzyme converts oxidized nicotinamide adenine dinucleotide $(\mathrm{NAD}^{+})$ to NADH, resulting in an absorbance change that is measured spectrophotometrically at $340~\mathrm{nm}$ .
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Imprecision was evaluated with human urine based quality control materials on the VITROS 5,1 FS Chemistry System following CLSI Protocol EP5 and CLSI protocol EP12.
Imprecision for BENZ: Semi-Quantitative
| System | Conventional Units (ng/mL) and SI Units (μg/L) | | | Within Lab CV%** | No. Observ. | No. Days |
| --- | --- | --- | --- | --- | --- | --- |
| | Mean Conc. | Within Day SD* | Within Lab SD** | | | |
| VITROS 5,1 FS | 157 | 8.3 | 15.1 | 9.6 | 84 | 22 |
| | 233 | 10.1 | 20.5 | 8.8 | 84 | 22 |
| | 264 | 12.8 | 19.1 | 7.2 | 82 | 22 |
| | 397 | 17.2 | 31.3 | 7.9 | 82 | 22 |
| | 626 | 37.2 | 58.0 | 9.3 | 84 | 22 |
* Within Day imprecision was determined using one or two runs per day with two replicates per run.
** Within Lab imprecision was determined using a single lot of reagents with one analyzer and four calibrations.
Qualitative imprecision was assessed using test fluids targeted at $\pm 25\%$ of each cutoff.
| System | Cutoff Level (ng/mL & g/L) | Test Fluid at ± 25% Cutoff | Number of Observations | Number of Correct Results |
| --- | --- | --- | --- | --- |
| VITROS 5,1 FS | 200 | -25% | 84 | 84 |
| | 200 | +25% | 82 | 82 |
| | 300 | -25% | 84 | 84 |
| | 300 | +25% | 82 | 82 |
#### b. Linearity/assay reportable range:
The sponsor followed CLSI EP6-A in determining the linear range of their device. The low and high concentration pools were mixed to give 13 admixtures of intermediate BENZ concentrations. Three determinations of all pools were made together with three determinations each of VITROS Chemistry Products
{4}
DAT Performance Verifiers I, II, III, IV, and V. This experiment was performed three times, once with each of three VITROS BENZ Reagent lots on the VITROS 5,1 FS Chemistry System. A linear regression was performed and the results indicated acceptable linearity across the range of 33 to 869 ng/mL. This linearity determination in conjunction with determination of the limit of quantitation was used to establish the reportable range of the VITROS BENZ assay to be 85 to 800 ng/mL.
## Analytical Recovery of Semi-Quantitative Results
Eight admixtures were prepared from two human urine pools. The BENZ values for the admixtures were verified by GC/MS. Percent recovery was calculated using the concentration obtained by the VITROS Chemistry Products BENZ Assay versus the GC/MS value.
Recovery of Lormetazepam
| GC/MS (ng/mL) | VITROS BENZ Assay (ng/mL) | % Recovery |
| --- | --- | --- |
| 661 | 642 | 97.1 |
| 557 | 539 | 96.7 |
| 465 | 454 | 97.6 |
| 364 | 378 | 103.9 |
| 281 | 288 | 102.4 |
| 190 | 199 | 104.9 |
| 142 | 158 | 111.0 |
| 101 | 99 | 97.8 |
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The assigned values for the calibrators and controls are traceable to the Cerilliant lormetazepam standard catalogue L-910 and are verified by GC/MS. Real time and accelerated stability studies were conducted; protocols and acceptance criteria were described and found to be acceptable. These studies support the manufacturer's stability claims. Real time studies are ongoing.
d. Detection limit:
The detection limit was determined according to protocol recommendations in CLSI EP-17 on three different lots of reagent and one instrument platform. The claimed lower limit for VITROS BENZ is 85 ng/mL.
e. Analytical specificity:
The specificity of the VITROS BENZ assay for various benzodiazepines and structurally similar compounds was estimated by generating a dose response curve for each of the compounds listed below. The quantity (ng/mL) of compound that produces a value equivalent to the lormetazepam quantity (ng/mL) at each cutoff value is listed below. The combined effects of more than one compound detected in a sample may cause levels lower than those listed below to produce a value approximately equivalent to or greater than the cutoff value.
5
{5}
Substances that Cross-react with BENZ
| Compound | Quantity (ng/mL) equivalent to 200 ng/mL of lormetazepam | % cross-reactivity * | Quantity (ng/mL) equivalent to 300 ng/mL of lormetazepam | % cross-reactivity * |
| --- | --- | --- | --- | --- |
| nordiazepam | 70 | 285.7 | 100 | 300.0 |
| prazepam | 90 | 222.2 | 120 | 250.0 |
| tetrazepam | 105 | 190.5 | 138 | 217.4 |
| ketazolam | 108 | 185.2 | 142 | 211.3 |
| diazepam | 118 | 169.5 | 160 | 187.5 |
| N-desalkyflurazepam | 120 | 166.7 | 182 | 164.8 |
| medazepam | 120 | 166.7 | 182 | 164.8 |
| a-hydroxyalprazolam | 120 | 166.7 | 155 | 193.5 |
| flurazepam | 130 | 153.8 | 171 | 175.4 |
| a-hydroxytriazolam | 140 | 142.9 | 185 | 162.2 |
| nitrazepam | 145 | 137.9 | 210 | 142.9 |
| flunitrazepam | 152 | 131.6 | 215 | 139.5 |
| alprazolam | 155 | 129.0 | 200 | 150.0 |
| temazepam | 165 | 121.2 | 230 | 130.4 |
| norfludiazepam | 195 | 102.6 | 260 | 115.4 |
| halazepam | 209 | 95.7 | 274 | 109.5 |
| clonazepam | 224 | 89.3 | 408 | 73.5 |
| clobazam | 240 | 83.3 | 440 | 68.2 |
| oxazepam | 270 | 74.1 | 380 | 78.9 |
| 7-aminoflunitrazepam | 300 | 66.7 | 650 | 46.2 |
| N-desmethyldiazepam | 300 | 66.7 | 490 | 61.2 |
| demoxepam | 400 | 50.0 | 550 | 54.5 |
| bromazepam | 422 | 47.4 | 749 | 40.1 |
| chlordiazepoxide | 1230 | 16.3 | 2900 | 10.3 |
| 7-aminoclonazepam | 1800 | 11.1 | 4000 | 7.5 |
| norchlordiazepoxide | 1900 | 10.5 | 3750 | 8.0 |
| oxazepam glucuronide | >10,000 | <2 | >10,000 | <3 |
| lorazepam glucuronide | >10,000 | <2 | >10,000 | <3 |
* The VITROS BENZ Assay cutoff value (ng/mL) divided by the amount of cross-reactant (ng/mL) that produces a value equivalent to the cutoff value, multiplied by 100.
Interfering Substances:
Known Interfering Substances for BENZ
| Cutoff Value (ng/mL) | Interferent* | Interferent Concentration | | Bias (ng/mL)** |
| --- | --- | --- | --- | --- |
| 200 | desipramine | 10 mg/dL | 330 μmol/L | +57 |
| | dextromethorphan | 10 mg/dL | 368 μmol/L | +47 |
| | dicyclomine | 10 mg/dL | 289 μmol/L | +47 |
| | diethylpropion | 10 mg/dL | 487 μmol/L | +61 |
| | ethacrynic acid | 10 mg/dL | 330 μmol/L | +73 |
| | imipramine | 10 mg/dL | 357 μmol/L | +88 |
| | indomethacin | 2.5 mg/dL | 70 μmol/L | +65 |
{6}
Known Interfering Substances for BENZ
| Cutoff Value (ng/mL) | Interferent* | Interferent Concentration | | Bias (ng/mL)** |
| --- | --- | --- | --- | --- |
| | phenyltoloxamine | 2.5 mg/dL | 98 μmol/L | +76 |
| | phenylbutazone | 0.5 mg/dL | 324 μmol/L | +56 |
| | promethazine | 2.5 mg/dL | 20 μmol/L | +49 |
| | sertraline | 0.5 mg/dL | 16 μmol/L | +51 |
| | tripelannamine | 1.0 mg/dL | 39 μmol/L | +68 |
| | tripolidine | 10 mg/dL | 359 μmol/L | +41 |
| | NaCl | 1500 mg/dL | 207 mmol/L | +46 |
| 300 | desipramine | 10 mg/dL | 330 μmol/L | +67 |
| | dextromethorphan | 10 mg/dL | 368 μmol/L | +61 |
| | dicyclomine | 10 mg/dL | 379 μmol/L | +75 |
| | diethylpropion | 10 mg/dL | 487 μmol/L | +61 |
| | doxylamine | 10 mg/dL | 370 μmol/L | +80 |
| | ethacrynic acid | 10 mg/dL | 330 μmol/L | +80 |
| | imipramine | 10 mg/dL | 357 μmol/L | +107 |
| | indomethacin | 5.0 mg/dL | 140 μmol/L | +72 |
| | phenyltoloxamine | 2.5 mg/dL | 98 μmol/L | +63 |
| | phenylbutazone | 10 mg/dL | 324 μmol/L | +70 |
| | sertraline | 10 mg/dL | 327 μmol/L | +152 |
| | tripelannamine | 1.0 mg/dL | 39 μmol/L | +64 |
* The degree of interference at concentrations other than those listed might not be predictable from these results. Other interfering substances may be encountered in the patient population.
** The bias is an estimate of the maximum difference observed.
f. Assay cut-off:
The stated cutoffs of this assay are either 200 ng/mL or 300 ng/mL.
2. Comparison studies:
a. Method comparison with predicate device:
A total of 115 human urine samples were assayed using the VITROS Chemistry Products BENZ Reagent and GC/MS or LC/MS reference method for lormetazepam. Percent agreement was evaluated at assay cutoff values of 200 ng/mL and 300 ng/mL.
GC/MS Reference Method Comparison for BENZ
| | | GC/MS or LC/MS Reference | | | | %Agreement | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Cutoff Value (ng/mL) | | Low Negative | Near Cutoff Negative | Near Cutoff Positive | High Positive | %Agreement Negative | %Agreement Positive | %Agreement Overall |
| 200 | | (<-50%) <100 ng/mL | (-50% to cutoff) 100-200 ng/mL | (cutoff to +50%) 200-300 ng/mL | (>+50%) >300 ng/mL | 76.8 | 93.7 | 83.5 |
| | VITROS Positive | 7 | 9 | 10 | 33 | | | |
| | VITROS Negative | 52 | 3 | 0 | 1 | | | |
{7}
GC/MS Reference Method Comparison for BENZ
| | | GC/MS or LC/MS Reference | | | | %Agreement | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Cutoff Value (ng/mL) | | Low Negative | Near Cutoff Negative | Near Cutoff Positive | High Positive | %Agreement Negative | %Agreement Positive | %Agreement Overall |
| 300 | | (<-50%) <150 ng/mL | (-50% to cutoff) 150-300 ng/mL | (cutoff to +50%) 300-450 ng/mL | (>+50%) >450 ng/mL | 86.1 | 88.9 | 87.0 |
| | VITROS Positive | 6 | 5 | 8 | 24 | | | |
| | VITROS Negative | 55 | 13 | 3 | 1 | | | |
Summary of Discordant Results: GC/MS or LC/MS
| Cutoff Value (ng/mL) | VITROS BENZ Assay (ng/mL) | Reference (ng/mL) |
| --- | --- | --- |
| 200 | 87 | 415 oxazepam & 26 temazepam on LC/MS |
| | 108 | 395 alprazolam, 21 alpha-OH-alprazolam, and 168 7-amino-clonazepam on LC/MS |
| | 173 | 184 alprazolam & 5 alpha-OH-alprazolam on LC/MS |
| | 209 | 237 oxazepam on LC/MS |
| | 210 | 284 7-amino-clonazepam on LC/MS |
| | 216 | 101 alpha-OH-alprazolam & 41 alprazolam on LC/MS |
| | 224 | 303 7-amino-clonazepam on LC/MS |
| | 243 | 90 oxazepam, 26 temazepam, & 4 nordiazepam on LC/MS |
| | 248 | 38 alprazolam and 64 alpha-OH-alprazolam on LC/MS |
| | 265 | 155 oxazepam, 23 temazepam, & 8 nordiazepam on LC/MS |
| | 303 | 284 7-amino-clonazepam on LC/MS |
| | 303 | 343 7-amino-clonazepam on LC/MS |
| | 319 | 106 alpha-OH-alprazolam on GC/MS |
| | 342 | 90 oxazepam, 20 temazepam, & 15 nordiazepam on LC/MS |
| | 422 | 26 oxazepam & 6 of alpha-OH-alprazolam on LC/MS |
| | >RR | 237 oxazepam on GC/MS |
| | >RR | 71 alprazolam and 51 alpha-OH-alprazolam on LC/MS |
| | >RR | < 75 on GC/MS |
| | >RR | 76 oxazepam on GC/MS |
Summary of Discordant Results: GC/MS or LC/MS
| Cutoff Value (ng/mL) | VITROS BENZ Assay (ng/mL) | Reference (ng/mL) |
| --- | --- | --- |
| 300 | 87 | 415 oxazepam &26 temazepam on LC/MS |
| | 108 | 395 alprazolam, 21 alpha-OH-alprazolam, and 168 7-amino-clonazepam on LC/MS |
| | 250 | 200 temazepam, 104 oxazepam, & 42 nordiazepam on LC/MS |
| | 277 | 169 oxazepam, 127 temazepam, & 50 nordiazepam on LC/MS |
| | 303 | 284 7-amino-clonazepam on LC/MS |
| | 303 | 343 7-amino-clonazepam on LC/MS |
| | 319 | 106 alpha-OH-alprazolam on GC/MS |
| | 342 | 211 oxazepam, 31 temazepam, & 15 nordiazepam on LC/MS |
| | 342 | 90 oxazepam, 20 temazepam, & 15 nordiazepam on LC/MS |
| | 422 | 26 oxazepam & 6 of alpha-OH-alprazolam on LC/MS |
| | 466 | 118 alpha-OH-alprazolam & 38 alprazolam on LC/MS |
| | >RR | < 75 on GC/MS |
| | >RR | 76 oxazepam on GC/MS |
| | >RR | 237 oxazepam on GC/MS |
| | >RR | 71 alprazolam and 51 alpha-OH-alprazolam on LC/MS |
{8}
A total of 115 human urine samples were assayed using the VITROS Chemistry Products BENZ Reagent and a commercially available immunoassay method, for benzodiazepine. Percent agreement was evaluated at assay cutoff values of 200 ng/mL and 300 ng/mL.
To challenge performance at the 200 ng/mL cutoff value, 44 of the 115 samples tested had concentrations within +/- 50% of the cutoff value, 27 samples below the cutoff value and 17 above the cutoff value.
To challenge performance at the 300 ng/mL cutoff value, 39 of the 115 samples tested had concentrations within +/- 50% of the cutoff value, 30 samples below the cutoff value and 9 above the cutoff value.
Commercial Method Comparison for BENZ
| | | Commercial Method** | | | | %Agreement | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Cutoff Value (ng/mL) | | Low Negative | Near Cutoff Negative | Near Cutoff Positive | High Positive | %Agreement Negative | %Agreement Positive | %Agreement Overall |
| 200 | | (<-50%) <100 ng/mL | (-50% to cutoff) 100-200 ng/mL | (cutoff to +50%) 200-300 ng/mL | (>+50%) >300 ng/mL | 100.0 | 95.2 | 97.4 |
| | VITROS Positive | 0 | 0 | 14 | 45 | | | |
| | VITROS Negative | 26 | 27 | 3* | 0 | | | |
| 300 | | (<-50%) <150 ng/mL | (-50% to cutoff) 150-300 ng/mL | (cutoff to +50%) 300-450 ng/mL | (>+50%) >450 ng/mL | 98.6 | 93.3 | 96.5 |
| | VITROS Positive | 0 | 1* | 6 | 36 | | | |
| | VITROS Negative | 40 | 29 | 3* | 0 | | | |
b. Matrix comparison:
Not applicable; this device is for use with urine only.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
Not applicable.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
{9}
10
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.