MICROSCAN MICROSTREP PLUS PANEL

K062276 · Dade Behring, Inc. · LRG · Aug 23, 2006 · Microbiology

Device Facts

Record IDK062276
Device NameMICROSCAN MICROSTREP PLUS PANEL
ApplicantDade Behring, Inc.
Product CodeLRG · Microbiology
Decision DateAug 23, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1640
Device ClassClass 2

Indications for Use

To determine bacterial antimicrobial agent susceptibility

Device Story

MicroScan MICroSTREP plus® Panel is a miniaturized broth dilution susceptibility test. Input: bacterial colonies grown on solid media. Process: panels rehydrated with Mueller-Hinton broth, lysed horse blood, and HEPES; inoculated with standardized organism suspension; incubated 20-24 hours at 35°C. Output: minimum inhibitory concentration (MIC) of Levofloxacin (0.12-16 mcg/ml). Operation: manual visual reading or automated reading via MicroScan® WalkAway instrument. Used in clinical laboratories by trained personnel. Output assists clinicians in selecting appropriate antimicrobial therapy for streptococcal infections.

Clinical Evidence

Bench testing only. Evaluation conducted using stock and CDC challenge strains. Performance of instrument read method compared to CLSI frozen reference panel results. Results showed 100% Essential Agreement for Levofloxacin. Reproducibility and precision testing with WalkAway instrument and Quality Control testing demonstrated acceptable performance.

Technological Characteristics

Miniaturized broth dilution panel. Reagents: Mueller-Hinton broth, 2-5% lysed horse blood, 50 mM HEPES. Levofloxacin concentrations: 0.12-16 mcg/ml. Readout: manual visual or automated via MicroScan® WalkAway instrument. Incubation: 20-24 hours at 35°C +/- 1°C in non-CO2 incubator.

Indications for Use

Indicated for determining quantitative/qualitative antimicrobial susceptibility of aerobic streptococci, including Streptococcus pneumoniae, Streptococcus pyogenes, Group C/F streptococci, Group G streptococci, Streptococcus agalactiae, and Viridans group streptococci. For prescription use.

Regulatory Classification

Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K062276 ## 510(k) Summary Information: ## AUG 2 3 2006 | Device Manufacturer: | Dade Behring Inc. | |----------------------|----------------------------------------------------------------------------------------------------------| | Contact name: | Maureen Mende, Regulatory Affairs Group Manager | | Phone/Fax: | 916-374-3174/916-374-3144 | | Date prepared: | July 25, 2006 | | Product Name: | Microdilution Minimum Inhibitory Concentration (MIC) Panels | | Trade Name: | MicroScan MICroSTREP plus® Panel | | Intended Use: | To determine bacterial susceptibility to Levofloxacin | | Indication for Use: | For determining antimicrobic susceptibility with aerobic streptococci including Streptococcus pneumoniae | | Predicate device: | MicroScan® MICroSTREP plus® Panel (K020556) | ## 510(k) Summary: The MicroScan MICroSTREP plus® Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of aerobic streptococci, including Streptococcus pneumoniae. After inoculation, panels are incubated for 20 - 24 hours at 35°C +/- 1°C in a non-CO2 incubator, and read according to the Package Insert. The antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test. Various antimicrobial agents are diluted in water, buffer or minute concentrations of broth to concentrations bridging the range of clinical interest. Panels are rehydrated with 115 ul Mueller-Hinton broth supplemented with 2-5% lysed horse blood (LHB) and buffered with 50 mM HEPES, after inoculation of the broth with a standardized suspension of the organism in saline. After incubation in a non-CO2 incubator for 20-24 hours, the minimum inhibitory concentration (MIC) for the test organism is manually read by observing the lowest antimicrobial concentration showing inhibition of growth. Additionally, the panels may be incubated in and read by a MicroScan® WalkAway instrument. The proposed instrument read method for the MicroScan MICroSTREP plus® Panel demonstrated substantially equivalent performance with streptococcal isolates when compared with an expected result generated on a CLSI frozen Reference Panel, as defined in the FDA document "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA", dated February 5, 2003. This Premarket Notification (510[k]) presents data in support of reading the MICroSTREP plus Panel with Levofloxacin on the MicroScan® WalkAway instrument. {1}------------------------------------------------ The external evaluation was conducted with stock and CDC Challenge strains. The external evaluations were designed to confirm the acceptability of the proposed instrument read method for the MICroSTREP plus® Panel by comparing its performance with Expected Results determined before the evaluation. The MICroSTREP plus® Panel demonstrated acceptable performance with an overall Essential Agreement of 100% for Levofloxacin instrument read results compared with the Expected Result. Instrument reproducibility testing demonstrated acceptable reproducibility and precision with Levofloxacin and the WalkAway® instrument. Quality Control testing demonstrated acceptable results for Levofloxacin. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three overlapping shapes that resemble ribbons or stylized human figures. Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Ms. Maureen Mende Regulatory Affairs Group Manager Dade Behring, Inc. 2040 Enterprise Boulevard West Scaramento, California 95691 AUG 2 3 2006 Re: k062276 Trade/Device Name: MicroScan MICroSTREP plus® Panel Levofloxacin (0.12 - 16 mcg/ml) Regulation Number: 21 CFR § 866.1640 Regulation Name: Antimicrobial susceptibility test powder Regulatory Class: II Product Code: LRG, LTT Dated: July 25, 2006 Received: August 7, 2006 Dear Ms. Mende: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Sally anton Sally A. Hoivat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indication for Use Statement 510(k) No .: 4062276 (To be assigned by FDA) Device Name: MicroScan MICroSTREP plus® Panel Levofloxacin (0.12 to 16 mcg/ml) Intended Use To determine bacterial antimicrobial agent susceptibility Indications for Use: The MicroScan MICroSTREP plus® Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of aerobic streptococci, including Streptococcus pneumoniae. After inoculation, panels are incubated for 20 -- 24 hours at 35°C +/- 1°C in a non-CO2 incubator, and read visually according to the Package Insert. Additionally, the panels may be incubated in and read by a MicroScan® WalkAway instrument. This particular submission is for the addition of instrument read capability of the antimicrobial Levofloxacin, at concentrations of 0.12 to 16 mcg/ml on the MicroScan MICroSTREP plus® Panel. The organisms which may be used for Levofloxacin susceptibility testing on this panel are: > Streptococcus pneumoniae (including penicillin-resistant strains) Streptococcus pyogenes Streptococcus (Group C/F) Streptococcus (Group G) Streptococcus agalactiae Viridans group streptococci Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Histoph Statestic Devices (OIVD) Office of In Vitro Diagnostic Device Evaluation and Safety 510(k)________________________________________________________________________________________________________________________________________________________________________
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...