← Product Code [IZQ](/productcode/IZQ) · K062265

# CT EXPRES III CONTRAST MEDIA DELIVERY SYSTEM (K062265)

_Swiss Medical Care S.A. · IZQ · Dec 26, 2006 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K062265

## Device Facts

- **Applicant:** Swiss Medical Care S.A.
- **Product Code:** [IZQ](/productcode/IZQ.md)
- **Decision Date:** Dec 26, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1650
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The CT Exprés III™ Contrast Media Delivery System (CMDS) is indicated for controlled automatic administration, on the venous side, of contrast media (CM) to human subjects while undergoing examination by means of a computed tomography (CT) scanner. The system consists of the CT Exprés III™ Instrument, the Bottle Spike, the Day Set III, the Patient Set, accessories and detachable parts. This device is not intended for injection of CM for coronary arteriography, or for any other use for which the device is not indicated. This device is only to be operated by and under quasi-continuous supervision of qualified medical staff in an appropriate licensed health care facility.

## Device Story

Programmable, software-driven, electromechanical high-pressure contrast media injector; uses rotary peristaltic cassette for positive pressure. Inputs: user-programmed flow rates (0.5-9.9 mL/s), volumes (0-300 mL), and injection sequences (up to 5 phases). Operates in clinical CT suites; managed by qualified medical staff. Features remote control panel, bottle insulators for passive temperature control, and multi-dose fluid management. Safety mechanisms include air-in-line sensors, pressure sensors (8 bar/120 psi limit), and automatic flow rate limiting for small-gauge needles. Output: controlled delivery of contrast media/saline to patient. Benefits: enables precise, multi-phase contrast administration for enhanced CT imaging; supports wider range of needle gauges compared to predicate.

## Clinical Evidence

No clinical data. Substantial equivalence is based on descriptive information, operational characteristics, and bench testing of the device's mechanical and software-controlled injection parameters.

## Technological Characteristics

Electromechanical, software-driven peristaltic pump. Materials: proprietary disposables (Bottle Spike, Day Set III, Patient Set). Energy: electric. Form factor: main unit with remote control panel, pedestal pole, and bottle insulators. Connectivity: standalone. Pressure limit: 8 bar (120 psi). Flow rate: 0.5-9.9 mL/s. Supports 16-27 G needles.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- Premica™-CT Contrast Media Delivery System ([K983314](/device/K983314.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Swiss Medical Care -

K062265

## 510(k) SUMMARY SWISS MEDICAL CARE S.A. CT EXPRÉS III ™ CONTRAST MEDIA DELIVERY SYSTEM (PER 21 CFR 807.92)

#### 1. SUBMITTER NAME AND ADDRESS

Swiss Medical Care S.A. Avenue de Sévelin 28 CH-1004 Lausanne SWITZERLAND Tel: +41 (0)21 623 60 30 Fax: +41 (0)21 623 60 31

Contact Person: Ms. Cécile Boyer, Manager, Regulatory Affairs Email: c.boyer@swissmedcare.com

#### 2. DEVICE NAME

CT Exprés III™ Contrast Media Delivery System (CMDS) Trade Name: Common/Usual Name: Automatic injector for contrast media Iniector Classification Name: Injector, Contrast medium, automatic (21 CFR 870.1650, Product Code IZQ)

#### 3. PREDICATE DEVICE/S

Premica™-CT Contrast Media Delivery System - K983314

#### DEVICE DESCRIPTION 4.

The CT Exprés III™ Contrast Media Delivery System (CMDS) is a programmable, software driven, electromechanical, high-pressure contrast media injection system.

The positive pressure necessary for the injection is generated by a disposable rotary peristaltic cassette, located inside one of the disposable (Patient Set).

The CT Exprés III™ Contrast Media Delivery System consists of the main following elements:

- . The instrument, i.e. the main unit (injector unit and power supply), the main control panel, and associated cables.
- . Accessories, i.e. the handswitch.
- Detachable parts, i.e. the remote control panel, the bottle insulators, and the pedestal pole. .
- Three associated proprietary disposables creating a complete fluidic pathway from the . bottle of CM to the patient, i.e. the "Bottle Spike", the "Day Set III" and the "Patient Set".

The printer and ceiling mount are provided as optional devices for use in combination with the CT Exprés III™.

DEC 26 2006

:

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains the text "Swiss Medical Care" followed by a black square with a white line through it. The text is in a simple, sans-serif font and is left-aligned. The black square is positioned to the right of the text, acting as a visual element or logo.

#### INTENDED USE 5.

The CT Exprés III™ Contrast Media Delivery System (CMDS) is indicated for controlled automatic administration, on the venous side, of contrast media (CM) to human subjects while undergoing examination by means of a computed tomography (CT) scanner.

The system consists of the CT Exprés III™ Instrument, the Bottle Spike, the Day Set III, the Patient Set, accessories and detachable parts.

This device is not intended for injection of CM for coronary arteriography, or for any other use for which the device is not indicated.

This device is only to be operated by and under quasi-continuous supervision of qualified medical staff in an appropriate licensed health care facility.

#### 6. BASIS FOR SUBSTANTIAL EQUIVALENCE

Information supplied in this premarket notification includes descriptive information about the intended use, operation, and technological characteristics. A side-by-side comparison of the CT Exprés III™ Contrast Media Delivery System with the CT Exprés™ Contrast Media Delivery System is provided in Table J-1 below:

| Characteristics                                                                               | CT Exprés III™ CMDS                                      | Premica™-CT CMDS<br>(K983314)                        |
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------|------------------------------------------------------|
| Indication                                                                                    |                                                          |                                                      |
| Indicated for controlled<br>administration of contrast media for<br>computed tomography scans | Yes                                                      | Yes                                                  |
| Not intended for use with children<br>under 16 years of age                                   | No                                                       | Yes                                                  |
| Physical Design                                                                               |                                                          |                                                      |
| Remote panel                                                                                  | Yes                                                      | Yes                                                  |
| Weight (without fluid containers)                                                             | Injector : 8.5 kg (18 lbs)<br>Console : 1.7 kg (3.7 lbs) | Injector : 8 kg (17.6 lbs)<br>Console : 1.7 kg (lbs) |
| Contrast media temperature control                                                            | Bottle insulators                                        | Bottle insulators                                    |
| Single patient use disposables                                                                | Patient Set                                              | Patient Set                                          |
| Multiple patient solutions                                                                    | Yes                                                      | Yes                                                  |
| Designed to prevent reuse of<br>disposables                                                   | Yes                                                      | Yes                                                  |
| Operational Characteristics                                                                   |                                                          |                                                      |
| Programmable flow rate                                                                        | 0.5 - 9.9 mL/s                                           | 0.2 - 9.9 mL/s                                       |
| Programmable injection volume per<br>injection                                                | 0 - 300 mL                                               | 0 - 300 mL                                           |
| Maximum injection duration                                                                    | 9 min 59 sec                                             | 9 min 59 sec                                         |

Table H-1: Comparison of CT Exprés III™ Contrast Media Delivery System with CT Exprés™ Contrast Media Delivery System

CONFIDENTIAL Appendix H, page 2

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image contains the text "Swiss Medical Care" in a simple, sans-serif font. To the right of the text, there are two small, solid black squares, one slightly above and to the left of the other. The overall impression is that of a logo or heading for a medical care provider.

## Table H-1: Comparison of CT Exprés III™ Contrast Media Delivery System with CT Exprés TM Contrast Media Delivery System (Continued)

| Characteristics                 | CT Exprés III™ CMDS | Premica™-CT CMDS<br>(K983314) |
|---------------------------------|---------------------|-------------------------------|
| Contrast media container volume | 50-500 mL           | 50-500 mL                     |
| Pressure limit                  | 8 bar (ca. 120 psi) | 8 bar (ca. 120 psi)           |
| Test injection default          | Yes                 | Yes                           |
| Injection capabilities          | Up to 3 phases      | Up to 5 phases                |
| Saline flush                    | Yes                 | Yes                           |
| Connecting tubing               | Length = 1.5 m      | Length = 1.5 m                |
| Access types                    | Flexible            | Flexible                      |
| Access gauge                    | 16-27 G             | 16-22 G                       |
| Interphase delay                | 0-99 sec            | 0-99 sec                      |
| Scan delay                      | 0-99 sec            | 0-99 sec                      |
| Injection protocol storage      | 100 protocols       | 100 protocols                 |

## Indications for use

Both the CT Exprés III™ and Premica™-CT contrast media delivery system are indicated for the controlled administration of contrast media for computed tomography scans.

The CT Exprés III™ can be used with all patients. The Premica™-CT labelling indicates that the device can not be used with patients under 16 years of age.

## Physical design

Both systems consist of several elements.

Both systems have remote panels to allow easier user control of the systems.

Both systems provide single patient use disposables ("Patient Set"). Both systems have disposables with design features that prevent their reuse.

Both systems draw their contrast media or saline from multi-dose containers (bottles or soft pouches). Contrast media container volumes are identical for the two systems (50 - 500 mL).

## Operational characteristics

Programmable flow rate ranges are nearly identical for both systems (0.5 - 9.9 mL/s for the CT Exprés III™, 0.2-9.9 mL/s).

The programmable volume of contrast media per injection is also identical (0 - 300 mL/injection).

Programmable flow rate ranges and programmable volume ranges being identical, the injection durations are also identical for the two systems.

July 17, 2006

CONFIDENTIAL Appendix H, page 3

{3}------------------------------------------------

## Swiss Medical Care = =

The CT Exprés III™ can be programmed for injection sequences of one to five steps:

- 1. Saline (pre-flush)
- 2. Contrast media (phase one)
- 3. Saline (intermediate flush)
- 4. Contrast media (phase two)
- 5. Saline (post-flush)

The Premica™-CT can be programmed for injection sequences of up to three steps:

- 1. Contrast media (phase one)
- 2. Contrast media (phase two)
- 3. Saline (post-flush)

Both systems incorporate pressure sensors with the limit for triggering a pressure alarm (occlusion alarm) at 8 bar (ca. 120 psi).

Both systems incorporate Air-In-Line sensors for patient safety.

Since contrast media temperature can change the fluid viscosity, both systems provide for passive temperature control in the form of bottle insulators to reduce temperature loss from the heated contrast.

One difference between the two systems is the ability of the CT Exprés III™ to simultaneously accommodate two bottles of contrast media and one container with saline, whereas the Premica™-CT can accommodate only two fluid containers (two bottles of contrast media OR one bottle with contrast media and one container with saline).

Another difference between the two systems is the ability of the CT Expres III™ to accept not only the range of needle sizes (16-22 G) normally used for injecting contrast media, but also very small needle sizes (23-27 G). Note that, when using the very small needle sizes (23-27 G). the programmable injection rates may be limited to values of only a few mL/s because the injection pressure that CT Exprés III™ is generating is limited to 8 bar (ca. 120 psi). As a safety feature, CT Exprés III™ will automatically display information for the user when an injection rate is programmed above the limit injection rate. Under these circumstances, the CT Expres III™ will not function. This makes the CT Expres III™ suitable for use with patients or injection conditions where very small needle sizes are preferred.

The Premica™-CT can only accept needle sizes in the range of 16 - 22 G, making it unsuitable for use with patients or injection conditions where very small needle sizes are preferred.

These differences are explained for in the CT Exprés III™ operator's manual and software programming capabilities and do not raise any new questions of safety or effectiveness.

### 7. PREPARATION DATE OF 510(K) SUMMARY

July 26, 2006

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a circular logo with text around the perimeter and a stylized eagle in the center. The text reads "DEPARTMENT OF HEALTH & HUMAN". The eagle is depicted with three thick, curved lines forming its body and wings, giving it a modern and abstract appearance.

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

Swiss Medical Care S.A. % Mr. Robert L. Aromando, Jr. Managing Partner K Street Associates, LLC PMB 237, 203 Main Street FLEMINGTON NJ 08822

Re: K062265

Trade/Device Name: Swiss Medical Care S.A. CT EXPRESS III™ Contrast Media Delivery System Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic Injector and Syringe Regulatory Class: II Product Code: IZQ

Dated: December 7, 2006 Received: December 8, 2006

Dear Mr. Aromando:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/4/Picture/11 description: The image shows a circular logo with the text "1906-2006" at the top. The letters "FDA" are prominently displayed in the center, with the word "Centennial" written below. Three stars are arranged in a row beneath the word "Centennial", and the phrase "Science in Service of Health" is written along the bottom curve of the circle. The logo appears to commemorate the centennial of the Food and Drug Administration.

Protecting and Promoting Public Health

{5}------------------------------------------------

## Page 2 --

Please be advised that FDA's issuance of a substantial equivalence determination does not mean I loase of actived that i Drimination that your device complies with other requirements of the Act that I Dri hes intact and regulations administered by other Federal agencies. You must or any I catal statutes and regerements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI K I at 607), lacemig (21 OFF regulation (21 CFR Part 820); and if applicable, the electronic form in the quint) of ovisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) I his icher will anow you to oogin mains of substantial equivalence of your device to a legally premarket notheadon: "The President for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific at Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology)                      | 240-276-0120 |
| Other          |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under (21 CF F at 60 : 377): " roo of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely yours,

Nancy C. Hogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# Indications for Use

510(k) Number (if known):

K062265

## Device Name: SWISS MEDICAL CARE S.A. CT EXPRÉS III™ CONTRAST MEDIA DELIVERY SYSTEM

Indications for Use:

The CT Exprés III™ Contrast Media Delivery System (CMDS) is indicated for controlled automatic administration, on the venous side, of contrast media (CM) to human subjects while undergoing examination by means of a computed tomography (CT) scanner.

The system consists of the CT Exprés III™ Instrument, the Bottle Spike, the Day Set III, the Patient Set, accessories and detachable parts.

This device is not intended for injection of CM for coronary arteriography, or for any other use for which the device is not indicated.

This device is only to be operated by and under quasi-continuous supervision of qualified medical staff in an appropriate licensed health care facility.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Barbara Buchholz for NCB

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K061265

Swiss Medical Care S.A. 510(K) July 28, 2006

CONFIDENTIAL

---

**Source:** [https://fda-staging.innolitics.com/device/K062265](https://fda-staging.innolitics.com/device/K062265)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
