In vitro test for the quantitative determination of glucose in serum, plasma, urine and cerebrospinal fluid (CSF) on COBAS INTEGRA systems. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus and idiopathic hypoglycemia.
Device Story
In vitro diagnostic reagent system for COBAS INTEGRA analyzers; utilizes hexokinase enzymatic method to measure glucose concentration in serum, plasma, urine, and CSF. Reagents (R1 and R2) react with patient samples to produce measurable signals; system quantifies glucose levels. Used in clinical laboratory settings by trained personnel. Output provides quantitative glucose values to clinicians for diagnosing and monitoring carbohydrate metabolism disorders, including diabetes mellitus and idiopathic hypoglycemia. Benefits include accurate, standardized glucose monitoring for patient management.
Clinical Evidence
Bench testing only. Analytical performance validated via precision studies (within-run and between-run CVs), lower detection limits, and interference testing (hemolysis, icterus, lipemia, gammopathy). No clinical prospective or retrospective studies were required for this modification.
Technological Characteristics
Enzymatic hexokinase assay. Reagents: R1 (TRIS, ATP, Mg2+, NADP) and R2 (Mg2+, HEPES, HK, G6PDH). Analyzed on COBAS INTEGRA systems. Standardized against Isotope Dilution Mass Spectrometry. Connectivity: Integrated with COBAS INTEGRA platform. Storage: 2-8°C.
Indications for Use
Indicated for quantitative glucose determination in serum, plasma, urine, and CSF to aid in diagnosis and treatment of carbohydrate metabolism disorders, including diabetes mellitus and idiopathic hypoglycemia.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
{0}
K062239
SEP 11 2006
# 510(k) Summary – COBAS INTEGRA Glucose HK Gen. 3
## Introduction
According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.
## Submitter name, address, contact
Roche Diagnostics
9115 Hague Rd
Indianapolis IN 46250
(317) 521-3723
Contact person: Theresa (Tracy) Ambrose Bush
Date prepared:
## Device Name
Proprietary name: COBAS INTEGRA Glucose HK New Formulation
Common name: Gluc2
Classification name: Glucose Test System
## Device Description
The cassette COBAS INTEGRA Glucose HK New Formulation contains an in vitro diagnostic reagent system intended for use on COBAS INTEGRA SYSTEMS for the quantitative determination of glucose in serum, plasma, urine, and cerebrospinal fluid (CSF).
The test principle is an enzymatic reference method with hexokinase.
## Intended use
In vitro test for the quantitative determination of glucose in serum, plasma, urine and cerebrospinal fluid (CSF) on COBAS INTEGRA systems.
## Predicate Device
We claim substantial equivalence to the COBAS INTEGRA Glucose HK Liquid cleared as K972250.
## Substantial equivalency – Similarities
The table below indicates the similarities between the modified COBAS INTEGRA Glucose HK New Formulation test and its predicate device (COBAS INTEGRA Glucose HK Liquid, K972250).
{1}
| Feature | Predicate device: Glucose HK Liquid (K972250) | Modified device: Glucose HK New formulation |
| --- | --- | --- |
| General | | |
| Intended Use/Indications for Use | The cassette COBAS INTEGRA Glucose HK Liquid contains an in vitro diagnostic reagent system intended for use on COBAS INTEGRA systems for the quantitative determination of glucose concentration in serum, plasma, urine and cerebrospinal fluid (CSF). Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus and idiopathic hypoglycemia. | In vitro test for the quantitative determination of glucose in serum, plasma, urine and cerebrospinal fluid (CSF) on COBAS INTEGRA systems. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus and idiopathic hypoglycemia. |
| Specimen type | Serum, plasma, urine, CSF | Same |
| Test principle | | |
| Reference method | Enzymatic reference method with hexokinase. | Same |
| Reagent information | | |
| Stability - shelf life and on-board | 2-8 °C until expiration date
COBAS INTEGRA 400
8 weeks at 10 to 15°C
COBAS INTEGRA 700/800
8 weeks at 8°C | Same |
| Calibrator | Calibrator f.a.s.
Interval: each lot | Same |
| Quality control | Serum and plasma:
Precinorm U and Precinorm U Plus Precipath U and Precipath U Plus Urine:
Quantitative urine controls
CSF:
Quantitative CSF controls
Interval: 24 hrs recommended | Same |
| Traceability | Standardized against Isotope Dilution Mass Spectrometry | Same |
| Performance characteristics | | |
{2}
| Measuring range | 0-40 mmol/L (0-720 mg/dL)
Extended measuring range with recommended post dilution factor of 10: 0-400 mmol/L (0-7200 mg/dL) | 0-40 mmol/L (0-720 mg/dL)
Extended measuring range with recommended post dilution factor of 10: 0 -400 mmol/L (0-7200 mg/dL) |
| --- | --- | --- |
| Expected values (literature reference)
Additional values are referenced in the method sheet | Plasma (fasting): 3.88-6.38 mmol/L
Urine:
1^{st} morning urine 0.3-1.1 mmol/L
24 h urine 0.11-0.50 mmol/24h
Serum/plasma:
Adults 4.11-5.89 mmol/L | Plasma (fasting): 3.88-6.38 mmol/L
Urine:
1^{st} morning urine 0.3-1.1 mmol/L
24 h urine 0.3-0.96 mmol/L
Same |
| Endogenous interferences | Hemolysis no significant interferences
Icterus no significant interferences
Lipemia no significant interferences | Same |
## Substantial equivalency – Differences
The table below indicates the similarities between the modified COBAS INTEGRA Glucose HK New Formulation test and its predicate device (COBAS INTEGRA Glucose HK Liquid, K972250).
| Feature | Predicate device: Glucose HK Liquid (K972250) | Modified device: Glucose HK New Formulation |
| --- | --- | --- |
| Reagent information | | |
| R1
R2 | Mono reagent in vial A and B (liquid)
100 mmol/L MOPS, 12 mmol/L ATP, 6 mmol/L NAD⁺, 10 mmol/L Mg²⁺, =50 μcat/L HK(yeast), =50 μcat/L G6PDH (microbial), 0.09% Sodium azide, pH 7.1 | R1:
100 mmol/L TRIS, 1.7 mmol/L ATP, 4 mmol/L Mg²⁺, 1 mmol/L NADP, pH 7.8 |
| | | R2:
4.0 mmol/L Mg²⁺, 30 mmol/L HEPES, =130 μcat/L HK (yeast), =250 μcat/L G6PDH (microbial), pH 7.0 |
| Performance characteristics | | |
{3}
| Precision | Serum and plasma:
Within run:
1.7% @ 5.3 mmol/L
0.72% @ 33.2 mmol/L
Between run:
2.6% @ 5.3 mmol/L
1.5% @ 33.2 mmol/L
Urine application
Within run:
1.7% @ 1.7 mmol/L
1.8% @ 37.1 mmol/L
Between run:
4.3% @ 1.7 mmol/L
2.9% @ 37.1 mmol/L
CSF application
Within run:
1.6% @ 1.7 mmol/L
1.8% @ 3.3 mmol/L
Between run:
2.3% @ 1.7 mmol/L
1.9% @ 3.3 mmol/L | Serum and plasma:
Within run CV%:
1.8 % @ 4.54 mmol/L
1.6% @ 13.5 mmol/L
Between run:
2.1% @ 4.54 mmol/L
2.0% @ 13.5 mmol/L
Urine application
Within run:
1.2% @ 1.63 mmol/L
1.1% @ 16.3 mmol/L
Between day:
1.2% @ 1.63 mmol/L
1.1% @ 16.3 mmol/L
CSF application
Within run:
0.87% @ 3.43 mmol/L
1.3% @ 1.72 mmol/L
Between run:
0.91% @ 3.43 mmol/L
1.4% @ 1.72 mmol/L |
| --- | --- | --- |
| Lower detection limit | Serum and plasma:
0.033 mmol/L
Urine application:
0.22 mmol/L
CSF application:
0.023 mmol/L | Serum, plasma, urine and CSF:
0.03 mmol/L |
| Exogenous Interferences | Falsely low results may be caused by elevated pyruvates levels
Gammopathy, in particular IgM, may cause unreliable results in rare cases | In rare cases gammopathy, in particular type IgM (Waldenstrom’s macroglobulinemia) may cause unreliable results. |
{4}
## Proposed Labeling
Proposed labeling sufficient to describe the device, its intended use, and the directions for use can be found in Section V. We believe the proposed version of the device labeling presented contains all of the technical information required per 21 CFR 809.10.
## Validation and Design Control
Development activities were conducted under appropriate design control procedures and the overall product specifications were met. The Declaration of Conformity with Design Controls and Results of Risk Analysis are provided in Section 5.1. Analytical Performance.
## Confidentiality
Roche Diagnostics Corporation requests that the FDA not disclose the nature or existence of this submission until the substantial equivalence decision has been reached.
## Closing
Modification of the COBAS INTEGRA Glucose HK assay, resulting in the Glucose HK New Formulation assay, did not affect the intended use or indications for use of the device as described in the labeling, nor did it alter the fundamental scientific technology of the device. Therefore, we trust the information provided in this Special 510(k) will support a decision of substantial equivalence of the COBAS INTEGRA Glucose HK Gen.3 to the predicate.
If you have any questions or require further information, please do not hesitate to contact this office.
- Phone: (317) 521-3723
- FAX: (317) 521-2324
- email: tracy.bush@roche.com
{5}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
2098 Gaither Road
Rockville MD 20850
Ms. Theresa Abrose Bush
Roche Diagnostics
9115 Hague Rd.
Indianapolis, IN 46250
SEP 11 2006
Re: k062239
Trade/Device Name: Glucose HK New Formulation (Gluc2)
Regulation Number: 21 CFR 862.1345
Regulation Name: Glucose test system
Regulatory Class: Class II
Product Code: CFR
Dated: August 1, 2006
Received: August 2, 2006
Dear Ms. Abrose Bush:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA’s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act’s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{6}
Page 2 –
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Alberto Gutierrez, Ph.D.
Director
Division of Chemistry and Toxicology
Office of In Vitro Diagnostic Device
Evaluation and Safety
Center for Devices and
Radiological Health
Enclosure
{7}
# Indications for Use
510(k) Number (if known):
Device Name: Glucose HK New Formulation (Gluc2)
Indications For Use:
In vitro test for the quantitative determination of glucose in serum, plasma, urine and cerebrospinal fluid (CSF) on COBAS INTEGRA systems. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus and idiopathic hypoglycemia.
Prescription Use XXXX
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Office of In Vitro Diagnostic Device
Evaluation and Safety
510(k) K062239
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.