← Product Code [LLZ](/productcode/LLZ) · K062237

# CADI, VERSION 4 (K062237)

_Synca Marketing, Inc. · LLZ · Nov 20, 2006 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K062237

## Device Facts

- **Applicant:** Synca Marketing, Inc.
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Nov 20, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

CADI is a dental imaging software for general dental and maxillo-facial diagnostic. It controls capture, display, treatment, analysis and saving of digital images from x-ray units, intraoral or extraoral dental camera, digital cameras, and images acquired by digitizing film with a scanner, in order to enable a dentist to diagnose the dental health of a patient. CADI is a prescription use only product, intended to be used by dental professionals only.

## Device Story

CADI Version 4 is a modular dental imaging software platform; inputs include digital x-rays (sensors/phosphor plates), intraoral/extraoral cameras, and scanned film images. Software modules provide image enhancement, sorting, and diagnostic tools; specific features include 16-bit x-ray viewing, automatic intraoral image sorting by tooth number, cosmetic whitening simulation, and orthocephalometric tracing (DFO) with point location tools. Used in dental clinics by dental professionals; output displayed on workstation monitors to assist in clinical diagnosis and patient treatment planning. Benefits include streamlined image management, enhanced diagnostic visualization, and patient communication via cosmetic simulations.

## Clinical Evidence

No clinical data provided; device relies on bench testing and functional verification of software modules for dental imaging and analysis.

## Technological Characteristics

Modular software architecture; supports 16-bit image processing; compatible with standard digital x-ray sensors, phosphor plates, intraoral/extraoral cameras, and scanners. Features include image enhancement algorithms, automated sorting, cosmetic simulation, and cephalometric tracing tools. Operates on standard dental workstation hardware.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows a logo for the Department of Health. The logo features a stylized human figure with three wavy lines representing the body and head. The words "DEPARTMENT OF HEALTH" are arranged vertically along the left side of the logo, with the letters oriented to follow the curve of the logo.

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

NOV 2 0 2006

Mr. Stephane Arsenault Official Correspondent Synca Marketing, Inc. 337 Marion Street Le Gardeur, Quebec, J5Z 4W8 CANADA

Re: K062237

Trade/Device Name: CADI Version 4 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: October 25, 2006 Received: October 26, 2006

Dear Mr. Arsenault:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the cnactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/0/Picture/10 description: The image is a circular seal or logo. The text "1906 - 2006" is at the top of the seal. The letters "FDA" are in the center of the seal, with the word "Centennial" underneath. There are four stars at the bottom of the seal.

ag and Promoting Public 9

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean I that FDA has made a determination that your device complies with other requirements of the Act that I Dri has made a and regulations administered by other Federal agencies. You must of any I cacial statutes and registerients, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI N Far 6077, adomig (21 CFR Part Paulation (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) This icael with anow you're ough times of substantial equivalence of your device to a legally premarket notification: "The Presults in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advise to your and of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology | 240-276-0115 |
|----------------|---------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)         | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology)                     | 240-276-0120 |
| Other          |                                 | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under (21 CF F r at 60 : 31). " v of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrl/industry/support/index.litml.

Sincerely yours,

Nancy C. Hogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Section 4 Indications for Use Statement

CADI is a dental imaging software for general dental and maxillo-facial diagnostic. It controls capture, display, treatment, analysis and saving of digital images from x-ray units, intraoral or extraoral dental camera, digital cameras, and images acquired by digitizing film with a scanner, in order to enable a dentist to diagnose the dental health of a patient.

CADI is a modular software, meaning that it offers dentistry professionals the choice to activate specific modules, depending on the type of image sources they use. These modules are as follows:

Digital x-ray module: This module features a 16-bit x-ray viewer and imaging tools used to enhance the image quality of x-rays taken by sensors or phosphor plate systems.

Intraoral Camera Module: CADI automatically enhances images captured by intraoral cameras and sorts them by tooth number.

Digital Camera Module:

This module allows importing digital camera images, and sorts them in order in a patient file. Printouts are automatically sorted in AAO order, or any custom order. This module also crops and prints images.

Cosmetic and Whitening Module:

CADI simulates whitening results in minutes to help build an in-office program for patients. It also provides cosmetic makeover simulations using a library of natural smile images.

Natural Smiles Library: Smiles library that sorts images in a logical order, including images for a range of ethnic backgrounds.

Orthocephalometric Tracing (DFO) Module:

CADI DFO uses point location tools and a tutor window to help achieve cephalometric tracings. It includes most current and popular tracing analysis, and allows the user to create his/her own analysis.

CADI is a prescription use only product, intended to be used by dental professionals only.

Stephane Arsenault

October 25, 2006

Prescription Use
(Per 21 CFR 801.109)
(Division Sign-Off)
Division of Reproductius

Division of Reproductive, Abdon and Radiological Devi 510(k) Number

Image /page/2/Picture/19 description: The image shows the word "SYNCA" in all capital letters. To the right of the word is a circular graphic with a solid black circle on top of a target-like design. The target design consists of three concentric circles, with the innermost circle being white and the other two being black.

337 Marion Street. Le Gardeur, QC, Canada, J5Z 4W8 1-800-667-9622 Fax : 1-800-554-1021 email : synca@synca.com

---

**Source:** [https://fda-staging.innolitics.com/device/K062237](https://fda-staging.innolitics.com/device/K062237)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
