K062219 · Prodrive Systems, Inc. · EFB · Aug 4, 2006 · Dental
Device Facts
Record ID
K062219
Device Name
PRODRIVE SYSTEMS REPLACEMENT TURBINE
Applicant
Prodrive Systems, Inc.
Product Code
EFB · Dental
Decision Date
Aug 4, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Attributes
3rd-Party Reviewed
Indications for Use
The ProDrive Replacement Turbine is intended for use by authorized persons in the practice of dentistry. The ProDrive Replacement Turbine is intended to be used intraorally by trained dental professionals for drilling and preparation of dental cavities for restoration, such as fillings.
Device Story
ProDrive Replacement Turbine is a high-speed air-powered turbine for existing dental handpieces. It utilizes a push-button spindle mechanism with unique geometry to grip ProDrive carbide and diamond cutting instruments. This mechanism locks the bur to prevent accidental disengagement and allows for an extended bur position to improve visibility and utility. The device is used by dental professionals in clinical settings. It is compatible only with ProDrive-specific burs, which feature a unique shank geometry. The turbine is constructed from stainless steel. It functions as a replacement component for refurbishing dental handpieces, providing improved performance and durability compared to standard friction-grip systems.
Clinical Evidence
Bench testing only. The device was evaluated against ISO 7785-1:1997(E) standards and FDA guidance for dental handpieces. Testing included extraction force, resistance to corrosion, torque, noise level, eccentricity, speed, air pressure, stall torque, and sterilization validation (250 cycles). Results indicate the device functions as well as or better than the predicate device.
Technological Characteristics
High-speed air-powered turbine; stainless steel construction; push-button spindle mechanism with unique geometry for proprietary bur locking; compatible with ProDrive carbide/diamond instruments; meets ISO 7785-1:1997(E) performance standards; sterilization validated for 250 cycles.
Indications for Use
Indicated for use by trained dental professionals for intraoral drilling and preparation of dental cavities for restoration, such as fillings.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for "PRODRIVE SYSTEMS". The logo is in black and white, with the word "PRODRIVE" in large, bold letters. Below the word "PRODRIVE" is the word "SYSTEMS" in smaller letters. The letter "O" in "PRODRIVE" is replaced with a circle.
K06 2219
#### 510(k) SUMMARY
#### ProDrive Replacement Turbine ProDrive Systems Inc.
AUG - 4 2006
#### Owned by:
ProDrive Systems Inc. 812 Commerce Drive Ogdensburg, NY 13669 USA
#### Contact Person:
Clive Hooton, Vice President Quality ProDrive Systems Inc. 812 Commerce Drive Ogdensburg, NY 13669 USA
#### Date of Summary:
June 28, 2006
#### Device Name:
ProDrive Replacement Turbine Proprietary Name: Common Name: turbine assembly for dental handpiece Classification Name: Handpiece, Air-powered, Dental Product Code: EFB ("Dental handpiece and accessories", 21CFR 872.4200)
#### Predicate Device:
Proprietary Name: 430 Series High Speed handpeice 510(k) Number: K960719 Product Code: EFA*
* Note that this handpiece is incorrectly identified in the CDRH database as EFA (handpiece, belt and/or gear driven, dental). It is, in fact, an air-powered handpiece (EFB), as is the device presented in this submission.
5
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for "PRODRIVE SYSTEMS". The word "PRODRIVE" is in large, bold, sans-serif font, with a circle in place of the letter "O". Below "PRODRIVE" is the word "SYSTEMS" in a smaller, sans-serif font.
#### Description of Device:
The ProDrive Replacement Turbine is a high-speed air-powered turbine designed to be installed in existing, legally marketed dental handpieces.
The ProDrive Replacement Turbine is appropriate for use with ProDrive carbide and diamond cutting instruments. A push-button spindle mechanism is used to grip the shanks of the cutting instruments. The unique geometry of the bur shank and turbine spindle securely locks the bur in the turbine eliminating the slipping that can occur over time with traditional burs and turbines. In addition to improved performance and durability, the secure locking mechanism enhances the safety of the ProDrive Replacement Turbine by reducing the risk of accidental disengagement of burs from the handpiece during operation.
When fully seated, the length that the ProDrive bur extends from the handpiece is the same as that of the standard bur/turbine assembly. In addition, the ProDrive system incorporates the ability to index the bur to a second, extended, position effectively improving on the visibility angle and the utility of the handpiece without adversely affecting its performance.
Because of the unique geometry of the ProDrive Replacement Turbine it is appropriate for use only in combination with ProDrive carbide and diamond cutting instruments. Standard round-shank burs will not fit into the ProDrive chuck, preventing their unintended use. ProDrive burs, however, will fit in traditional friction-grip chucks and will meet appropriate performance requirements when used in this way.
With the exception of the unique drive-lock spindle, the ProDrive Replacement Turbine consists of commercially available replacement parts currently used to refurbish dental handpieces in the United States. The spindle is composed entirely of stainless steel. The predicate device spindle is also manufactured from stainless steel. Therefore, the products are considered to be identical in materials used in their manufacture.
ProDrive burs are composed of tungsten vanadium stainless steel, coated with carbide or diamond cutting surfaces. The ProDrive burs are machined for ProDrive Systems Inc. by the bur manufacturer to the unique shank geometry required by the ProDrive drive-lock mechanism. In all other respects the burs are identical to commercially available burs currently distributed in the
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for "PRODRIVE SYSTEMS". The word "PRODRIVE" is written in all capital letters, with a circle in place of the letter "O". The word "SYSTEMS" is written in smaller capital letters below the word "PRODRIVE".
United States. The use of stainless steel in the manufacture of ProDrive spindles and burs ensures biocompatibility and resistance to corrosion.
The ProDrive Replacement Turbine meets or exceeds all of the applicable performances standards outlined in ISO 7785-1:1997(E). Extensive testing indicates that after 250 sterilization cycles the ProDrive Replacement Turbine continues to meet all performance requirements.
#### Intended Use of Device:
The ProDrive Replacement Turbine is intended for use by authorized persons in the practice of dentistry. The intended use of the ProDrive Replacement Turbine is identical to that of the Predicate Device.
## Comparison of Technological Characteristics to Predicate Device:
The following table compares the features of the ProDrive Replacement Turbine to the Predicate Device:
| TECHNOLOGICAL CHARACTERISTIC | COMPARISON TO PREDICATE |
|--------------------------------------------------|-------------------------|
| Intended use | Identical |
| Indications for use | Identical |
| Target population | Identical |
| Anatomical sites | Identical |
| Where used | Identical |
| Energy used and/or delivered | Similar |
| Human factors | Similar |
| Design | Similar |
| Performance | Similar |
| Standards met | Similar |
| Materials | Identical |
| Biocompatibility | Identical |
| Compatibility with environment and other devices | Similar |
| Sterility | Similar |
| Electrical safety | Identical |
| Mechanical safety | Similar |
| Chemical Safety | Identical |
| Thermal safety | Similar |
| Radiation safety | Identical |
### Table 1: Substantial Equivalence Table
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "PRODRIVE SYSTEMS". The word "PRODRIVE" is in large, bold, sans-serif font, with a circle replacing the letter "O". Below "PRODRIVE" is the word "SYSTEMS" in a smaller, lighter font.
## Conclusions Drawn from Technical Comparison:
The ProDrive Replacement Turbine is essentially the same as the predicate device in terms of its intended use, operating principles and materials.
The ProDrive Replacement Turbine combined with ProDrive burs conform to the ISO 7785-1:1997(E) standard and the FDA Guidance Document on Dental Handpieces, which include testing and performance limits on the following characteristics:
●
- Extraction force .
- . Resistance to corrosion
- Torque .
.
♥
- Noise level .
- Eccentricity Speed
- Air pressure Sterility validation .
- Stall torque
- Reprocessing .
The performance testing of the ProDrive Replacement Turbine indicates that it functions as well as or better than the predicate device in all aspects examined. Therefore, we conclude that the ProDrive Replacement Turbine is both safe and effective for its intended use.
8
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
ProDrive Systems, Incorporated C/O Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, Minnesota 55313
AUG - 4 2006
Re: K062219
Trade/Device Name: ProDrive Systems Replacement Turbine Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EFB Dated: July 31, 2006 Received: August 2, 2006
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendie comments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III If your device is classes to such additional controls. Existing major regulations affecting (1 wr 1), it a can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) _ You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Swer Quare
Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): Not yet assigned
The ProDrive Replacement Turbine Device Name:
To be used intraorally by trained dental professionals for Indications For Use: drilling and preparation of dental cavities for restoration, such as fillings.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Super Quary
on Sign-Cif) ്ന of Anesthesiology, General Hospital, Jon Control, Dental Devices
. (1) Number:
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.