The Central Monitoring System is intended to conduct centralized monitoring of adult, pediatric and neonatal vital sign information from multiple monitors in hospitals or medical institutions, the monitoring parameters include ECG, NIBP, SpO2. PR, RESP, IBP, TEMP, CO2 and AG. It is not intended for home use.
Device Story
HYPERVISOR VI Central Monitoring System is a medical information system for centralized patient monitoring in hospitals. It connects via TCP/IP network to Mindray bedside monitors (PM-8000 Express, PM-9000 Express, VS-800) to collect, display, and export physiological data. Inputs include ECG, RESP, CO2, PLETH, IBP, and AG waveforms, plus HR, RR, SpO2, PR, SYS, DIA, MEAN, TEMP, CO2, and AwRR parameters. The system supports up to 32 bedside monitors across two displays. It provides audible and visual alarms for bedside monitor alerts and system errors. Database technology stores alarm events for clinical review. Healthcare providers use the centralized display to manage multiple patients simultaneously, reducing manual monitoring burden and facilitating clinical decision-making through integrated data access.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
PC-based system running Windows XP Professional. Connectivity via TCP/IP network protocol. Supports up to 32 bedside monitors. Displays physiological waveforms and parameters. No ST segment or arrhythmia analysis. No wireless/telemetry capability.
Indications for Use
Indicated for centralized monitoring of adult, pediatric, and neonatal patients in hospitals or medical institutions. Not for home use.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
Agilent Technologies Information Center (AIC) Software (K011093)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
The assigned 510(k) number is: K062194
APR 1 1 2007
# Submitter:
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China Tel: +86 755 2658 2888 Fax: +86 755 2658 2680
- Contact Person: .
- Li Dongling
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
- Date Prepared: .
July 14, 2006
Name of the device:
- Trade/Proprietary Name: HYPERVISOR VI Central Monitoring System
- Common Name: Central Monitoring System
#### . Classification
| 21 CFR 870.2300 | Cardiac monitor (including cardiotachometer and rate alarm) | Class II |
|-----------------|----------------------------------------------------------------------|----------|
| 21 CFR 870.1130 | Non-Invasive blood pressure measurement System | Class II |
| 21 CFR 870.1110 | Blood pressure computer | Class II |
| 21 CFR 880.2910 | Clinical Electronic Thermometers -<br>Temperature Monitor with Probe | Class II |
| 21 CFR 870.2700 | Oximeter, Pulse | Class II |
| 21 CFR 868.1400 | Carbon Dioxide Gas Analyzer | Class II |
| 21 CFR 868.1720 | Oxygen Gas Analyzer | Class II |
{1}------------------------------------------------
062194
## Legally Marketed Predicate Device:
Agilent Technologies Information Center (AIC) Software K011093
#### Description:
Central Monitoring System is a kind of medical information system widely applied in clinical monitoring, whose operation requires specific hardware environment such as a PC platform, the PC and Operation System is dedicated. Depending on the specific hardware platform, the Central Monitor System constitutes a monitoring network system together with software, the Operation System and work together with bedside monitors. It can collect, and export the monitoring information of the patient transferred via network from the bedside monitor. It can also display and export the integrated information from bedside monitors so as to free-up doctors from clinical monitoring work and create centralized monitoring management.
The HYPERVISOR VI Central Monitoring System is a medical information system. This System includes a high-performance PC hardware system, the application software and Windows operation system (Windows XP). Through applying Mindray specified protocol, the Central Monitoring System obtains physiological information from cleared bedside monitors manufactured by Mindray, includes: PM-8000 Express, PM-9000 Express and VS-800.
The Central Monitoring System can maximally support the simultaneous networking of 32 bedside monitors. Also, TCP/IP protocol is applied to ensure reliable transmission of network data.
The Central Monitoring System offers centralized display of physiological information of many patients who are currently monitored simultaneously. The software can support two displays simultaneously, and can display information from up to 16 bedside monitors on a display and 32 bedside monitors on two displays simultaneously. And if any of the bedside monitors gives off alarm, prompt information will be immediately displayed by means of audible /words for medical personnel's attention.
The physiological waveforms displayed on the central monitoring system include to: ECG waveforms, RESP waveforms, CO2 waveforms, PLETH waveforms, IBP waveforms, and AG waveforms. The physiological parameters displayed on the central monitoring system include: HR, RR, SpO2, PR, SYS, DIA, MEAN, TEMP, CO2, AwRR.
{2}------------------------------------------------
K062194
Central Monitoring System makes an announcement for the physiological parameters based on the bedside monitor. The software also performs notification directly for alarms from the bedside monitor and triggers technical alarms for the error and failure of itself. Moreover database technology is applied to store these alarm events for the convenience of doctor's review and analysis.
#### Statement of intended Use:
The Central Monitoring System is intended to conduct centralized monitoring of adult, pediatric and neonatal vital sign information from multiple monitors in hospitals or medical institutions, the monitoring parameters include ECG, NIBP, SpO2. PR, RESP, IBP, TEMP, CO2 and AG. It is not intended for home use.
#### Comparison of Technological Characteristics:
The HYPERVISOR VI Central Monitoring System is substantially equivalent to currently marketed predicate device. Both HYPERVISOR VI Central Monitoring System and the M3150 predicate device free-up doctors from clinical monitoring work and create centralized monitoring management.
The notable difference between the technical specifications of the HYPERVISOR VI and M3150 is shown as following:
- 1. M3150 predicate can be used for ST segment and arrhythmia analysis, while the Central Monitoring System can not.
- 2. M3150 can be used for Telemetry and wireless monitoring, while Central Monitoring System can not;
- 3. The PC basic requirement is different, such as the operation system for Central Monitoring System is Windows XP professional, while M3150 is Windows NT4.0; the host CPU of Central Monitoring System is PIV 2.0 G, while the M3150 is PII 400+MHz;
- 4. For Central Monitoring System, the maximum number of connecting bedside monitors is 32, while for M3150, the number is 16.
### Conclusion:
The conclusions drawn from testing and validation of the HYPERVISOR VI Central Monitoring System demonstrates that the device is as safe, as effective, and performs as well as the legally marketed predicate device, the M3150 Information Center (K#011093, by Agilent Co., Ltd).
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing the body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
#### Public Health Service
ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# APR 1 1 2007
Shenzhen Mindray Bio-Medical Electronics Co. Ltd. c/o Ms. Susan D. Goldstein-Falk Official Correspondent mdi Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, New York 11021
Re: K062194
> Trade/Device Name: Hypervisor VI Central Monitoring System Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm) Regulatory Class: Class II (two) Product Code: MHX Dated: March 9, 2007 Received: March 12, 2007
Dear Ms. Goldstein-Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
#### Page 2 - Ms. Susan D. Goldstein-Falk
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Blymmma for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Attachment #
Exhibit #E
510(k) Number (if known): K062194 Page 1 of 1
Device Name: HYPERVISOR VI Central Monitoring System Indications For Use:
The Central Monitoring System is intended to conduct centralized monitoring of adult, pediatric and neonatal vital sign information from multiple monitors in hospitals or medical institutions, the monitoring parameters include ECG, NIBP, SpO2, PR, RESP, IBP, TEMP, CO2 and AG. It is not intended for home use.
Prescription Use X (Per 21 CFR 801 Subpart D) Over-The Counter Use OR (21 CFR 807 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Blummenauer
ular Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.