DIMENSION VISTA SYSTEM DRUGS OF ABUSE POSITIVE AND NEGATIVE CONTROLS, MODELS KC515 AND KC516
K062191 · Dade Behring, Inc. · DIF · Aug 21, 2006 · Clinical Toxicology
Device Facts
Record ID
K062191
Device Name
DIMENSION VISTA SYSTEM DRUGS OF ABUSE POSITIVE AND NEGATIVE CONTROLS, MODELS KC515 AND KC516
Applicant
Dade Behring, Inc.
Product Code
DIF · Clinical Toxicology
Decision Date
Aug 21, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3280
Device Class
Class 1
Indications for Use
The Dimension Vista™ System Drugs of Abuse Positive Control is an in vitro diagnostic product intended as quality control product for the Amphetamines /Methamphetamines (AMPH), Barbiturates (BARB), Benzodiazepines (BENZ), Cocaine Metabolite (COC), Methadone (METH), Opiates (OPI), Phencyclidine (PCP), and Cannabinoids (THC) methods on the Dimension Vista™ System. The Dimension Vista™ System Drugs of Abuse Negative Control is an in vitro diagnostic product intended as quality control product for the Amphetamines /Methamphetamines (AMPH), Barbiturates (BARB), Benzodiazepines (BENZ), Cocaine Metabolite (COC), Methadone (METH), Opiates (OPI), Phencyclidine (PCP), and Cannabinoids (THC) methods on the Dimension Vista™ System.
Device Story
Liquid, human urine-based quality control materials; used to monitor performance of drugs of abuse assays on Dimension Vista System. Positive control contains analytes above assay cutoff; negative control contains analytes below cutoff. Ready-to-use vials; no preparation required. Healthcare providers use these controls to verify assay accuracy and precision during routine clinical toxicology testing. Values assigned via GC/MS reference testing.
Clinical Evidence
No clinical data. Bench testing only. Value assignment performed via GC/MS reference testing; stability protocols and acceptance criteria were evaluated and found acceptable.
Technological Characteristics
Liquid, human urine-based multi-analyte control. Constituents: D-methamphetamine, Secobarbital, Nordiazepam, Benzoylecgonine, Methadone, Morphine, Phencyclidine, 11-nor-Δ⁹-THC-9-COOH. Traceable to GC/MS reference testing. Designed for use on Dimension Vista System.
Indications for Use
Indicated for use as quality control material for drugs of abuse testing (AMPH, BARB, BENZ, COC, METH, OPI, PCP, THC) on the Dimension Vista System. For prescription use.
Regulatory Classification
Identification
A clinical toxicology control material is a device intended to provide an estimation of the precision of a device test system and to detect and monitor systematic deviations from accuracy resulting from reagent or instrument defects. This generic type of device includes various single, and multi-analyte control materials.
Predicate Devices
Dimension® Drugs of Abuse Positive Control (k950138)
Dimension® Drugs of Abuse Negative Control (k950138)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k062191
B. Purpose for Submission:
510(k) premarket notification package to manufacture and market the Dimension Vista™ System Drugs of Abuse Positive Control (UDAT CON(+)-KC515) and Negative Control (UDAT CON(-)-KC516).
C. Measurand:
Control Materials for Amphetamine/Methamphetamine, Barbiturate, Benzodiazepines, Cocaine Metabolite, Methadone, Opiates, Phencyclidine, and Cannabinoids.
D. Type of Test:
Control Materials
E. Applicant:
Dade Behring, Inc.
F. Proprietary and Established Names:
Dimension Vista™ Drugs of Abuse Positive Control
Dimension Vista™ Drugs of Abuse Negative Control
G. Regulatory Information:
1. Regulation section:
862.3280 - Clinical toxicology control material
2. Classification:
Class I
3. Product code:
{1}
Drug mixture control materials (DIF)
4. Panel:
Toxicology (91)
# H. Intended Use:
1. Intended use(s):
See Indication(s) for use below.
2. Indication(s) for use:
The Dimension Vista™ System Drugs of Abuse Positive Control is an in vitro diagnostic product intended as quality control product for the Amphetamines /Methamphetamines (AMPH), Barbiturates (BARB), Benzodiazepines (BENZ), Cocaine Metabolite (COC), Methadone (METH), Opiates (OPI), Phencyclidine (PCP), and Cannabinoids (THC) methods on the Dimension Vista™ System.
The Dimension Vista™ System Drugs of Abuse Negative Control is an in vitro diagnostic product intended as quality control product for the Amphetamines /Methamphetamines (AMPH), Barbiturates (BARB), Benzodiazepines (BENZ), Cocaine Metabolite (COC), Methadone (METH), Opiates (OPI), Phencyclidine (PCP), and Cannabinoids (THC) methods on the Dimension Vista™ System.
3. Special conditions for use statement(s):
For prescription use
4. Special instrument requirements:
Dimension Vista System
# I. Device Description:
The Dimension Vista™ System Drugs of Abuse Positive Control is a liquid, multi-analyte, human urine-based product for the following constituents at above the cutoff levels of the methods:
| Analyte | Constituent |
| --- | --- |
| Amphetamine/Methamphetamine | D-methamphetamine |
| Barbiturate | Secobarbital |
| Benzodiazepines | Nordiazepam |
| Cocaine Metabolite | Benzoylecgonine |
| Methadone | Methadone |
{2}
3
| Analyte | Constituent |
| --- | --- |
| Opiates | Morphine |
| Phencyclidine | Phencyclidine |
| Cannabinoids | 11-nor-Δ⁹-THC-9 –COOH |
The kit consists of six vials which are ready for use (no preparation is required). The volume per vial is 2.5 mL.
The Dimension Vista™ System Drugs of Abuse Negative Control is a liquid, multi-analyte, human urine-based product containing the following constituents below the cutoff levels of the methods:
| Analyte | Constituent |
| --- | --- |
| Amphetamine/Methamphetamine | D-methamphetamine |
| Barbiturate | Secobarbital |
| Benzodiazepines | Nordiazepam |
| Cocaine Metabolite | Benzoylecgonine |
| Methadone | Methadone |
| Opiates | Morphine |
| Phencyclidine | Phencyclidine |
| Cannabinoids | 11-nor-Δ⁹-THC-9 –COOH |
The kit consists of six vials which are ready for use (no preparation is required). The volume per vial is 2.5 mL
## J. Substantial Equivalence Information:
1. Predicate device name(s):
Dimension® Drugs of Abuse Positive Control
Dimension® Drugs of Abuse Negative Control
2. Predicate 510(k) number(s):
k950138
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Analytes | Amphetamines /Methamphetamines (AMPH), Barbiturates | Amphetamines /Methamphetamines (AMPH), Barbiturates |
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | (BARB), Benzodiazepines (BENZ), Cocaine Metabolite (COC), Methadone (METH), Opiates (OPI), Phencyclidine (PCP), and Cannabinoids (THC). | (BARB), Benzodiazepines (BENZ), Cocaine Metabolite (COC), Methadone (METH), Opiates (OPI), Phencyclidine (PCP), and Cannabinoids (THC). |
| Form | Liquid | Liquid |
| Traceability | GC/MS | GC/MS |
| Matrix | Human urine based | Human urine based |
| Number of levels | One positive, one negative | One positive, one negative |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended use | The Dimension Vista™ System Drugs of Abuse Control is an in vitro diagnostic product intended as quality control product for the Amphetamines /Methamphetamines (AMPH), Barbiturates (BARB), Benzodiazepines (BENZ), Cocaine Metabolite (COC), Methadone (METH), Opiates (OPI), Phencyclidine (PCP), and Cannabinoids (THC) methods on the Dimension VistaTM System. | The Drugs of Abuse Positive Control is an in vitro diagnostic product intended as quality control product for the following methods packaged in Flex® reagent cartridges: Amphetamines /Methamphetamines (AMPH), Barbiturates (BARB), Benzodiazepines (BENZ), Cocaine Metabolite (COC), Methadone (METH), Opiates (OPI), Phencyclidine (PCP), and Cannabinoids (THC). |
# K. Standard/Guidance Document Referenced (if applicable):
FDA Guidance for Industry - Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators; Final
FDA Guidance for Industry and FDA Staff - Use of Symbols on Labels and in Labeling of In Vitro Diagnostic Devices Intended for Professional Use
FDA Guidance for Industry - Points to Consider Guidance Document on
{4}
Assayed and Unassayed Quality Control Material
FDA Guidance for Industry and FDA Staff – Bundling Multiple Devices or Multiple Indications in a Single Submission; Draft
CEN 13640 Stability testing of In-Vitro Diagnostic Devices
ISO 14971:2000 Medical devices - Application of risk management to medical devices
L. Test Principle:
Not Applicable
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility: Not Applicable
b. Linearity/assay reportable range: Not Applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability
The assigned values of the Dimension Vista™ System Drugs of Abuse Positive Control and the Dimension Vista™ System Drugs of Abuse Negative Control are traceable to Gas Chromatography / Mass Spectrometry (GC/MS) Reference Testing.
Value Assignment
Calculated quantities of D-Methamphetamine, Secobarbital, Nordiazepam, Benzoylecgonine, Methadone, Morphine, Phencyclidine, and 11-nor-Δ⁹-THC-9-COOH high purity stock solutions are added to drug free normal human urine to target concentrations for each of the Negative and Positive Controls. The Negative and Positive Bulk Products are tested by GC/MS and any analyte that is out of the acceptable GC/MS range is adjusted and retested. Values are assigned to the controls once the GC/MS results are in the acceptable ranges.
Stability
Protocols and acceptance criteria were described and found to be acceptable.
d. Detection limit:
5
{5}
Not Applicable
e. Analytical specificity:
Not Applicable
f. Assay cut-off:
Not Applicable
2. Comparison studies:
a. Method comparison with predicate device:
Not Applicable
b. Matrix comparison:
Not Applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not Applicable
4. Clinical cut-off:
Not Applicable
5. Expected values/Reference range:
Not Applicable
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.