← Product Code [JWY](/productcode/JWY) · K062190

# SENSITITRE HAEMOPHILUS / STREPTOCOCCUS PNEUMONIAE (HP) MIC PLATES (K062190)

_Trek Diagnostic Systems, Inc. · JWY · Aug 31, 2006 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K062190

## Device Facts

- **Applicant:** Trek Diagnostic Systems, Inc.
- **Product Code:** [JWY](/productcode/JWY.md)
- **Decision Date:** Aug 31, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1640
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

The Sensititre HP Susceptibility System is an in vitro diagnostic product for clinical susceptibility testing of fastidious organisms. This 510(k) is for the addition of Tigecycline in the dilution range of 0.004 - 8 ug/ml for testing Streptococcus spp. isolates on the Sensititre HP susceptibility system. The approved primary "Indications for Use" and clinical significance of Tigecycline is for: Aerobic facultative Gram-postive microorganisms Streptoccus agalactiae Streptococcus pyogenes Streptococcus anginosus grp. (includes S. anginosus, S. intermedius, and . S. constellatus)

## Device Story

Sensititre HP Susceptibility System is an in vitro diagnostic device used in clinical laboratories to determine the Minimum Inhibitory Concentration (MIC) of antimicrobial agents against fastidious organisms. This specific submission adds Tigecycline (0.004-8 µg/ml) to the existing plate format for testing Streptococcus spp. The system utilizes microbroth dilution; organisms are inoculated into wells containing varying concentrations of the antibiotic. After incubation, growth is assessed to determine the MIC. Results assist clinicians in selecting appropriate antibiotic therapy for patients with bacterial infections. The device provides standardized, quantitative susceptibility data, aiding in the management of infections caused by aerobic facultative Gram-positive microorganisms.

## Clinical Evidence

Bench testing only. Performance was evaluated by comparing the Sensititre HP system results for Tigecycline against a reference broth microdilution method. Testing included clinical isolates of Streptococcus spp. (S. agalactiae, S. pyogenes, S. anginosus group). Results demonstrated acceptable essential agreement and categorical agreement between the subject device and the reference method.

## Technological Characteristics

Microbroth dilution susceptibility plate. Contains dehydrated antimicrobial agent (Tigecycline) in specific concentration ranges. Designed for use with the Sensititre automated or manual reading systems. Standardized for in vitro diagnostic use in clinical microbiology laboratories.

## Regulatory Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

## Predicate Devices

- Sensititre® Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC susceptibility plates ([K062190](/device/K062190.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Cynthia C. Knapp Director Lab Services TREK Diagnostic Systems, Inc. 982 Keynote Circle, Suite 6 Brooklyn Heights, Ohio 44145

AUG 3 1 2006

k062190 Re:

Trade/Device Name: Sensititre® Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC susceptibility plates, for Tigecycline 0.004-8 µg/ml for Streptococcus spp.

Regulation Number: 21 CFR § 866.1640 Regulation Name: Antimicrobial Susceptibility Test Powder Regulatory Class: II Product Code: JWY, LRG Dated: July 26, 2006 Received: July 31, 2006

Dear Ms. Knapp:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{1}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Sally, artton

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known): K 062190

Device Name: Sensititre® Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC susceptibility plates, for Tigecycline 0.004-8μg/ml for Streptococcus spp.

Indications For Use:

The Sensititre HP Susceptibility System is an in vitro diagnostic product for clinical susceptibility testing of fastidious organisms.

This 510(k) is for the addition of Tigecycline in the dilution range of 0.004 - 8 ug/ml for testing Streptococcus spp. isolates on the Sensititre HP susceptibility system. The approved primary "Indications for Use" and clinical significance of Tigecycline is for:

Aerobic facultative Gram-postive microorganisms

Streptoccus agalactiae Streptococcus pyogenes Streptococcus anginosus grp. (includes S. anginosus, S. intermedius, and . S. constellatus)

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Freddie thi. Poole
Division Sign-Off

**Division Sign-Off**

Office of In Vitro Diagnostic Device Evaluation and Safety

100k)

.

---

**Source:** [https://fda-staging.innolitics.com/device/K062190](https://fda-staging.innolitics.com/device/K062190)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).