← Product Code [PSZ](/productcode/PSZ) · K062109

# BINAXNOW INFLUENZA A & B TEST (K062109)

_Binax, Inc. · PSZ · Nov 9, 2006 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K062109

## Device Facts

- **Applicant:** Binax, Inc.
- **Product Code:** [PSZ](/productcode/PSZ.md)
- **Decision Date:** Nov 9, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3328
- **Device Class:** Class 2
- **Review Panel:** Microbiology
- **Attributes:** Real-World Evidence, Pediatric

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K062109 · Nov 9, 2006 | BINAXNOW INFLUENZA A & B TEST | Binax, Inc. | Archived clinical samples from physician offices, clinics, and hospitals | Retrospective clinical samples were used to evaluate the performance (sensitivity and specificity) of the BinaxNOW Influenza A & B Test compared to culture/DFA, specifically supporting the use of nasal wash/aspirate specimens. | Retrospective study; Archived clinical samples; Clinical performance |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Retrospective Study; Retrospective evaluation of frozen clinical samples | Symptomatic patients; Sample Size: 293; Number of Sites: Multiple (physician offices, clinics, and hospitals in US and Sweden) | Culture/DFA | Sensitivity and specificity for detection of Flu A and Flu B |

## Indications for Use

The BinaxNOW® Influenza A & B Test is an in vitro immunochromatographic assay for the qualitative detection of influenza A and B nucleoprotein antigens in nasopharyngeal (NP) swab, nasal swab, and nasal wash/aspirate specimens. It is intended to aid in the rapid differential diagnosis of influenza A and B viral infections. Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other management decision.

## Device Story

Lateral flow immunochromatographic assay; detects influenza A and B nucleoprotein antigens in nasopharyngeal swabs, nasal swabs, and nasal wash/aspirate specimens. Device provides qualitative visual results to aid differential diagnosis of influenza infections. Used in clinical settings (physician offices, clinics, hospitals); operated by healthcare professionals. Output is visual interpretation of test lines; assists clinicians in patient management decisions. Benefits include rapid identification of viral infection.

## Clinical Evidence

Prospective study (n=1183) and retrospective study (n=293) compared device to cell culture/DFA. Prospective study: 52% pediatric, 48% adult. Flu A sensitivity ranged 75-83% and specificity 93-99% depending on sample type. Flu B sensitivity ranged 43-53% and specificity 94-100%. No performance differences observed based on age, gender, or sample type.

## Technological Characteristics

Lateral flow immunochromatographic assay; utilizes monoclonal antibodies immobilized on a membrane support as three distinct lines. Form factor is a book-shaped hinged cardboard test device. No external energy source required. Manual interpretation of visual lines.

## Regulatory Identification

An influenza virus antigen detection test system is a device intended for the qualitative detection of influenza viral antigens directly from clinical specimens in patients with signs and symptoms of respiratory infection. The test aids in the diagnosis of influenza infection and provides epidemiological information on influenza. Due to the propensity of the virus to mutate, new strains emerge over time which may potentially affect the performance of these devices. Because influenza is highly contagious and may lead to an acute respiratory tract infection causing severe illness and even death, the accuracy of these devices has serious public health implications.

## Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) The device's sensitivity and specificity performance characteristics or positive percent agreement and negative percent agreement, for each specimen type claimed in the intended use of the device, must meet one of the following two minimum clinical performance criteria:
(i) For devices evaluated as compared to an FDA-cleared nucleic acid based-test or other currently appropriate and FDA accepted comparator method other than correctly performed viral culture method:
(A) The positive percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent.
(B) The negative percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent.
(ii) For devices evaluated as compared to correctly performed viral culture method as the comparator method:
(A) The sensitivity estimate for the device when testing for influenza A must be at the point estimate of at least 90 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 80 percent. The sensitivity estimate for the device when testing for influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent.
(B) The specificity estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent.
(2) When performing testing to demonstrate the device meets the requirements in paragraph (b)(1) of this section, a currently appropriate and FDA accepted comparator method must be used to establish assay performance in clinical studies.
(3) Annual analytical reactivity testing of the device must be performed with contemporary influenza strains. This annual analytical reactivity testing must meet the following criteria:
(i) The appropriate strains to be tested will be identified by FDA in consultation with the Centers for Disease Control and Prevention (CDC) and sourced from CDC or an FDA-designated source. If the annual strains are not available from CDC, FDA will identify an alternative source for obtaining the requisite strains.
(ii) The testing must be conducted according to a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(iii) By July 31 of each calendar year, the results of the last 3 years of annual analytical reactivity testing must be included as part of the device's labeling. If a device has not been on the market long enough for 3 years of annual analytical reactivity testing to have been conducted since the device received marketing authorization from FDA, then the results of every annual analytical reactivity testing since the device received marketing authorization from FDA must be included. The results must be presented as part of the device's labeling in a tabular format, which includes the detailed information for each virus tested as described in the certificate of authentication, either by:
(A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where the analytical reactivity testing data can be found; or
(B) In the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.
(4) If one of the actions listed at section 564(b)(1)(A)-(D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services (HHS) determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate. The procedure and location of testing may depend on the nature of the emerging virus.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.

## Predicate Devices

- BinaxNOW Influenza A & B ([K041049](/device/K041049.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE

A. 510(k) Number: K062109

B. Purpose for Submission:
Expand the Indications for Use claim to add nasal swabs specimens, modify cautions statement, include 2 additional influenza A strains in the Analytical Reactivity claim, support use of additional transport media and update the labeling in compliance with FDA Guidance

C. Analyte:
Influenza Type A nucleoprotein antigens

D. Type of Test:
Lateral flow immunochromatographic assay

E. Applicant:
Binax, Inc.

F. Proprietary and Established Names:
BinaxNOW Influenza A &amp; B

G. Regulatory Information:
1. Regulation section:
21 CFR Part 866.3330
2. Classification:
Antigens, CF (including CF Control), Influenza virus A, B, C
3. Product Code:
GNX
4. Panel:
83 Microbiology

H. Intended Use:
1. Intended use(s):
The BinaxNOW® Influenza A &amp; B Test is an in vitro immunochromatographic assay for the qualitative detection of influenza A and B nucleoprotein antigens in nasopharyngeal (NP) swab, nasal swab, and nasal wash/aspirate specimens. It is intended to aid in the rapid differential diagnosis of influenza A and B viral infections. Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other management decision.

{1}

Page 2 of 7

Caution: Assay sensitivity for nasal wash/aspirate samples was determined primarily using archived specimens.
Users may wish to establish the sensitivity of these specimens on fresh samples.

2. Indication(s) for use: NA
3. Special condition for use statement(s): Prescription use only
4. Special instrument Requirements: NA

I. Device Description:

See: http://www.fda.gov/cdrh/reviews/K041049.pdf

J. Substantial Equivalence Information:

Predicate device name(s): BinaxNOW Influenza A &amp; B

1. Predicate K number(s): K041049
2. Comparison with predicate:

A total of 1183 prospective specimens collected from children (less than 18 years of age) and adults (18 years or older) were evaluated in the BinaxNOW® Influenza A &amp; B Test and compared to culture/DFA. Evaluated specimens include nasopharyngeal, nasal, and throat swabs and nasal wash/aspirates collected from patients presenting with influenza-like symptoms. Forty-three percent (43%) of the population tested was male, 57% female, 52% pediatric (&lt; 18 years), and 48% adult (≥ 18 years). No differences in test performance were observed based on patient age or gender. A/H3 and A/H1 were the predominant influenza subtypes observed during this time.

K. Standard/Guidance Document Referenced (if applicable): NA

L. Test Principle:

See: http://www.fda.gov/cdrh/reviews/K041049.pdf

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

{2}

Page 3 of 7

A blind study of the BinaxNOW® Influenza A &amp; B Test was conducted at 3 separate sites using panels of blind coded specimens containing negative, low positive, and moderate positive samples. Participants tested each sample multiple times on 3 different days. There was 96.8% (242/250) agreement with expected test results, with no significant differences within run (replicates tested by one operator), between run (3 different days), between sites (3 sites), or between operators (6 operators).

b. Linearity/assay reportable range: NA

c. Traceability, Stability, Expected values (controls, calibrators, or method): NA

d. Detection limit:

The BinaxNOW® test limit of detection (LoD), defined as the concentration of influenza virus that produces positive BinaxNOW® test results approximately 95% of the time, was identified by evaluating different concentrations of inactivated Flu A/Beijing and inactivated Flu B/Harbin in the BinaxNOW® test.

Twelve (12) different operators each interpreted 2 devices run at each concentration for a total of 24 determinations per level. The following results identify a concentration of 1.03 × 10² ng/ml as the LoD for Flu A/Beijing and 6.05 × 10¹ ng/ml for Flu B/Harbin.

e. Assay cut-off:

NA

Comparison studies:

f. Method comparison with predicate device:

See J.3 above

g. Matrix comparison: NA

2. Clinical studies:

BinaxNOW® Influenza A &amp; B Test Performance vs. Cell Culture / DFA – Prospective Study

A total of 1183 prospective specimens collected from children (less than 18 years of age) and adults (18 years or older) were evaluated in the BinaxNOW® Influenza A &amp; B Test and compared to culture/DFA. Evaluated specimens include nasopharyngeal, nasal, and throat swabs and nasal wash/aspirates

{3}

Page 4 of 7

collected from patients presenting with influenza-like symptoms. Forty-three percent (43%) of the population tested was male, 57% female, 52% pediatric (&lt; 18 years), and 48% adult (≥ 18 years). No differences in test performance were observed based on patient age or gender. A/H3 and A/H1 were the predominant influenza subtypes observed during this time.

BinaxNOW® A &amp; B Test performance by sample type versus cell culture/DFA, including 95% confidence intervals, is listed below.

BinaxNOW® Influenza A &amp; B Test Performance vs. Cell Culture/DFA for Detection of Flu A

|   | Test Sensitivity |   |   |   | Test Specificity  |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Sample | +/+ | -/+ | % Sens | 95% CI | -/- | +/- | % Spec | 95% CI  |
|  NP Swab | 53 | 16 | 77% | 65-86% | 278 | 3 | 99% | 97-100%  |
|  Nasal Swab | 85 | 17 | 83% | 74-90% | 378 | 16 | 96% | 93-98%  |
|  Overall | 162 | 53 | 75% | 69-81% | 947 | 21 | 98% | 97-99%  |

BinaxNOW® Influenza A &amp; B Test Performance vs. Cell Culture/DFA for Detection of Flu B

|   | Test Sensitivity |   |   |   | Test Specificity  |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Sample | +/+ | -/+ | % Sens | 95% CI | -/- | +/- | % Spec | 95% CI  |
|  NP Swab | 2 | 2 | 50% | 9-91% | 346 | 0 | 100% | 99-100%  |
|  Nasal Swab | 9 | 4 | 69% | 39-90% | 481 | 2 | 100% | 98-100%  |
|  Overall | 13 | 17 | 43% | 26-62% | 1150 | 3 | 100% | 99-100%  |

# BinaxNOW® Influenza A &amp; B Test Performance vs. Cell Culture / DFA – Retrospective Study

A total of 293 retrospective frozen clinical samples were evaluated in the BinaxNOW® Influenza A &amp; B Test and compared to culture/DFA. All clinical samples were collected from symptomatic patients at multiple physician offices, clinics and hospitals located in the Southern, Northeastern and Midwestern regions of the United States and from one hospital in Sweden. Fifty-three percent (53%) of the population tested was male, 47% female, 62% pediatric (&lt;18 years) and 38% adult (≥ 18 years). Nasal wash/aspirate specimens comprised approximately 61% of the samples tested, while NP swabs represented 39%. No differences in test performance were observed based on patient age and gender or based on sample type tested.

BinaxNOW® A &amp; B Test performance by sample type versus cell culture/DFA, including 95% confidence intervals, is listed below.

{4}

Page 5 of 7

BinaxNOW® Influenza A &amp; B Test Performance vs. Cell Culture/DFA for Detection of Flu A

|   | Test Sensitivity |   |   |   | Test Specificity  |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Sample | +/+ | -/+ | % Sens | 95% CI | -/- | +/- | % Spec | 95% CI  |
|  NP Swab | 19 | 8 | 70% | 50-86% | 77 | 9 | 90% | 81-95%  |
|  Wash/Aspirate | 51 | 6 | 89% | 78-96% | 117 | 6 | 95% | 89-98%  |
|  Overall | 70 | 14 | 83% | 73-90% | 194 | 15 | 93% | 88-96%  |

BinaxNOW® Influenza A &amp; B Test Performance vs. Cell Culture/DFA for Detection of Flu B

|   | Test Sensitivity |   |   |   | Test Specificity  |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Sample | +/+ | -/+ | % Sens | 95% CI | -/- | +/- | % Spec | 95% CI  |
|  NP Swab | 0 | 0 | N/A | N/A | 111 | 2 | 98% | 93-100%  |
|  Wash/Aspirate | 8 | 7 | 53% | 27-78% | 155 | 10 | 94% | 89-97%  |
|  Overall | 8 | 7 | 53% | 27-78% | 266 | 12 | 96% | 92-98%  |

## Analytical Sensitivity:

The BinaxNOW® test limit of detection (LOD), defined as the concentration of influenza virus that produces positive BinaxNOW® test results approximately 95% of the time, was identified by evaluating different concentrations of inactivated Flu A/Beijing and inactivated Flu B/Harbin in the BinaxNOW® test.

Twelve (12) different operators each interpreted 2 devices run at each concentration for a total of 24 determinations per level. The following results identify a concentration of $1.03 \times 10^{2} \, \mathrm{ng/ml}$ as the LOD for Flu A/Beijing and $6.05 \times 10^{1} \, \mathrm{ng/ml}$ for Flu B/Harbin.

|  Influenza A/Beijing  |   |   |
| --- | --- | --- |
|  Concentration (ng/ml) | # Detected | % Detected  |
|  1.03 x 10² (LOD) | 23/24 | 96  |
|  5.60 x 10¹ (Cut-off) | * | 50  |
|  Influenza B/Harbin  |   |   |
| --- | --- | --- |
|  Concentration (ng/ml) | # Detected | % Detected  |
|  6.05 x 10¹ (LOD) | 23/24 | 96  |
|  2.42 x 10¹ (Cut-off) | 11/24 | 46  |

{5}

Page 6 of 7

|  3.27 x 10^{1} (High Neg) | 4/24 | 17  |
| --- | --- | --- |
|  True Negative | 0/24 | 0  |
|  1.51 x 10^{1} (High Neg) | 6/24 | 25  |
| --- | --- | --- |
|  True Negative | 0/24 | 0  |

*Linear regression was used to calculate a line equation, which was then used to project the cutoff concentration of Flu A/Beijing.

a. Other clinical supportive data (when a and b are not applicable):

3. Clinical cut-off: NA

4. Expected values/Reference range:

The prevalence of influenza varies from year to year, with outbreaks typically occurring during the fall and winter months. The rate of positivity found in influenza testing is dependent on many factors including the method of specimen collection, the test method used, geographic location, and the disease prevalence in specific localities. Type A viruses are typically associated with most serious influenza epidemics, while Type B are typically milder. In multi-center clinical studies conducted by Binax outside the U.S. during the 2004 respiratory season and in the US during the 2004-2005 respiratory season, the average prevalence of influenza A (as determined by viral cell culture) was 18%. The average prevalence of influenza B was 3%.

{6}

Page 7 of 7

N. Conclusion:
The submitted material in this premarket notification is complete and supports a substantial equivalence decision

---

**Source:** [https://fda-staging.innolitics.com/device/K062109](https://fda-staging.innolitics.com/device/K062109)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
