EDIA ANTI-CCP
Device Facts
| Record ID | K062045 |
|---|---|
| Device Name | EDIA ANTI-CCP |
| Applicant | Euro-Diagnostica AB |
| Product Code | NHX · Immunology |
| Decision Date | Dec 4, 2006 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 866.5775 |
| Device Class | Class 2 |
| Attributes | Real-World Evidence |
Real-World Evidence
| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
|---|---|---|---|---|---|
| K062045 · Dec 4, 2006 | EDIA ANTI-CCP | Euro-Diagnostica AB | Retrospective clinical serum samples from RA patients; Retrospective clinical serum samples from non-RA disease patients; Retrospective clinical serum samples from healthy blood donors | Retrospective clinical samples were used to determine the assay cut-off, perform ROC analysis, and establish clinical sensitivity and specificity. | Retrospective clinical samples; Clinical sensitivity; Clinical specificity; Cut-off determination |
Clinical Evidence
| Study Design | Population | Comparator | Key Endpoints |
|---|---|---|---|
| Retrospective analysis of clinical samples | 416 RA patients, 531 non-RA disease patients, and 262 healthy controls; Sample Size: 1209 | Not applicable for this study | Assay cut-off determination via ROC analysis |
| Retrospective method comparison | 416 RA patients and 262 healthy blood donors; Sample Size: 678 | Euro-Diagnostica Immunoscan RA anti-CCP (predicate) | Positive/Negative percent agreement |
Indications for Use
The EDIA™ anti-CCP test kit is an enzyme-linked immunosorbent assay (ELISA) for detection and semi-quantitation of IgG antibodies to Cyclic Citrullinated Peptides (CCP) in human sera and plasma. The assay is used to detect antibodies in a single specimen. The results of the assay are to be used as an aid to the diagnosis of Rheumatoid Arthritis (RA), in conjunction with other laboratory and clinical findings. The analysis should be performed by trained laboratory professionals. For in vitro diagnostic use.
Device Story
EDIA™ anti-CCP is an ELISA test kit for detecting IgG antibodies to Cyclic Citrullinated Peptides in human serum or plasma. Input: patient serum/plasma specimen. Principle: wells coated with CCP antigen capture specific antibodies; alkaline phosphatase-labeled anti-human IgG conjugate binds captured antibodies; substrate solution added to produce color intensity proportional to antibody concentration. Output: absorbance (optical density) measured and calculated against a calibrator curve to provide results in arbitrary units (U/mL). Used in clinical laboratories by trained professionals. Results aid physicians in diagnosing Rheumatoid Arthritis when combined with other clinical and laboratory findings.
Clinical Evidence
Clinical performance was evaluated using 416 RA patients and 793 non-RA controls (including healthy donors and various autoimmune/infectious disease states). Clinical sensitivity was 76% (95% CI: 72.1%-80.3%). Clinical specificity was 98.6% (782/793). Method comparison against the predicate device (n=678) showed 98.4% positive percent agreement and 99.4% negative percent agreement, with an overall agreement of 99.1%.
Technological Characteristics
ELISA-based immunoassay. Wells coated with Cyclic Citrullinated Peptides. Detection via alkaline phosphatase-labeled anti-human IgG conjugate and substrate solution. Measurement via optical density (absorbance). Standalone laboratory test kit. No specific software algorithm or connectivity described; manual or automated plate reader processing.
Indications for Use
Indicated for the detection and semi-quantitation of IgG antibodies to Cyclic Citrullinated Peptides (CCP) in human serum and plasma specimens to aid in the diagnosis of Rheumatoid Arthritis (RA). For use by trained laboratory professionals.
Regulatory Classification
Identification
A rheumatoid factor immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the rheumatoid factor (antibodies to immunoglobulins) in serum, other body fluids, and tissues. Measurement of rheumatoid factor may aid in the diagnosis of rheumatoid arthritis.
Predicate Devices
- Immunoscan RA anti-CCP test kit