K062016 · Choice Smart Health Care Co. , Ltd. · FLL · Oct 27, 2006 · General Hospital
Device Facts
Record ID
K062016
Device Name
RAPID DIGITAL THERMOMETER, MODEL RDT -18 SERIES
Applicant
Choice Smart Health Care Co. , Ltd.
Product Code
FLL · General Hospital
Decision Date
Oct 27, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The Rapid Digital Thermometer is an electronic thermometer used to measure body temperature in oral, axillaries (underarm use), and rectal. It is an electronic thermometer used for clinical temperature measurement. It is intended for use at home and hospital environment for both children and adult.
Device Story
Rapid Digital Thermometer (RDT-18-XY) uses a thermistor sensor to measure body temperature. Powered by a 1.5V button battery; device captures thermal signals and converts them to digital temperature readings. Used in home and hospital environments by patients or clinicians. Output displayed on device screen; assists in monitoring patient health status and fever detection.
Clinical Evidence
Bench testing only. Performance validated against ASTM E1112 standards including temperature range, accuracy, and resolution tests. Biocompatibility testing conducted per ISO 10993-1:2003 (skin irritation, in vitro cytotoxicity, sensitivity). Electrical safety and EMC testing performed per IEC 60601-1 and IEC 60601-1-2.
Technological Characteristics
Thermistor-based electronic thermometer. Powered by 1.5V button battery. Compliant with ASTM E1112 (performance), IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and ISO 10993-1:2003 (biocompatibility).
Indications for Use
Indicated for clinical temperature measurement in children and adults in home and hospital settings via oral, axillary, or rectal routes.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
OC1 2 7 2006
## 510(K) SUMMARY
2016 (0.152)
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92, and the relevant 510(K) submission guidance.
The assigned 510(K) number is:
#### 1. Submitter's Identifications:
Mr. Mok Chi Wing Choice Smart Health Care Company Limited. RM 1901, CC WU BUILDING 302 HENNESSY ROAD, WANCHAI Hong Kong Phone (852) 28934110 Fax: (852)28934318
Date Summary Prepared: March 15, 2006
#### 2. Name of the device:
Rapid Digital Thermometer, Model RDT-18-XY (X=0-9, Y=1-9) Classification Name: Thermometer, Electronic, Clinical
#### 3. Predicate Device Information and Substantial Equivalence: Electronic Thermometer GT010706 (K021052).
#### 4. Device Description:
The Rapid Digital Thermometer is an electronic thermometer by using a thermistor as the temperature sensor.
The thermometer uses a 1.5V button battery for operation.
#### 5. Intended Use:
The Rapid Digital Thermometer is an electronic thermometer used to measure body temperature in oral, axillaries (underarm use), and rectal.
#### 6. Comparison to the 510(k) Cleared Device (Predicate Device):
{1}------------------------------------------------
# K412416 (0.25A2)
The Rapid Digital Thermometer (RDT-18 series) has the same intended use and technological characteristics as the cleared device of Electronic Thermometer GT010706 (K021052). Although there are slight differences between the new device and the legally marketed one, these differences do not affect the safety, performance of the subject device. So the new device is substantial equivalence to the selected predicate device.
#### 7. Discussion of Non-Clinical Tests Performed Determination of Substantial Equivalence are as follows:
Both the predicted device and the Rapid Digital Thermometer (RDT-18 series) are in compliance to applicable voluntary standards. Various performance testing data which conducted according to ASTM E1112 standards, such as temperature range test, accuracy test, resolution test, cleaning test, demonstrate the same safety and effectiveness as that of cleared device.
Also both devices conform to IEC 60601-1, IEC 60601-1-2 requirements, and as well as ISO 10993-1:2003 biocompatibility testing on skin irritation, in vitro cytotoxicity and sensitivity.
Guidance documents included the "FDA Guidance on the Content of Premark Notification (510(k)) Submission for Clinical Electronic Thermometers"
#### 8. Conclusions:
The Rapid Digital Thermometer (RDT-18 series) has the same intended use and technological characteristics as the predicted device. Various performance testing data which conducted according to ASTM E1112 standards, such as temperature range test, accuracy test, resolution test, cleaning test, demonstrate the same safety and effectiveness as that of cleared device. In the other words, the Rapid Digital Thermometer is substantial equivalence to predicted device.
{2}------------------------------------------------
Public Health Service
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The eagle is facing to the right. The logo is black and white.
Food and Drug Administration 200 Corporate Boulevard Rockville MD 20850
Choice Smart Health Care Cornpany Limited C/O Mr. Marc M. Mouser Responsible Third Party Official Underwriters Laboratories, Incorporated Laboratory and Testing 2600 NW Lake Road Camas, Washington 98607-9526
OCT 2 7 2006
Re: K062016
Trade/Device Name: Rapid Digital Thermometer, Model RDT-18-XY(X=0-9, Y=1-9) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: October 10, 2006 Received: October 12, 2006
Dear Mr. Mouser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Mouser
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely rours,
Susek Rynoe
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
i6a416
### Indication for Use Statement
510(k) Number (if known):
Rapid Digital Thermometer, Model RDT-18-XY(X=0-9, Y=1-9) Device name:
Indications for Use:
It is an electronic thermometer used for clinical temperature measurement. It is intended for use at home and hospital environment for both children and adult.
Prescription Use (Per 21CFR 801.109) OR
Over-The-Counter Use Y
(PLESE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Chm Ch
n of Anesthesiology, General Hospits 1, on Control, Dental Devices
K02.016
Page18 of 69
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.