VITEK 2 AST-YS FLUCONAZOLE

K061945 · bioMerieux, Inc. · NGZ · Sep 19, 2006 · Microbiology

Device Facts

Record IDK061945
Device NameVITEK 2 AST-YS FLUCONAZOLE
ApplicantbioMerieux, Inc.
Product CodeNGZ · Microbiology
Decision DateSep 19, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1640
Device ClassClass 2

Indications for Use

VITEK® 2 Yeast Fluconazole is designed for antifungal susceptibility testing of Candida species. VITEK 2 Yeast Fluconazole is a quantitative test. It is intended for use with the VITEK 2 System as a laboratory aid in the determination of in vitro susceptibility to antifungal agents. The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 System for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, S. pneumoniae and clinically significant yeast.

Device Story

VITEK® 2 Yeast panels test Candida species susceptibility to fluconazole (1-64 µg/mL). Isolated colonies are diluted in 0.45% saline; loaded into VITEK® 2 AST test cards containing dried antifungal drugs and media. System automatically fills, seals, and incubates cards. Optical scanner measures light attenuation to monitor growth over 10-36 hours. Computer determines MIC values and interpretive categories based on growth kinetics. Used in clinical laboratories by technicians. Output provides MIC and susceptibility category (Susceptible, S-Dose Dependent, Resistant). Facilitates clinical decision-making regarding antifungal therapy.

Clinical Evidence

External evaluation conducted using fresh and stock clinical isolates and stock challenge strains. Performance compared against CLSI reference method incubated at 24 and 48 hours. Results showed 97.6% overall Essential Agreement at 24 hours and 97.1% overall Essential Agreement at 48 hours. Reproducibility and Quality Control testing yielded acceptable results.

Technological Characteristics

Miniaturized card-based system for antimicrobial susceptibility testing. Employs automated optical growth monitoring during incubation. Rehydrates antifungal media with standardized yeast isolate suspensions. Operates within the VITEK 2 System platform. Quantitative MIC determination via automated photometric detection.

Indications for Use

Indicated for the testing of Candida species for fluconazole antifungal susceptibility.

Regulatory Classification

Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # RINSTRUCTIONS FOR COMPLETING 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: K061945. B. Purpose for Submission: For testing *Candida species* for fluconazole antifungal susceptibility testing. C. Measurand: Fluconazole at 1-64 µg/mL D. Type of Test: Growth based quantitative and qualitative susceptibility test E. Applicant: bioMerieux, Inc. F. Proprietary and Established Names: Vitek®2 Yeast Fluconazole. G. Regulatory Information: 1. Regulation section: 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility Test 866.1640 Antimicrobial Susceptibility Test 2. Classification: Class II 3. Product code: NGZ Susceptibility testing -antifungal {1} 4. Panel: Microbiology (83) H. Intended Use: 1. Intended use(s): Fluconazole at 1-64 µg/mL is intended for antifungal susceptibility testing of Candida species on the Vitek® 2 Fungal Susceptibility Card for use with the Vitek® 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant yeasts to antifungal agents when used as instructed in the Online Product Information. 2. Indication(s) for use: Fluconazole at 1-64 µg/mL is indicated for the testing of Candida species. 3. Special conditions for use statement(s): Prescription use only. 4. Special instrument requirements: Not Applicable I. Device Description: VITEK® 2 Yeast panels are designed for testing Candida species. The panels contain concentrations of antifungal drugs equivalent by efficacy to standard method concentrations in µg/mL dried with microbiological media. Isolated colonies of Candida species are diluted to a standardized concentration in 0.45% saline before being used to rehydrate the antifungal drug concentration in the card. The VITEK® 2 automatically fills, seals and places the card into the incubator/reader. The VITEK® 2 monitors the growth of each well in the card over a defined period of time (up to 36 hours). The computer determines when a well demonstrates growth based on attenuation of light measured by an optical scanner. At the completion of the incubation cycle a report is generated that contains the MIC value along with the interpretive category result for each antifungal contained on the card. The final concentration in the card of the inoculum is ~10⁶ CFU/ml. A final dilution of organism can be performed using a manual procedure or an automated feature but there are no manual readings possible. J. Substantial Equivalence Information: 1. Predicate device name(s): {2} Sensititre® YeastOne 2. Predicate K number(s): K991810 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | Quantitative and qualitative susceptibility for colonies of Candida spp. | Same | | Incubation | 35°C | Same | | Inoculation | Isolated colonies of Candida spp. | Same | | Differences | | | | Item | Device | Predicate | | Technology | Automated growth based detection using attenuation of light measured by optical scanner. | Broth micro dilution – growth based with colorimetric growth indicator for manual readings | | Format | VITEK® 2 AST test card with dried antifungals | Micro tray with dried antifungals | | Medium | VITEK® 2 Yeast Base Broth | Sensititre® yeast susceptibility inoculum broth | | Time to results | 10 to 36 hours | 24-25 hours | ## K. Standard/Guidance Document Referenced (if applicable): "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA" CLSI standard M27-A2 "Reference Method for Broth Dilution Antifungal Susceptibility Testing of Yeasts; Approved Standard". ## L. Test Principle: Automated growth based detection using attenuation of light measured by an optical scanner. ## M. Performance Characteristics (if/when applicable): 1. Analytical performance: {3} a. Precision/Reproducibility: Reproducibility was conducted at three sites on ten yeast isolates with testing performed over three days in triplicate. The testing was performed using both the manual dilution mode and the automated dilution mode. The reproducibility was >95% with +/- one dilution observations. b. Linearity/assay reportable range: Not Applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): Quality Control testing was performed on each day of clinical testing on the quality control ATCC strains recommended in the CLSI standard M27. The following table represents the frequency of the results in both the reference method and the Vitek® 2 Yeast Fluconazole cards and the acceptable range. Quality control was performed on the Vitek®2 on the Vitek®2 Yeast Fluconazole card using both the manual dilution and the automated method. | ORGANISM | Conc in ug/ml | Reference | | Conc in ug/ml | Auto dilution VITEK® 2 | Manual dilution VITEK® 2 | | --- | --- | --- | --- | --- | --- | --- | | | 0.5 | 24 h | 48 h | | | | | C. krussi ATCC 6258 Range 24-h-8-64 ug/ml 48h-16-128 μg/mL | 1 | | | ≤ 1 | | | | | 2 | | | 2 | | | | | 4 | | | 4 | | | | | 8 | 29 | | 8 | 2 | 3 | | | 16 | 105 | | 16 | 50 | 67 | | | 32 | 10 | 74 | 32 | 30 | 3 | | | ≥ 64 | 2 | 82 | ≥ 64 | | 1 | | | | | | | | | | C. parapsilosis ATCC 22019 Range 0.5-4 ug/ml | 0.5 | | | | | | | | 1 | 37 | 42 | ≤ 1 | 4 | 34 | | | 2 | 116 | 121 | 2 | 76 | 39 | | | 4 | | | 4 | | | | | 8 | | | 8 | | | | | 16 | | | 16 | | | | | 32 | | | 32 | | | At least one Quality control organism was in control in the reference on all days. Quality Control results for the Vitek®2 using either inoculation dilution method demonstrated that the Vitek®2 system could produce the expected quality control results. The DensiCheck instrument was used to standardize the inoculum with instrument standardization weekly with acceptable performance at all times. {4} Verification of the DensiCheck instrument was performed prior to the study on five instruments. d. Detection limit: Not Applicable e. Analytical specificity: Not Applicable f. Assay cut-off: Not Applicable 2. Comparison studies: a. Method comparison with predicate device: Performance was established on the VITEK® 2 System for Candida spp. at multiple clinical laboratories. The CLSI reference method as described in the CLSI document M27 "Method for Broth Dilution Antifungal Susceptibility Testing of Yeasts" was compared to the Vitek® 2 result. A total of 346 isolates of Candida species were tested on the Vitek®2 using the automated inoculation dilution feature. For the evaluation, the 46 C. krusei tested were not used in the analysis since these are reported as resistant to fluconazole regardless of the MIC achieved. An additional 75 challenge isolates of Candida spp. were tested using the automated inoculum dilution method and added to the data for analysis. EA was calculated when the results for the reference method and the Vitek® 2 were within +/- two doubling dilutions of the antifungal drug. Purity checks were performed daily during the clinical study. Clinical results on the Vitek®2 were read when the instrument determined the positive control was growing sufficiently to obtain a reading. In this study this occurred as early as 10 hours and a maximum time of 31 hours with an average of 13.5 hours. The following table presents the performance of the Vitek®2 as compared to the reference method when read after both 24 hour and 48 hour readings. | | | +/- 2 dilutions | | +/- 1 dilution | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | total | # EA | % EA | # EA | % EA | %CA | #R | min | maj | vmj | | Fluconazole | | | | | | | | | | | | Candida spp. 24 h | 375 | 366 | 97.6 | 360 | 96.0 | 96.8 | 12 | 12 | 0 | 0 | | Candida spp. 48h | 375 | 364 | 97.1 | 342 | 91.2 | 86.7 | 14 | 49 | 0 | 1 | | | | | | | | | | | | | EA-Essential Agreement CA-Category Agreement R-resistant isolates maj-major discrepancies vmj-very major discrepancies min- minor discrepancies {5} An evaluation was also performed using the +/- one dilution for EA to the reference method. The CA would be the dame for either method of comparison. The 24 hour reading of the reference method agreed very well with the Vitek®2 results with either the +/- 1 dilution or the +/- 2 dilutions evaluation. There was only one additional vmj with the 48 reference readings but as shown the number of minor errors did increase. The challenge data was also performed using the manual method of inoculation preparation. The comparison between the data generated with the challenge, reproducibility and quality control using both inoculation methods was very good with no apparent trending. The no growth rate was ≤ 5%. b. Matrix comparison: Not Applicable 3. Clinical studies: a. Clinical Sensitivity: Not Applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not Applicable. 4. Clinical cut-off: Not Applicable. 5. Expected values/Reference range: The expected range for Quality Control testing and the interpretive criteria are the same as what is currently recommended and is included in the package insert. Fluconazole – interpretive criteria for Candida species. Susceptible ≤ 8 µg/mL; S- Dose Dependent 16-32 µg/mL; Resistant ≥ 64 µg/mL Fluconazole for C. krusei will be reported as resistant with the following comment generated automatically 6 {6} "Fluconazole is considered resistant to Candida krusei" **N. Proposed Labeling:** The labeling is sufficient and it satisfies the requirements of 21 CFR section 809.10. **O. Conclusion:** The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 7
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...