INDOMITABLE MAGNETIC RESONANCE IMAGING SCANNER ACCESSORY
Applicant
Fonar Corp.
Product Code
LNH · Radiology
Decision Date
Jul 28, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The Indomitable Magnetic Resonance Imaging System is indicated for use in producing images of multiple planes in the head and body. These images correspond to the distribution of hydrogen nuclei exhibiting nuclear magnetic resonance (NMR) and depend for their contrast upon NMR parameters [hydrogen nuclei concentration and flow velocity, T1 (spin-lattice relaxation time) and T2 (spin-spin relaxation time)]. As a result of the acquisition and processing of the NMR data, these images display the internal structure of the head and body, and when interpreted by a trained physician, can vield diagnostically useful information. WARNING: This device is limited by U.S. Federal law to investigational use for indications not in the indications statement. Under the requirements of the law, the non-indicated applications can be used only under an Institutional Review Board approved protocol for a non-significant risk device or an Investigational Device Exemption application approved by the FDA for a significant risk device. The procedures to be followed, under the sponsorship of FONAR Corporation, are determined by the current guidelines established by the FDA, which should provide the IRB with sufficient guidance to determine the level of risk for a MRI device. WARNING: U.S. Federal law restricts the sale, distribution and use of this device by or on the order of a physician.
Device Story
Indomitable MRI scanner uses magnetic resonance to image head and body internal structures. Modification increases gradient subsystem current limits from 200A to 300A, raising maximum gradient amplitude from 12 mT/m to 20.27 mT/m and slew rate from 19.73 mT/m/ms to 33.34 mT/m/ms. System utilizes Sympulse scanner software to control gradient current amplifiers and ensure operation within NORMAL mode safety limits. Operated by trained professionals in clinical settings; output images used for diagnostic decision-making. Benefits include improved gradient performance while maintaining patient safety regarding peripheral nerve stimulation (PNS).
Clinical Evidence
Clinical trial conducted per IEC 60601-2-33 guidelines to assess Peripheral Nerve Stimulation (PNS) risks. Study involved human volunteers exposed to increased gradient current limits (300A). No volunteers experienced PNS sensations. Results confirm safe operation in NORMAL mode.
Technological Characteristics
0.6T iron-core electromagnet; three orthogonal quadrupolar linear gradients; 30-ton closed-loop liquid chiller; copper bar coils with hollow-core cooling. Gradient amplifiers rated for 300A max current. Software-controlled gradient limits (20.27 mT/m max amplitude, 33.34 mT/m/ms max slew rate). Operates in NORMAL mode. Passive shimming.
Indications for Use
Indicated for producing multi-planar images of the head and body based on hydrogen nuclei NMR parameters (concentration, flow velocity, T1, T2) for diagnostic interpretation by physicians.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
7/5/2006
## Fonar Corporation 110 Marcus Drive Melville, New York 11747-4292
JUL 2 8 2006
June 30, 2006
# Special 510(k) Summary - Device Modification
### Submitter Information:
| Company | FONAR Corporation<br>Registration Number 2432211<br>110 Marcus Drive, Melville, New York 11747-4292 |
|----------|---------------------------------------------------------------------------------------------------------------------|
| Contact: | Luciano Bonanni Executive Vice President<br>Phone: (631) 694-2929 Fax: (631) 753-5150<br>E-mail: lbonanni@fonar.com |
### Device Designation:
| Trade Name: | Indomitable Magnetic Resonance Imaging Scanner Accessory |
|-----------------------|----------------------------------------------------------|
| Common Name: | Magnetic Resonance Imaging Scanner (MRI Scanner) |
| Classification: | System, Nuclear Magnetic Resonance Imaging (NMR/MRI) |
| Product Code: | LNH |
| Class: | 2 |
| Tier: | 2 |
| C.F.R. Section | 892.1000 |
| Classification Panel: | Radiology |
Reason for Submission: Change in operational limits for gradient subsystem.
FONAR Corporation has modified the existing control limits for the Indomitable scanner gradient subsystem to allow the maximum gradient current to increase by 50 percent, with a corresponding increase in the maximum slew rate for the gradient waveforms. All other subsystems remain essentially unchanged from their previously approved (K002490) state.
#### Device Modifications
This modified gradient subsystem follows the same basic design principles as our previously approved gradient systems. It functions on the same basic operating and physical principles as the predicate device, the Indomitable gradient system (K002490). It is constructed with the same manufacturing materials, and is fabricated in the same fashion to the previously approved gradients.
The primary differences of this modification to the gradient subsystem are the software limits for maximum gradient current value and the increased slew rate that accompanies the changed current value. It is, in summary, a significant equivalent of FONAR's currently approved predicate device.
{1}------------------------------------------------
7/5/2006
## Description of Device
## Gradient Subsystem
The gradients provide the ability of the system to accurately determine spatial location within the imaging volume. They are also used to help establish control of other imaging factors such as slice thickness and field-of-view. The gradients used on the Indomitable magnet, which are unchanged in this submission have the same overall diameter, and are still comprised of the same three quadrupolar (linear) gradients installed on all Fonar scanners.
One pair is placed in each of the orthogonal imaging planes (X, Y, and Z), developing a typical readout gradient of 1.2 mT/m. Note that while the typical gradient remains at 1.2 mT/m for this submission, the value of maximum gradient change has increased from a value of 12 mT/m to a new maximum of 20.27 mT/m. The gradient linearity remains unchanged at better than 2% over a 28 cm DSV with a stability of + 0.01%.
#### Gradient Current Amplifiers
The gradient current amplifier power required does not affect the actual gradient strength (which is still specified at a typical value of 1.2 mT/m with a typical amplifier power dissipation of 12 kW). The amplifiers for the Indomitable are still designed for accurate control of the current output to the gradient coils with very low distortion. and are rated for a maximum current output of 300 Amps. The gradient amplifiers are unchanged from previous submissions.
## Gradient System Control
The operating characteristics for the hardware of the gradient system described above are controlled by the Indomitable's Sympulse scanner operating software. The software is programmed to operate the scanner system in a safe and efficient manner, and is substantially equivalent to the current operating software in the Indomitable system as approved in K002490. The software has always included "safety limits", which were established through laboratory testing and clinical imaging.
Prior to this submission, the maximum current value allowed by software for the gradient current amplifiers was 200 Amps. This setting effectively limited the maximum gradient amplitude (dB/dt) to 12 mT/m for the system. This rate of change produced a maximum slew rate of 19.73 mT/m/ms and allowed the Indomitable scanner to operate in the NORMAL mode. Testing performed for this submission has shown that an increase of the maximum current from the previous maximum of 200 Amps to the new maximum value of 300 Amps would still allow the scanner to be operated in the NORMAL mode, since no PNS was detected in the test subjects.
Therefore, the maximum gradient current for the gradient amplifiers will now be established by the Sympulse operating software at a value of 300 Amps. This will translate to a maximum dB/dt of 20.27 mT/m with a corresponding maximum slew rate of 33.34 mT/m/ms. The scanner will continue to operate in the NORMAL mode.
{2}------------------------------------------------
## Magnet and Gradient Subsystem Comparisons
The tables of information presented below compare and summarize the common specifications for the Indomitable magnets and gradient subsystems as configured in the original submission (K002490) and in this revision. These tables show that the specifications compare favorably and therefore are substantially equivalent.
| MAGNET CONFIGURATIONS | FROM 510(K) K002490 | THIS SUBMISSION |
|----------------------------------------------------|--------------------------------------------------------------------------|--------------------------------------------------------------------------|
| Magnet Specification | Indomitable Electromagnet | Indomitable Electromagnet |
| Type | Iron-core Electromagnet | Iron-core Electromagnet |
| Field Strength | $0.6 T \pm 5%$ | $0.6 T \pm 5%$ |
| Operational Mode | Normal Operating Mode for both dB/dt and SAR | Normal Operating Mode for both dB/dt and SAR |
| Calculated SAR | 0.76 W/kg | 0.76 W/kg |
| Power Consumption | 98 KVA | 98 KVA |
| Magnet Coil Winding Material | Copper Bar with hollow core cooling | Copper Bar with hollow core cooling |
| Cooling | 30 ton closed loop liquid chiller | 30 ton closed loop liquid chiller |
| Stability (magnet drift) | long-term < 2 ppm/hr - short term < . 3 ppm/min | long-term < 2 ppm/hr - short term < . 3 ppm/min |
| Field Homogeneity | 7 ppm within 30 cm DSV - 1 ppm within 20 cm DSV | 7 ppm within 30 cm DSV - 1 ppm within 20 cm DSV |
| Shimming | passive only | passive only |
| Open Gap™ Configuration | 18" transverse clearance - 8' vertical axis | 18" transverse clearance - 8' vertical axis |
| Fringe Field<br>(5 Gauss Line -from magnet center) | 9.77 ft vertically - 13 ft to either side<br>9.3 ft toward front or rear | 9.77 ft vertically - 13 ft to either side<br>9.3 ft toward front or rear |
| Pole Cap Eddy Current Compensation | steel pole with inlaid laminates<br>and self-shield gradients | steel pole with inlaid laminates<br>and self-shield gradients |
| Outer Dimensions (h x 1 x w) | 128" x 142" x 87" | 128" x 142" x 87" |
| Weight (w/ base) | 290,000 lbs. (Approx.) | 290,000 lbs. (Approx.) |
| GRADIENT SUBSYSTEM | FROM 510(K) K002490 | THIS SUBMISSION |
|----------------------|--------------------------------------------------|----------------------------------------------------|
| Magnet Specification | Indomitable Electromagnet | Indomitable Electromagnet |
| Gradient Type | three orthogonal - quadrupolar (linear) X, Y, Z | three orthogonal - quadrupolar (linear) X, Y, Z |
| Gradient Strength | typical readout gradient 1.2 mT/m (12 mT/m max) | typical readout gradient 1.2 mT/m (20.27 mT/m max) |
| Typical Rise Time | typical 0.75 msec per 15 A (1.2 mT/m equivalent) | typical 0.75 msec per 15 A (1.2 mT/m equivalent) |
| Maximum Slew Rate | 19.73 mT/m/ms | 33.34 mT/m/ms |
| Stability | $\pm$ 0.01% | $\pm$ 0.01% |
| Heat Dissipation | 4 kW typical; 12 kW max (Air Cooled) | 4 kW typical; 12 kW max (Air Cooled) |
{3}------------------------------------------------
7/5/2006
## Indications for Use
The Indomitable Magnetic Resonance Imaging System is indicated for use in producing images of multiple planes in the head and body. These images correspond to the distribution of hydrogen nuclei exhibiting nuclear magnetic resonance (NMR) and depend for their contrast upon NMR parameters [hydrogen nuclei concentration and flow velocity, T1 (spinlattice relaxation time) and T2 (spin-spin relaxation time)]. As a result of the acquisition and processing of the NMR data, these images display the internal structure of the head and body, and when interpreted by a trained physician, can vield diagnostically useful information.
WARNING: This device is limited by U.S. Federal law to investigational use for indications not in the indications statement.
Under the requirements of the law, the non-indicated applications can be used only under an Institutional Review Board approved protocol for a non-significant risk device or an Investigational Device Exemption application approved by the FDA for a significant risk device. The procedures to be followed, under the sponsorship of FONAR Corporation, are determined by the current guidelines established by the FDA, which should provide the IRB with sufficient guidance to determine the level of risk for a MRI device.
WARNING: U.S. Federal law restricts the sale, distribution and use of this device by or on the order of a physician.
## Description of Safety and Substantial Equivalence
MRI effectiveness parameters such as spatial resolution, geometric distortion, specification volume, image uniformity, and slice spacing are unchanged by the modifications to the gradient subsystem. A series of clinical and non-clinical tests were performed to demonstrate the safety and effectiveness of the system, and to verify substantial equivalence to the predicate device. All testing was conducted in accordance with current FDA guidance documents and international standards. Results from all testing demonstrate substantial equivalence to the predicate device.
#### Non-Clinical Testing
The following specific parameters have been identified by the FDA as being pertinent to patient safety and thereby have been highlighted for inclusion in 510(k) submissions. The majority of these concerns are specifically tested by comparison to the ISO and NEMA standards, and where utilized are described individually.
- a. Static Field Strength: 0.6T ± 5%
- b. Peak and A-Weighted acoustic noise: 98.3 DBA ± 0.05
- c. Description of operational modes of the system: Normal Mode Only
- d. Maximum SAR for transmit coil: (Unaffected for this submission)
e. Maximum dB/dt (pulsing X, Y and Z gradients): 20.27 mT/m
f. Potential emergency conditions and shutdown: (Unaffected for this submission)
g. Biocompatibility of materials: (Unaffected for this submission)
{4}------------------------------------------------
## Clinical Testing
A voluntary clinical trial was conducted under the testing guidelines described in the ISO standard IEC 60601-2-33, which describes the criteria for experimental studies of human subjects to directly determine limits related to minimizing PNS for any given type of gradient system. All volunteers participated in the study under informed consent. All volunteers completed the test. During the testing process, no volunteer experienced any sensations associated with PNS. This result meets the criteria for continued operation in the normal mode for the Indomitable scanner with the new maximum current limits.
## Summary
The results of the testing, combined with the additional material within this submission has demonstrated that the modified gradient subsystem of the Indomitable magnet is substantially equivalent to the previously approved gradient subsystem of the predicate device (K002490) under the current conditions for intended use, with no adverse effects to patient health or safety reported.
{5}------------------------------------------------
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
JUL 2 8 2006
Mr. Luciano Bonanni Executive Vice President Fonar Corporation 110 Marcus Drive MELVILLE NY 11747-4292
Re: K061930
Trade/Device Name: Indomitable Magnetic Resonance Imaging Scanner Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: July 5, 2006 Received: July 7, 2006
Dear Mr. Bonanni:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/5/Picture/9 description: The image is a black and white circular logo. The logo commemorates a centennial, with the years 1906-2006 displayed at the top. The letters "FDA" are prominently featured in the center of the logo. The word "Centennial" is written below the letters "FDA".
Protesting and Promoting Public Health
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## STATEMENT OF INDICATIONS FOR USE
Applicant: FONAR CORPORATION 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: Indomitable Magnetic Resonance Imaging Scanner
Indications For Use:
The Indomitable Magnetic Resonance Imaging System is indicated for use in producing images of multiple planes in the head and body. These images correspond to the distribution of hydrogen nuclei exhibiting nuclear magnetic resonance (NMR) and depend for their contrast upon NMR parameters [hydrogen nuclei concentration and flow velocity, T1 (spin-lattice relaxation time) and T2 (spin-spin relaxation time)]. As a result of the acquisition and processing of the NMR data, these images display the internal structure of the head and body, and when interpreted by a trained physician, can yield diagnostically useful information.
WARNING: This device is limited by U.S. Federal law to investigational use for indications not in the indications statement.
Under the requirements of the law, the non-indicated applications can be used only under an Institutional Review Board approved protocol for a non-significant risk device or an Investigational Device Exemption application approved by the FDA for a significant risk device. The procedures to be followed, under the sponsorship of FONAR Corporation, are determined by the current guidelines established by the FDA, which should provide the IRB with sufficient guidance to determine the level of risk for a MRI device.
WARNING: U.S. Federal law restricts the sale, distribution and use of this device by or on the order of a physician.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER LINE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Per 21 CFR 801.109)
(Optional Format 1-2-96)
Nancy Broughton
Division Sign-C Division of Reproductive, and Radiological Devic 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.