DIMENSION VISTA ENZYME 1 CALIBRATOR (ENZ 1 CAL - KC310)
K061923 · Dade Behring, Inc. · JIX · Aug 25, 2006 · Clinical Chemistry
Device Facts
Record ID
K061923
Device Name
DIMENSION VISTA ENZYME 1 CALIBRATOR (ENZ 1 CAL - KC310)
Applicant
Dade Behring, Inc.
Product Code
JIX · Clinical Chemistry
Decision Date
Aug 25, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1150
Device Class
Class 2
Indications for Use
The ENZ 1 CAL is an in vitro diagnostic product for the calibration of Amylase (AMY), Gamma-Glutamyl Transferase (GGT), Lactate Dehydrogenase (LDH), Lipase (LIP), and Pseudocholinesterase (PCHE) methods on the Dimension Vista™ System.
Device Story
ENZ 1 CAL is a liquid, multi-analyte, bovine serum albumin-based calibrator kit; contains amylase (human saliva), GGT (bovine kidney), LDH (chicken heart), lipase (porcine pancreas), and pseudocholinesterase (horse serum). Kit includes six vials (three of Calibrator A, three of Calibrator B); ready-to-use. Used in clinical laboratory settings by trained personnel to calibrate the Dimension Vista System. Output is a set of assigned values used to establish the calibration curve for specific enzyme assays on the instrument. Enables accurate quantitative measurement of patient enzyme levels, supporting clinical diagnosis and monitoring of related metabolic or organ-specific conditions.
Clinical Evidence
No clinical data. Bench testing only. Traceability of assigned values was verified using the Dimension Vista System calibrated with an approved Master Pool. Stability testing (real-time) was conducted to support a 12-month shelf life at 4°C, with specific stability claims for on-board (1 day) and open-vial (30 days) storage.
Technological Characteristics
Liquid, multi-analyte calibrator. Matrix: Bovine serum albumin base. Analytes: Amylase (human saliva), GGT (bovine kidney), LDH (chicken heart), lipase (porcine pancreas), PCHE (horse serum). Traceability: Master Pool, Dimension clinical chemistry system. Stability: 12-month shelf life; 31-day open-vial stability (refrigerated); 7-day on-instrument stability. Complies with CEN 13640 and ISO 14971.
Indications for Use
Indicated for the calibration of Amylase, GGT, LDH, Lipase, and Pseudocholinesterase assays on the Dimension Vista System. For professional, prescription use only.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k061923
B. Purpose for Submission:
New Device
C. Measurand:
Calibrator materials for Amylase (AMY), Gamma-Glutamyl Transferase (GGT), Lactate Dehydrogenase (LDH), Lipase (LIP), and Pseudocholinesterase (PCHE)
D. Type of Test:
Not Applicable
E. Applicant:
Dade Behring, Inc.
F. Proprietary and Established Names:
Dimension Vista Enzyme 1 Calibrator, (KC310)
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| Calibrator, Multi-Analyte Mixture (JIX) | Class II | 21 CFR 862.1150, Calibrator. | 75 Clinical Chemistry (CH) |
H. Intended Use:
1. Intended use(s):
See below indications(s) for use.
2. Indication(s) for use:
{1}
The ENZ 1 CAL is an in vitro diagnostic product for the calibration of Amylase (AMY), Gamma-Glutamyl Transferase (GGT), Lactate Dehydrogenase (LDH), Lipase (LIP), and Pseudocholinesterase (PCHE) methods on the Dimension Vista™ System
3. Special conditions for use statement(s):
For Prescription Use
4. Special instrument requirements:
Dimension Vista™ System
I. Device Description:
ENZ 1 CAL is a liquid, multi-analyte, bovine serum albumin based product containing amylase (human saliva), gamma-glutamyl transferase (bovine kidney), lactate dehydrogenase (chicken heart), lipase (porcine pancreas), and pseudocholinesterase (horse serum). The kit consists of six vials, three vials of Calibrator A, and three vials of Calibrator B which are ready for use (no preparation is required). The volume per vial is 2.5 mL.
J. Substantial Equivalence Information:
| | Device | Predicate Devices | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Item | Dimension Vista™ System Chemistry 1 Calibrator | Chem I Calibrator K860021 (URCA-k862359) | Chemistry II Calibrator k861700 | Cholesterol Calibrator k861700 | Troxine Calibrator k862359 | Thyroxine Uptake Calibrator k862359 |
| Intended Use | The CHEM I CAL is an in vitro diagnostic product for the calibration of Calcium (CA), Cholesterol (CHOL), Creatinine (CREA), Glucose (GLU), Lactic Acid (LA), Magnesium (MG), Thyroxine (T4), Tyrosine Uptake (TU), Blood Urea Nitrogen (BUN)2 and Uric Acid (URCA) methods on the Dimension Vista Systems. | The Dimension® Chemistry I Calibrator is an in vitro diagnostic product used to calibrate the Dimension® clinical chemistry system for the Calcium (CA), Creatinine (CREA), Glucose (GLU/GLUC), Lactic Acid (LA), Urea Nitrogen (BUN) and Uric Acid (URCA) methods. | Chem II Calibrator is and in vitro diagnostic product to be used to calibrate the Dimension® clinical chemistry system for the Magnesium (MG), phosphorus (PHOS) and Triglycerides (TRIG) methods. | The Dimension® Cholesterol Calibrator is an in vitro diagnostic product to be used to calibrate the Dimension® clinical chemistry system for the Cholesterol (CHOL) method. | The Dimension® Thyroxine Calibrator is and in vitro diagnostic product to be used to calibrate the Dimension® clinical chemistry system for the Thyroxine (T4) method. | The Dimension® Thyroxine Uptake is an in vitro diagnostic product to be used to calibrate the Dimension® clinical chemistry system for the Thyroxine (T4) method. |
{2}
3
| | Device | Predicate Devices | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Item | Dimension Vista™ System Chemistry 1 Calibrator | Chem I Calibrator K860021 (URCA-k862359) | Chemistry II Calibrator k861700 | Cholesterol Calibrator k861700 | Troxine Calibrator k862359 | Thyroxine Uptake Calibrator k862359 |
| Analytes | Calcium (CA, Cholesterol (CHOL), Creatinine (CREA), Glucose (GLU), Lactic Acid (LA), Magnesium (MG), Thyroxine (T4), Thyronine Uptake (TU), Blood Urea Nitrogen (BUN) and Uric Acid (URCA) | Calcium (CA, Creatinine (CREA), Glucose (GLU), Lactic Acid (LA), Blood Urea Nitrogen (BUN) and Uric Acid (URCA) | Magnesium (MG) | Cholesterol (CHOL) | Thyroxin (T4) | Thyronine Uptake (TU) |
| Form | Liquid | Lyophilized | Liquid | Lyophilized | Lyophilized | Lyophilized |
| Trace-a bility | Bun – NIST SRM 912², CA – Abell Kendall (CDC-NCEP), CREA-NIST SRM 914, GLU-NIST SRM 917, LA-Lactic Acid-lithium salt A-Grade, MG-NIST SRM 929A, T4-USP, TU-Calculated value, URCA-NIST SRM 913 | BUN- NIST SRM 912, CA-NIST SRM 915, CREA-NIST SRM 915, CREA-NIST SRM 914, GLU-NIST SRM 917, LA- Lactic Acid- lithium salt A-Grade, URCA-NIST SRM 913 | MG- NIST SRM 924A | CHOL-NIST SRM 911 | T4-Thyroxine Master Pool | TU- Thyronine Uptake Master Pool |
| Matrix | Bovine serum albumin based product. | Bovine serum albumin based product. | Pure magnesium dissolved in a dilute solution of HCL, reagent grade potassium dihydrogen phosphate and reagent grade glycerol | Bovine serum albumin based product | Human serum base product | Human serum base product |
| Number of Levels | Two levels | Three levels | Three levels | Three levels | Five levels | Five levels |
K. Standard/Guidance Document Referenced (if applicable):
| STANDARDS |
| --- |
| Title and Reference Number |
| Stability Testing of In Vitro Diagnostic Reagents (13640) |
| Medical devices - Application of risk management to medical devices (14971:2000) |
{3}
4
| GUIDANCE | | | |
| --- | --- | --- | --- |
| Document Title | Office | Division | Web Page |
| Guidance for Industry - Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators; Final | OIVD | | http://www.fda.gov/cdrh/ode/calibrator.html |
| Guidance for Industry and FDA Staff - Use of Symbols on Labels and in Labeling of In Vitro Diagnostic Devices Intended for Professional Use | | | http://www.fda.gov/cdrh/ocd/guidance/4444.html |
L. Test Principle:
Not Applicable
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Not Applicable
b. Linearity/assay reportable range:
Not Applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability: The assigned values of the Enzyme 1 Calibrator were verified on a Dimension® Vista™ System calibrated with an approved Master Pool. Master Pool values were assigned on multiple Dimension® clinical chemistry instruments. According to the sponsor the traceability of the assigned values of the Enzyme 1 Calibrator was standardized to the below table of assigned values.
| Analyte | Reference System | Master Pool Levels |
| --- | --- | --- |
| Amylase (AMY) | Dimension® clinical chemistry system | 3 |
| γ-Glutamyl Transferase (GGT) | Dimension® clinical chemistry system | 3 |
| Lactic Dehydrogenase (LDH) | Dimension® clinical chemistry system | 3 |
{4}
| Lipase (LIP) | Dimension® clinical chemistry system | 4 |
| --- | --- | --- |
| Pseudocholinesterase (PCHE) | Dimension® clinical chemistry system | 3 |
Stability: The target shelf life for the Dimension Vista™ Chemistry I Calibrator is 12 months. The sponsor’s shelf life was determined by comparing results of the product stored at 4°C with control stored at -20°C. Shelf-life stability (expiration) dating assignment at commercialization reflects the real-time data on file at Dade Behring, Inc.
A vial punctured by the instrument and stored on board has a stability claim of one day.
An open vial not on instrument, but recapped and stored is a refrigerator has a stability claim of 30 days.
d. Detection limit:
Not applicable
e. Analytical specificity:
Not applicable
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
{5}
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Not applicable
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalent decision.
6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.