FUSION

K061912 · Burke, Inc. · INI · Aug 1, 2006 · Physical Medicine

Device Facts

Record IDK061912
Device NameFUSION
ApplicantBurke, Inc.
Product CodeINI · Physical Medicine
Decision DateAug 1, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.3800
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Motorized five-wheeled battery powered scooter intended for medical purposes by disabled persons and other non-medical purposes.

Device Story

Fusion is a battery-operated, indoor/outdoor motorized scooter. It utilizes a frame, motor, transaxle, digital controller, and thumb-operated throttle. The device features a patent-pending frame design with three primary wheels and two spring-loaded trailing arm caster wheels for stability and maneuverability. The frame is separable for transport; batteries are removable. Accessories include a front light and basket. Operated by the user, the device provides mobility assistance. It is intended for use by disabled persons for medical purposes and for non-medical use. The device is designed to offer the indoor maneuverability of smaller scooters with the traction and stability of larger models.

Clinical Evidence

No clinical data. Bench testing only, including compliance with ANSI/RESNA standards for static/dynamic stability, braking, energy consumption, dimensions, speed/acceleration, and fatigue strength.

Technological Characteristics

Battery-powered motorized scooter. Features a frame with three primary wheels and two spring-loaded trailing arm caster wheels. Includes digital controller, motor, transaxle, and thumb-operated throttle. Complies with ANSI/RESNA WC/01-06, WC/08-10, WC/15, and WC/93 standards. No software-based algorithmic processing described.

Indications for Use

Indicated for disabled persons requiring a motorized scooter for mobility, as well as for non-medical purposes.

Regulatory Classification

Identification

A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K0619/2 # Fusion 510(k) Summary AUG 0 1 2006 | Submitter's Name & Address: | Burke, Inc.<br>1800 Merriam Lane<br>Kansas City, KS 66105<br>Phone: 913-722-5653<br>Fax: 913-722-2614 | |--------------------------------------|-------------------------------------------------------------------------------------------------------| | Contact Person: | DuWayne E Kramer<br>Official Corresponde-- | | Date Prepared: | 06-22-06 | | Name of Device and Proprietary Name: | Fusion | | Common or Usual Name: | 3 Wheel Power Scooter | | Class: | II | | Classification Name: | Vehicle, Motorized 3-Wheeled | | Product Code: | INI | #### Device Description: The Fusion is an indoor/outdoor scooter that is battery operated. Fusic standard technology seating, thumb operated throttle, digital controller rarger, motor, brake, transaxle, frame material and till we have used for years on our - - Saver units. The Fusion frame separates for transport and has removable batteries anc تحتوب Accessories include a front light and a basket. #### Comparison to Predicate Devices: The Fusion is substantially equivalent to our PaceSaver Little Junior (K.S.25 122) and PaceSaver Espree Scooter (K965123) when comparing maneuverability, performance and technological characteristics. While the frame material s the same, Fusion's key change is in the frame design. PaceSaver Fusion uses 3 primary wheels similar to the predicate device design. However, while the predicate devices يڪ اس fixed-in-place anti-tip wheels for tip-over safety, the Fusion uses two caster wheels on spring-loaded trailing arms. Fusion's patent-pending frame design provides the indoor manesterability of our small scooter with the traction and stability of our larger standard scooters. #### Intended Use: Motorized five-wheeled battery powered scooter intended for medical purposes by disabled persons and other non-medical purposes. {1}------------------------------------------------ ## Non-Clinical Testing: Compliance with applicable Testing Standards is as follows: ANSI/RESNA WC/01 Determination of Static Stability ANSI/RESNA WC/02 Determination of Dynamic Stability ANSI/RESNA WC/03 Effectiveness of Brakes ANSI/RESNA WC/04 Determination of Energy Consumption - Theoretical Range ANSI/RESNA WC/05 Overall Dimensions, Mass, and Turning Space ANSI/RESNA WC/06 Determination of Maximum Speed, Acceleration, and Retardation of Electric Wheelchairs ANSI/RESNA WC/08 Test Methods for Static, Impact, and Fatigue Strengths ANSI/RESNA WC/09 Climatic Tests ANSI/RESNA WC/10 Obstacle Climbing ANSI/RESNA WC/15 Documentation and Labeling ANSI/RESNA WC/93 Maximum Overall Dimensions ## Discussion of Clinical Testing Performed: Clinical testing is not appropriate and not required for the proposed device. Clinical testing was not performed. ## Conclusions: The PaceSaver Fusion has the same intended use and substantially equivalent technological, physical, and operational characteristics as the predicate devices PaceSaver Little Jr. (K925122) and PaceSaver Espree Scooter (K965123). Moreover, the non-clinical testing and comparison to predicate devices demonstrates that the differences do not cause any questions as to the safety or effectiveness of the proposed device. The Fusion has passed the necessary testing to recognized consensus standards and is considered to be safe for user operation. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the department's name around the perimeter. Inside the circle is a stylized image of a human figure with three lines extending above it, representing the department's mission to protect the health of all Americans and provide essential human services. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # AUG () 1 2006 Burke, Inc. % Mr. DuWayne E. Kramer, Jr. President 1800 Merriam Lane Kansas City, Kansas 66106 Re: K061912 Trade/Device Name: PaceSaver Fusion Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized three-wheeled vehicle Regulatory Class: II Product Code: INI Dated: June 30, 2006 Received: July 6, 2006 Dear Mr. Kramer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Cor.trols) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. {3}------------------------------------------------ Page 2- Mr. DuWayne E. Kramer, Jr. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Hubert Leuner wo Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ To: Elmar Einberg Re.: K061912 Fusion # Indications for Use 510(k) Number (if known): K061912 Device Name: PaceSaver Fusion Indications For Use: Motorized five-wheeled battery powered scooter intended for medical purposes by disabled persons and other non-medical purposes. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use X (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Deyice Evaluation (ODE) Huler Leener Division of General, Restorative, and Neurological Devices Page 1 of _ 510(k) Number K061912 4 - 2
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...