BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-PIPERACILLIN-TAZOBACTAM (GP) 1/4-128/4 UG/ML

K061869 · Becton, Dickinson & CO · LON · Aug 18, 2006 · Microbiology

Device Facts

Record IDK061869
Device NameBD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-PIPERACILLIN-TAZOBACTAM (GP) 1/4-128/4 UG/ML
ApplicantBecton, Dickinson & CO
Product CodeLON · Microbiology
Decision DateAug 18, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2

Indications for Use

The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.

Device Story

The BD Phoenix System is an automated microbiology platform for rapid identification and antimicrobial susceptibility testing (AST) of bacterial isolates. It utilizes molded polystyrene panels containing 136 micro-wells with dried biochemicals and antimicrobial agents. Pure culture samples are inoculated into panels, which are then incubated at 35°C within the instrument. The system performs continuous monitoring, taking readings every 20 minutes. It uses a redox indicator to detect bacterial growth in the presence of antimicrobial agents, combined with turbidity measurements. The system software interprets these readings to determine MIC values and categorical interpretations (S, I, or R). Results are provided to clinicians to guide antimicrobial therapy. The system is intended for use in clinical laboratory settings by trained personnel.

Clinical Evidence

Performance evaluated by comparing Phoenix System results to CLSI reference broth microdilution method using clinical, stock, and challenge isolates. Study assessed Essential Agreement (EA) and Category Agreement (CA). Intra-site reproducibility >90% and inter-site reproducibility >95% for Gram-positive isolates. Performance data for piperacillin-tazobactam demonstrated substantial equivalence to reference methods.

Technological Characteristics

Automated, growth-based AST system. Materials: molded polystyrene trays with 136 micro-wells. Sensing: colorimetric oxidation-reduction (redox indicator). Energy: electrical (instrumentation). Connectivity: barcode scanning, automated data recording. Software: rule-based 'EXPERT' system. Sterilization: not specified.

Indications for Use

Indicated for in vitro quantitative determination of antimicrobial susceptibility (MIC) of Gram-negative and Gram-positive bacteria (Staphylococcus, Enterococcus, Streptococcus) from pure culture. Specifically for piperacillin-tazobactam (1/4-128/4 µg/mL) testing in Gram-positive panels. Contraindicated for methicillin/oxacillin-resistant Staphylococcus isolates.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k061869 B. Purpose for Submission: Addition of piperacillin – tazobactam to the BD Phoenix™ Automated Microbiology System C. Measurand: Piperacillin – tazobactam 1/4 – 128/4 µg/mL D. Type of Test: Antimicrobial Susceptibility Test (AST) (Qualitative) colorimetric oxidation-reduction, growth-based E. Applicant: Becton, Dickinson &amp; Company F. Proprietary and Established Names: BD Phoenix™ Automated Microbiology System – Piperacillin – tazobactam 1/4 – 128/4 µg/mL G. Regulatory Information: 1. Regulation section: 21 CFR 866.1645 Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System 2. Classification: II 3. Product code: LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation 4. Panel: 83 Microbiology H. Intended Use: 1. Intended use(s): Piperacillin - tazobactam at 1/4 – 128/4 µg/mL on the Phoenix™ Gram Positive (GP) ID/AST or AST only panel is intended for use with the BD Phoenix Automated Microbiology System for in vitro quantitative determination of {1} antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus, and Streptococcus. 2. Indication(s) for use: This application is indicated for the addition of the antimicrobial agent piperacillin – tazobactam at concentrations of 1/4 – 128/4 µg/mL to Gram-positive ID/AST or AST only for testing Staphylococcus spp. and Enterococcus spp. only. 3. Special conditions for use statement(s): For prescription use only 4. Special instrument requirements: Not Applicable I. Device Description: The BD Phoenix™ Automated Microbiology System includes instrumentation and software, sealed and self-inoculating molded polystyrene trays with 136 micro-wells containing dried reagents, and specific inoculum broth formulations for ID and AST Indicator. The organism to be tested must be a pure culture and be preliminarily identified as gram positive or gram negative. Colonies are then suspended in broth, and equated to a 0.5 McFarland with the recommendation to use the BD CrystalSpec™ Nephelometer. A further dilution is made into an AST broth, which contains an AST indicator, prior to inoculating the panel. The AST broth is a cation-adjusted formulation of Mueller-Hinton broth containing 0.01% Tween 80. After adding the indicator solution to the AST inoculum the color is blue and after inoculation and incubation changes to pink then to colorless as reduction in the panel well proceeds. Inoculated panels are barcode scanned and loaded into the BD Phoenix™ Automated Microbiology System instrument where the panels are continuously incubated at 35°C. The AST has a final inoculum of 5 × 10⁵ CFU/ml. The instrument incubates, reads and records the results of the biochemical substrates and antimicrobial agents and interprets the reactions to give an ID of the isolate and MIC value and category interpretation of the antimicrobial agents. Organisms growing in the presence of a given antimicrobic agent reduce the indicator, signaling organism growth and resistance to the antimicrobic agent. Organisms killed or inhibited by a given antimicrobic do not cause reduction of the indicator and therefore do not produce a color change. Additional interpretation is done using software driven "EXPERT" System using rules derived from the NCCLS documentation. Readings are taken every 20 minutes with an AST result available between 4-16 hours. This is only an autoread result; no manual readings are possible with this system. J. Substantial Equivalence Information: {2} 1. Predicate device name(s): VITEK® System 2. Predicate 510(k) number(s): N50510 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | 1. Intended Use | Intended for the in vitro rapid identification (ID) and quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most bacteria. | Same | | 2. Isolates | Isolated colonies from culture used | Isolated colonies from culture used | | 3. Result Reported | Report results as minimum inhibitory concentration (MIC) and categorical interpretation (SIR) | Report results as minimum inhibitory concentration (MIC) and categorical interpretation (SIR) | | 4. Incubation Time | <16 hours | <16 hours | | 5. Type of Test | Automated | Automated | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | 1. Results achieved | Results are determined from serial twofold dilutions of antimicrobial agents | Results are determined from extrapolation of doubling dilutions | | 2. Sample Preparation | Inoculum density equated to 0.5 McFarland standard | Inoculum density equated to 1.0 McFarland standard | | 3. Technology | Automated growth based enhanced by use of a redox indicator (colorimetric oxidation-reduction) to detect organism growth. | Automated growth based with detection using an attenuation of light measured by an optical scanner. | # K. Standard/Guidance Document Referenced (if applicable): "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test Systems; Guidance for Industry and FDA"; CLSI M7 (M100-S16) "Methods for {3} Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard." L. Test Principle: The system employs conventional, colorimetric, fluorogenic and chromogenic substrates to identify the genus and species of the isolate. The AST portion of the BD Phoenix™ Automated Microbiology System is a broth based microdilution method that utilizes a redox indicator (colorimetric oxidation-reduction) to enhance detection of organism growth. The MIC is determined by comparing growth in wells containing serial two-fold dilutions of an antibiotic to the growth in “growth control wells” which contains no antibiotic. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Thirteen isolates were evaluated for site to site and inter site reproducibility demonstrating &gt;95% reproducibility. The ten isolate study described in the guidance document was used (10 organisms tested 3 times on 3 days at 3 sites). b. Linearity/assay reportable range: Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): The FDA and CLSI recommended QC isolate, S. aureus ATCC 29213 was tested on every test occasion with the reference method and the BD Phoenix™. The reference method QC results were in range for every day tested. The BD Phoenix™ was tested a sufficient number of times to demonstrate that the system can produce QC results in the recommended ranges most of the time. Since most of the results of S. aureus ATCC 29213 were not on-scale, two additional QC organisms, E. faecalis ATCC 29212 and E. faecalis ATCC 51299 were also tested to provide on scale results. The reference method was in range for every day tested. 4 {4} Piperacillin - Tazobactam GP QC Table | ORGANISM | conc. (μg/mL) | Reference | | BD Phoenix™ | | | --- | --- | --- | --- | --- | --- | | | | | | | | | S. aureus ATCC 29213 Expected Range: 1/4 - 2/4 μg/mL | ≤1 | | 248 | | 247 | | | 2 | | 4 | | 3 | | | | | | | | | | | | | | | | | | | | | | | E. faecalis ATCC 29212 Expected Range: ≤1/4 - 2/4 μg/mL | ≤1 | | | | 1 | | | 2 | | 58 | | 241 | | | 4 | | 191 | | 2 | | | 8 | | | | 2 | | | | | | | | | E. faecalis ATCC 51299 Expected Range: 2/4 - 8/4 μg/mL | 2 | | 28 | | 3 | | | 4 | | 134 | | 177 | | | 8 | | 35 | | 6 | | | 16 | | | | 2 | | | 32 | | | | 1 | | | >32 | | | | 5 | Inoculum density control: The organism suspension density of the ID broth was equivalent to a 0.5 McFarland standard using the BBL™ CrystalSpec™ Nephelometer which was verified each day of testing. Internal data was used to demonstrate that the use of the BBL™ CrystalSpec™ Nephelometer would produce reproducible results. Five different instruments were used. d. Detection limit: Not Applicable e. Analytical specificity: Not Applicable f. Assay cut-off: Not Applicable 2. Comparison studies: a. Method comparison with predicate device: The CLSI recommended broth dilution reference panel was prepared according to the CLSI recommendation. Clinical testing was performed at four sites. The testing included both fresh clinical isolates and stock isolates along with a challenge set with known results. A comparison was provided to the reference method with the following agreement. {5} GP Accuracy Summary Clinical and Challenge | Report Group | EA Tot | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | CA N | CA % | #R | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Enterococci | 758 | 721 | 95.1 | 548 | 530 | 96.7 | - | - | - | - | - | - | | Staphylococci (MSS only) | 585 | 565 | 96.6 | 19 | 1 | 5.3 | 585 | 100 | 0 | N/A | 0 | 0 | EA-Essential Agreement vmj – very major discrepancies CA-Category Agreement maj – major discrepancies R-resistant isolates N/A – No minor errors possible since there is no intermediate category Essential agreement (EA) is when the BD Phoenix™ panels agree with the reference test panel results exactly or within one doubling dilution of the reference method. Category agreement (CA) is when the BD Phoenix™ panel result interpretation (SIR) agrees exactly with the reference panel result interpretation. Evaluable EA is when the MIC result is on scale for both the BD Phoenix™ and the reference and have on-scale EA. Only essential agreement can be calculated with Enterococci report group since there are no interpretive criteria. Also tested in the clinical study were 584 Methicillin Resistant Staphylococci (MRS). However, MRS will be reported as piperacillin – tazobactam resistant regardless of MIC value. There were no errors for the Methicillin Susceptible Staphylococci (MSS) group. A trend was observed with the S. aureus group where the reference method appears to be more resistant than the BD Phoenix™. b. Matrix comparison: Not Applicable 3. Clinical studies: a. Clinical Sensitivity: Not Applicable b. Clinical specificity: Not Applicable c. Other clinical supportive data (when a. and b. are not applicable): Not Applicable 4. Clinical cut-off: Not Applicable 5. Expected values/Reference range: {6} Staphylococcus spp. $\leq 8/4(S)$, $\geq 16/4(R)$ **N. Proposed Labeling:** The expected value range, interpretive criteria and QC for gram positive panels are included in the package insert. The labeling is sufficient and satisfies the requirements of 21 CFR Part 809.10. **O. Conclusion:** The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 7
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...