HYPERSPECTRAL TISSUE OXYGENATION MEASUREMENT SYSTEM, MODEL OXY VU1
K061848 · Hypermed, Inc. · MUD · Nov 8, 2006 · Cardiovascular
Device Facts
Record ID
K061848
Device Name
HYPERSPECTRAL TISSUE OXYGENATION MEASUREMENT SYSTEM, MODEL OXY VU1
Applicant
Hypermed, Inc.
Product Code
MUD · Cardiovascular
Decision Date
Nov 8, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Indications for Use
The OxyVu-1 Hyperspectral Tissue Oxygenation (HTO) Measurement System is intended for use by healthcare professionals as a non-invasive tissue oxygenation measurement system that reports an approximate value of: - oxygen saturation (HT-Sat), - oxyhemoglobin level (HT-Oxy), and - deoxyhemoglobin (HT-Deoxy) level in superficial tissue. The OxyVu-1 system displays two-dimensional color-coded images of tissue oxygenation of the scanned surface and reports hyperspectral tissue oxygenation measurements for selected tissue regions. The OxyVu-1 system is indicated for use to determine oxygenation levels in superficial tissues for patients with potential circulatory compromise.
Device Story
OxyVu-1 is a hyperspectral imaging system for non-invasive assessment of superficial tissue oxygenation. Input: hyperspectral imaging cube collected via instrument head (broadband illuminator, camera, spectral filter). Processing: spectral analysis at each point in a 2D scanned area to calculate oxygen saturation (HT-Sat), oxyhemoglobin (HT-Oxy), and deoxyhemoglobin (HT-Deoxy) levels. Output: 2D color-coded map of tissue oxygenation and numeric measurements for selected regions. Used in clinical settings by healthcare professionals. System includes console (cart, electronics, CPU, monitor) and instrument head with support arm. Check pads and targets used for calibration and image registration. Output assists clinicians in evaluating tissue perfusion and circulatory status, potentially aiding in the management of patients with circulatory compromise.
Clinical Evidence
The document mentions that substantial equivalence is supported by comparable results in clinical testing, but does not provide specific study design, sample sizes, or quantitative performance metrics (e.g., sensitivity, specificity, AUC).
Technological Characteristics
System utilizes hyperspectral imaging technology. Components: console (CPU, monitor, electronics), hyperspectral instrument head (broadband illuminator, camera, spectral filter), and single-use check pads/targets. Performs spectral analysis at specific wavelengths to derive HT-Sat, HT-Oxy, and HT-Deoxy. Output is 2D color-coded image and numeric data. Connectivity includes keyboard/pointing device interface. No specific material standards or software architecture details provided.
Indications for Use
Indicated for determining oxygenation levels in superficial tissues for patients with potential circulatory compromise.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the text "HyperMed" in a bold, sans-serif font. There are two overlapping line graphs above the text. The graphs have peaks and valleys, and they appear to be related to the text. The overall image has a technical or scientific feel.
NOV - 8 2006
510(k) Summary OxyVu-1 Hyperspectral Tissue Oxygenation Measurement System (June 28, 2006)
### Submittal information:
Post-approval contact: Jenny Freeman, MD, President, Hypermed, Inc. Hypermed, Inc. 305 Second Avenue Suite B Waltham, MA 02451
Phone: 781-647-0009
510(k) prepared by Chas Burr, President, Chas Burr Q/R Systems, Inc. Phone: 781-789-1646 Fax: 781-723-2422
### Device name and classification
Proprietary Name: OxyVu-1 Hyperspectral Tissue Oxygenation Measurement System Hyperspectral Tissue Oxygenation Measurement System Common Name: Classification Name: Tissue Saturation Oximeter Classification Panel: Cardiovascular CFR Section: 21 CFR 870.2700 Class: II Product Code: MUD
## Substantial Equivalence
The Oxy Vu-1 system is substantially equivalent to the Inspectra Tissue Spectrometer System, Model 325 manufactured by Hutchinson Technology, Inc. The Inspectra system was cleared in 510(k)'s K053618, K042020, K023933, and K012759.
### Device Description
The OxyVu-1 system is based on hyperspectral imaging technology. The
{1}------------------------------------------------
technology performs spectral analysis at each point in a two-dimensional scanned area producing an image displaying information derived from the analysis. For the OxyVu-1 system, the spectral analysis determines in superficial tissues approximate values of oxygen saturation (HT-Sat), oxyhemoglobin levels (HToxy), and deoxyhemoglobin levels (HT-deoxy). The OxyVu-1 system displays the tissue oxygenation in a two-dimensional, color-coded image.
The system consists of:
- System console: cart, system electronics, CPU, monitor, keyboard, pointing . device and printer.
- Hyperspectral instrument head with support arm: broadband illuminator, . camera and spectral filter for collecting hyperspectral imaging cube.
- Single use OxyVu Check Pads and Targets: The OxyVu Check Pad is used . to perform an instrument check prior to patient measurements. The OxyVu Target is placed within the intended field of view and is used as a fiduciary mark for image registration and for focusing.
### Intended Use
The OxyVu-1 Hyperspectral Tissue Oxygenation (HTO) Measurement System is intended for use by healthcare professionals as a non-invasive tissue oxygenation measurement system that reports an approximate value of:
- . oxygen saturation (HT-Sat),
- . oxyhemoglobin level (HT-Oxy), and
- . deoxyhemoglobin (HT-Deoxy) level
in superficial tissue. The OxyVu-1 system displays two-dimensional color-coded images of tissue oxygenation of the scanned surface and reports hyperspectral tissue oxygenation measurements for selected tissue regions.
The OxyVu-1 system is indicated for use to determine oxygenation levels in superficial tissues for patients with potential circulatory compromise.
### Comparison with the Predicate Device
| | OxyVu-1 | Inspectra Model 325 |
|--------------------------|---------------------------------------------------------------------------------|---------------------|
| Measures | Oxygen saturation<br>Oxyhemoglobin level<br>Deoxyhemoglobin level | Oxygen saturation |
| Method of<br>Measurement | Spectral analysis at specific wavelengths of light returned from target tissue. | |
| Output Display | Numeric<br>Two-dimensional color map of<br>approximate tissue oxygenation | Numeric |
page 2 of 3
{2}------------------------------------------------
# Similarities and Differences
Oth devices use spectral analysis to determine oxygenation levels in near-surface tissues. The Clyckly, I systems Both devices use spectral analysis to decomment on the hemoglobin. The OxyVu-l system a display numeric values of the approximate oxygen and deoxyhemoglobin levels necessary for the oxygen saturation calculation.
The hyperspectral scanning method used by the OxyVu-1 system provides two-dimensional mapping of color-coded oxygenation levels.
# Basis of Substantial Equivalence
Based on equivalent intended uses and technologies and on comparable results in clinical testing, the Oxy Vu-1 Hyperspectral Tissue Oxygenation Measurement System is substantially equivalent to the Inspectra Model 325 Tissue Spectrometer System.
Page 20 ge 3d 3
{3}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV - 8 2006
HyperMed, Inc. c/o Mr. Chas Burr Chas Burr Q/R Services, Inc. 11 Mystic Avenue Winchester, MA 01890-2920
Re: K061848
Trade Name: Oxy Vu-1 Hyperspectral Tissue Oxygenation (HTO) Measurement System Regulation Number: 21 CFR 870.2700 Regulation Name: Tissue saturation oximeter Regulatory Class: Class II Product Code: MUD Dated: October 11, 2006 Received: October 13, 2006
Dear Mr. Burr:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Chas Burr
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for Use
510(k) Number (if known): __ 対応 Kole1048
Device Name: OxyVu-1 Hyperspectral Tissue Oxygenation Measurement System
Indications for Use:
The OxyVu-1 Hyperspectral Tissue Oxygenation (HTO) Measurement System is intended for use by healthcare professionals as a non-invasive tissue oxygenation measurement system that reports an approximate value of:
- oxygen saturation (HT-Sat), �
- oxyhemoglobin level (HT-Oxy), and .
- deoxyhemoglobin (HT-Deoxy) level
in superficial tissue. The OxyVu-1 system displays two-dimensional color-coded images of tissue oxygenation of the scanned surface and reports hyperspectral tissue oxygenation measurements for selected tissue regions.
The OxyVu-1 system is indicated for use to determine oxygenation levels in superficial tissues for patients with potential circulatory compromise.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH/Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K061848
page 1 of 1
Page 17
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.