K061755 · Esaote Europe B.V. · IYO · Aug 7, 2006 · Radiology
Device Facts
Record ID
K061755
Device Name
MYLAB15/20 NEW INDICATIONS ULTRASOUND SYSTEM
Applicant
Esaote Europe B.V.
Product Code
IYO · Radiology
Decision Date
Aug 7, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Esaote's MyLab15/MyLab20 is a compact console ultrasound system intended to be used by a physician to perform general diagnostic ultrasound studies including Fetal, Abdominal, Pediatric, Small organ, Neonatal Cephalic, Cardiac, Transrectal, Transvaginal, Peripheral Vascular, Intraoperative: Abdominal, Other: Urological, Musculoskeletal (Conventional and Superficial).
Device Story
Compact console ultrasound system; performs general diagnostic studies. Inputs: ultrasonic signals via various probes (linear, convex, endocavity, mechanically driven 3D convex array). Processing: B-Mode, M-Mode, PW Doppler, Color Flow Mapping, Tissue Enhancement Imaging (TEI); produces real-time 2D, 3D (manual), and 4D (automatic with BC432P probe) images. Used in clinical settings by physicians. Output: visual display on 15" CRT/LCD or 19" LCD monitors; supports digital archival, VCR, and video printing. Assists clinicians in diagnostic visualization of anatomy and pathology; facilitates clinical decision-making through real-time imaging and measurement/analysis (M&A) capabilities.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence is based on technological comparison and adherence to established safety standards for diagnostic ultrasound systems.
Technological Characteristics
Compact console ultrasound; 15" CRT/LCD or 19" LCD monitor. Transducers: linear, convex, endocavity, mechanically driven 3D convex array. Frequencies: 2.7-15 MHz. Modes: B, M, PW Doppler, Color Flow, Amplitude Doppler, TEI, 3D/4D. Electrical safety: EN60601-1. Connectivity: digital archival, VCR/printer support. Optional ECG. Software-based image processing.
Indications for Use
Indicated for physicians performing general diagnostic ultrasound studies on fetal, pediatric, and adult patients. Applications include abdominal, small organ (thyroid, breast, testicles), neonatal cephalic, cardiac, transrectal, transvaginal, peripheral vascular, musculoskeletal, and intraoperative abdominal imaging.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
510(k) Summary NIvLab15/20 Esaote Europe BV
K061755
510(k) Summary
AUG - 7 2006
The following safety and effectiveness summary has been prepared pursuant to requirement for 510(k) summaries specified in 21CFR 1807.92(a).
807.92(a){ })
## Submitter Information
| Carri Graham, Official Correspondent | 11460 N Meridian St., Ste 150 |
|--------------------------------------|-------------------------------|
| Carmel, Indiana 46032 | |
| Phone: | (317) 569-9500 |
| Facsimile: | (317) 569-9520 |
| Contact Person: | Carri Graham |
|-----------------|---------------|
| Date: | June 21, 2006 |
| 807.92(a)(2) | |
|-------------------------|-----------------------------------------------------------------------------------------------------|
| Trade Name: | MyLab15/20 New Indications Ultrasound System |
| Common Name: | Ultrasound Imaging System |
| Classification Name(s): | Ultrasonic pulsed echo imaging system 892.1560<br>Ultrasonic pulsed Doppler imaging system 832.1550 |
| Classification Number: | 90IYO<br>90IYN |
| 807.92(a)(3) | Predicate Device(s) |
|--------------|---------------------|
|--------------|---------------------|
| K014168 | Technos Esaote, S.p.A. |
|---------|------------------------------------------|
| K043588 | MyLab15/20 Ultrasound System Pie Medical |
| K053154 | MyLab15/20 Just3D/4D Pie Medical |
Additional Substantial Equivalence Information is provided in the following substantial Equivalence Comparison Table.
{1}------------------------------------------------
510(k) Summary MyLab15/20 Esaote Europe BV
K061755
807.92(a)(4)
#### Device Description
The MyLab15/20 is a compact console ultrasound system used to perform general diagnostic ultrasound studies. Its primary modes of operation are: B-Mode, M-Mode, PW Doppler and Color Flow Mapping and Tissue Enhancement Imaging (TEI). MyLab15/20 is able to produce real time 2D images and 3D images (in manual mode) with all probes. The system, in combination with the probe BC432P, offers the possibility to also produce automatic 3D and real time 4D images
#### 807.92(a)(5)
### Intended Use(s)
Esaote's MyLab15/MyLab20 is a compact console ultrasound system intended to be used by a physician to perform general diagnostic ultrasound studies including Fetal, Abdominal, Pediatric, Small organ, Neonatal Cephalic, Cardiac, Transrectal, Transvaginal, Peripheral Vascular, Intraoperative: Abdominal, Other: Urological, Musculoskeletal (Conventional and Superficial).
{2}------------------------------------------------
510(k) Summary MyLab15/20 Esaote Europe BV
K061755
# Comparison Chart for Substantial Equivalence
| General<br>Characteristics | Esaote MyLab15/20<br>K043588, K053154 | Esaote Technos<br>K014168 | Esaote MyLab15/20<br>Current 510(k) |
|------------------------------|---------------------------------------|---------------------------|-------------------------------------|
| Applications | | | |
| Intraoperative:<br>Abdominal | No | Yes | Yes |
| Other: Urological | No | Yes | Yes |
{3}------------------------------------------------
510(k) Summary
MyLab15/20
Esaote Europe BV
| | 1<br><br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
|----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| -----------------------------------<br>Un 12 1 1 1 1 1 1 1 1 1 | 400000<br>annono | Frenta Mul 2115/20 |
| General<br>Characteristics | Esaote MyLab15/20<br>K043588, K053154 | Esaote Technos<br>K014168 | Esaote MyLab15/20<br>Current 510(k) |
|----------------------------------|-------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| Transducer Type | | | |
| Linear | Yes | Yes | Yes |
| Convex | Yes | Yes | Yes |
| 2D Freq MHz | 2.7 - 15 | 2.8 - 12.5 | 2.7 - 15 |
| Multifrequency | Yes | Yes | Yes |
| Special probes | • Endocavity probe<br>• Mechanically<br>Driven 3D Convex<br>Array | • Endocavity<br>probe<br>• Mechanically<br>Driven 3D<br>Convex Array<br>• CW Doppler<br>Probe | • Endocavity probe<br>• Mechanically<br>Driven 3D Convex<br>Array |
| Biopsy attachments | | | |
| Convex | Yes | Yes | Yes |
| Linear | Yes | Yes | Yes |
| Imaging modes | | | |
| Real Time 2D | Yes | Yes | Yes |
| M-mode | Yes | Yes | Yes |
| PW Doppler | Yes | Yes | Yes |
| CW Doppler | No | Yes | No |
| CFM Doppler | Yes | Yes | Yes |
| Amplitude Doppler | Yes | Yes | Yes |
| Triplex | Yes | Yes | Yes |
| 3D/4D | Yes | Yes | Yes |
| Monitor size (inches) | • 15" CRT<br>monitor<br>• 15" LCD | 15" Color VGA<br>CRT Monitor | • 15" CRT<br>monitor<br>• 15" LCD<br>• 19" LCD |
| ECG | Optional | Optional | Optional |
| Digital archival<br>capabilities | Yes | Yes | Yes |
| VCR & Video<br>printers | Yes | Yes | Yes |
| M&A capabilities | Yes | Yes | Yes |
| Safety | | | |
| Electrical safety | EN60601-1 | EN60601-1 | EN60601-1 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and body. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
AUG - 7 2006
Esaote Europe BV % Ms. Carri Graham Consultant The Anson Group 11460 N Meridian St, Ste 150 CARMEL IN 46032
Re: K061755
·
Trade Name: MyLab 15/MyLab 20 Systems Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: IYO, IYN, and ITX Dated: June 19, 2006 Received: June 21, 2006
Dear Ms. Graham:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the MyLab 15/MyLab 20 Systems, as described in your premarket notification:
Image /page/4/Picture/10 description: The image shows a logo for the FDA Centennial 1906-2006. The logo is circular and contains the letters FDA in the center. The words "Centennial" are below the letters FDA. The text below the logo reads "Protecting and Promoting Public Health."
{5}------------------------------------------------
#### Transducer Model Number
| IOE323 |
|----------|
| CA123 |
| C5-2 R13 |
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. may publish hat FDA's issuance of a substantial equivalence determination does not mean r loase be devices and i bre broundevice complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any 1 with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI It Fat 607); adoning (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket I notification. The FDA finding of substantial equivalence of your device to a legally marketed nonification. - 11.0 results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain
{6}------------------------------------------------
Page 3 - Ms. Graham
other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
If you have any questions regarding the content of this letter, please contact Ralph Shuping at (301) 594-1212.
Sincerely yours,
Daniel B. Leyron
for
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known):
K 061755
MyLab15/MyLab20 Systems Device Name:
Indications For Use:
Esaote's MyLab15/MyLab20 is a compact console ultrasound system intended to be used by a physician to perform general diagnostic ultrasound studies including Fetal, Abdominal, Pediatric, Small organ, Neonatal Cephalic, Cardiac, Transrectal, Transvaginal, Peripheral Vascular, Musculoskeletal (Conventional and Superficial), Intraoperative: Abdominal, and Other: Urological.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel A. Severson
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K061755
.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the text "K061755" in a handwritten style on the top line. Below that, the text "MyLab15/20 Systems" is printed in a clear, sans-serif font. The handwritten text appears to be a code or identifier, while the printed text likely refers to a product or system name.
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) | |
|----------------------------------|---|---|---|-----|-----|---------------|-------------------|------------------------|--------------------|---------------------|--|
| Ophthalmic | | | | | | | | | | | |
| Fetal | | P | P | P | | P | P | | P [2] | P[3], P[4],<br>P[5] | |
| Abdominal | | P | P | P | | P | P | | P [2] | P[3], P[4],<br>P[5] | |
| Intraoperative Abdominal | | N | N | N | | N | N | | N[2] | N[3], N[4] | |
| Intraoperative Neurological | | | | | | | | | | | |
| Pediatric | | P | P | P | | P | P | | P [2] | P[3], P[4],<br>P[5] | |
| Small Organ (specify) [1] | | P | P | P | | P | P | | P [2] | P[3], P[4] | |
| Neonatal Cephalic | | P | P | P | | P | P | | P [2] | P[3], P[4] | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | P | P | P | | P | P | | P [2] | P[3] | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | P | P | P | | P | P | | P [2] | P[3], P[4] | |
| Transvaginal | | P | P | P | | P | P | | P [2] | P[3], P[4] | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | P | P | P | | P | P | | P [2] | P[3], P[4] | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | | P | P | | P [2] | P[3], P[4] | |
| Musculo-skeletal Superficial | | P | P | P | | P | P | | P [2] | P[3], P[4] | |
| Other: Urological | | N | N | N | | N | N | | N[2] | N[3], N[4] | |
N=new indication; P=previously cleared by FDA; E= added under Appendix E Additional Comments:
[1] Small organs include Thyroid, Breast and Testicles.
[2] Applicable combined modes: B+M+PW+ CFM+Amplitude Doppler
[3] Tissue Enhancement Imaging (TEI)
[4] 3D Imaging
(5) 4D Imaging
Daniel H. Ingram
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number .
{9}------------------------------------------------
K061755
IOE323
| Clinical Application | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-------------|-------------|---------------------------|------------------------------|------------------------------|-----------------------|--------------------|
| | A | B | M | PWD<br>(PW) | CWD<br>(CW) | Color<br>Doppler<br>(CFM) | Amplitude<br>Doppler<br>(PD) | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | N | N | N | | N | N | | N (2) | N (3) |
| Intraoperative (Abdominal) | | N | N | N | | N | N | | N (2) | N (3) |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) [1] | | N | N | N | | N | N | | N [2] | N [3] |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | | N | N | | N [2] | N [3] |
| Laparoscopic | | | | | | | | | | |
| Muscolo-skeletal<br>Conventional | | N | N | N | | N | N | | N [2] | N [3] |
| Muscolo-skeletal<br>Superficial | | N | N | N | | N | N | | N [2] | N [3] |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
[1] Small organs include Thyroid, Breast and Testicles.
(2) Applicable combined modes: B+M+PW+CFM+PD
[3] Tissue Enhancement Imaging (TEI)
iplease do not write below this line. Continute on another Page if needed)
concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
David be. Semon
(Division Sign-Off) Division of Reproductive, Abdoo and Radiological Devices 510(k) Numb
ート アルバイト (1) 2017 - 10:20
{10}------------------------------------------------
K061755
CA123
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-------------|-------------|---------------------------|------------------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD<br>(PW) | CWD<br>(CW) | Color<br>Doppler<br>(CFM) | Amplitude<br>Doppler<br>(PD) | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | N | N | N | | N | N | | N (2) | N [3] |
| Abdominal | | N | N | N | | N | N | | N (2) | N (3) |
| Intraoperative (Adominal) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) [1] | | N | N | N | | N | N | | N (2) | N [3] |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | N | N | N | | N | N | | N (2) | N [3] |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | | N | N | | N (2) | N [3] |
| Laparoscopic<br>Muscolo-skeletal | | | | | | | | | | |
| Conventional<br>Muscolo-skeletal | | | | | | | | | | |
| Superficial | | | | | | | | | | |
| Other: Urological | | N | N | N | | N | N | | N (2) | N [3] |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
[1] Small organs include Thyroid, Breast and Testicles.
[2] Applicable combined modes: B+M+PW+CFM+PD
[3] Tissue Enhancement Imaging (TEI)
(PLEASE DO NOT WRITE BELOW THIS LIKE. CONTINUE OF ANOTHER PAGE IF KELDED)
concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
David A. Severson
(Division Sign-Off) Division of Reproductive, Abdo and Radiological Devices 510(k) Number
ા ર
{11}------------------------------------------------
# Ko61755
C5-2 R13
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-------------|-------------|---------------------------|------------------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD<br>(PW) | CWD<br>(CW) | Color<br>Doppler<br>(CFM) | Amplitude<br>Doppler<br>(PD) | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | N | N | N | | N | N | | N [2] | N [3] |
| Abdominal | | N | N | N | | N | N | | N [2] | N [3] |
| Intraoperative (Adominal) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) [1] | | N | N | N | | N | N | | N [2] | N [3] |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | N | N | N | | N | N | | N [2] | N [3] |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | | N | N | | N [2] | N [3] |
| Laparoscopic | | | | | | | | | | |
| Muscolo-skeletal<br>Conventional | | | | | | | | | | |
| Muscolo-skeletal<br>Superficial | | | | | | | | | | |
| Other: Urological | | N | N | N | | N | N | | N [2] | N [3] |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
[1] Small organs include Thyroid, Breast and Testicles.
[2] Applicable combined modes: B+M+PW+CFM+PD
[3] Tissue Enhancement Imaging (TEI)
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEBOED)
concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
David A. Styrone
(Division Sign-Off) Division of Reproductive, Abdo and Radiological Devices 510(k) Number
16
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.