K061716 · Prestige Ameritech · FXX · Aug 14, 2006 · General, Plastic Surgery
Device Facts
Record ID
K061716
Device Name
PRESTIGE AMERITECH FACE MASK
Applicant
Prestige Ameritech
Product Code
FXX · General, Plastic Surgery
Decision Date
Aug 14, 2006
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
The following Prestige Ameritech Surgical Masks are single use disposable devices intended to be worn in the operating room as well as dental, isolation and other medical procedures to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material.
Device Story
Prestige Ameritech Surgical Masks are three-layer, flat-folded, pleated disposable face masks. Construction includes an inner layer of cellulose or spunbonded polypropylene, a middle meltblown polypropylene filter, and an outer layer of medical-grade tissue or spunbonded polypropylene. Variants include optional anti-fog plastic splash shields. Masks are secured via tie strips or latex-free elastic earloops and feature a malleable aluminum nosepiece for fit. Used in operating rooms, dental, and isolation settings to protect wearers and patients from cross-contamination of fluids and particulates. Bench testing confirms performance against fluid resistance, filtration efficiency, and flammability standards.
Clinical Evidence
Bench testing only. Performance evaluated against ASTM F1862 (Fluid Resistance), ASTM F2299 (Particulate Filtration Efficiency: 98.5% at 0.1 microns), ASTM F2101 (Bacterial Filtration Efficiency: 99.6%), MIL M36954C (Differential Pressure: 2.6), and 16CFR 1610 (Flammability Class 1). Biocompatibility testing performed per ISO 10993-10.
Technological Characteristics
Three-layer construction: cellulose/spunbonded polypropylene inner, meltblown polypropylene filter, medical-grade tissue/spunbonded polypropylene outer. Features: malleable aluminum nosepiece, tie strips or latex-free elastic earloops, optional anti-fog plastic splash shield. Standards: ASTM F1862, ASTM F2299, ASTM F2101, MIL M36954C, 16CFR 1610, ISO 10993-10.
Indications for Use
Indicated for use by healthcare personnel and patients in operating rooms, dental, isolation, and medical settings to provide a barrier against microorganisms, body fluids, and particulates. Single-use disposable.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Predicate Devices
Tucker and Associates Surgical Face Mask (K022256)
Hong Ya Non-Woven Products Co. Ltd. Surgical Face Mask (K060776)
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990.
The assigned 510(k) number is: K061716
- Submitted for: Prestige Ameritech 7425 Airport Frwy. Fort Worth, TX. 76118 817 595-1131 Establishment Registration number: 3005022483 Dan Reese, President Contact Person: 7425 Airport Frwy. Fort Worth, TX. 76118 817 595-1131 Date Prepared: July 19, 2006 Proprietary Name: Prestige Ameritech Face Mask (multiple labels); Prestige Ameritech Earloop Surgical Mask-Blue, Pink, Green, White, Yellow, Peach Prestige Ameritech Earloop Surgical Mask with splash shield-Blue Green, White, Peach Prestige Ameritech Earloop Surgical Mask Tissue-Blue, Pink, White, Yellow Prestige Ameritech Tie on Surgical Mask-Blue, Pink, Green, White, Peach Prestige Ameritech Tie on Surgical Mask with splash shield-Blue, Green, White, Peach Common Name: Surgical Face Mask Classification Name: Mask, Surgical Classification Product Code: FXX Regulation Number: 878.4040 Predicate Devices: Tucker and Associates Surgical Face Mask K022256 Hong Ya Non-Woven Products Co. Ltd. Surgical Face Mask K060776
{1}------------------------------------------------
## KOG1716
Device Description: The Prestige Ameritech Surgical Mask device is composed of three layers of material flat folded and pleated to form the Mask. The inner layer is composed of cellulose or spunbonded polypropylene, the middle layer is a meltblown polypropylene filter material and the outer layer is a medical grade tissue or spunbonded polypropylene. Masks with splash shields have anti-fog plastic shields attached to masks. Masks are held in place on wearer with tie strips or latex free elastic loops and contain a malleable aluminum nosepiece strip. All of the materials used in this device are typical materials commonly used in the construction of Surgical Face Masks and are being used in current legally marketed devices.
Intended Use: The following Prestige Ameritech Surgical Masks are single use disposable devices intended to be worn in the operating room as well as dental, isolation and other medical procedures to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material.
Prestige Ameritech Earloop Surgical Mask-Blue, Pink, Green, White, Yellow, Peach Prestige Ameritech Earloop Surgical Mask with splash shield-Blue Green, White, Peach Prestige Ameritech Earloop Surgical Mask Tissue-Blue, Pink, White, Yellow Prestige Ameritech Tie on Surgical Mask-Blue, Pink, Green, White, Peach Prestige Ameritech Tie on Surgical Mask with splash shield-Blue. Green, White, Peach
| Performance<br>Characteristics | Test Method | Results<br>K022256 | Results<br>K060776 | Results<br>Prestige<br>Ameritech<br>Device |
|--------------------------------------|----------------|--------------------------|-------------------------|--------------------------------------------|
| Fluid Resistance | ASTM<br>F1862 | No Visual<br>Penetration | Fluid<br>Resistant | Fluid<br>Resistant |
| Particulate Filtration<br>Efficiency | ASTM<br>F2299 | 2.0<br>microns | 96.8% at 0.1<br>microns | 98.5% at 0.1<br>Microns |
| Bacterial Filtration<br>Efficiency | ASTM<br>F2101 | 97.9% | 99.9% | 99.6% |
| Differential Pressure | Mil<br>M36954C | 1.8 | 2.34 | 2.6 |
| Flammability Class | 16CFR 1610 | 2 | 1 | 1 |
### Comparison To Predicate Devices:
{2}------------------------------------------------
Discussion of Non-Clinical tests Performed for Determination of Substantial Equivalence:
- 1) Fluid Resistance-ASTM F1862
- 2) Particulate Filtration Efficiency-ASTM F2299
- 3) Bacterial Filtration Efficiency-ASTM F2101
- 4) Differential Pressure-MIL M36954C
- 5) Flammability Testing-16CFR 1610
- 6) Irritation and Sensitization-ISO 10993-10
Conclusions: The Prestige Ameritech Face Masks have the same intended use and technological characteristics as the predicate devices K022256 and K060776. Moreover, bench testing contained in this submission demonstrates that the technological characteristics do not raise any new question of safety or effectiveness. Therefore, the Prestige Face Mask (multiple labels) is substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle is encircled by the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 1 4 2006
Mr. Dan Reese President Prestige Ameritech 7425 Airport Freeway Fort Worth, Texas 76118
Rc: K061716
Trade/Device Name: Prestige Ameritech Face Mask (Multiple Labels) Regulation Number: 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: FXX Dated: July 17, 2006 Received: July 20, 2006
Dear Mr. Reese:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting , your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Lederal Register.
{4}------------------------------------------------
#### Page 2 - Mr. Reesc
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sutte Y. Michael Dms.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# K061716
### Indications for Use
510(k) Number (if known):
**Device Name:**
Device Name;
Indications For Use:
The following Prestige Ameritech Surgical Masks are single use disposable devices intended to be worn in the operating room as well as dental, isolation and other medical procedures to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material.
Prestige Ameritech Earloop Surgical Mask-Blue, Pink, Green, White. Yellow, Peach Prestige Ameritech Earloop Surgical Mask with splash shield-Blue Green, White, Peach Prestige Ameritech Earloop Surgical Mask Tissue-Blue, Pink, White, Yellow Prestige Ameritech Tie on Surgical Mask-Blue, Pink, Green, White, Peach Prestige Ameritech Tie on Surgical Mask with splash shield-Blue, Green, White, Peach
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Shale H. Munday, R.D. 8/9/04
(Sign-Off)
ion of Anesthesiology, General Hospital,
ion Control, Dental Devices
) Number_
Page 1 of *_*_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.