← Product Code [GCJ](/productcode/GCJ) · K061666

# CELLVIZIO WITH CONFOCAL MINIPROBES, MODEL F-600; COLOFLEX AND GASTROFLEX (K061666)

_Mauna Kea Technologies · GCJ · Aug 24, 2006 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K061666

## Device Facts

- **Applicant:** Mauna Kea Technologies
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Aug 24, 2006
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex) is a confocal laser imaging system with fiber optic probes that is intended to allow confocal laser imaging of the internal microstructure of tissues in the anatomical tract, i.e., gastrointestinal or pulmonary, accessed by the endoscope.

## Device Story

Cellvizio F-600 is a confocal laser imaging system for in vivo microscopic tissue visualization. Input: laser light transmitted through miniaturized fiber optic probes (Coloflex, Gastroflex, Alveoflex) inserted into the working channel (≥2.8 mm) of standard endoscopes. Operation: laser scanning unit generates light; fiber bundle transmits light to tissue and collects backscattered signal; proprietary software on remote computer processes signals into real-time images. Output: microscopic tissue images displayed on a flat panel monitor. Used in clinical settings by physicians during endoscopic procedures. Provides real-time visual feedback of internal microstructure to assist in clinical assessment of tissue.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on technological comparison and design validation of the F-600 system relative to the predicate F-400 system.

## Technological Characteristics

Confocal laser scanning system; fiber optic probe-based imaging; compatible with standard endoscopes (≥2.8 mm working channel). Components: Laser Scanning Unit, remote computer, flat panel display, and miniaturized fiber optic probes. Connectivity: remote computer processing. Software-based image reconstruction.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Pentax Confocal Laser System ([K042740](/device/K042740.md))
- Pentax EC-3870CILK Confocal Video Colonoscope ([K042741](/device/K042741.md))
- superDimension Bronchus ([K042438](/device/K042438.md))

## Reference Devices

- F-400 system ([K051585](/device/K051585.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Ko61664

#### Mauna Kea Technologies Special 510(k): F-600 System

#### Premarket Notification Special 510(k) Summary 10.0

In accordance with the Safe Medical Devices Act of 1990 and 21 CFR 807.92, a Premarket Notification 510(k) Summary for the F-600 system Premarket Notification is provided below

### 10.1 Submitter Information

Company Name and Address:

Mauna Kea Technologies 9, rue d'Enghien 75010 Paris France

Contact Name:

Fouad Tarabah, PharmD. Director of Quality and Regulatory Affairs Mauna Kea Technologies Telephone: +33 1 70 08 09 61 Fax: +33 1 48 24 12 18 e-mail: fouad@maunakeatech.com

Date Prepared: May 2006

#### 10.2 Name of Device

Proprietary Name: Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex)

Classification Name: Endoscope and/or Accessories

## 10.3 Predicate Device(s) Information

Pentax Confocal Laser System, K042740

Pentax EC-3870CILK Confocal Video Colonoscope, K042741

superDimension Bronchus, K042438

#### 10.4 Device Description

The Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex) is a confocal microscope with a fiber optic probe which allows in vivo visual inspection of tissues with a microscopic resolution during an endoscopic procedure. The F-600 system has been designed to allow real-

{1}------------------------------------------------

61666

#### Mauna Kea Technologies Special 510(k): F-600 System

time observations of tissues. The device is based on a common laser scanning technology adapted for imaging through a bundle of optical fibers and is thus composed of several elements: a Laser Scanning Unit, proprietary software running on a remote computer, a flat panel display and Miniaturized Fiber Optic Probes. The F-600 system can be used with any legally marketed endoscope with a working channel of 2.8 mm or greater.

### 10.5 Intended Use

The Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex) is a confocal faser system that is intended to allow confocal laser imaging of the internal microstructure of tissues in the anatomical tract, i.e., gastrointestinal or respiratory, accessed by the endoscope.

### 10.6 Comparison to Predicate Device(s)

The comparison to the predicate devices was based on a review of the F-400 system information included in the company's 510(k) Premarket Notification K051585 and information concerning the predicate devices that was available in the public domain. The F-400 system is the Mauna Kea Technologie's legally marketed device (K051585) which serves as the unmodified device for the Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex) special 510(k) submission.

{2}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# AUG 2 4 2006

Mauna Kea Technologies % Fouad Tarabah Director of Quality and Regulatory Affairs 9, Rue D'Enghien Paris, France 75010

Re: K061666

Trade/Device Name: Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: June 20, 2006 Received: June 25, 2006

Dear Fouad Tarabah:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product

{3}------------------------------------------------

Page 2 -- Fouad Tarabah

radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Vabare Preelup

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Mauna Kea Technologies Special 510(k): F-600 System

# ODE Indications Statement

Special 510(k) Number (if known): K061666

Device Name: Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex)

Indications for Use:

The Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex) is a confocal laser imaging system with fiber optic probes that is intended to allow confocal laser imaging of the internal microstructure of tissues in the anatomical tract, i.e., gastrointestinal or pulmonary, accessed by the endoscope.

Prescription Use: ____________________________________________________________________________________________________________________________________________________________

AND/OR

Over-the-Counter Use:

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Tarbar Buchen for wwn
(Division Sign-Off)

Division of General, Restorative, and Neurological Devices

510(k) Number K061666

F-600 System Special 510(k) July 17, 2006

Confidential Page 10 of 63

---

**Source:** [https://fda-staging.innolitics.com/device/K061666](https://fda-staging.innolitics.com/device/K061666)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).