COULTER LH 750 HEMATOLOGY ANALYZER, MODEL 6605632

K061574 · Beckman Coulter, Inc. · GKZ · Jul 21, 2006 · Hematology

Device Facts

Record IDK061574
Device NameCOULTER LH 750 HEMATOLOGY ANALYZER, MODEL 6605632
ApplicantBeckman Coulter, Inc.
Product CodeGKZ · Hematology
Decision DateJul 21, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5220
Device ClassClass 2

Indications for Use

The COULTER LH 750 Hematology Analyzer is a quantitative, automated hematology analyzer and leukocyte differential counter for in vitro diagnostic use in clinical laboratories. The COULTER LH 750 Hematology Analyzer provides automated reticulocyte analysis and enumeration of nucleated red blood cells (NRBCs) as well as an automated method for enumeration of RBCs and WBCs in body fluids.

Device Story

Automated hematology analyzer; performs complete blood cell analysis, leukocyte differential, reticulocyte analysis, NRBC enumeration, and body fluid cell counts. Inputs: blood or body fluid samples. Principle: Coulter Principle (electrical impedance) for cell counting/sizing; VCS technology (volume, conductivity, light scatter) for WBC differential/reticulocyte analysis; single beam photometer for hemoglobinometry. Operation: automated diluting/mixing; manual mode for body fluids. Output: quantitative hematology results. Used in clinical laboratories by trained personnel. Output used by clinicians for diagnostic decision-making regarding hematological status.

Clinical Evidence

No clinical data provided. Substantial equivalence supported by analytical performance data regarding MCV accuracy specifications.

Technological Characteristics

Automated hematology analyzer utilizing Coulter Principle (impedance) and VCS (Volume, Conductivity, Light Scatter) technology. Includes single beam photometer for hemoglobinometry. Reagent-based system (diluents, lytic reagents, cleaners). Software version 2B or higher. Connectivity/integration via standard laboratory workflows.

Indications for Use

Indicated for in vitro diagnostic use in clinical laboratories for quantitative automated hematology analysis, leukocyte differential counting, reticulocyte analysis, nucleated red blood cell (NRBC) enumeration, and enumeration of RBCs and WBCs in body fluids.

Regulatory Classification

Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

Predicate Devices

Submission Summary (Full Text)

{0} # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY INSTRUMENT ONLY TEMPLATE A. 510(k) Number: K061574 B. Purpose for Submission: To obtain clearance for a labeling modification to the COULTER LH 750 Hematology Analyzer to reflect a revised MCV accuracy specification. C. Manufacturer and Instrument Name: Beckman Coulter, Inc. COULTER LH 750 Hematology Analyzer D. Type of Test or Tests Performed: Automated differential cell counter E. System Descriptions: 1. Device Description: The COULTER LH 750 is an automated hematology analyzer capable of supplying a complete blood cells analysis and includes a differential leukocyte cell count. The product also provides automated reticulocyte analysis and enumeration of nucleated red blood cells (NRBCs) as well as an automated method for the enumeration of RBCs and WBCs in body fluids. 2. Principles of Operation: The COULTER LH 750 Hematology Analyzer utilizes the Coulter Principle for enumerating and sizing blood cells, automatic diluting and mixing for sample processing and a sing beam photometer for hemoglobinometry. It used COULTER VCS (volume, conductivity, light scatter) technology for WBC Differential analysis and classification and reticulocyte analysis. The analyzer uses a reagent system consisting of an isotonic diluent, lytic reagents to lyse the red cells without significantly affecting the white cells and an instrument cleaner. All systems include reagents used for reticulocyte staining and analysis. {1} For body fluids analysis, the specimen is aspirated into the LH 750 Analyzer via the manual mode and is diluted in separate WBC and RBC baths. The Coulter Method of counting cells is used to detect and measure changes in electrical resistance when a cell, suspended in a conductive diluent, passes through a small aperture. Each suspended cell acts as an insulator. As the cell passes through the aperture, it momentarily increases the resistance of the electrical path between two submerged electrodes, one located on each side of the aperture. The resistance generates an electrical pulse. The accumulation of electrical pulses are channelized, processed for coincidence correction, and multiplied by a calibration factor, yielding the WBC and RBC counts. The WBC result represents the TNC (total nucleated cell count) in the analysis of body fluids. *Note: Refer to previously cleared 510(k) submissions – K011342 & K030606 3. Modes of Operation: 4. Specimen Identification: 5. Specimen Sampling and Handling: 6. Calibration: 7. Quality Control: 8. Software: FDA has reviewed applicant’s Hazard Analysis and Software Development processes for this line of product types: Yes ________ or No ________ F. Regulatory Information: 1. Regulation section: 21 CFR 864.5220 2. Classification: Class II 3. Product code: GKZ 4. Panel: {2} 81 Hematology G. Intended Use: 1. Indication(s) for Use: The COULTER LH 750 Hematology Analyzer is a quantitative, automated hematology analyzer and leukocyte differential counter for in vitro diagnostic use in clinical laboratories. The COULTER LH 750 Hematology Analyzer provides automated reticulocyte analysis and enumeration of nucleated red blood cells (NRBCs) as well as an automated method for enumeration of RBCs and WBCs in body fluids. 2. Special Conditions for Use Statement(s): H. Substantial Equivalence Information: 1. Predicate Device Name(s) and 510(k) numbers: (a) COULTER LH 750 Hematology Analyzer K011342 (b) COULTER LH 750 Hematology Analyzer with Body Fluids Application K030606 2. Comparison with Predicate Device: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | | LH 750 (modified) | LH 750 (predicate K011342) | | No changes from previously cleared device except the MCV accuracy specification. | | | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | | LH 750 (modified) | LH 750 (predicate K011342) | | MCV Accuracy Specification | ■ Mean % Difference ±2.0%* *Due to the effect of temperature on red cell size, the specification applies to a temperature range of 70 to 80° F. ■ Performance characteristics provided for each reagent system at 60 and 90° F. | Mean % Difference ±2.0% | I. Special Control/Guidance Document Referenced (if applicable): J. Performance Characteristics: {3} *Note: Refer to previously cleared 510(k) submissions – K011342 & K030606 1. Analytical Performance: a. Accuracy: b. Precision/Reproducibility: c. Linearity: d. Carryover: e. Interfering Substances: 2. Other Supportive Instrument Performance Data Not Covered Above: K. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. L. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...