The KSEA Bipolar Electrotome is intended for use in combination with a resectoscope and a sheath by qualified surgeons to resect, ablate or coagulate tissue via a high frequency electrical current waveform during various endoscopic urological procedures. These procedures include bladder tumor diagnosis and resection, transurethral prostatic and bladder biopsy, transurethral prostatic resection, removal of urethral calculus, and treatment of vesical neck constriction.
Device Story
Manual surgical device; consists of working element, single-use cutting loop, and high-frequency cable. Used with 4 mm resectoscope and 24/26 Fr sheath. Operated by qualified surgeons in clinical/OR settings. Delivers high-frequency electrical current to resect, ablate, or coagulate tissue during endoscopic urological procedures. Provides visual/mechanical feedback to surgeon during tissue manipulation; facilitates precise tissue removal or hemostasis. Benefits patient by enabling minimally invasive urological interventions.
Clinical Evidence
Bench testing only.
Technological Characteristics
Manual surgical instrument; stainless steel construction; insulation materials; high-frequency electrical energy source; compatible with 4 mm resectoscope and 24/26 Fr sheaths.
Indications for Use
Indicated for patients undergoing endoscopic urological procedures, including bladder tumor diagnosis/resection, transurethral prostatic/bladder biopsy, transurethral prostatic resection, urethral calculus removal, and treatment of vesical neck constriction.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Karl Storz Endoscopy. The logo is in white text on a dark background. The word "STORZ" is in large, bold letters, with a circle in the middle of the "O". Below the word "STORZ" is the phrase "KARL STORZ ENDOSCOPY" in smaller letters.
K06154/
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.
| Applicant: | Karl Storz Endoscopy - America, Inc. |
|------------|--------------------------------------|
| | 600 Corporate Pointe Drive |
| | Culver City, CA 90230 |
| | (310) 338-8100 |
| | AUG 2 1 2006 |
|--|--------------|
|--|--------------|
Contact:
James A. Lee Senior Regulatory Affairs Specialist
Device Identification: Common Name: Bipolar Working Element/Cutting Loop/Cable
> Trade Name: (optional) The KSEA Bipolar Electrotome
Indication: The KSEA Bipolar Electrotome is intended for use in combination with a resectoscope and a sheath by qualified surgeons to resect, ablate or coagulate tissue via a high frequency electrical current waveform during various endoscopic urological procedures. These procedures include bladder tumor diagnosis and resection, transurethral prostatic and bladder biopsy, transurethral prostatic resection, removal of urethral calculus, and treatment of vesical neck constriction.
Device Description: The KSEA Bipolar Electrotome is a manual surgical device, featuring a working element, a single-use cutting loop, and a high frequency cable. It is designed to be used with a 4 mm resectoscope and a 24 or 26 Fr resectoscope sheath.
Substantial Equivalence: The KSEA Bipolar Electrotome is substantially equivalent to the predicate devices since the basic features, dimensions, stainless steel, insulation materials and intended uses are similar. The minor differences between the KSEA Bipolar Electrotome and the predicate devices raise no new issues of safety and effectiveness, as these differences have no effect on the performance, function or intended use of these devices.
Signed:
A:
James A. Lee, Ph.D. Senior Regulatory Affairs Specialist
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
AUG 2 1 2006
James A. Lee, Ph.D. Senior Regulatory Affairs Specialist Karl Storz Endoscopy-America, Inc. 600 Corporate Pointe CULVER CITY CA 90230
Re: K061541
Trade/Device Name: KSEA Bipolar Electrotome Regulation Number: 21 CFR §876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: II Product Code: FAS Dated: August 3, 2006 Received: August 4, 2006
Dear Dr. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your Bocton 9 rotty production is substantially equivalent (for the indications for reletect above and nave accemined arketed predicate devices marketed in interstate commerce use stated in the enotobare) to togally date of the Medical Device Amendments, or to devices that phot to May 20, 1970, and chances with the provisions of the Federal Food, Drug, and Cosmetic have bech recassified in accordance wif a premarket approval application (PMA). You may, Act (Act) that do flor roquire apple far the general controls provisions of the Act. The general therefore, mailier the device, saloject to an essent for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket IT your device is olassilled (500 a0010) within al controls. Existing major regulations affecting your Approval), it thay of subject to such adeitions, Title 21, Parts 800 to 898. In addition, FDA device can be found in the cements concerning your device in the Federal Register.
Image /page/1/Picture/10 description: The image shows a logo with the text "FDA Centennial 1906-2006". The logo is circular and contains the letters "FDA" in a stylized font. The text "Centennial" is written below the letters. The words "oling Public Health" are written in a decorative font below the logo.
{2}------------------------------------------------
Page 2 -
l,
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Prease be advised that I Dri- 3 issuates of a substition.
that FDA has made a determination that your device complies with other requirements of the Act that I DA has made a dolorimiation administered by other Federal agencies. You must of ally reach statutes and regaranents ancluding, but not limited to: registration and listing (21 Comply with an the Act 31equirements, morations, were and manufacturing practice requirements as set CFK Part 807), labeling (21 CFR Part 820); and if applicable, the electronic (200); and if applicable, the electronic (200 forth in the quality systems (QD) regarders (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) i his letter will anow you to begin manusing of substantial equivalence of your device to a legally premarket notification: "The PDA mialing of casion for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please IT you desire specific advice for your he following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology | 240-276-0115 |
|----------------|---------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" Also, picase note the regulation other general information on your responsibilities under (21CFK Fall 607.77). Tod maj octain offer and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Karl Storz Endoscopy. The logo consists of the word "STORZ" in large, bold letters, with a circular design in the middle of the word. Below the word "STORZ" is the text "KARL STORZ ENDOSCO". The logo is black and white.
510(k) Number (if known): K061541
Device Name: KSEA Bipolar Electrotome
Indications for Use: The KSEA Bipolar Electrotome is intended for use in combination with a resectoscope and a sheath by qualified surgeons to resect, ablate or coagulate tissue with a resercedope and a bitcher courtent waveform during various endoscopic urological via a nigh nequency ocedures include bladder tumor diagnosis and resection, procedures: These prossit and bladder biopsy, transurethral prostatic resection, removal of urethral calculus, and treatment of vesical neck constriction.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use:
OR Over-The-Counter Use:
(Per 21 CFR 801 Subpart D) (Per 21 CFR 801 Subpart C)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
| 510(k) Number | K061541 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.