A & D MEDICAL UM-101 DIGITAL BLOOD PRESSURE MONITOR

K061456 · A&D Engineering, Inc. · DXN · Jun 21, 2006 · Cardiovascular

Device Facts

Record IDK061456
Device NameA & D MEDICAL UM-101 DIGITAL BLOOD PRESSURE MONITOR
ApplicantA&D Engineering, Inc.
Product CodeDXN · Cardiovascular
Decision DateJun 21, 2006
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1130
Device ClassClass 2

Indications for Use

The A&D Medical UM-101 digital blood pressure monitor is intended for use by medical professionals for measuring the systolic and diastolic blood pressure and pulse rate.

Device Story

UM-101 is a digital blood pressure monitor for use by medical professionals. Device utilizes an upper-arm cuff inflated manually by the clinician. Principle of operation is the auscultatory method; clinician regulates deflation rate via a mechanical valve. During deflation, clinician uses a 'MARK' button to record systolic and diastolic pressure points. Device includes an internal counter to store measurement counts, retrievable via the 'MARK' button. Provides blood pressure and pulse rate readings to assist clinical decision-making.

Clinical Evidence

Bench testing only. Device demonstrated compliance with ANSI/AAMI SP-10, EN60601-1, EN60601-2-30, and EN60601-1-2/EN55011 electromagnetic compatibility standards.

Technological Characteristics

Non-invasive blood pressure measurement system. Components: manual inflation cuff, mechanical deflation valve, 'MARK' button for event recording, internal counter. Standards: ANSI/AAMI SP-10, EN60601-1, EN60601-2-30, EN60601-1-2, EN55011. Operates via auscultatory method.

Indications for Use

Indicated for medical professionals to measure systolic and diastolic blood pressure and pulse rate in patients.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K06145.6 # Attachment (D) 510(k) Summary #### DATE PREPARED 1. May 23, 2006 #### SPONSOR INFORMATION 2. A&D Engineering, Inc. Mr. Jerry Wang 1555 McCandless Drive, Milpitas, CA 95035 Tel: 408-518-5113 Fax: 408-635-2313 Email: jwang@andmedical.com #### DEVICE NAME 3. | Proprietary Name: | A&D Medical UM-101 Digital Blood Pressure Monitor | |----------------------|-------------------------------------------------------------------------------------| | Common/Usual Name: | Blood Pressure Monitor | | Classification name: | Non-invasive blood pressure measurement System<br>21 CFR 870-1130, Class II, 74DXN. | #### DEVICE DESCRIPTION AND INTENDED USE 4. The A&D Medical UM-101 digital blood pressure monitor is intended for use by medical professionals for measuring the systolic and diastolic blood pressure and pulse rate. #### PREDICATE DEVCIE 5. A&D LifeSource model UA-704 digital blood pressure monitor with FDA 510(k) K032499. #### TECHNOLOGICAL CHARACTERISTICS 6. UM-101 uses an inflated cuff which is wrapped around the upper arm. The cuff is inflated manually by the medical professionals. The systolic and diastolic blood pressures are determined by auscultatory method. The deflation rate is regulated by a {1}------------------------------------------------ mechanical valve controlled by the medical professionals. There is a quick release valve so that the pressure of the cuff can be completely released at any time during the measurement. During the deflation, the systolic and diastolic can be marked by the medical professionals through a "MARK" button. UM-101 has an internal counter so it remembers the number of measurements. The information can be retrieved by holding the "MARK" button. #### 7. DEVICE TESTING A&D Medical UM-101 digital blood pressure monitor meets the following standards: - . ANSI/AAMI SP-10 standard - European Directive 93/42 EEC for Medical Products . - . EN60601 General Safety Provisions - . EN60601-2-30 Particular Requirements for the Safety of BP Monitor - EN60601-1-2 and EN55011 Electromagnetic Compatibility . {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features an eagle with its wings spread, surrounded by the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" in a circular arrangement. The eagle is stylized with three distinct lines forming its body and wings. The overall design is simple and monochromatic. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 2 1 2006 A&D Engineering, Inc. c/o Jerry Wang Director of Engineering & QA 1555 McCandless Drive Milpitas, CA 959035 Re: K061456 Trade Name: UM-101 Digital Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: II (two) Product Code: DXN Dated: May 24, 2006 Received: May 25, 2006 Dear Mr. Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {3}------------------------------------------------ ### Page 2 -- Mr. Jerry Wang or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Bhimiman for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use K061456 510(k) Number (if known): Device Name: A&D Medical UM-101 Digital Blood Pressure Monitor Indications For Use: Measure blood pressure (systolic and diastolic) and pulse rate. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use × (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Blymmema Sign-Off) Division of Cardloyascular Devices 510(k) Number Page 1 of __ 1 __
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...