PRE-IMPRESSION CONDITIONING SOLUTION

K061427 · Dentsply Intl. · ELW · Jun 26, 2006 · Dental

Device Facts

Record IDK061427
Device NamePRE-IMPRESSION CONDITIONING SOLUTION
ApplicantDentsply Intl.
Product CodeELW · Dental
Decision DateJun 26, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3660
Device ClassClass 2

Indications for Use

The PRE-IMPRESSION CONDITIONING SOLUTION is indicated as a preparation conditioning solution for uniform surface wetting prior to elastomeric impression procedures.

Device Story

Pre-Impression Conditioning Solution is a liquid applied to dental substrates (sulcular tissue, tooth preparations, restorative materials, implant abutments) prior to taking an elastomeric impression. The solution reduces surface tension and contact angle, facilitating uniform wetting of the substrate. This allows the impression material to capture fine preparation details both supra- and subgingivally. Used in a dental clinic setting by a dentist or dental professional. The output is a conditioned, wettable surface that improves the accuracy of the final dental impression, benefiting the patient by ensuring better-fitting dental restorations.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Liquid conditioning solution. Components consist of materials previously used in legally marketed dental devices. No specific materials, energy sources, or software algorithms are described.

Indications for Use

Indicated for use as a preparation conditioning solution to provide uniform surface wetting on substrates including sulcular tissue, tooth preparations, ceramic, composite, or metal core build-up materials, and implant abutments or copings prior to the application of elastomeric impression materials.

Regulatory Classification

Identification

Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.

Special Controls

*Classification.* Class II (Special Controls).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K061427 # 510(k) SUMMARY JUN 2 6 2006 .NAME & ADDRESS: ్రాలు మూలాలు DENTSPLY International World Headquarters Susquehanna Commerce Ctr. 221 West Philadelphia Street York, PA 17405-0872 (717) 845-7511 (voice) (717) 849-4762 (fax) www.dentsply.com K-3 Helen Lewis CONTACT: DATE PREPARED: May 17, 2006 TRADE OR PROPRIETARY NAME: PRE-IMPRESSION CONDITIONING SOLUTION CLASSIFICATION NAME: Impression material, 872.3660 PREDICATE DEVICES: Xeno® III Dental Adhesive, K023776 Nupro® Prophylaxis Paste, K983966 Pretreatment Solution, K940685 Aquasil Easy Mix Putty Impression Material. K973781 #### DEVICE DESCRIPTION: The PRE-IMPRESSION CONDITIONING Solution is a product to be used in the impressiontaking process. When applied to substrates such as sulcular tissue; tooth preparations; ceramic, composite, or metal core build-up material; or implant abutments or copings, it provides a uniform surface condition with reduced surface tension prior to the application of an elastomeric impression material. This standardized surface with reduced contact angle on all impressed substrates allows capture of fine preparation detail both supra and subgingiyally. #### INTENDED USE: The PRE-IMPRESSION CONDITIONING SOLUTION is indicated as a preparation conditioning solution for uniform surface wetting prior to elastomeric impression procedures. ### TECHNOLOGICAL CHARACTERISTICS: All of the components found in PRE-IMPRESSION CONDITIONING SOLUTION have been used in legally marketed devices and are safe for dental use. We believe that the prior use of the components of PRE-IMPRESSION CONDITIONING SOLUTION in legally marketed devices and the data provided, support the safety and effectiveness of PRE-IMPRESSION CONDITIONING SOLUTION for the indicated use. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, with three wavy lines representing the branches of government. The caduceus is surrounded by the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA" in a circular arrangement. The logo is black and white. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # JUN 2 6 6 2006 Ms. Helen Lewis Director of Corporate Compliance and Regulatory Affairs DENTSPLY International Susquehanna Commerce Center West 221 West Philadelphia Street, Suite 60 York, Pennsylvania 17405-0872 Re: K061427 Trade/Device Name: Pre-Impression Conditioning Solution Regulation Number: 21 CFR 872.3660 Regulation Name: Impression Material Regulatory Class: II Product Codes: ELW and MVL Dated: May 17, 2006 Received: May 23, 2006 Dear Ms. Lewis: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices. good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Page 2 - Ms. Helen Lewis Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requiremants of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Paris (201): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product railition control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to 10gal/y marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Chiu S. Lin, PhD Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## INDICATIONS FOR USE STATEMENT (As Required by 21 CFR 807.87(e) 510(K) Number (if known): K061427 #### Device Name: PRE-IMPRESSION CONDITIONING SOLUTION Indications for Use: Preparation conditioning solution for uniform surface wetting prior to elastomeric impression procedures Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Sudor Runa esiology, General Hospital. K0611727
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...