Quantiva™ software system coregisters pairs of anatomic (CT) and functional (PET) volumetric image data and displays the fused images to provide additional combined anatomic plus functional image information to the diagnosing radiologist. The Quantiva™ is a software application intended to co-register and display 2D & 3D multimodality (CT & PET) medical images data. The medical practitioner can visualize, process, render, view, store, print and distribute DICOM 3.0 compliant medical image data within the system and/or across computer networks as distributed locations utilizing standard P.C. hardware. The volume and linear measurement functions are intended for evaluation and quantification of tumor measurements, location/displacement study, analysis and evaluation of both hard and soft tissue. The software also supports interactive segmentation of the region of interest, automated contouring of multi-slice ROI and labeling of avoidance structures during evaluation. Typical users of Quantiva™ are for trained professionals (including but not limited to: radiologists, clinicians and technicians). When interpreted by a trained physician, reviewed images may be used as an element for diagnosis. The Quantiva™ is indicated for use when it is necessary to acquire, record, review and distribute these images. The Quantiva™ is a prescription device. The labeling, instructions and user operations are designed for trained, licensed medical professionals.
Device Story
Quantiva™ is a software application for co-registering and fusing PET and CT volumetric images. It processes DICOM 3.0 compliant data on standard PC hardware. The system performs rigid and non-rigid image fusion, interactive segmentation, automated contouring of multi-slice regions of interest (ROI), and labeling of avoidance structures. It provides volume and linear measurement tools for tumor evaluation and tissue analysis. Operated by radiologists, clinicians, or technicians in clinical settings, the device enables visualization, rendering, and distribution of multimodal images across networks. The output assists physicians in clinical decision-making by providing combined anatomic and functional information for diagnostic purposes.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Software application for standard PC hardware. Performs rigid and non-rigid image fusion, interactive segmentation, and automated contouring. Supports DICOM 3.0 connectivity. Developed under Quality Systems.
Indications for Use
Indicated for trained medical professionals (radiologists, clinicians, technicians) to acquire, record, review, and distribute 2D/3D CT and PET medical images for diagnosis. Used for tumor measurement, location/displacement studies, and analysis of hard/soft tissue.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
# K611418 510(k) Summary
# Quantiva™
JUL 21 2006
| 1 May 2006 | | JUL 21 2006 |
|-----------------------------------------------------------------------------------|--|--------------------------------------------|
| Sponsor | | Consultant |
| Tomographix IP Ltd. | | Mr. Richard Keen |
| 33 Hazelton Avenue, Suite 88 | | Compliance Consultants |
| Toronto, Ontario M5R 2E3 | | 1151 Hope Street |
| Voice 416 638 3103 | | Stamford, CT 06907-1659 |
| Fax 416 924 5884 | | 203 329 2700 F 203 329 2345 |
| audrius@tomographix.com | | rkeen@fda-complianceconsultants.com |
| Proprietary Name: | | Quantiva™ |
| Common Name | | Quantiva™ |
| Device Classification Name | | System, Image Processing Radiological |
| Classification Name: | | Picture Archiving and Communication System |
| Product Code | | LLZ |
| Device Classification | | Class II |
| Establishment registration No. | | Not applicable (foreign manufacture) |
| Predicate Device | | Fusion7D, K033955, Mirada Solutions Ltd. |
| Trademark Notice: All Trademarks used other than those of Tomographix IP Ltd. are | | |
registered to their respective owners.
Confidentiality notice: All data contained in this application and all appendixes provided with his appendix or aided trade secrets or proprietary data which the sponsor requests are treated in accordance with law.
## Device Description
The Quantiva™ is a Class II software application intended to co-register anddisplay fused PET plus CT images enabling a qualified radiological technological technologist to visualize 2D & 3D multimodal (CT and PET) medical image data. The qualified user may process, render, view, store, and print DICOM 3.0 compliant medical image data within the system and/or across computer networks utilizing standard P.C. hardware and software.
#### Intended Use
Quantiva™ software system coregisters pairs of anatomic (CT) and functional (PET) volumetric image data and displays the fused images to provide additional combined anatomic plus functional image information to the diagnosing radiologist.
## Technological Characteristics and Substantial Equivalence
This system creates a rigid and non-rigid fusion of two common diagnostic images. This process results in more diagnostic information than is provided by current methods. The QuantivaTM software has benefited from design, development, testing and production procedures that conform to Quality Systems. Tomographix IP Ltd. has determined that the Quantiva™ software has fundamentally the same indications for use as the predicate device.
#### Performance Testing
Information submitted in this premarket notification for the Quantiva™ software includes results of performance testing.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring a symbol that resembles a person embracing or supporting another, with three curved lines above them.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
# JUL 2 1 2006
Tomographic IP Ltd. % Mr. Richard Keen Responsible Third Party Official Compliance Consultants 1151 Hope Street STAMFORD CT 06907-1659
Re: K061418
Trade/Device Name: Quantiva™ software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: May 1, 2006 Received: May 22, 2006
Dear Mr. Keen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/1/Picture/10 description: The image is a black and white circular logo. The logo has the letters "FDA" in large bold font in the center. Above the letters, the numbers "1906-2006" are printed. Below the letters, the word "Centennial" is printed in a cursive font. Three stars are printed below the word "Centennial". The logo is surrounded by a circular border with text printed around the border.
oting Public -
{2}------------------------------------------------
Page 2 --
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely vours.
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# 510(K) Number (If known): _ K061418_
# Device Name: Quantiva™ software
The Quantiva™is a software application intended to co-register and display 2D & 3D multimodality (CT & PET) medical images data. The medical practitioner can visualize, process, render, view, store, print and distribute DICOM 3.0 compliant medical image data within the system and/or across computer networks as distributed locations utilizing standard P.C. hardware.
The volume and linear measurement functions are intended for evaluation and quantification of tumor measurements, location/displacement study, analysis and evaluation of both hard and soft tissue. The software also supports interactive segmentation of the region of interest, automated contouring of multi-slice ROI and labeling of avoidance structures during evaluation.
Typical users of Quantiva™ are for trained professionals (including but not limited to: radiologists, clinicians and technicians). When interpreted by a trained physician, reviewed images may be used as an element for diagnosis.
The Quantiva™is indicated for use when it is necessary to acquire, record, review and distribute these images. The Quantiva™ is a prescription device. The labeling, instructions and user operations are designed for trained, licensed medical professionals.
David R. Legnon
(Division Sign-Off) Division of Reproductive, Abdomin and Radiological Devic 510(k) Number .
Prescription Use _ XXX (Part 21 CFR 801 Subpart D)
AND/OR
Over - The - Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.