EXPEDITIONARY DELPLOYABLE OXYGEN CONCENTRATION SYSTEM (EDOCS), MODELS 120B AND 30;MOBILE OXYGEN STORAGE TANK (MOST), MOD
Applicant
Pacific Consolidated Industries, LLC
Product Code
CAW · Anesthesiology
Decision Date
Jan 22, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Expeditionary Deployable Oxygen Concentration System (EDOCS-120B) is intended to generate and deliver USP grade 93% +/- 3% oxygen for supplemental oxygen use. This device is to be used only by trained personnel in disaster relief situations where bottled oxygen is not readily available. The Mobile Oxygen Storage Tank (MOST) is intended to store and dispense USP grade 93% (+7%/-3%) oxygen at 50psig nominal pressure for supplemental oxygen use. This device is to be used only by trained personnel in disaster relief situations where bottled oxygen is not readily available.
Device Story
EDOCS: Electromechanical system; draws ambient air; uses molecular sieve process to adsorb water/nitrogen; produces 93% ±3% USP grade oxygen; includes integral oxygen analyzer for purity control; output 30-500 LPM at 50-100 psi; fills high-pressure cylinders. MOST: Storage/distribution system; composite wound, brass-lined cylinders; distribution manifold in waterproof case; stores 10,000-13,400 liters oxygen; 225 LPM max flow; 3 output fittings. Both systems used in disaster relief/military deployments by trained personnel. Output provides supplemental oxygen to patients; aids clinical decision-making by ensuring oxygen availability in austere environments.
Clinical Evidence
Bench testing only. Verification and validation activities confirmed performance against US Army (USAMMA) and USAF (AFMESA) specifications, DOT-CFFC, NASA, and MIL-STD-810F standards. Testing covered purity, flow rate, electrical safety, mechanical integrity, and control functionality.
Technological Characteristics
EDOCS: Molecular sieve oxygen concentrator, compressor modules, oxygen collection cylinders. MOST: Composite wound, brass-lined cylinders, distribution manifold. Standards: USP 24-NF 19, DOT-CFFC, IEC 60601-1-2 (EMC), ISO 14971 (Risk), MIL-STD-810F. Integral oxygen analyzer for purity monitoring. Electromechanical operation.
Indications for Use
Indicated for supplemental oxygen therapy in disaster relief settings where bottled oxygen is unavailable. For use by trained personnel only. Contraindicated in oil-rich or flammable atmospheres.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
Predicate Devices
Deployable Oxygen Concentration System (DOCS) (K020330)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Pacific Consolidated Industries. The logo consists of a stylized triangle with the letters "PCI" embedded within it. To the right of the logo is the text "Pacific Consolidated Industries" in a bold, sans-serif font.
K061414
# 510(k) Summary
JAN 2 2 2007
# Expeditionary Deployable Oxygen Concentration System (EDOCS) & Mobile Oxygen Storage Tank (MOST)
| Type of FDA Submission | Traditional 510(k) |
|------------------------|-----------------------------------------------------------|
| Submitter Information | |
| Submitter's Name: | Pacific Consolidated Industries |
| Submitter's Address: | 12201 Magnolia Avenue<br>Riverside, California 92503-4820 |
| Owner/Operator ID: | 9049531 |
| Contact Person: | Gary W. Clark |
| Submitter's Phone: | (951) 479-0872 (Phone)<br>(951) 479-0861 (Fax) |
| Date of Preparation: | January 12, 2007 |
| Device Classification | |
| Classification Name: | Portable Oxygen Generator |
| Device Classification: | 868.5440 |
| Product Code: | CAW |
| Panel: | Anesthesiology and Respiratory Therapy Devices |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Pacific Consolidated Industries. The logo consists of a stylized triangle with the letters "P", "C", and "I" incorporated into the design. To the right of the logo is the text "Pacific Consolidated Industries" in a bold, sans-serif font.
# Expeditionary Deployable Oxygen Concentration System (EDOCS)
## Predicate Legally Marketed Device Equivalence
Substantial equivalence is claimed to the legally marketed predicate device previously submitted under K020330 - Deployable Oxygen Concentration System (DOCS).
### Safe Medical Devices Act (SMDA) Statement
The oxygen supplied by the oxygen concentrator is supplemental and is not considered to be life supporting or life sustaining. The failure of the device would not have serious health consequences to the user.
## Intended Use of Device
The Expeditionary Deployable Oxygen Concentration System (EDOCS-120B) is intended to . generate and deliver USP grade 93% +/- 3% oxygen for supplemental oxygen use. This device is to be used only by trained personnel in disaster relief situations where bottled oxygen is not readily available.
### Description of Device
The Expeditionary Deployable Oxygen Concentration System (EDOCS) is a family of devices that draw in normal air and produce oxygen. The EDOCS separates nitrogen from the air, producing an output of concentrated oxygen at 93 percent by the molecular sieve process. The final product contains not less than 90 percent and not more than 96 percent oxygen, by volume. The remainder of the product consists mostly of argon and nitrogen.
EDOCS employs an integral oxygen analyzer to provide control of oxygen purity.
The EDOCS device is an electromechanical device consisting primarily of a molecular sieve type oxygen concentrator, a compressor module(s), and one or more oxygen collection cylinders. The oxygen concentrator operates by adsorbing water and nitrogen from filtered air. Depending on the unit size, the resulting gas has increased oxygen at flow rates from 30 to 500 liters per minute at a pressure of 50 - 100 psi nominal and is capable of high pressure cylinder filling.
### Device Labeling
The EDOCS is clearly labeled with the following:
- 93% ±3% Oxygen .
- Contraindications, for example, oil-rich or flammable atmosphere .
- Rx Only
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Pacific Consolidated Industries. The logo consists of a stylized triangle on the left, with the words "Pacific Consolidated Industries" to the right of the triangle. The triangle is made up of three horizontal lines, with the top line being the shortest and the bottom line being the longest.
## Comparison of Technological Characteristics
The technological characteristics of these devices and their intended use to supply supplemental oxygen are the same as the predicate devices and raise no new questions of safety and effectiveness.
### Special Controls/Conformance to Recognized Standards
The Expeditionary Deployable Oxygen Concentration System (EDOCS) conforms to the recognized standard USP 24-NF 19 (through Supplement Four, December 31, 2001) and employs an integral oxygen analyzer to provide control of oxygen purity in accordance with the standard.
The EDOCS has been tested under IEC 60601-1-2 and has passed the required Electromagnetic Compatibility tests.
A Risk Analysis, as described in ISO 14971, has been conducted on the EDOCS and all identified risks have been managed to the degree possible via design, engineering controls or labeling.
#### Summary of Performance Testing
Verification and validation testing activities were conducted to establish the performance and reliability characteristics of the EDOCS, to demonstrate performance as intended and substantial equivalency to predicate device. Testing involved the following areas:
- Purity I
- 피 Flow Rate
- Electrical Safety ■
- 트 Mechanical
- 해 Controls
- 피 Device Performance
Acceptance criteria were based on US Army specifications (USAMMA), and those established in voluntary standards.
In all instances the device met all required performance criteria and functioned as intended, meeting the acceptance criteria.
#### Conclusions
In summary, Pacific Consolidated Industries has demonstrated that the EDOCS meets its specifications, is safe and effective for its intended use, and is substantially equivalent to the currently marketed devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Pacific Consolidated Industries. The logo consists of a stylized letter "A" on the left, followed by the company name "Pacific Consolidated Industries" in bold, sans-serif font. The letter "A" is formed by three horizontal lines stacked on top of each other, creating a triangular shape.
# Mobile Oxygen Storage Tank (MOST)
### Predicate Legally Marketed Device Equivalence
Substantial equivalence is claimed to the legally marketed predicate device previously submitted under K040738 - Mobile Oxygen Storage Tank (MOST).
# Safe Medical Devices Act (SMDA) Statement
The oxygen supplied by the MOST is supplemental and is not considered to be life supporting or life sustaining. The failure of the device would not have serious health consequences to the user.
# Description of Device
The Mobile Oxygen Storage Tank (MOST) is a oxygen storage and distribution system consisting of lightweight, high-pressure, composite wound, brass lined cylinders and a distribution manifold secured within a rugged waterproof case. The MOST contains 10,000 liters of oxygen when filled to 2,250 psig (13,400 liters at 3,000 psig). The MOST has a maximum flow capability of 225 LPM, which is equivalent to 75 LPM per Oxygen Output fitting (3 fittings per MOST). These three oxygen output flows would typically be subdivided further by the customer's equipment.
The small footprint, relatively light weight and easy carrying of the MOST system make it optimal for emergency use and military deployments. Eight handles, two on each long side of the MOST and two on the top at each end are provided for ease of transportation. The MOST contains the Apparatus Kit, which includes the equipment required to distribute the oxygen to up to three patients concurrently. The exterior dimensions of the MOST are: a length of 37.25", a width of 27.18", and height of 15.44." The MOST weighs approximately 200 pounds when charged to 2,250 psig.
# Intended Use of Device
The Mobile Oxygen Storage Tank (MOST) is intended to store and dispense USP grade 93% (+7%/-3%) oxygen at 50 psig nominal pressure for supplemental oxygen use. This device is to be used only by trained personnel in disaster relief situations where bottled oxygen is not readily available.
# Device Labeling
The MOST is clearly labeled with the following:
- 93% +7%1-3% Oxygen .
- Contraindications, for example, oil-rich or flammable atmosphere
- Rx Only
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Pacific Consolidated Industries. The logo consists of a stylized letter "A" on the left, followed by the company name in a sans-serif font. The letter "A" is formed by three horizontal lines connecting the two sides of the triangle.
# Comparison of Technological Characteristics
There are no significant changes in the MOST unit in this submission from that of the original submission.
## Special Controls/Conformance to Recognized Standards
The Mobile Oxygen Storage Tank (MOST) conforms to the DOT-CFFC standard.
A Risk Analysis, as described in ISO 14971, has been conducted on the MOST and all identified risks have been managed to the degree possible via design, engineering controls or labeling.
# Summary of Performance Testing
Verification and validation testing activities were conducted to establish the performance, reliability and safety characteristics of the MOST and to demonstrate performance as intended. Testing involved the following areas:
- Purity
- Flow Rate
- Mechanical
- 트 Controls
- Device Performance
- Safety
Acceptance criteria were based on USAF specifications (AFMESA), DOT-CFFC, NASA and MIL-STD-810F.
In all instances the device met all required performance criteria and functioned as intended, meeting the acceptance criteria.
#### Conclusions
In summary, Pacific Consolidated Industries has demonstrated that the MOST meets its specifications, is safe and effective in providing a portable supply of USP oxygen to the intended markets, and is substantially equivalent to the currently marketed device.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with its wings spread, clutching a staff with a snake wrapped around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 2 2007
Mr. Gary Clark Vice President-Regulatory Affairs/ Ouality Assurance Pacific Consolidated Industries, LLC 12201 Magnolia Avenue Riverside, California 92503-4820
Re: K061414
Trade/Device Name: Expeditionary Deployable Oxygen Concentration System (EDOCS-120B) Regulation Number: 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: II Product Code: CAW Dated: January 12, 2007 Received: January 16, 2007
Dear Mr. Clark:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. Clark
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Statement of Indications for Use
# Name: Expeditionary Deployable Oxygen Concentration System (EDOCS-120B)
The Expeditionary Deployable Oxygen Concentration System (EDOCS-120B) is intended to generate and deliver USP grade 93% +/- 3% oxygen for supplemental oxygen use. This device is to be used only by trained personnel in disaster relief situations where bottled oxygen is not readily available.
### Name: Mobile Oxygen Storage Tank (MOST)
The Mobile Oxygen Storage Tank (MOST) is intended to store and dispense USP grade 93% (+7%/-3%) oxygen at 50psig nominal pressure for supplemental oxygen use. This device is to be used only by trained personnel in disaster relief situations where bottled oxygen is not readily available.
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use X (Per 21 CFR 801.109)
Mell Zelen
Anesthesiology, General Hospital
Pain Control, Cranial Devices
Number: K061414
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.