MICROLIFE WRIST WATCH BLOOD PRESSURE MONITOR, MODEL BP3AX1-4U

K061403 · Microlife Intellectual Property GmbH · DXN · Jun 14, 2006 · Cardiovascular

Device Facts

Record IDK061403
Device NameMICROLIFE WRIST WATCH BLOOD PRESSURE MONITOR, MODEL BP3AX1-4U
ApplicantMicrolife Intellectual Property GmbH
Product CodeDXN · Cardiovascular
Decision DateJun 14, 2006
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1130
Device ClassClass 2

Indications for Use

Microlife Wrist Watch Blood Pressure Monitor, Model BP3AX1-4U is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a technique in which an inflatable cuff is wrapped around the wrist. The device can be used in connection with your personal computer (PC) running the Microlife Blood Pressure Analyzer (BPA) software. The memory data can be transferred to the PC by connecting the monitor via cable with the PC.

Device Story

The Microlife Wrist Watch Blood Pressure Monitor (Model BP3AX1-4U) is an automated non-invasive blood pressure monitor. It uses an inflatable wrist cuff and an electronic semiconductor pressure sensor to detect pressure variations, which are converted into electrical signals. The device employs the oscillometric method to calculate systolic and diastolic blood pressure and pulse rate. It features a MAM (Microlife Average Mode) function. The device is intended for use by adults. Users can transfer stored measurement data to a personal computer (PC) via a cable connection for analysis using the Microlife Blood Pressure Analyzer (BPA) software. This allows for long-term tracking and management of blood pressure data, aiding in clinical decision-making by providing a record of measurements.

Clinical Evidence

No new clinical testing was performed for the modified device. The manufacturer relied on clinical data from the predicate device (K040002), which was tested against the ANSI/AAMI SP10-2002 standard for electronic sphygmomanometers. The manufacturer concluded that the modifications (addition of PC-link functionality) did not affect clinical performance results.

Technological Characteristics

Non-invasive oscillometric blood pressure monitor; semiconductor pressure sensor; automatic cuff inflation; factory-set bleed valve for deflation; PC connectivity via cable; MAM (Microlife Average Mode) function; software-based algorithm for signal processing.

Indications for Use

Indicated for adult individuals to measure systolic and diastolic blood pressure and pulse rate via an inflatable wrist cuff.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K061403 P 1\3 # JUN 1 4 2006 Exhibit #1 # 510(K) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. The assigned 510(k) number is: #### 1. Submitter's Identification: Microlife Intellectual Property GmbH, Switzerland Max Schmidheiny-Strasse 201 9435 Heerbrugg / Switzerland Date Summary Prepared: May 18, 2006 Contact: Mr. Gerhard Frick ### 2. Name of the Device: Microlife Wrist Watch Blood Pressure Monitor, Model BP3AX1-4U #### 3. Information for the 510(k) Cleared Device (Predicate Device): Microlife Wrist Watch Blood Pressure Monitor, Model BP3AX1, K#040002. #### 4. Device Description: The Microlife Wrist Watch Blood Pressure Monitor, Model BP3AX1-4U is designed to measure the systolic and diastolic blood pressure and pulse rate of an individual by using a technique in which an inflatable cuff is wrapped around the wrist. Our method to define systolic and diastolic pressures is similar to the auscultatory method but uses an electronic semiconductor pressure sensor rather than stethoscope and mercury manometer. The sensor converts tiny alteration in cuff pressure to electrical signals; by analyzing those signals to define the systolic, diastolic and calculating pulse rate is a well known technique in the market called the "oscillometric method". The device can be used in connection with your personal computer (PC) running the Microlife Blood Pressure Analyzer (BPA) software. The memory data can be transferred to the PC by connecting the monitor via cable with the PC. {1}------------------------------------------------ ### ട. Intended Use: Microlife Wrist Watch Blood Pressure Monitor, Model BP3AX1-4U is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a technique in which an inflatable cuff is wrapped around the wrist. The device can be used in connection with your personal computer (PC) running the Microlife Blood Pressure Analyzer (BPA) software. The memory data can be transferred to the PC by connecting the monitor via cable with the PC. ### 6. Comparison to the 510(k) Cleared Device (Predicate Device): Both devices use the well-known oscillometric method within the software algorithm to determine the systolic and diastolic blood pressure and pulse rate. Wrist cuff is inflated automatically; deflate rate is controlled but a factory set bleed valve and the deflation pressures are transferred via tubing to a sensor in both units. They use semiconductor pressure sensor instead of capacitive pressure sensor to translate the pressure variations to electrical signals that can be interpreted by an integrating circuit. Moreover both devices have a MAM function. The difference between BP3AX1-4U and the predicate device is the addition of a PC function. The device can be used in connection with your personal computer (PC) running the Microlife Blood Pressure Analyzer (BPA) software. The memory data can be transferred to the PC by connecting the monitor via cable with the PC. ## 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows: Testing information demonstrating safety and effectiveness of the Microlife Wrist Watch Blood Pressure Monitor, Model BP3AX1-4U in the intended environment of use is supported by testing that was conducted in accordance with the FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND, which outlines Electrical, Mechanical and Environmental Performance Requirements. The following testing was conducted: - a. Reliability Test StorageTest - b. Reliability Test - Operating Test {2}------------------------------------------------ - Reliability Test Vibration Test ﻥ - d. Reliability Test Drop Test - e. Reliability Test Life Test - f. EMC Test - g. PC-link software BPA Test None of the testing demonstrated any design characteristics that violated the requirements of the Reviewer Guidance or resulted in any safety hazards. It was our conclusion that the Microlife Wrist Watch Automatic Blood Pressure Monitor, Model BP3AX1-4U tested met all relevant requirements of the aforementioned tests. 061403 3 3 ### 8. Discussion of Clinical Tests Performed: ANSI/AAMI SP10-2002 "National Standard for Manual, Electronic or Automated Sphygmomanometers" testing was performed on our predicate device. All relevant sections were addressed and testing conducted. The BP3AX1-4U met all relevant requirements of this standard, as applicable to our modified device. Repeat testing was not performed for the modified device, as clinical testing results were not affected by the changes to the modified device. ### 9. Software information: In keeping with current FDA policy on software level of concern, the modified device is consistent with a moderate level of concern. We provided software documentation in accordance with the FDA November 2005 document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices ". Moreover, the subject device requires the use of off-the-shelf software to operate the PC-link function, and we met all required elements as outlined in FDA's "Off the Shelf Software Guidance Document". ### 10. Conclusions: We have demonstrated that the Microlife Wrist Watch Blood Pressure Monitor, Model BP3AX1-4U, is as safe and effective as our predicate, the Microlife Wrist Watch Blood Pressure Monitor, Model BP3AX1 based on electrical, mechanical and environmental testing results as well as the FDA DCRND November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", and our "Risk Analysis", as supplied with this submission. {3}------------------------------------------------ Image /page/3/Picture/2 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, overlapping each other. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC - 7 2006 Microlife Intellectual Property GmbH c/o Ms. Susan D. Goldstein-Falk Official Correspondent mdi Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, New York 11021 # Re: K061403 Trade Name: Microlife Wrist Watch Blood Pressure Monitor, Model BP3AXI-4U Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: II Product Code: DXN Dated: May 18, 2006 Received: May 19, 2006 Dear Ms. Goldstein-Falk This letter corrects our substantially equivalent letter of June 14, 2006. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ # Page 2 - Ms. Susan D. Goldstein-Falk Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other requirements of the Free any 1 outs with all the Act's requirements, including, but not I coord agencies: - 1 ou nnd listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 manufacturing practice rollicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to continue marketing your device as described in your Section This locer will and w your will be FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), If you court opening at (240) 276-0120. Also, please note the preparation entitled, "Misbranding by reference to premarket notification" (21CFR Part 16 guallon chittion, "Miboranang of a virtuation on your responsibilities under the Act 807.27). Touring obtain of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health {5}------------------------------------------------ # Exhibit# B # Indications for Use | Page | 1 of 1 | |------|--------| |------|--------| | 510(k) Number (if known): | K061403 | |---------------------------|---------| |---------------------------|---------| Device Name: Microlife Wrist Watch Blood Pressure Monitor, Model BP3AX1-4U # Indications For Use: The Microlife Wrist Watch Blood Pressure Monitor, Model BP3AX1-4U with semiconductor pressure sensor is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a technique in which an inflatable cuff is wrapped around the wrist. The device can be used in connection with your personal computer (PC) running the Microlife Blood Pressure Analyzer (BPA) software. The memory data can be transferred to the PC by connecting the monitor via cable with the PC. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) × (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) - Concurrence of CDRH, Office of Device Evaluation (ODE) B/ummmma
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...