K061370 · GE Healthcare · JAK · May 30, 2006 · Radiology
Device Facts
Record ID
K061370
Device Name
CARDIQ FUSION
Applicant
GE Healthcare
Product Code
JAK · Radiology
Decision Date
May 30, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
CardIQ Fusion is intended to provide an optimized non-invasive application to analyze vascular anatomy and pathology, aid in the assessment of functional data e.g. PET perfusion, and aid in tailoring treatment plans based on the fused anatomical and functional information. Anatomical data could be from a set of Computed Tomography (CT) Angiographic images while functional data could be from PET, SPECT, or processed CT data. CardIQ Fusion is a software post-processing package for the Advantage Workstation (AW) platform, PET/CT and CT scanners and PACS reading stations. It is an additional tool for the analysis of 3D CT angiographic cardiac images/data providing a number of display, measurements and batch filming/archive features to study user-selected vessels. Also included is the capability to visualize reformatted CT/PET/SPECT perfusion and viability data. Finally, it provides different ways of visualizing CT anatomy fused with PET/SPECT functional information. With CardIQ Fusion, clinicians have the opportunity to overlay functional information over the (CT) anatomy of a patient's heart, and thus, they can potentially tailor their decision for that particular patient. CardIQ Fusion provides the visualization of the vessels in several different formats including 3D rendering, curved reformats. Once vessels are visualized, tools are available for sizing the vessel, analyzing calcified and non-calcified plaque to determine the densities of plaque within a vessel, measure areas of abnormalities within a vessel (like stenosis, plaque). Functional data could come from PET, SPECT, or processed CT data for perfusion information. The functional and anatomical data could come from the same scanner as in the case of PET/CT scanner; or they could come from separate scanners like stand alone NUC camera and stand alone CT scanner.
Device Story
CardIQ Fusion is a post-processing software package for Advantage Workstation (AW), PET/CT, CT scanners, and PACS. It inputs 3D CT angiographic cardiac images and functional data (PET, SPECT, or processed CT). The device transforms these inputs by overlaying functional perfusion/viability data onto anatomical CT heart structures. It provides 3D rendering, curved reformats, vessel sizing, and plaque density/stenosis analysis. Used by radiologists, cardiologists, and nuclear medicine physicians in clinical settings to visualize fused data; the output assists clinicians in tailoring patient-specific treatment plans. The device benefits patients by providing non-invasive, integrated anatomical and functional assessment of coronary vasculature.
Clinical Evidence
No clinical data provided; substantial equivalence is based on bench testing, software verification, and validation processes.
Technological Characteristics
Software post-processing package for Advantage Workstation (AW), PET/CT, CT scanners, and PACS. Features include 3D rendering, curved reformats, vessel sizing, and plaque analysis. Operates on standard clinical imaging hardware. Software-based analysis.
Indications for Use
Indicated for clinicians (radiologists, nuclear medicine doctors, cardiologists) to analyze cardiac vascular anatomy and pathology, assess functional data (PET/SPECT/CT perfusion/viability), and aid treatment planning via fused anatomical/functional visualization in patients undergoing cardiac imaging.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
MAY 3 0 2006
K061370
Phylof
## GE Healthcare
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h).
- Submitter Larry A. Kroger, Ph.D. Senior Regulatory Programs Manager Tel: (262) 544-3894 Fax: (262) 544-3202 GE Healthcare W-440 3000 North Grandview Blvd. Waukesha, WI 53188 USA Date Prepared: April 19, 2006
## PRODUCT IDENTIFICATION
| Name:<br>Manufacturer: | CardIQ Fusion<br>Classification Name: Accessory to Computed Tomography System per 21 CFR 892-1750<br>GE Medical Systems SCS<br>283, rue de la Minière<br>78533 Buc Cedex, FRANCE |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Distributor: | GE Healthcare, P.O. Box 414, Milwaukee, WI 53210 |
| Marketed Devices | CardIQ Fusion is substantially equivalent to the devices listed below: |
| Model:<br>Manufacturer:<br>Model:<br>Manufacturer:<br>Model:<br>�<br>Manufacturer: | CardIQ Pro, 510(k) #K041267<br>GE Medical Systems SCS, Buc, France<br>CT/PET Fusion, 510(k) # K010336<br>GE Medical Systems SCS, Buc, France<br>CardIO Function, 510(k) # K013422<br>GE Healthcare, 3200 North Grandview Blvd, Waukesha, WI 53188 USA |
| Model:<br>Manufacturer:<br>Model:<br>●<br>1 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - | ECToolbox with HeartFusion, 510(k) # K040141<br>Syntermed, INC, USA<br>Advanced Vessel Analysis II, 510(k) # K060779<br>CD Madinal Cristoma CCC Duo Eronoo |
- GE Medical Systems SCS, Buc, France Manufacturer: .
## Device Description:
CardIQ Fusion is a post processing analysis software package designed to assist Radiologists, Nuclear Medicine Doctors, Cardiologists, and other clinicians in the evaluation and assessment of heart including its coronary vasculature and perfusion conditions.
CardIQ Fusion is a software post-processing package for the Advantage Workstation (AW) platform, PET/CT and CT scanners and PACS reading stations. It is an additional tool for the
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows handwritten text. The text at the top appears to be a code or identifier, possibly "K061370". Below this, the text "page 2 of 3" is visible, indicating this is page 2 of a 3-page document. The handwriting is somewhat stylized and the ink is dark, providing clear contrast against the background.
analysis of 3D CT angiographic cardiac images/data providing a number of display, measurements and batch filming/archive features to study user-selected vessels. Also included is the capability to visualize reformatted CT/PET/SPECT perfusion and viability data. Finally, it provides different ways of visualizing CT anatomy fused with PET/SPECT functional information.
## Indications for Use:
CardIQ Fusion is intended to provide an optimized non-invasive application to analyze vascular anatomy and pathology, aid in the assessment of functional data e.g. PET perfusion, and aid in tailoring treatment plans based on the fused anatomical and functional information. Anatomical data could be from a set of Computed Tomography (CT) Angiographic images while functional data could be from PET, SPECT, or processed CT data.
CardIQ Fusion is a software post-processing package for the Advantage Workstation (AW) platform, PET/CT and CT scanners and PACS reading stations. It is an additional tool for the analysis of 3D CT angiographic cardiac images/data providing a number of display, measurements and batch filming/archive features to study user-selected vessels. Also included is the capability to visualize reformatted CT/PET/SPECT perfusion and viability data. Finally, it provides different ways of visualizing CT anatomy fused with PET/SPECT functional information.
With CardIQ Fusion, clinicians have the opportunity to overlay functional information over the (CT) anatomy of a patient's heart, and thus, they can potentially tailor their decision for that particular patient. CardIQ Fusion provides the visualization of the vessels in several different formats including 3D rendering, curved reformats. Once vessels are visualized, tools are available for sizing the vessel, analyzing calcified and non-calcified plaque to determine the densities of plaque within a vessel, measure areas of abnormalities within a vessel (like stenosis, plaque).
Functional data could come from PET, SPECT, or processed CT data for perfusion information. The functional and anatomical data could come from the same scanner as in the case of PET/CT scanner; or they could come from separate scanners like stand alone NUC camera and stand alone CT scanner.
#### Comparison with Predicate:
CardIO Fusion is substantially equivalent to the predicate devices listed below:
| Device Name | FDA Clearance Number |
|----------------------------|----------------------|
| CardIQ Pro | K041267 |
| CT PET Fusion | K010336 |
| CardIQ Function | K013422 |
| AVA II | K060779 |
| ECToolbox with HeartFusion | K040141 |
#### Adverse Effects on Health:
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows handwritten text that appears to be a combination of numbers and letters. The text includes "K061370" on the top line, followed by what looks like "page 303" on the second line. The handwriting is somewhat stylized, making it slightly difficult to read some of the characters.
The potential hazards are identified in a risk management summary (hazard analysis) and are controlled by:
- · Software Development, Validation and Verification Process to ensure performance to specifications, Federal Regulations and user requirements.
- · Adherence to industry and international standards.
#### Conclusions:
CardIQ Fusion does not result in any new potential safety risks and performs as well as devices currently on the market. GE considers features of the CardIQ Fusion to be equivalent to those of CardIQ Pro, CT PET Fusion, ECToolbox with HeartFusion, CardIQ Function, and Advanced Vessel Analysis II.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
MAY 3 0 2006
GE Healthcare % Mr. Neil E. Devine, Jr. Responsible Third Party Official Intertek Testing Services NA, Inc. 2307 East Aurora Rd, Unit B7 TWINSBURG OH 44087
Re: K061370
Trade/Device Name: CardIQ Fusion Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: May 16, 2006 Received: May 17, 2006
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your Section 910(x) prematice is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate winest the use stated in the encrosuly manation date of the Medical Device Amendments, or to devices that proof to May 26, 1770, the chaomistic date the no visions of the Federal Food, Drug, and Cosmetic liave bech recassified in accordines not a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general therefore, matter the device, babyer ex- ex- general registration, listing of devices, good Controls provisions of the necessariat misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket IT your device is classified (soc accres) additional controls. Existing major regulations affecting your Apploval), it they be subject to out Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/3/Picture/10 description: The image is a black and white circular logo. The logo commemorates the centennial anniversary from 1906 to 2006. The letters "FDA" are prominently displayed in the center of the logo. The word "Centennial" is written below the letters.
oting Public Healt
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); Habeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director. Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
# 510(k) Number (if known): _ K 0 4 137 0
Device Name: CardIQ Fusion
Indications for Use:
ರ
CardIQ Fusion is intended to provide an optimized non-invasive application to analyze vascular anatomy and pathology, aid in the assessment of functional data e.g. PET perfusion, and aid in tailoring treatment plans based on the fused anatomical and functional information. Anatomical data could be from a set of Computed Tomography (CT) Angiographic images while functional data could be from PET, SPECT, or processed CT data.
CardIQ Fusion is a software post-processing package for the Advantage Workstation (AW) platform, PET/CT and CT scanners and PACS reading stations. It is an additional tool for the analysis of 3D CT angiographic cardiac images/data providing a number of display, measurements and batch filming/archive features to study user-selected vessels. Also included is the capability to visualize reformatted CT/PET/SPECT perfusion and viability data. Finally, it provides different ways of visualizing CT anatomy fused with PET/SPECT functional information.
With CardIQ Fusion, clinicians have the opportunity to overlay functional information over the (CT) anatomy of a patient's heart, and thus, they can potentially tailor their decision for that particular patient. CardIQ Fusion provides the visualization of the vessels in several different formats including 3D rendering, curved reformats. Once vessels are visualized, tools are available for sizing the vessel, analyzing calcified and non-calcified plaque to determine the densities of plaque within a vessel, measure areas of abnormalities within a vessel (like stenosis, plaque),
Functional data could come from PET, SPECT, or processed CT data for perfusion information. The functional and anatomical data could come from the same scanner as in the case of PET/CT scanner; or they could come from separate scanners like stand alone NUC camera and stand alone CT scanner.
Prescription Use Over-The-Counter Use ANDIOR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel A. Dorman
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number_ K061370
Page_1_of__1_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.