K061350 · Pck Electronic Industry & Trade Co., Ltd. · LNS · Dec 22, 2006 · Gastroenterology, Urology
Device Facts
Record ID
K061350
Device Name
LITHODIAMOND ULTRA
Applicant
Pck Electronic Industry & Trade Co., Ltd.
Product Code
LNS · Gastroenterology, Urology
Decision Date
Dec 22, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5990
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LithoDiamond ULTRA is intended to fragment urinary stones in the kidney (renal pelvis and renal calyces) and ureter (upper, middle, and lower ureter).
Device Story
Extracorporeal shock wave lithotripter; fragments urinary stones. Inputs: X-ray imaging (via C-arm) and optional ultrasound localization. Operation: Therapy unit generates shock waves via interchangeable electromagnetic or electrohydraulic heads; patient positioned on motorized table. Output: Fragmented stones; visualized by clinician via X-ray monitors. Used in clinical settings by physicians. Device includes therapy unit, isocentric C-arm, and control interface master with server PC. Benefits: Non-invasive stone fragmentation. Modification of Stonelith V5 (K011106); adds dual-modality therapy, motorized shock head, and C-arm imaging.
Clinical Evidence
Bench testing (shock wave characteristics, localization accuracy, road testing) and clinical performance testing conducted per FDA guidance. Device met all specifications and functioned as intended, providing reasonable assurance of safety and effectiveness.
Technological Characteristics
System comprises therapy unit and control interface master. Therapy unit: shock wave generator (electromagnetic/electrohydraulic), motorized patient table, isocentric C-arm, X-ray module (monoblock/collimator), imaging module, and optional ultrasound localization. Control interface: X-ray monitors, control panel, remote monitoring connector, release buttons, foot switch, and server PC. Connectivity: Networked components within housing. Software-controlled.
Indications for Use
Indicated for fragmentation of urinary stones in the kidney (renal pelvis and renal calyces) and ureter (upper, middle, and lower ureter).
Regulatory Classification
Identification
An extracorporeal shock wave lithotripter is a device that focuses ultrasonic shock waves into the body to noninvasively fragment urinary calculi within the kidney or ureter. The primary components of the device are a shock wave generator, high voltage generator, control console, imaging/localization system, and patient table. Prior to treatment, the urinary stone is targeted using either an integral or stand-alone localization/imaging system. Shock waves are typically generated using electrostatic spark discharge (spark gap), electromagnetically repelled membranes, or piezoelectric crystal arrays, and focused onto the stone with either a specially designed reflector, dish, or acoustic lens. The shock waves are created under water within the shock wave generator, and are transferred to the patient's body using an appropriate acoustic interface. After the stone has been fragmented by the focused shock waves, the fragments pass out of the body with the patient's urine.
Special Controls
*Classification.* Class II (special controls) (FDA guidance document: “Guidance for the Content of Premarket Notifications (510(k)'s) for Extracorporeal Shock Wave Lithotripters Indicated for the Fragmentation of Kidney and Ureteral Calculi.”)
United Medical Systems' Piezolith 300 Lithotripter
Submission Summary (Full Text)
{0}------------------------------------------------
K06/350
Page 1 of 3
#### 510(k) SUMMARY
#### PCK's LithoDiamond ULTRA
DEC 2 2 2006
## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
PCK Electronic Industry And Trade Company, LTD, Inc. 1 Organize Sanayi Bolgesi Orhan Isik Cad. No:4 Sincan 06935 Ankara TURKEY
OR
Phone: (+90) (312) 2672046 Facsimile: (+90) (312) 2670609
Contact Person:
Jonathan S. Kahan, Esq. Hogan & Hartson, L.L.P 555 13th Street, NW Washington, DC 20004
(202) 637-5794 Phone: Facsimile: (202) 637-5910
As Regulatory Counsel to PCK Electronic Industry and Trade Company, LTD, Inc.
Cengiz Kabakcı General Manager PCK Electronic Industry and Trade Company, LTD, Inc. 1 Organize Sanayi Bolgesi Orhan Isik Cad. No:4 Sincan 06935 Ankara TURKEY
Phone: 90 (312) 267-2046 Facsimile: 90 (312) 267-0609
Filing Date: May 15, 2006
#### Name of Device and Name/Address of Sponsor
| Proprietary Device Name: | LithoDiamond ULTRA |
|-----------------------------|-----------------------------------------------------|
| Common/Generic Device Name: | Extracorporeal Shock Wave<br>Lithotripter |
| Classification Name: | Lithotripter, extracorporeal shock-wave, urological |
{1}------------------------------------------------
VC61350
Page 2 of =
rns Product Code: Class II Regulatory Class: 21 CFR 876.5990 Regulation Number: PCK Electronic Industry and Trade Sponsor Company, LTD, Inc. 1 Organize Sanayi Bolgesi Orhan Isik Cad. No:4 Sincan 06935
#### Predicate Devices
The LithoDiamond ULTRA Extracorporeal Shock Wave Lithotripter is substantially equivalent to the following currently marketed devices:
Ankara TURKEY
- PCK Stonelith V5 .
- Dornier Compact Delta .
- Siemens Lithostar Modularis .
- Storz MODULITH Model SLX-F2 .
- United Medical Systems' Piezolith 300 Lithotripter .
#### Intended Use / Indications for Use
The LithoDiamond ULTRA is intended to fragment urinary stones in the kidney (renal pelvis and renal calyces) and ureter (upper, middle, and lower ureter).
#### Technological Characteristics
The LithoDiamond ULTRA consists of: (1) a Therapy Unit; and (2) a Control Interface Master.
The Therapy Unit consists of the following submodules: (a) shock wave generator with therapy head; (b) movement unit (patient table and X-ray Isocentric C-Arm; (c) X-ray Component; (d) optional Ultrasound Localization System ("ULS") and (e) accessories. The X-ray Component of the Therapy Unit consists of the following modules: (a) X-ray module (monoblock and collimator), where the X-rays are generated; (b) Imaging Module, where the X-ray Image is detected and processed; (c) Electronics Module, containing control modules and power supplies: and (d) one or two high resolution black and white monitors. The X-ray
{2}------------------------------------------------
K061350
page 3 of 3.
Module and Imaging Module are mounted on a turnable C-Arm. The Electronics Module is mounted within the LithoDiamond ULTRA's Housing.
The Control Interface Master contains: (1) one or two X-ray Monitors; (2) the Control Panel; (3) the Remote Monitoring Connector Box; (4) the Release Buttons for X-ray and Shock Wave Generator; (5) the Foot Switch for the X-ray; and (6) a server PC accommodating the Software.
The LithoDiamond ULTRA is a modification to PCK's cleared Stonelith V5 device (K011106). The principal design changes to the Stonelith V5 lithotripter to form the LithoDiamond ULTRA lithoptriper are as follows: (1) use of interchangeable shock heads to provide electromagnetic and electrohydraulic therapy (in contrast to the existing StoneLith V5 device which delivers only electrohydraulic therapy); (2) replacement of the fixed imaging arm ("U" arm) (used with the Stonelith V5) with an Isocentric C-Arm; (3) changes in the software (used with the Stonelith V5) to accommodate the use of both electrohydraulic and electromagnetic therapy as well as the C-arm; (4) the addition of a motorized shock head; and (5) certain minor changes in the dimensions of the support table.
#### Performance Data
In accordance with FDA's Guidance for the Content of Premarket Notifications (510(k)s) for Extracorporeal Shock Wave Lithotripters Indicated for the Fragmentation of Kidney and Uretal Calculi (August 9, 2000), PCK conducted the following types of performance testing: Shock Wave Characteristics; Localization Accuracy; Road Testing; and Clinical Performance Testing. In all instances, the LithoDiamond ULTRA met its specifications and functioned as intended. The laboratory and clinical data provide reasonable assurance of the safety and effectiveness of the LithoDiamond ULTRA for the extracorporeal fragmentation of urinary stones in the kidney (renal pelvis and renal calvces) and ureter (upper, middle, and lower ureter).
#### Substantial Equivalence
The LithoDiamond ULTRA has the same or similar intended uses and indications for use; similar principles of operation; and similar technological characteristics as its predicate devices. The technological differences between the LithoDiamond ULTRA and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the LithoDiamond ULTRA is as safe and effective as the predicate devices. Thus, the LithoDiamond ULTRA is substantially equivalent to legally marketed extracorporeal shock wave lithotripters.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
PCK Electronic Industry and Trade Company, LTD c/o Jonathan S. Kahan Hogan & Hartson LLP Columbia Square 555 Thirteenth Street, NW WASHINGTON DC 20004
# DEC 2 2 2006
Re: K063150
Trade/Device Name: LithoDiamond® ULTRA Extracorporeal Shock Wave Lithotripter Regulation Number: 21 CFR \$876.5990 Regulation Name: Extracorporeal shock wave lithotripter Regulatory Class: II Product Code: LNS Dated: October 25, 2006 Received: October 25, 2006
Dear Mr. Kahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/3/Picture/10 description: The image is a circular logo or seal. The text "1906-2006" is at the top of the logo. The letters "FDA" are in the center of the logo, with the word "Centennial" underneath. The text around the edge of the circle is difficult to read.
ing and Promoting Public .
{4}------------------------------------------------
Page 2 -
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxx. | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use Statement
510(k) Number: (KOG)350 ____ )
Device Name: LithoDiamond ULTRA
Indications for Use:
To fragment urinary stones in the kidney (renal pelvis and renal calyces) and ureter (upper, middle, and lower ureter).
Prescription Use X (Part 21 C.F.R. 801 Subpart D) AND/OR
Over-The-Counter Use (21 C.F.R. 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
David R. Ingram
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devic 510(k) Number .
Page of
\\\DC - 25846/0001 - 2293324 v6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.