← Product Code [NBW](/productcode/NBW) · K061328

# THINK POSITIVE DIABETES MANAGEMENT SYSTEM (T+DIABETES SYSTEM) (K061328)

_E-San Limited · NBW · Jun 13, 2006 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K061328

## Device Facts

- **Applicant:** E-San Limited
- **Product Code:** [NBW](/productcode/NBW.md)
- **Decision Date:** Jun 13, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1345
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The intended use of the t+ Diabetes System is for transmission and storage of blood glucose data obtained from a LifeScan OneTouch® Ultra® Blood Glucose Monitoring System.

## Device Story

e-San Bluetooth Cradle acts as a wireless communication accessory for LifeScan OneTouch Ultra Blood Glucose Monitoring System; connects via 3.5mm stereo plug to meter serial port; utilizes Bluetooth V1.2 to transmit glucose measurement data to Bluetooth-enabled cellular phones (e.g., Nokia 6230). Device enables patient diary application on phone; facilitates data transfer to remote internet-based database for storage and display. Intended for home use by patients. Battery-powered with power-saving features (auto-off, remote-shutdown, RF TX power control). Device provides transparent serial cable replacement; does not perform glucose testing itself.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Bluetooth cradle; Bluetooth V1.2 wireless transmission; connects to LifeScan OneTouch® Ultra® Blood Glucose Monitoring System; compatible with Bluetooth-enabled cellular phones; data transmission to remote database via internet.

## Regulatory Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

## Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- t+ Diabetes System ([K052343](/device/K052343.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY INSTRUMENT ONLY TEMPLATE

A. 510(k) Number:
k061328

B. Purpose for Submission:
Addition of software allowing data transfer to 3rd party servers

C. Manufacturer and Instrument Name:
e-San Limited, Think Positive (t+) Diabetes Management System (e-San Bluetooth Cradle)

D. Type of Test or Tests Performed:
This device does not directly test glucose. This is an accessory to a previously cleared meter (LifeScan OneTouch® Ultra® Blood Glucose Monitoring System k002134).

E. System Descriptions:

1. Device Description:
The device is the e-San Bluetooth Cradle and it is connected to a LifeScan OneTouch® Ultra® Blood Glucose Monitoring System and uses short-range low power wireless transmission (Bluetooth V1.2) to send the data to a Bluetooth compatible cellular telephone such as the Nokia Model 6230 and provides a patient diary software application. The t+ Diabetes System enables users to store and display data on the cellular phone, and to send data from the cellular telephone to a remote database for storage and display via the internet.

The Cradle is battery powered and fits over and plugs into a 3.5 mm 3-wire stereo socket that is in the end of the OneTouch® Ultra® device.

The Cradle is to be sold over-the-counter (OTC), as it plugs onto OneTouch® Ultra® glucose meter which is also an OTC device.

2. Principles of Operation:
Glucose Monitoring System - Subject of k002134.

The e-San Bluetooth Cradle provides transparent serial cable replacement between LifeScan OneTouch® Ultra® Blood Glucose Monitoring System and a Bluetooth enabled cellular phone, e.g. Nokia 6600, which allows patient's measurement data to be downloaded to a cellular telephone.

It is interfaced to the serial for the LifeScan OneTouch® Ultra® Blood Glucose Monitoring System via 3.5mm stereo plug.

{1}

It is battery operated with a 6 month life span at 6 uses per day. Power-saving mechanisms are employed in order to maximize batter life, e.g. auto-off, remote-shutdown, RF TX power control, etc.

The product is designed to match the ergonomics, size and weight of the OneTouch® Ultra® Blood Glucose Monitoring System while maintain sound mechanical support for the serial interface connector.

In order to minimize the size of the device, it utilizes an internal Bluetooth antenna.

3. Modes of Operation:
Wireless data transfer of stored glucose measurements.

4. Specimen Identification:
Subject of k002134.

5. Specimen Sampling and Handling:
Subject of k002134.

6. Calibration:
Subject of k002134.

7. Quality Control:
Subject of k002134.

8. Software:
FDA has reviewed applicant's Hazard Analysis and Software Development processes for this line of product types:
Yes ☐ X ☐ or No ☐

F. Regulatory Information:

1. Regulation section:
21 CFR § 862.1345, Glucose Test System

2. Classification:
Class II

3. Product code:
NBW

4. Panel:
75 (Clinical Chemistry)

G. Intended Use:

1. Indication(s) for Use:
The e-San Bluetooth Cradle is intended to be used by patients at home. It is physically connected to a LifeScan OneTouch® Ultra® Blood Glucose Monitoring System and wirelessly sends the signals (via Bluetooth V1.2) to a Bluetooth enabled cellular phone such as the Nokia 6230.

The e-San Bluetooth Cradle serves as the remote communication link between the LifeScan OneTouch® Ultra® Blood Glucose Monitoring System and a cellular telephone.

2

{2}

The t+ Diabetes System enables users to store and display data on the cellular phone, and to send data from the cellular telephone to a remote database for storage and display via the internet.

2. Special Conditions for Use Statement(s):
None.

H. Substantial Equivalence Information:

1. Predicate Device Name(s) and 510(k) numbers:
k052343

2. Comparison with Predicate Device:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Cradle | Cradle sends data to cellular phone | Cradle sends data to cellular phone  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Cradle | Cradle sends data to cellular phone which can then send to a 3^{rd} party server for data storage and display. | Predicate does not have this capability.  |

I. Special Control/Guidance Document Referenced (if applicable):
None referenced.

J. Performance Characteristics:

1. Analytical Performance:
a. Accuracy:
Subject of k002134.
b. Precision/Reproducibility:
Subject of k002134.
c. Linearity:
Subject of k002134.
d. Carryover:
Subject of k002134.
e. Interfering Substances:
Subject of k002134.

2. Other Supportive Instrument Performance Data Not Covered Above:

{3}

K. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

L. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

4

---

**Source:** [https://fda-staging.innolitics.com/device/K061328](https://fda-staging.innolitics.com/device/K061328)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com