K061310 · Ziehm Imaging, Inc. · JAK · Jul 7, 2006 · Radiology
Device Facts
Record ID
K061310
Device Name
ZIEHM VARIO 3D DIGITAL MOBILE IMAGING SYSTEM
Applicant
Ziehm Imaging, Inc.
Product Code
JAK · Radiology
Decision Date
Jul 7, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
The ZIEHM VARIO and ZIEHM VARIO 3-D C-arms provide fluoroscopic and spot film imaging of the patient during diagnostic, Interventional and surgical procedures. It is intended to be used whenever the Clinician benefits from Iso-centric positioning and/or intraoperatively generated 3D imaging of high contrast objects (bones and joints), complex anatomical structures such as cervical, thoracic, and lumber regions of the spine and joint fractures of the upper and lower extremities. The design includes features intended for performing vascular, angiographic, cholangiography, endoscopic, urologic, orthopedic, neurologic, and critical care and emergency room procedures. At the discretion of a physician the device may be used for other imaging applications. Radiographic film examinations can be made with an accessory cassette device attached to the Image Intensifier.
Device Story
Mobile C-arm x-ray system; provides fluoroscopic, spot film, and intraoperative 3D imaging. Inputs: x-ray radiation; outputs: 2D/3D digital images displayed on dual LCD monitors. Components: mobile stand with high-frequency generator, x-ray tube, image intensifier, laser localizers; monitor cart with digital image memory, processing, and touch interface. Used in hospitals, clinics, and ORs by physicians, surgeons, cardiologists, radiologists, and technologists. Supports iso-centric positioning for 3D reconstruction of high-contrast anatomical structures. Integrates with Computer Aided Surgery (CAS) systems via interface; supports DICOM 3 and external media storage. Benefits: enables real-time intraoperative visualization of complex anatomy and fractures, facilitating surgical guidance and diagnostic accuracy.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing and compliance with recognized international safety and performance standards for diagnostic x-ray systems and medical electrical equipment.
Technological Characteristics
Mobile C-arm; high-frequency x-ray generator; image intensifier; dual flat-panel LCD monitors. Connectivity: DICOM 3, RS-170/CCIR video, CAS interface. Standards: 21 CFR 1020.30/31, UL 60601-1, IEC 60601-1 series (safety, EMC, radiation protection), IEC 60825-1 (laser). Features integrated laser light localizers and iso-centric positioning mechanism.
Indications for Use
Indicated for patients requiring fluoroscopic and spot film imaging during diagnostic, interventional, and surgical procedures, specifically for high-contrast objects (bones/joints), complex spinal anatomy (cervical, thoracic, lumbar), and extremity fractures. Used in vascular, angiographic, cholangiography, endoscopic, urologic, orthopedic, neurologic, critical care, and emergency procedures.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
SIEMENS Medical Systems AG ISO-C 3D Mobile Surgical C-arm System (K003266)
SIEMENS Medical Solutions USA, ISO-C 3D Mobile C-arm System (K032280)
Submission Summary (Full Text)
{0}------------------------------------------------
#### 510(k) Summarv
# Kob 1310
In accordance with the requirements of 21 CFR \$807.92 the following summary of 510(k) safety and effectiveness information for the Ziehm Vario / 3D Family is being submitted.
#### Date:
April 29, 2006
#### Name of Submitter:
JUL - 7 2006
Ziehm Imaging, Inc. 4181 Latham Street Riverside, CA 92501 (951) 718-2020
#### Corresponding Official:
Richard Westrich. V.P. Product Development, Regulatory Affairs
#### Device Proprietary Name:
Ziehm Vario 3D Mobile Imaging System
#### Classification Name:
System, X-ray, Fluoroscopic, Image-Intensified or System, X-ray, Mobile
#### Common/Usual Names:
Digital Mobile C-Arm, Fluoroscopic imaging System, Digital Mobile Imaging System
#### Substantial Equivalence:
The ZIEHM VARIO 3D is substantially equivalent to the following devices that are currently marketed:
- SIEMENS Medical Systems AG ISO-C 3D Mobile Surgical C-arm System 510(k) K003266 ー
- SIEMENS Medical Solutions USA, ISO-C 3D Mobile C-arm System 510(k) K032280 -
These devices are mobile C-arm x-ray systems intended for fluoroscopic and 3d imaging. The systems include high-voltage x-ray generator and control, x-ray tube, image intensifier, and monitor cart/workstation, with video image displays, 2D and 3D digital image processing and innage storage capability, Computer Aided Surgery (CAS) interface, as well as conventional sports may storage
{1}------------------------------------------------
#### Device Description:
Indications for Use
Kó6 13/0
The ZIEHM VARIO and ZIEHM VARIO 3-D C-arms provide fluoroscopic and spot film imaging of the patient during diagnostic, Interventional and surgical procedures. It is intended to be used whenever the Clinician benefits from Iso-centric positioning and/or intraoperatively generated 3D imaging of high contrast objects (bones and joints), complex anatomical structures such as cervical, thoracic, and lumber regions of the spine and joint fractures of the upper and lower extremities. The design includes features intended for performing vascular, angiographic, cholangiography, endoscopic, urologic, orthopedic, neurologic, and critical care and emergency room procedures. At the discretion of a physician the device may be used for other imaging applications.
Radiographic film examinations can be made with an accessory cassette device attached to the Image Intensifier.
#### User Characteristics:
The device is intended to be used by health care professionals such as physicians, surgeons, cardiologists, radiologists and technologists in hospitals, out-patient clinics and other clinical environments. It is expected that the device will be used on a nearly daily basis. Ziehm Imaging applications specialists and/or qualified site personnel provide on site operator training in the proper use of the device.
#### Device Description
The ZIEHM VARIO 3D has two main units; Mobile Stand and Monitor Cart workstation. The Mobile Stand C-arm consists of a high frequency generator, x-ray tube assembly, image intensifier, and user. touch control interface, C-Profile supporting the generator and image intensifier and Integrated Laser light localizers in the image receptor. The Mobile Stand C-profile provides fixed distance mounting of the generator and image receptor allowing the user rotational and linear movements for positioning the c-arm at various angles and distances for visualization of patient's anatomical structures in addition the device has variable ISO Centric operation, and integrated Iso-centric operations for performing 3D imaging.
The monitor cart workstation supports dual flat panel LCD display monitors, digital image memory device, imaging capture, image processing, and touch control user interface. External Video connection is provided with RS-170 video timing for domestic market, CCIR for International markets. The ZIEHM VARIO 3D also provides optional peripheral connections for such devices as video printers, DICOM 3, CAS connection for image guided surgery and external media storage devices.
{2}------------------------------------------------
## 1661310
The ZIEHM VARIO 3D series mobile x-ray systems were designed to comply with applicable portions of the following standards and regulations for product safety requirements:
- Federal Performance Standard for Diagnostic X-ray Systems 21 CFR 1020.30, 21 CFR -1020.31 and 21 CFR 1020.30
- --UL 60601-1 Medical Electrical Equipment
- IEC 6060'1-1, Medical Electrical Equipment, General Requirements for l Safety
- IEC 60601-1-2, Medical Electrical Equipment, General Requirements for Safety, m Electromagnetic Compatibility
- IEC 60601-1-3, Medical Electrical Equipment, Radiation Protection in Diagnostic X-ray -Equipment
- IEC 60601-1-4, General requirements for safety, Programmable electrical medical --systems.
- IEC 60601-2-7, Medical Electrical Equipment, Safety of HV/X-ray -Generators
- IEC 60601-2-28, Medical Electronic, Particular Requirements for Safety of ﺳﮯ X-ray Source Assemblies, and X-ray Tube Assemblies.
- IEC 60601-2-32, Medical Electrical Equipment, Safety of Associated X-ray --Equipment
- -IEC 60825-1, Safety of Laser Products, Equipment Safety , Requirements, and User Guide
- 93/42/EEC Annex 1 Essential Requirements of the Medical — Devices Directive
- । DIN ISO 14971
#### Conclusion:
Standards:
The ZIEHM VARIO 3D does not raise new questions of safety or effectiveness and is substantially equivalent to the current legally marketed devices SIEMENS ISO-C 3D K003266, and SIEMENS ISO-C 3D K032280.
End of 510(k) Summary
Richard Westrich
Vice President Product Development and Regulatory Affairs Ziehm Imaging, Inc. Riverside, CA 92501
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
JUL - 7 2006
Mr. Richard Westrich Vice President of Product Development and Regulatory Affairs Ziehm Imaging, Inc. 4181 Latham Street RIVERSIDE CA 92501
Re: K061310
Trade/Device Name: Ziehm Vario 3D Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Code: JAA and IZL Dated: April 29, 2006 Received: May 10, 2006
Dear Mr. Westrich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/3/Picture/10 description: The image shows a circular logo with the text "1906 - 2006" at the top. The letters "FDA" are prominently displayed in the center of the logo, with the word "Centennial" written below. The phrase "Promoting Public" is visible at the bottom of the image.
{4}------------------------------------------------
Page 2 -
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050,
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Nancy C. hogdon
Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Ziehm Imaging logo. The logo consists of a circular graphic on the left and the text "ziehm imaging" on the right. The graphic appears to be a stylized representation of a person inside a circle.
### Indications For Use Statement
Applicant:
Ziehm Imaging, Inc.
510(k) Number (if known): K 00 13 10
Device Name: ZIEHM VARIO 3D
Indications for Use:
The ZIEHM VARIO and ZIEHM VARIO 3-D C-arms provide fluoroscopic and spot film imaging of the patient during diagnostic, Interventional and surgical procedures. It is intended to be used whenever the Clinician benefits from Iso-centric positioning and/or intraoperatively generated 3D imaging of high contrast objects (bones and joints), complex anatomical structures such as cervical, thoracic, and lumber regions of the spine and joint fractures of the upper and lower extremities. The design includes features intended for performing vascular, angiographic, cholangiography, endoscopic, urologic, orthopedic, neurologic, and critical care and emergency room procedures. At the discretion of a physician the device may be used for other imaging applications.
Radiographic film examinations can be made with an accessory cassette device attached to the Image Intensifier.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy Snowdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.