GLUCOPLUS BLOOD GLUCOSE MONITORING SYSTEM

K061234 · Glucoplus, Inc. · NBW · Oct 30, 2006 · Clinical Chemistry

Device Facts

Record IDK061234
Device NameGLUCOPLUS BLOOD GLUCOSE MONITORING SYSTEM
ApplicantGlucoplus, Inc.
Product CodeNBW · Clinical Chemistry
Decision DateOct 30, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Indications for Use

The Glucoplus™ Blood Glucose Test System is comprised of Control Solutions and Test Strip biosensors for use only with the Glucoplus™ Blood Glucose Meter. It is for quantitative measurement of the concentration of glucose in capillary whole blood taken from the fingerstick by people with diabetes at home and/or by healthcare professionals in a clinical setting as an aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.

Device Story

GlucoPlus Blood Glucose Monitoring System measures glucose in capillary whole blood. System includes hand-held meter, test strips, and control solutions. User inserts strip, applies finger-tip blood, and meter performs electrochemical assay using glucose oxidase and potassium ferricyanide. Meter software converts electrical current increase into plasma-equivalent glucose concentration, displaying result on LCD. Used at home by patients or in clinical settings by healthcare professionals. Provides quantitative glucose values to aid diabetes management. Benefits include rapid monitoring of blood glucose levels to assess treatment effectiveness.

Clinical Evidence

Performance validated via bench testing and clinical studies. Precision evaluated per CLSI EP5-T2; within-run CV 2.1-3.9%, between-day CV 2.5-3.3%. Linearity confirmed (R=0.999) across 44-596 mg/dL. Consumer study (n=120) and POC study (n=100) compared GlucoPlus to YSI reference and predicate; results showed high correlation (r=0.985-0.997). Interference testing identified potential positive bias for L-dopa, tetracycline, and uric acid.

Technological Characteristics

Electrochemical biosensor using glucose oxidase enzyme. Power: 3V lithium battery (CR 2032). Connectivity: RS232 or USB interface for data download. Measuring range: 40-600 mg/dL. Hematocrit range: 30-55%. Sample volume: 1.5 µL. Test time: <15 seconds.

Indications for Use

Indicated for quantitative measurement of glucose concentration in capillary whole blood from fingersticks by people with diabetes at home or by healthcare professionals in clinical settings to monitor diabetes control effectiveness. Not for diagnosis or screening of diabetes mellitus. Not for use on neonates.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k061234 B. Purpose for Submission: New device C. Measurand: Glucose, home glucose monitoring test D. Type of Test: Quantitative E. Applicant: GlucoPlus Inc. F. Proprietary and Established Names: GlucoPlus Blood Glucose Monitoring System G. Regulatory Information: 1. Regulation section: 21 CFR §862.1345, Blood Glucose Test System, Over-the-Counter 21 CFR §862.1660, Quality Control Material (assayed and unassayed) 2. Classification: Class II Class I, reserved 3. Product code: NBW, system, test, blood glucose, over the counter CGA, glucose oxidase, glucose JJX, single (specified) analyte controls (assayed and unassayed) 4. Panel: Clinical Chemistry (75) H. Intended Use: 1. Intended use(s): The GlucoPlus is intended to measure glucose quantitatively in capillary whole blood taken from the fingerstick. 2. Indication(s) for use: The GlucoPlus Blood Glucose Test System is comprised of Control Solutions and Test Strip biosensors for use only with the GlucoPlus™ Blood Glucose Meter. It is for quantitative measurement of the concentration of glucose in capillary whole {1} blood taken from the fingerstick by people with diabetes at home and or by healthcare professionals in a clinical setting as an aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates. 3. Special conditions for use statement(s): For use only with capillary whole blood taken from the finger This product is intended for over-the-counter and point-of-care use. 4. Special instrument requirements: GlucoPlus™ Blood Glucose Meter I. Device Description: The GlucoPlus Blood Glucose Monitoring System consists of a hand-held blood glucose meter, test strips, and control materials. The user enters a code for each lot of test strips manually; the meter software contains specific calibration information for each code. The meter is turned on by strip insertion; the user then supplies finger-tip blood or control solution to the strip and the meter starts the assay, which completes in 15 seconds. The meter's software converts the results read off the test strip into a plasma glucose concentration and displays the value on the meter's LCD screen. J. Substantial Equivalence Information: 1. Predicate device name(s): k964630 2. Predicate 510(k) number(s): Bayer Glucometer Elite Blood Glucose Meter 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | GlucoPlus | Ascensia ELITE | | Intended Use | Blood glucose monitoring for home and point-of-care | Same | | System Components | Meter, calibration code strip, test strip, check strip, battery, control solutions | Same | | Specimen | Capillary blood | Same, and cleared for arterial and neonatal specimens | | Test Principle/ Enzyme/ Mediator | Electrochemical/ Glucose oxidase/ Potassium ferricyanide | Same/ same/ same | | Calibration | Automatic (code strip) | Same | | Stability | Closed strips and controls: 18 months Opened strips and controls: 3 months | Same | {2} 3 | Operating Range | 50 – 104° F, relative humidity 20 - 80%, | Same | | --- | --- | --- | | Differences | | | | Item | GlucoPlus | Ascensia ELITE | | | | | | Sample Volume | 1.5 ul | 2.0 ul | | Test Time | Within 15 seconds | 30 seconds | | Hematocrit Range | 30 – 55% | 20 – 60% | | Test Range | 40 – 600 mg/dL | 20 – 600 mg/dL | | Memory capacity | 100 test results | 20 test results | ## K. Standard/Guidance Document Referenced (if applicable): | Area of Study | Reference Procedure | Procedure Title | | --- | --- | --- | | Method Comparison/Anticoagulant Studies | CLSI EP9-T | Method Comparisons and Bias Estimations Using Patient Samples. | | Precision | CLSI EP5-T2 | User Evaluation of Precision Performance of Clinical Chemistry Devices | | Linearity | CLSI EP6-P2 | Evaluation of the Linearity of Quantitative Methods | | Interferences/Cross-Reactivity | CLSI EP7-P | Interference Testing in Clinical Chemistry | | Guidance | ISO 14971 | Medical Devices – Application of risk management to medical devices | | | ISO 15197 | In vitro diagnostic test systems – requirements for blood glucose systems | ## L. Test Principle: The test is based on the release of electrical potential after a two-step reaction where glucose and ferricyanide, in the presence of glucose oxidase, are converted into gluconolactone and ferrocyanide. Ferrocyanide, when electrical current is applied, becomes ferricyanide and releases electrons; the increase in current measured by the meter is proportional to the glucose concentration. ## M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Precision testing in the manufacturer’s facility was based on CLSI EP5-T2. {3} Within-run precision was established using whole blood samples from non-diabetic volunteers spiked or glycolyzed to yield 6 glucose concentrations ranging from 40 – 500 mg/dL. Results were compared to YSI values. One strip lot was tested. Within-run precision was tested on twenty replicates of each glucose level. Between-day precision was tested by testing two levels of control materials (approximately 100 and 300 mg/dL) tested four times a day over 20 days. Testing results are summarized in the tables below: GlucoPlus: Within Run Precision | Glucose Conc. by YSI (mg/mL) | Mean Glucose by GlucoPlus (mg/dL) (n=20) | Standard Deviation | % CV | | --- | --- | --- | --- | | 43.5 | 43.7 | 1.7 | 3.9 | | 96.2 | 90.5 | 2.1 | 2.4 | | 138 | 132.6 | 3.4 | 2.6 | | 207 | 198.0 | 4.2 | 2.1 | | 330 | 327.4 | 7.3 | 2.2 | | 507 | 531.6 | 14.8 | 2.8 | GlucoPlus: Between Day Precision | Control Solution | Mean (mg/dL) (n=80) | Std Dev | % CV | | --- | --- | --- | --- | | Level 1 (~100 mg/dL) | 106 | 3.5 | 3.3 | | Level 2 (~300 mg/dL) | 298 | 7.6 | 2.5 | b. Linearity/assay reportable range: Heparinized venous whole blood was spiked or glycolyzed to six concentrations between 44 mg/dL and 596 mg/dL (confirmed by YSI) then tested with the GlucoPlus and YSI. Six samples were tested for each method for each of three lots. Regression analysis showed a linear relationship between the GlucoPlus and the YSI method: $y = 0.972x + 2.27$ , $R = 0.999$ (all lots combined). c. Traceability, Stability, Expected values (controls, calibrators, or methods): Control materials are traceable to YSI standards which are traceable to NIST SRM #917A (Clinical Dextrose). Real-time shelf life studies performed by the manufacturer indicate that control solutions and the unopened test strips have at least a 17 month life-span. Once the control solution or strip container is opened, they have been shown to be stable for 3 months. d. Detection limit: See linearity study above; the meter will report 'LO' for values less than 40 mg/dL. {4} # e. Analytical specificity: Assay interferents were tested in a dose-response manner following CLSI EP7-A guidelines. Aliquots of the blood were supplemented with glucose to a final concentration of $100\mathrm{mg / dL}$ and measured on a YSI analyzer. The interferent was prepared with an appropriate solvent, and spiked into the 100 mg/dL blood. A control pool was prepared by supplementing the blood with solvent minus the interferent. A series of four to five levels that included the maximum concentration of the substance that would be expected to be encountered in clinical practice were used for each interferent. The sponsor defined interference as the level of compound that caused $\geq \pm 10\%$ difference between the interferent-spiked sample and a sample that did not contain interferent. The table below shows the effect of common interferents at the upper end of normal or therapeutic levels on GlucoPlus test levels: Interference at High-Normal or High Therapeutic Levels GlucoPlus System | Interferent | Upper End Therapeutic or Normal Range (mg/dL) | Highest concentration tested with no interference (mg/dL) | | --- | --- | --- | | Acetaminophen | 2 mg/dL | 2 | | Bilirubin | 1.2 mg/dL | 5 | | Cholesterol | 300 mg/dL | 500 | | Creatinine | 1.5 mg/dL | 30 | | Dopamine | N/A | 10 | | Ephedrine | 2 mg/dL | 10 | | Ibuprofen | 4.2 mg/dL | 20 | | L-dopa | 0.3~10 mg/dL | 1 (positive bias) | | Methyl dopa | 0.5 mg/dL | 1 | | Salicylate | 30 mg/dL | 30 | | Tetracycline | 4 mg/dL | 1 (positive bias) | | Tolazamide | 2.5 mg/dL | 3 | | Tolbutamide | 10 mg/dL | 10 | | Triglycerides | 190 mg/dL | 300 | | Ascorbic acid | 2 mg/dL | 2.25 | | Uric acid | 7.7 mg/dL | 5 (positive bias) | These results suggest that L-dopa, tetracycline, and uric acid at physiologically relevant levels may cause blood samples to produce an artificially high value. Care should be exercised by people with gout, {5} undergoing dialysis, taking chemotherapy or radiation treatments, with Parkinson's disease, and those taking tetracycline or related antibiotics. ## Hematocrit Effect: The effect of sample hemoglobin variation on the GlucoPlus system was tested experimentally by preparing samples of known hematocrit (Hct) and spiking aliquots of these samples with three different levels of glucose. These samples were run on the GlucoPlus and YSI; there was less than an ± 14% bias across the claimed range of 30 ~ 55% Hct. The meter was tested at different altitudes to assess the effect of low oxygen levels on meter performance. No effect on performance was found when the two control solutions were tested up to 8800 ft. above sea level. Higher elevations were not tested. The sponsor presented data that supported using the test system between 10°C to 40°C and 20 – 80% relative humidity. f. Assay cut-off: Not applicable. ## 2. Comparison studies: a. Method comparison with predicate device: **GlucoPlus Consumer Studies:** The consumer study was performed at one site with 120 lay-users and a technician. The lay-users ranged in age, education, and were about equally divided between males and females. The native language of most of the participants was English. Labeling materials were in English. Each participant performed their own fingerstick and tested their blood using the instructions in the User's Guide. A trained technician then performed another fingerstick and tested the blood on the same meter. Capillary blood was collected and measured on a YSI analyzer. Samples ranged from 57 – 481 mg/dL. Linear regression analysis of the data yielded the results below: Lay users vs. YSI y = 0.96x - 7.9 r = 0.985 Technician vs. YSI y = 0.95x - 6.6 r = 0.989 **GlucoPlus POC Studies:** The point-of-care (POC) study was performed at three sites on a total of 100 diabetic participants by one technician at each site. Each participant was tested twice – once each with two GlucoPlus meters – and once with the predicate meter. Linear regression analysis of the data yielded the results below: GlucoPlus: POC Study Results | Location | n | GlucoPlus vs. Predicate | r | Sample Range (mg/dL) | | --- | --- | --- | --- | --- | | Home Medical Dept. | 58 | y = 1.05x - 6.3 | 0.990 | 59 – 341 | | Metabolism Dept. | 62 | y = 1.03x - 6.1 | 0.986 | 56 – 392 | {6} | Internal Dept. | 80 | y = 1.05x – 4.8 | 0.997 | 38 – 548 | | --- | --- | --- | --- | --- | | Total | 200 | y = 1.04x – 5.8 | 0.994 | 38 - 548 | In this study, 28 of 28 samples that were $&lt;75 \mathrm{mg/dL}$ were $\pm15 \mathrm{mg/dL}$ and 170 of 172 samples that were $\geq 75 \mathrm{mg/dL}$ were within $\pm20\%$ of the reference value. b. Matrix comparison: This System is cleared for use with fingerstick capillary samples. The meter’s software adjusts the whole-blood glucose reading to a plasma-equivalent reading. 3. Clinical studies: a. Clinical Sensitivity: See Method Comparison Studies section above. b. Clinical specificity: Not applicable. c. Other clinical supportive data (when a. and b. are not applicable): Not applicable. 4. Clinical cut-off: Not applicable. 5. Expected values/Reference range: The normal fasting adult glucose range for a non-diabetic is $70 - 105\mathrm{mg/dL}$. One to two hours after a meal, normal blood glucose levels should be less than 140 mg/dL. A medical professional should determine the range that is appropriate for diabetes patients. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...