ORAL B PROFESSIONALCARE SERIES, ORAL B ADVANCEPOWER SERIES
Applicant
Braun GmbH
Product Code
JEQ · Dental
Decision Date
May 31, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6865
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
To promote good oral hygiene including plaque removal and treating and preventing gingivitis.
Device Story
Rechargeable electric toothbrush; consists of charging unit and rechargeable power unit. Uses oscillating/rotating technology; small round brush head with flagged bristles and power tips for interdental penetration. Oscillation frequency 63-73 Hz; vibration frequency 340 Hz (3800 strokes/min); oscillation angle 42°-60°. Inductively charged; thermoplastic enclosure; anti-slip handle. Used by patients in home setting for daily oral hygiene. Mechanical action facilitates plaque removal and gingivitis management. Wear indicator function included on bristles.
Clinical Evidence
Multiple controlled clinical studies evaluated safety, plaque removal, gingivitis, and bleeding. Results demonstrate effectiveness in treating and preventing gingivitis. Bench testing confirmed product integrity and compliance with UL 1431 standards.
Technological Characteristics
Oscillating/rotating powered toothbrush; Polyamid (PA6.12) bristles; thermoplastic enclosure; inductive charging; 100-120 V, 50-60 Hz supply. Complies with UL 1431. Features wear indicator colorant (FD&C blue No.2).
Indications for Use
Indicated for individuals seeking to promote oral hygiene, remove plaque, and treat or prevent gingivitis.
Regulatory Classification
Identification
A powered toothbrush is an AC-powered or battery-powered device that consists of a handle containing a motor that provides mechanical movement to a brush intended to be applied to the teeth. The device is intended to remove adherent plaque and food debris from the teeth to reduce tooth decay.
{0}------------------------------------------------
#### 510(k) SUMMARY
Image /page/0/Picture/1 description: The image shows a sequence of handwritten characters, which appear to be alphanumeric. The characters are 'K061199', written in a bold, somewhat stylized font. The writing is dark against a light background, making it easily readable.
5.0
Applicant:
Application Correspondent:
Braun GmbH Frankfurter Strasse 145 61476 Kronberg Germany
MAY 3 1 2006
Amanda Heffernan Manager, Requlatory Affairs Clinical Research & Dental affairs The Gillette Company 800 Boylston Street Prudential Tower Building. Floor 46 Boston, MA 02199-8004
Telephone: 617.463.9532 Fax: 617.463.9578 Email: amanda heffernan@gillette.com
Oral-B® AdvancePower™Series powered toothbrush
Oral-B® ProfessionalCare™ Series/
| Date summary prepared: | November 3, 2005 |
|------------------------|------------------|
|------------------------|------------------|
Proprietary name of device:
Generic /classification name: Toothbrush, Powered
Product code (classification): JEQ (Class 1, 21 C.F.R. 872.6865)
#### Legally Marketed Predicate Devices:
- Sonicare® Advance Toothbrush (K040416) . Philips Oral Healthcare, Inc.
- . Oral-B® ProfessionalCare™Series/Oral-B® AdvancePower™Series Braun GmbH
## Device Description and Technological Characteristics:
Oral-B® ProfessionalCare/Oral-B® AdvancePower is a rechargeable electric toothbrush comprised of a charging unit and a rechargeable power unit. The rechargeable power unit is inductively charged and therefore electrically safe in accordance with the Second Edition of the standard for Personal Hygiene and Health Care appliances, UL 1431. The Oral-B® ProfessionalCare/Oral-B® AdvancePower electric toothbrush has an oscillating/rotating technology, a small round brush head that aids tooth-to-tooth cleaning. The brush head is designed to facilitate deeper penetration of interdental areas. The action of the brush head has an oscillation range frequency between 63 to 73 Hz with a vibration frequency of 340 Hz (3800 strokes per minute) and an oscillating range angle of 42° to 60° depending on the toothbrush model.
The brush head design is a small circular brush head and features an arrangement of flagged bristles and power tips. The power tips have tufts of longer bristles on either side of the brush head in the outer ring. These power tips have a wear indicator function and are colored with the FDA approved colorant FD&C blue No.2.
{1}------------------------------------------------
### 510(k) SUMMARY Cont'd
The flagged bristle tufts have tetralocular filaments and are designed to increase the brushing action on tooth surfaces. The bristle material is Polyamid or PA6.12. The toothbrush features a round handle with anti-slip characteristics for better control in wet conditions and offers a variable range of speeds to best meet individual needs.
While power toothbrushes are 510(k) exempt device, we believe the expanded indication for use of treating and preventing gingivitis may exceed the limitation for 510(k) exemption. The expanded indications for use (i.e., treating and preventing gingivitis) are the only changes from the currently marketed devices Oral-B® ProfessionalCare ™Series/Oral-B® AdvancePower™Series. There are no changes to the design, materials or manufacturing process of these devices.
#### Indications for Use:
To promote good oral hygiene, including plaque removal and treating and preventing gingivitis.
#### Testing:
Oral-B® ProfessionalCare ™ Series/Oral-B® AdvancePower™ Series powered toothbrushes have been tested in numerous controlled clinical studies. These trials evaluated oral soft and hard tissue for safety, plaque, gingivitis and bleeding. Collectively, these studies demonstrate that the Oral-B® ProfessionalCare™Series/ Oral-B® AdvancePower™Series powered toothbrush is effective at treating and preventing gingivitis.
Quality assurance testing on the Oral-B® ProfessionalCare ™Series/Oral-B® AdvancePower™Series electric toothbrush has been conducted to ensure the integrity of products.
Oral-B® Professional Care™ Series/Oral-B® AdvancePower™ Series products are rechargeable battery operated toothbrushes. The toothbrush handles are inductively charged so there is no electrical connection between the charger and handle. All electrical components are housed within thermoplastic enclosures and the product is provided with a Listed 1flexible supply cord terminating in a parallel blade attachment plug for connection to a nominal 100-120 V, 50-60 Hz supply source.
Oral-B® Professional Care™ Series/Oral-B® AdvancePower™ Series products are evaluated and comply with the applicable requirements to bear the Underwriters Laboratories Inc. Mark (UL 1431).
#### Conclusions:
The results from these tests support the safety and effectiveness of Oral-B® ProfessionalCare™ Series/Oral-B® AdvancePower™ Series power toothbrush and its substantial equivalence to the predicate device without raising new safety or effectiveness issues.
{2}------------------------------------------------
## 510(k) SUMMARY Cont'd
## Bibliography
- 1. Cronin M, Dembling W, Warren P, King D.; A 3-month clinical investigation comparing the safety and efficacy of a novel electric toothbrush [Braun Oral-B 3D Plaque Remover] with a manual toothbrush. Am J Dent 1998; 11: (Spec Iss):S17-S21.
- 2. Haffajee A, Thompson M, Torresyap G, Guerrero D and Socransky S.; Efficacy of manual and powered toothbrush (1). Effect on clinical parameters. J Clin Periodontol 2001: 28: 937-946
- 3. Warren P, Cugini M.A., Marks P, King D.; Safety, efficacy and acceptability of a new power toothbrush: A 3-month comparative clinical investigation. Am J Dent 2001; 14: 3-7
- Rosema N.A., Timmerman M.F., Piscaer M., Strate J., Warren P.R., Van der Belden U., Van del বা Weijden G.A.; An oscillating/pulsating electric toothbrush versus a high-frequency electric toothbrush in the treatment of gingivitis.
- 5. Van der Weijden G.A., Timmerman M.K., Piscaer M., ljzerman Y., Warre P.R., Van der Velden U.; A comparison of the efficacy of a novel electric toothbrush and a manual toothbrush in the treatment of gingivitis. Am J Dent 1998; 11: (Spec Iss):S23-S28.
- 6. Van der Weijden G.A., Timmerman M.K., Piscaer M., ljzerman Y., Warre P.R., Van der Velden U.; A clinical comparison of three powered toothbrushes. J Clin Periodontol 2002; 29: 1042-1047.
- 7 . Isaacs R.L., Beiswanger B.B., Rosenfield S.T., Crawford J.L., Mau M.S., Eckert G.J., Warren P.R.; A crossover clinical investigation of the safety and efficacy of a new oscillating/rotating electric toothbrush and a high frequency electric toothbrush. Am J Dent 1998; 11: No.1: 7-12
- 8. Conforti N.J. , Chaves E.S., Leibman J.P., Warren P.R., Cugini M.A.; A comparative 3month clinical investigation of the safety and efficacy of a battery-operated and a rechargeable oscillating-rotating power toothbrush. Am J Dent 2001; 14: 59-62.
{3}------------------------------------------------
Image /page/3/Picture/11 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird with three stylized lines representing its wings.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# MAY 3 1 2006
Braun GmbH C/O Ms. Amanda Heffernan Manager, Regulatory Affairs The Gillette Company 800 Boylston Street Prudential Tower Building FL 46 Boston, Massachusetts 02199-8004
Re: K061199
Trade/Device Name: Oral-B® Professional Care™ Series/Oral-B® AdvancePower™ Series Regulation Number: 872.6865 Regulation Name: Powered Toothbrush Regulatory Class: I Product Code: JEQ Dated: March 20, 2006 Received: March 28, 2006
Dear Ms. Heffernan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2 - Ms. Heffernan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Oral-B® Professional Care™ Series/Oral-B® AdvancePower™ Series
Indications for Use: To promote good oral hygiene including plaque removal and treating and preventing gingivitis.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Runge
Anot Anasthesiology, General Hospital, .Jn Control, Dental Devices
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Page 1 of 1 in section 4.0
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.