K061193 · Abbott Laboratories, Inc. · CGX · Jun 23, 2006 · Clinical Chemistry
Device Facts
Record ID
K061193
Device Name
CREATININE
Applicant
Abbott Laboratories, Inc.
Product Code
CGX · Clinical Chemistry
Decision Date
Jun 23, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1225
Device Class
Class 2
Indications for Use
A creatinine test system is a device intended to measure creatinine levels in serum. plasma, and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes
Device Story
Liquid, ready-to-use, two-reagent kit (Sodium Hydroxide, Picric Acid) for quantitative measurement of creatinine in human serum, plasma, or urine. Used on Abbott Aeroset and Architect c8000 Systems. Principle: alkaline pH reaction of creatinine with picrate to form creatinine-picrate complex; rate of absorbance increase at 500 nm proportional to creatinine concentration. Operated by laboratory professionals in clinical settings. Output provides quantitative creatinine levels to assist clinicians in diagnosing/treating renal disease and monitoring dialysis.
Clinical Evidence
Bench testing only. Performance established via precision studies (CLSI EP5-A, EP10-A), linearity, limit of detection/quantitation (CLSI EP17-A), and interference testing. Method comparison against predicate showed high correlation (r=0.995-0.999) for serum and urine across both Aeroset and Architect systems.
Technological Characteristics
Quantitative colorimetric assay; Modified Jaffe method. Reagents: Sodium Hydroxide (0.8 mol/L) and Picric Acid (25.0 mmol/L). Energy source: photometric detection at 500 nm. Form factor: liquid reagent kit for automated clinical chemistry analyzers. Connectivity: integrated with Abbott Aeroset and Architect c8000 systems.
Indications for Use
Indicated for patients requiring creatinine measurement in serum, plasma, or urine for diagnosis and treatment of renal diseases, monitoring renal dialysis, and as a calculation basis for other urine analytes. For professional use.
Regulatory Classification
Identification
A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.
Predicate Devices
Roche creatinine assay (k941837)
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k061193
B. Purpose for Submission:
New device
C. Measurand:
Creatinine
D. Type of Test:
Quantitative, Colorimetric (absorbance change at 500 nm)
E. Applicant:
Abbott Laboratories
F. Proprietary and Established Names:
Abbott Laboratories Clinical Chemistry Creatinine
G. Regulatory Information:
1. Regulation section:
21 CFR § 862.1225
2. Classification:
Class II
3. Product code:
CGX
4. Panel:
75 (Clinical Chemistry)
{1}
H. Intended Use:
1. Intended use(s):
A creatinine test system is a device intended to measure creatinine levels in serum, plasma and urine.
2. Indication(s) for use:
A creatinine test system is a device intended to measure creatinine levels in serum, plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.
3. Special conditions for use statement(s):
For professional use.
4. Special instrument requirements:
Abbott Laboratories Aeroset and Architect c8000 Systems
I. Device Description:
The Aeroset and Architect c8000 creatinine assay is a liquid, ready-to-use, two-reagent kit. R1 contains Sodium Hydroxide at a concentration of 0.8 mol/L and R2 contains Picric Acid at a concentration of 25.0 mmol/L. The same reagent kit may be used on either the Aeroset or Architect c8000 Systems.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Roche creatinine assay
2. Predicate 510(k) number(s):
k941837
3. Comparison with predicate:
{2}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Assay | Same | Automated Quantitative Measurement of Creatinine |
| Methodology | Same | Modified Jaffe |
| Matrices | Same | Serum, Urine, Plasma |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Reportable Range | Serum: 0.20 – 37.00 mg/dL | Serum: 0.20 – 25.00 mg/dL |
| | Plasma: 0.20 – 37.00 mg/dL | Plasma: 0.20 – 25.00 mg/dL |
| | Urine: 5.0 – 740.0 mg/dL | Urine: 0.2 – 650 mg/dL |
# K. Standard/Guidance Document Referenced (if applicable):
CLSI (formerly NCCLS) Document EP9-A2: Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline – Second Edition
CLSI (formerly NCCLS) Document EP5-A: Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline
CLSI (formerly NCCLS) Document EP17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline
CLSI (formerly NCCLS) Document EP10-A: Preliminary Evaluation of Quantitative Clinical Laboratory Methods; Approved Guideline
# L. Test Principle:
Abbott Clinical Chemistry Creatinine is an in vitro diagnostic assay for the quantitation of creatinine in human serum, plasma, or urine. At an alkaline pH, creatinine in the sample reacts with picrate to form a creatinine-picrate complex. The rate of increase in absorbance at 500 nm due to the formation of this complex is directly proportional to the concentration of creatinine in the sample.
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
a. Precision/Reproducibility:
Serum Application: The total precision as well as the precision for each component of variation (between-day, between-run, and within-run) was estimated for creatinine. Two control levels (Level 1 and Level 2) at normal and abnormal analyte concentrations were tested. These controls were evaluated over 20 days, two runs per day, and two replicates per run. Precision
{3}
was reported as the total percent CV. This study was conducted in accordance with CLSI Document EP5-A. Summary results were as follows:
## AEROSET Serum Precision
| Control | | Level 1 | Level 2 |
| --- | --- | --- | --- |
| N | | 80 | 80 |
| Mean (mg/dL) | | 1.20 | 4.66 |
| Within Run | SD | 0.01 | 0.03 |
| | %CV | 0.76 | 0.57 |
| Between Run | SD | 0.01 | 0.04 |
| | %CV | 0.81 | 0.80 |
| Between Day | SD | 0.06 | 0.14 |
| | %CV | 4.83 | 3.02 |
| Total | SD | 0.06 | 0.15 |
| | %CV | 4.95 | 3.18 |
## ARCHITECT Serum Precision
| Control | | Level 1 | Level 2 |
| --- | --- | --- | --- |
| N | | 80 | 80 |
| Mean (mg/dL) | | 1.27 | 4.81 |
| Within Run | SD | 0.02 | 0.02 |
| | %CV | 1.74 | 0.44 |
| Between Run | SD | 0.01 | 0.04 |
| | %CV | 1.01 | 0.92 |
| Between Day | SD | 0.03 | 0.06 |
| | %CV | 2.36 | 1.15 |
| Total | SD | 0.04 | 0.07 |
| | %CV | 3.10 | 1.54 |
Urine Application: Five day precision studies were conducted on the AEROSET and ARCHITECT c8000 Systems in accordance with CLSI Document EP10-A. This study was intended to supplement data obtained from the twenty-day serum precision study and provides a limited assessment of the performance of the assay with the urine matrix. Two control levels (Level 1 and Level 2) at normal and abnormal analyte concentrations were tested for the urine application. These controls were evaluated over five days, two runs per day, and five replicates per run. Precision was reported as the total percent CV.
{4}
5
# AEROSET Urine Precision
| Control | | Level 1 | Level 2 |
| --- | --- | --- | --- |
| N | | 50 | 50 |
| Mean (mg/dL) | | 42.28 | 92.75 |
| Within Run | SD | 0.96 | 1.80 |
| | %CV | 2.26 | 1.94 |
| Between Run | SD | 0.14 | 1.01 |
| | %CV | 0.32 | 1.09 |
| Between Day | SD | 0.32 | 0.85 |
| | %CV | 0.77 | 0.92 |
| Total | SD | 1.02 | 2.23 |
| | %CV | 2.41 | 2.41 |
# ARCHITECT Urine Precision
| Control | | Level 1 | Level 2 |
| --- | --- | --- | --- |
| N | | 50 | 50 |
| Mean (mg/dL) | | 42.59 | 93.91 |
| Within Run | SD | 0.37 | 0.73 |
| | %CV | 0.87 | 0.78 |
| Between Run | SD | 0.11 | 0.56 |
| | %CV | 0.25 | 0.60 |
| Between Day | SD | 0.12 | 0.14 |
| | %CV | 0.28 | 0.15 |
| Total | SD | 0.40 | 0.93 |
| | %CV | 0.94 | 0.99 |
## b. Linearity/assay reportable range:
The sponsor’s reportable range for serum and plasma is 0.20 – 37.00 mg/dL and 5.0 – 740.0 mg/dL for urine.
Linearity of the assay across the measuring range was established by measuring serum and urine samples at concentrations spanning the measuring range of the assay in four replicates. At least one level was included which exceeded the desired linear range. To measure the highest concentration sample, the analyzer’s “linear high” and “>” error codes were suppressed. The
{5}
percent recovery for each sample was determined by dividing the mean observed result by the expected value.
The sponsor also provided data demonstrating equivalence between the analyzer's auto-dilution feature and manual dilutions.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
This assay utilizes the Abbott Multiconstituent Calibrator which was cleared under k981706.
Shelf life, calibration, and on-board stability study protocols and acceptance criteria were described and found to be acceptable.
d. Detection limit:
The functional sensitivity of the Creatinine assay was determined based on the Limit of Quantitation (LOQ). To determine the LOQ, test levels near the linear low for the Creatinine assay were run in replicates of 10, on three instruments, two runs per instrument. The limit of quantitation was defined as the lowest concentration of analyte which has imprecision less than or equal to 20% CV. An internal verification study supported an LOQ of 0.10 mg/dL for the serum application and 2.0 mg/dL for the urine application. The (Limit of Detection (LOD) testing for Creatinine was performed using a study design based on CLSI EP17-A. An internal verification study supported an LOD of 0.05 mg/dL for the serum application, and 1.0 mg/dL for the urine application. The proportions of false positives (α) and false negatives (β) were less than 5% and the limit of blank (LOB) was 0.0224 mg/dL for serum and 0.3732 mg/dL for urine.
Conclusions: The LOQ for Creatinine serum is 0.10 mg/dL and the LOD is 0.05 mg/dL. The LOQ for Creatinine urine is 2.0 mg/dL and the LOD is 1.0 mg/dL.
Limit of Detection (Serum)
| | Instrument | N | Limit of Blank (mg/dL) | SD | Limit of Detection (mg/dL) |
| --- | --- | --- | --- | --- | --- |
| AEROSET | 1 | 20 | 0.0153 | 0.0064 | 0.0258 |
| | 2 | 20 | 0.0197 | 0.0100 | 0.0362 |
| | 3 | 20 | 0.0081 | 0.0067 | 0.0191 |
| ARCHITE | 1 | 20 | 0.0108 | 0.0071 | 0.0224 |
| | 2 | 20 | 0.0224 | 0.0046 | 0.0300 |
| | 3 | 20 | 0.0224 | 0.0051 | 0.0307 |
{6}
7
Limit of Quantitation (Serum)
| | Instrument | N | Mean (mg/dL) | SD | % CV |
| --- | --- | --- | --- | --- | --- |
| AEROSET | 1 | 20 | 0.0721 | 0.0093 | 12.8771 |
| | 2 | 20 | 0.0713 | 0.0127 | 17.8165 |
| | 3 | 20 | 0.0684 | 0.0073 | 10.6171 |
| ARCHITECT | 1 | 20 | 0.0368 | 0.0072 | 19.6566 |
| | 2 | 20 | 0.0457 | 0.0058 | 12.7092 |
| | 3 | 20 | 0.0506 | 0.0057 | 11.2252 |
Limit of Detection (Urine)
| | Instrument | N | Limit of Blank (mg/dL) | SD | Limit of Detection |
| --- | --- | --- | --- | --- | --- |
| AEROSET | 1 | 20 | 0.2557 | 0.1770 | 0.5508 |
| | 2 | 20 | 0.3732 | 0.1673 | 0.6484 |
| | 3 | 20 | 0.2833 | 0.1301 | 0.4973 |
| ARCHITECT | 1 | 20 | 0.3024 | 0.2009 | 0.6329 |
| | 2 | 20 | 0.2856 | 0.2204 | 0.6482 |
| | 3 | 20 | 0.2509 | 0.2466 | 0.6565 |
Limit of Quantitation (Urine)
| | Instrument | N | Mean (mg/dL) | SD | % CV |
| --- | --- | --- | --- | --- | --- |
| AEROSET | 1 | 20 | 1.8140 | 0.1262 | 6.9579 |
| | 2 | 20 | 1.7137 | 0.1741 | 10.1572 |
| | 3 | 20 | 0.7340 | 0.1413 | 19.2540 |
| ARCHITECT | 1 | 20 | 1.8112 | 0.1465 | 8.0874 |
| | 2 | 20 | 2.0207 | 0.2300 | 11.3821 |
| | 3 | 20 | 1.8596 | 0.2018 | 10.8516 |
{7}
e. Analytical specificity:
Interferents, which may falsely elevate or reduce the concentration of an analyte, were tested. Human serum samples at the medical decision level of the analyte and urine samples were spiked with various levels of interferents. Four replicates of each interferent level and four replicates of an unspiked reference sample were run. The percent recovery was determined by dividing the mean result of each interferent sample by the mean result of the reference sample. Testing was performed using the AEROSET System only.
The percent interference was within ± 10% difference for serum samples containing 30 mg/dL bilirubin; 2,000 mg/dL hemoglobin; 1,000 mg/dL Intralipid; 3.0 mg/dL ascorbate; 300 mg/dL glucose; and 10.6 g/dL protein at Medical Decision Level 1.
The percent interference was within ± 10% difference for serum samples containing 30 mg/dL bilirubin; 2,000 mg/dL hemoglobin; 1,000 mg/dL Intralipid; 3.0 mg/dL ascorbate; 600 mg/dL glucose; and 14.7 g/dL protein at Medical Decision Level 2.
Interfering Substances for Serum - Level 1
| Interfering Substance | Interfering Substance Concentration | Target (mg/dL) | Observed (% of Target) |
| --- | --- | --- | --- |
| Bilirubin | 30 mg/dL | 1.549 | 98.11 |
| | 60 mg/dL | 1.549 | 71.87 |
| Hemoglobin | 1,000 mg/dL | 1.403 | 104.79 |
| | 2,000 mg/dL | 1.403 | 108.86 |
| Intralipid | 750 mg/dL | 1.425 | 99.44 |
| | 1,000 mg/dL | 1.425 | 98.39 |
| Ascorbate | 1.5 mg/dL | 1.522 | 99.49 |
| | 3.0 mg/dL | 1.522 | 98.83 |
| Glucose | 300 mg/dL | 1.522 | 107.08 |
| | 600 mg/dL | 1.522 | 115.92 |
| Protein | 10.6 g/dL | 1.537 | 107.97 |
| | 14.3 g/dL | 1.537 | 115.37 |
{8}
Interfering Substances for Serum - Level 2
| Interfering Substance | Interfering Substance Concentration | Target (mg/dL) | Observed (% of Target) |
| --- | --- | --- | --- |
| Bilirubin | 30 mg/dL | 5.334 | 94.91 |
| | 60 mg/dL | 5.334 | 74.71 |
| Hemoglobin | 1,000 mg/dL | 4.701 | 102.07 |
| | 2,000 mg/dL | 4.701 | 103.29 |
| Intralipid | 750 mg/dL | 4.619 | 98.99 |
| | 1,000 mg/dL | 4.619 | 99.10 |
| Ascorbate | 1.5 mg/dL | 5.230 | 99.61 |
| | 3.0 mg/dL | 5.230 | 99.71 |
| Glucose | 300 mg/dL | 4.999 | 101.46 |
| | 600 mg/dL | 4.999 | 103.16 |
| Protein | 10.8 g/dL | 5.571 | 99.13 |
| | 14.7 g/dL | 5.571 | 99.08 |
The percent difference was within $\pm 10\%$ for urine samples containing 6.25 mL/dL acetic acid (8.5 N), 200 mg/dL ascorbate, 250 mg/dL boric acid, 1,000 mg/dL glucose, 2.5 mL/dL hydrochloric acid (6 N), 5.0 mL/dL nitric acid (6 N), 50 mg/dL protein, 1.25 g/dL sodium carbonate, 400 mg/dL sodium fluoride, and 60 mg/dL sodium oxalate.
Interfering Substances for Urine
| Interfering Substance | Interfering Substance Concentration | Target (mg/dL) | Observed (% of Target) |
| --- | --- | --- | --- |
| Acetic Acid (8.5 N) | 6.25 ml/dL | 95.543 | 100.10 |
| Ascorbate | 200 mg/dL | 90.533 | 99.75 |
| Boric Acid | 250 ml/dL | 95.035 | 100.14 |
| Glucose | 1,000 mg/dL | 94.128 | 100.39 |
| Hydrochloric Acid (6 N) | 2.5 ml/dL | 95.043 | 100.75 |
| Nitric Acid (6 N) | 5.0 ml/dL | 95.338 | 99.82 |
| Protein | 50 mg/dL | 95.470 | 101.93 |
| Sodium Carbonate | 1.25 g/dL | 95.805 | 99.96 |
| Sodium Fluoride | 400 mg/dL | 96.150 | 99.78 |
| Sodium Oxalate | 60 mg/dL | 95.678 | 100.50 |
{9}
f. Assay cut-off:
Not Applicable.
2. Comparison studies:
a. Method comparison with predicate device:
AEROSET:
The sponsor compared 117 serum samples ranging from 0.69 to 38.36 mg/dL (as measured by the predicate method). Samples with concentrations greater than 25 mg/dL were diluted X2 on the predicate method but run undiluted on the Aeroset. Linear regression produced a slope of 0.98, a y-intercept of -0.18, and a correlation coefficient of 0.995.
The sponsor also compared 50 urine samples ranging from 7.14 to 401.72 mg/dL (as measured by the predicate method). Linear regression produced a slope of 0.94, a y-intercept of -3.66, and a correlation coefficient of 0.999.
ARCHITECT:
The sponsor compared 117 serum samples ranging from 0.69 to 38.36 mg/dL (as measured by the predicate method). Samples with concentrations greater than 25 mg/dL were diluted X2 on the predicate method but run undiluted on the Architect. Linear regression produced a slope of 0.96, a y-intercept of -0.24, and a correlation coefficient of 0.999
The sponsor also compared 50 urine samples ranging from 7.14 to 401.72 mg/dL (as measured by the predicate method). Linear regression produced a slope of 0.93, a y-intercept of -4.16, and a correlation coefficient of 0.999.
b. Matrix comparison:
Serum, lithium heparin plasma, and sodium heparin plasma are recommended as acceptable matrices in the sponsor’s labeling. To demonstrate comparable performance, the sponsor compared serum from a plain glass tube (baseline) with serum from a gel tube, lithium heparin plasma, lithium heparin plasma from a gel tube, and sodium heparin plasma. All of the recommended matrices showed acceptable performance.
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable.
b. Clinical specificity:
Not Applicable.
{10}
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
Not Applicable.
5. Expected values/Reference range:
The sponsor cites reference ranges from Tietz Fundamentals of Clinical Chemistry, 5th ed, 2001:975.
| SERUM/PLASMA | Range (mg/dL) | Range (μmol/L) |
| --- | --- | --- |
| Cord | 0.6 to 1.2 | 53 to 106 |
| Newborn, 1 to 4 days | 0.3 to 1.0 | 27 to 88 |
| Infant | 0.2 to 0.4 | 18 to 35 |
| Child | 0.3 to 0.7 | 27 to 62 |
| Adolescent | 0.5 to 1.0 | 44 to 88 |
| Adult, Male | 0.7 to 1.3 | 62 to 115 |
| Adult, Female | 0.6 to 1.1 | 53 to 97 |
| URINE | Range (mg/kg/day) | Range (μmol/kg/day) |
| --- | --- | --- |
| Infant | 8 to 20 | 71 to 177 |
| Child | 8 to 22 | 71 to 194 |
| Adolescent | 8 to 30 | 71 to 265 |
| Adult, Male | 14 to 26 | 124 to 230 |
| Adult, Female | 11 to 20 | 97 to 177 |
(Declines with age to 10 mg/kg/day at age 90)
The sponsor recommends that each laboratory determine its own reference ranges based upon its particular locale and patient characteristics.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.