CLINIQA LIQUID QC LIPID CONTROL LEVELS 1 & 2;LINICAL LIPID CALIBRATION VERIFIERS LEVELS A-E FOR OLYMPUS AU SYSTEMS;QALIB
Applicant
Cliniqa Corporation
Product Code
JJY · Clinical Chemistry
Decision Date
May 25, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Indications for Use
CLINIQA Liquid QC Lipid Controls Levels 1 and 2 are intended for use as assayed quality control material for Apolipoprotein A-1, Apolipoprotein B, Cholesterol (total), High Density Lipoprotein, Low Density Lipoprotein, and Triglyceride analysis. CLINIQA LiniCAL Lipid Calibration Verifiers Levels A - E for Olympus AU Systems are assayed, liquid, quality control products which may be used to evaluate the performance of the Olympus AU Systems for Cholesterol (total), High Density Lipoprotein, Low Density Lipoprotein, and Triglyceride at five useful concentrations. CLINIQA QALIBRATE - Lipid Calibrator is intended for use in assay method calibration for clinical chemistry analyzers listed in the product insert.
Device Story
CLINIQA Lipid products consist of human serum-based liquid controls, calibration verifiers, and calibrators. These products are used in clinical laboratories to monitor the performance of automated chemistry analyzers and calibrate lipid assays (Apolipoprotein A-1, Apolipoprotein B, Cholesterol, HDL, LDL, Triglycerides). The products are ready-to-use, containing preservatives and stabilizers. Calibration verifiers are specifically designed for Olympus AU systems to assess performance across the reportable range (Levels A-E). Healthcare providers use the output (control/calibration results) to verify the accuracy and precision of patient sample testing, ensuring reliable clinical decision-making regarding lipid management.
Clinical Evidence
No clinical data. Bench testing only. Stability was validated using the Arrhenius model of accelerated elevated temperature studies to predict storage stability at 2-8°C. Value assignment was performed using consensus testing across multiple laboratories with a minimum of 12 data points per lot.
Technological Characteristics
Liquid, assayed, multi-level control and calibration materials. Formulated for specific clinical chemistry analytes (Apolipoprotein A-1, B, Cholesterol, HDL, LDL, Triglycerides). Designed for use with clinical chemistry analyzers (e.g., Olympus AU systems).
Indications for Use
Indicated for use as quality control material and calibrators for lipid-related clinical chemistry assays. Intended for professional use in clinical laboratory settings on automated chemistry analyzers.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k061182
B. Purpose for Submission:
New Device
C. Measurand:
Lipids
D. Type of Test:
Not Applicable.
E. Applicant:
Cliniqa Corporation
F. Proprietary and Established Names:
CLINIQA Liquid QC Lipid Controls Levels 1 and 2
CLINIQA LiniCAL Lipid Calibration Verifiers Levels A - E for Olympus AU Systems
CLINIQA QALIBRATE Lipid Calibrator
G. Regulatory Information:
1. Regulation section:
21 CFR § 862.1660 (Quality Control Material)
21 CFR § 862.1150 (Calibrator)
2. Classification:
Class I (Quality Control Material)
Class II (Calibrator)
3. Product code:
JJY, multi-analyte controls, all kinds (assayed and unassayed)
JIX, calibrator, multi-analyte mixture
4. Panel:
Clinical Chemistry (75)
H. Intended Use:
1. Intended use(s):
See Indication(s) for use section below.
{1}
2. Indication(s) for use:
CLINIQA Liquid QC Lipid Controls Levels 1 and 2 are intended for use as assayed quality control material for Apolipoprotein A-1, Apolipoprotein B, Cholesterol (total), High Density Lipoprotein, Low Density Lipoprotein, and Triglyceride analysis.
CLINIQA LiniCAL Lipid Calibration Verifiers Levels A - E for Olympus AU Systems are assayed, liquid, quality control products which may be used to evaluate the performance of the Olympus AU Systems for Cholesterol (total), High Density Lipoprotein, Low Density Lipoprotein, and Triglyceride at five useful concentrations.
CLINIQA QALIBRATE - Lipid Calibrator is intended for use in assay method calibration for clinical chemistry analyzers listed in the product insert.
3. Special conditions for use statement(s):
Prescription Use Only
4. Special instrument requirements:
The calibration verifiers are to be used with the Olympus AU systems only. The liquid QC controls and lipid calibrator may be used with a number of automated chemistry analyzers which are listed in the product insert.
I. Device Description:
CLINIQA Liquid QC Lipid Controls Levels 1 and 2 are human serum based, containing constituents of human origin, in liquid form, and are ready to use. Preservatives, stabilizers and sodium azide have been added to maintain product integrity.
CLINIQA LiniCAL Lipid Calibration Verifiers Levels A - E for Olympus AU Systems are human serum based, in liquid form, and are ready to use. The product is manufactured without glycol's thereby affording reproducible pipetting characteristics and minimizing undesirable matrix effects. Preservatives, stabilizers and sodium azide have been added to maintain product integrity. Constituent concentrations in Level A are at the low end to allow assessment of the lower limit of the reportable range. Constituent concentrations in Level E are at the high end and are designed to challenge the upper limit of the reportable range. Levels B, C, and D provide intermediate constituent concentrations over the reportable range.
CLINIQA QALIBRATE Lipid Calibrator is human serum based, containing constituents of human origin and purified chemicals and is a liquid calibrator that
2
{2}
does not require reconstitution. Preservatives, stabilizers and sodium azide have been added to maintain product integrity.
# J. Substantial Equivalence Information:
1. Predicate device name(s): MAS Lipid Control, LiniCAL Chemistry Calibration Verifiers Levels A - E for Olympus AU Systems, Roche Calibrator for Automated Systems.
2. Predicate 510(k) number(s): k030679, k033162, and k011658 respectively.
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Lipid Control/Calibrator Verifier/Calibrator | Lipid Control/Calibrator Verifier/Calibrator |
| Origin | Human serum | Human serum |
| Analyte | Control: Apolipoprotein A-1, Apolipoprotein B, Cholesterol (total), High Density Lipoprotein, Low Density Lipoprotein, and Triglyceride | Control: Apolipoprotein A-1, Apolipoprotein B, Cholesterol (total), High Density Lipoprotein, Low Density Lipoprotein, and Triglyceride |
| | Calibrator Verifier: Triglyceride | Calibrator Verifier: Triglyceride |
| | Calibrator: High Density Lipoprotein, Low Density Lipoprotein, Triglyceride | Calibrator: High Density Lipoprotein, Low Density Lipoprotein |
| Stability (opened) | Control: 30 days at °C | Control: 30 days at °C |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Analyte | Control: Not Present | Control: High Density Lipoprotein (precipitate) Phospholipids |
| | Calibrator Verifier: Cholesterol (total), High Density Lipoprotein, Low Density Lipoprotein | Calibrator Verifier: Albumin, BUN, Calcium, Creatinine, Lactate, Magnesium, Phosphorus, |
{3}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | Calibrator: Cholesterol (total), Triglyceride | Total Protein, Glucose, Iron, Sodium, Chloride, Potassium |
| | | Calibrator: Apolipoprotein A-1, Apolipoprotein B |
| Stability (opened) | Calibrator Verifier: 30 days at 2-8 °C | Calibrator Verifier: 14 days at 2-8 °C |
| | Calibrator: 7 days at 2-8 °C | Calibrator: 5 days at 2-8 °C |
K. Standard/Guidance Document Referenced (if applicable):
None were reference.
L. Test Principle:
Not applicable.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Not applicable.
b. Linearity/assay reportable range:
Not applicable.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability
The sponsor has not provided any information regarding the traceability of the values assigned to the product. Assays used to establish the assignment of values are performed in at least two laboratories. A minimum of 12 data points are used to determine the mean (expected) value. Within and between assay standard deviations and coefficient of variations (CV) are calculated for each set of data. The Dixon Method for removing outliers from data sets of 3 through 25 observations are used to analyze data sets with CVs greater than 10%. No more than 10% of a set of data are to be removed as a statistical outlier. The resulting data are averaged to obtain a representative expected value for each constituent. All values are assigned with the instrument manufacturer's reagents available at the time of assay.
{4}
5
# Stability
Stability characteristics of the CLINIQA Liquid QC Lipid Controls Levels 1 and 2, the CLINIQA LiniCAL Lipid Calibration Verifiers Levels A - E for Olympus AU Systems, and the CLINIQA QALIBRATE Lipid Calibrator were determined using the Arrhenius model of accelerated elevated temperature studies to predict estimated storage stability at $2 - 8^{\circ}\mathrm{C}$. All samples were tested on the Roche Integra and Olympus AU 400 chemistry analyzers. The data submitted supports an open vial stability of 30 days and an unopened vial stability of 2 years at $2 - 8^{\circ}\mathrm{C}$ for the Cliniqa Liquid QC Lipid Controls Levels 1 and 2 and the Cliniqa LiniCAL Lipid Calibration Verifiers Levels A - E for Olympus AU Systems. The data submitted supports an open vial stability of 7 days and an unopened vial stability of 1 year at $2 - 8^{\circ}\mathrm{C}$ for the Cliniqa Qalibrate Lipid Calibrator. Real time stability studies are ongoing.
# Expected Values
Expected values are derived from replicate analyses of representative samples of the product and are specific to each lot. Consensus testing data used to establish the expected values were derived from multiple laboratories. A minimum of 12 data points will be used to determine the mean (expected) value. All values are assigned with the instrument manufacturer's reagents available at the time of assay. The values for the specific analyzers are provided in the product insert.
d. Detection limit:
Not applicable.
e. Analytical specificity:
Not applicable.
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable.
b. Matrix comparison:
Not applicable.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
{5}
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
5. Clinical cut-off:
Not applicable.
6. Expected values/Reference range:
Not applicable.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.