K061101 · Diagnostic Hybrids, Inc. · GNW · Nov 20, 2006 · Microbiology
Device Facts
Record ID
K061101
Device Name
D3 ULTRA DFA RESPIRATORY VIRUS SCREENING & ID KIT
Applicant
Diagnostic Hybrids, Inc.
Product Code
GNW · Microbiology
Decision Date
Nov 20, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3330
Device Class
Class 1
Indications for Use
The Diagnostic Hybrids, Inc. D3 Ultra DFA (direct fluorescent antibody) RESPIRATORY VIRUS SCREENING & ID KIT is intended for the qualitative detection and identification of the Influenza A, Respiratory Syncytial Virus (RSV), Adenovirus, Parainfluenza 1, Parainfluenza 2 and Parainfluenza 3 virus in respiratory specimens. by either direct detection or cell culture method, by immunofluorescence using fluoresceinated monoclonal antibodies (MAbs). It is recommended that specimens found to be negative after examination of the direct specimen result be confirmed by cell culture. Negative results do not preclude respiratory virus infection and should not be used as the sole basis for diagnosis, treatment or other management decisions.
Device Story
The D3 Ultra DFA Respiratory Virus Screening & ID Kit is an in vitro diagnostic tool for identifying seven respiratory viruses in clinical specimens or cell cultures. The kit utilizes a screening reagent containing a blend of fluorescein-labeled murine monoclonal antibodies (MAbs) and seven individual virus-specific DFA reagents. In a laboratory setting, a technician fixes respiratory specimens or cell cultures onto slides with acetone, applies the DFA reagents, and incubates. The process uses Evan's Blue as a counter-stain. A fluorescence microscope (FITC filter) is used to visualize results; infected cells exhibit apple-green fluorescence, while uninfected cells appear dull red. If direct specimen testing is negative, cell culture confirmation is recommended. The device assists clinicians in identifying viral pathogens, supporting appropriate patient management and infection control decisions.
Clinical Evidence
Clinical performance was evaluated using 849 respiratory specimens (fresh and frozen) and 81 banked clinical isolates. Studies compared the subject kit to the predicate device using both direct specimen (DS) and cell culture (CC) methods. Results showed high agreement: PPA for fresh specimens was 95.5% (Screen+) and 100% for individual viruses; NPA ranged from 96.7% to 100%. Frozen specimen studies showed 100% PPA and 100% NPA across all analytes. No clinical sensitivity/specificity in the traditional sense was calculated; performance was established via method comparison.
Technological Characteristics
The device consists of fluorescein-labeled murine monoclonal antibodies (MAbs) in a buffered, stabilized aqueous solution with Evan's Blue counter-stain and 0.1% sodium azide preservative. It requires a fluorescence microscope with FITC filter (excitation 490 nm, emission 520 nm). The kit includes antigen control slides and wash solution concentrate. It is a manual, non-automated, in vitro diagnostic test.
Indications for Use
Indicated for qualitative detection and identification of Influenza A, Influenza B, RSV, Adenovirus, and Parainfluenza 1, 2, and 3 in respiratory specimens (NP aspirates, washes, swabs, BAL, tracheal aspirates) via direct detection or cell culture. For prescription use only. Negative results do not rule out infection.
Regulatory Classification
Identification
Influenza virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to influenza in serum. The identification aids in the diagnosis of influenza (flu) and provides epidemiological information on influenza. Influenza is an acute respiratory tract disease, which is often epidemic.
Predicate Devices
DFA Respiratory Virus Screening & ID Kit (K022713)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k061101
B. Purpose for Submission:
New device
C. Measurand:
Influenza A, Influenza B, Respiratory Syncytial Virus (RSV), Adenovirus, Parainfluenza 1, Parainfluenza 2 and Parainfluenza 3 virus in respiratory specimens
D. Type of Test:
Direct detection or cell culture method, by immunofluorescence using fluoresceinated monoclonal antibodies (MAbs)
E. Applicant:
Diagnostic Hybrids, Inc.
F. Proprietary and Established Names:
D³ Ultra DFA Respiratory Virus Screening & ID Kit
Common Name: DFA (Direct Fluorescent Antibody) test kit for the identification of 7 common respiratory viruses (Influenza A, Influenza B, Respiratory Syncytial Virus (RSV), Adenovirus, Parainfluenza 1, Parainfluenza 2 and Parainfluenza 3 virus) in patient specimens and cell cultures
G. Regulatory Information:
1. Regulation section:
866.3330 Influenza virus serological reagents
2. Classification:
Class I
3. Product code:
GNW
4. Panel:
Microbiology (83)
H. Intended Use:
1. Intended use(s):
The Diagnostic Hybrids, Inc. D3 Ultra DFA (direct fluorescent antibody) RESPIRATORY VIRUS SCREENING & ID KIT is intended for the qualitative detection and identification of the Influenza A, Influenza B, Respiratory Syncytial Virus (RSV), Adenovirus, Parainfluenza 1, Parainfluenza 2 and Parainfluenza 3 virus in respiratory specimens, by either direct detection or cell culture method, by immunofluorescence using fluoresceinated monoclonal antibodies (MAbs). It is recommended that specimens found to be negative after examination of the direct specimen result be confirmed by cell culture. Negative results do not preclude respiratory virus infection and should not be used as the sole basis for diagnosis, treatment or other management decisions.
2. Indication(s) for use:
The Diagnostic Hybrids, Inc. D3 Ultra DFA (direct fluorescent antibody) RESPIRATORY VIRUS SCREENING & ID KIT is intended for the qualitative detection and identification of the Influenza A, Influenza B, Respiratory Syncytial Virus (RSV), Adenovirus, Parainfluenza 1, Parainfluenza 2 and Parainfluenza 3
{1}
virus in respiratory specimens, by either direct detection or cell culture method, by immunofluorescence using fluoresceinated monoclonal antibodies (MAbs). It is recommended that specimens found to be negative after examination of the direct specimen result be confirmed by cell culture. Negative results do not preclude respiratory virus infection and should not be used as the sole basis for diagnosis, treatment or other management decisions.
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Fluorescence microscope with the correct filter combination for FITC (excitation peak = 490 nm, emission peak = 520nm).
I. Device Description:
The Diagnostic Hybrids, Inc. D3 Ultra DFA RESPIRATORY VIRUS SCREENING & ID KIT includes a DFA Screening Reagent that contains a blend of murine monoclonal antibodies (MAbs) directed against seven respiratory viruses (Influenza A, Influenza B, Respiratory Syncytial Virus, Adenovirus, Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3) plus seven separate DFA Reagents, each consisting of MAb blends directed against a single respiratory virus. The kit can be used for direct specimen or cell culture screening and final virus identification.
Kit Components:
- Respiratory Virus DFA Screening Reagent - a mixture of fluorescein labeled murine monoclonal antibodies directed against respiratory viral antigens of Influenza A, Influenza B, Respiratory Syncytial Virus (RSV), Adenovirus, Parainfluenza 1, Parainfluenza 2 and Parainfluenza 3. The buffered, stabilized, aqueous solution contains Evan's Blue as a counter-stain and 0.1% sodium azide as preservative.
- Influenza A DFA Reagent - fluorescein labeled murine monoclonal antibodies directed against antigens produced by Influenza A virus infected cells. The buffered, stabilized, aqueous solution contains Evan's Blue as a counter-stain and 0.1% sodium azide as preservative.
- Influenza B DFA Reagent - fluorescein labeled murine monoclonal antibodies directed against antigens produced by Influenza B virus infected cells. The buffered, stabilized, aqueous solution contains Evan's Blue as a counter-stain and 0.1% sodium azide as preservative.
- RSV DFA Reagent - fluorescein labeled murine monoclonal antibodies directed against antigens produced by RSV infected cells. The buffered, stabilized, aqueous solution contains Evan's Blue as a counter-stain and 0.1% sodium azide as preservative.
- Adenovirus DFA Reagent - fluorescein labeled murine monoclonal antibodies directed against antigens produced by Adenovirus infected cells. The buffered, stabilized, aqueous solution contains Evan's Blue as a counter-stain and 0.1% sodium azide as preservative.
- Parainfluenza 1 DFA Reagent - fluorescein labeled murine monoclonal antibodies directed against antigens produced by Parainfluenza 1 infected cells. The buffered, stabilized, aqueous solution contains Evan's Blue as a counter-stain and 0.1% sodium azide as preservative.
2
{2}
- Parainfluenza 2 DFA Reagent - fluorescein labeled murine monoclonal antibodies directed against antigens produced by Parainfluenza 2 infected cells. The buffered, stabilized, aqueous solution contains Evan's Blue as a counter-stain and 0.1% sodium azide as preservative.
- Parainfluenza 3 DFA Reagent - fluorescein labeled murine monoclonal antibodies directed against antigens produced by Parainfluenza 3 infected cells. The buffered, stabilized, aqueous solution contains Evan's Blue as a counter-stain and 0.1% sodium azide as preservative.
- Respiratory Virus Antigen Control Slides - five individually packaged control slides containing wells with cell culture derived positive and negative control cells. Each positive well is identified as to the virus infected cells present, i.e., Influenza A, Influenza B, Respiratory Syncytial Virus (RSV), Adenovirus, Parainfluenza 1, Parainfluenza 2 and Parainfluenza 3. The Negative well contains uninfected cells. Each slide is intended to be stained only one time.
- Normal Mouse Gamma Globulin DFA Reagent - a mixture of fluorescein labeled murine gamma globulin that has been shown to be unreactive with any of the listed respiratory viruses. The buffered, stabilized, aqueous solution contains Evan's Blue as a counter-stain and 0.1% sodium azide as preservative.
- Wash Solution Concentrate - a 40X concentrate consisting of Tween 20 and 4% sodium azide (after dilution to 1X in water, the concentration of sodium azide in the solution is 0.1%) in Phosphate Buffered Saline.
- Mounting Fluid - an aqueous, buffered, stabilized solution of glycerol and 0.1% sodium azide.
J. Substantial Equivalence Information:
1. Predicate device name(s):
DFA Respiratory Virus Screening & ID Kit
2. Predicate 510(k) number(s):
K022713
3. Comparison with predicate:
The similarities to predicate device are in indicated use, operating principle, basic design, materials and formulation.
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | For the qualitative detection and identification of the respiratory viruses, Influenza A, Influenza B, Respiratory Syncytial Virus (RSV), Adenovirus, Parainfluenza 1, Parainfluenza 2 and Parainfluenza 3 virus by either direct detection or cell culture method | For the qualitative detection and identification of the respiratory viruses, Influenza A, Influenza B, Respiratory Syncytial Virus (RSV), Adenovirus, Parainfluenza 1, Parainfluenza 2 and Parainfluenza 3 virus in respiratory specimens, by either direct detection or cell culture method |
| Basic principle | DFA (Direct Fluorescent Antibody) test - Immunofluorescence using | DFA (Direct Fluorescent Antibody) test - Immunofluorescence using |
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | fluoresceinated monoclonal antibodies (MAbs) | fluoresceinated monoclonal antibodies (MAbs) |
| Antibody | Blend of murine monoclonal antibodies (MAbs) directed against seven respiratory viruses | Blend of murine monoclonal antibodies (MAbs) directed against seven respiratory viruses |
| Instrumentation (required but not provided) | Fluorescence microscope with the correct filter combination for FITC (excitation peak = 490 nm, emission peak = 520nm). | Fluorescence microscope with the correct filter combination for FITC (excitation peak = 490 nm, emission peak = 520nm). |
| Sample type | Respiratory specimens | Respiratory specimens |
K. Standard/Guidance Document Referenced (if applicable):
N/A
L. Test Principle:
The test kit uses viral antigen-specific murine monoclonal antibodies that are directly labeled with fluorescein, for rapid detection and identification of respiratory viruses. The cells to be tested, either derived from a clinical specimen or cell culture, are fixed in acetone. The DFA Screening Reagent is added to the cells to determine the presence of viral antigens. After incubating at 35°C to 37°C, the stained cells are rinsed with the diluted Wash Solution, a drop of the supplied Mounting Fluid is added and a coverslip is placed on the prepared cells. The cells are examined using a fluorescence microscope. Virus infected cells will be stained with viral specific apple-green fluorescence when stained with the DFA Screening Reagent while uninfected cells will contain no fluorescence but will be stained red by the Evan’s Blue counter-stain. If on examination of a direct stained specimen, no fluorescent-stained cells are found and all the cells stain red from the Evan’s Blue, it is recommended that the specimen be cultured and stained using the DFA Screening Reagent.
If fluorescent cells are seen, the particular virus is identified using the separate DFA Reagents on new, separate cell preparations. Cell preparations are fixed in acetone. The individual virus DFA Reagents are added to the cell preparations. After incubating at 35° to 37°C, the stained cells are rinsed with the diluted Wash Solution, a drop of the supplied Mounting Fluid is added and a coverslip is placed on the stained cells. The cells are examined using a fluorescence microscope for the presence of viral specific apple-green fluorescence, by which the unknown respiratory virus is identified.
Interpretation of results:
It is recommended that controls be examined first to ensure proper test performance before examination of the specimens. A positive reaction is one in which bright apple-green fluorescence is observed in the infected cells. Uninfected cells will stain dull red due to the Evan’s Blue counter-stain included in the DFA Reagent.
{4}
Fluorescent staining pattern of respiratory virus infected cells:
The "typical" apple-green fluorescence staining pattern for each virus is as follows: Influenza A and B: The fluorescence is cytoplasmic, nuclear or both. Cytoplasmic staining is often punctate with large inclusions while nuclear staining is uniformly bright.
Respiratory Syncytial Virus: The fluorescence is cytoplasmic and punctate with small inclusions in the syncytia.
Adenovirus: The fluorescence is cytoplasmic and punctate or bright nuclear or both. Parainfluenza 1, 2, 3: The fluorescence is cytoplasmic and punctate with irregular inclusions. Types 2 and 3 cause the formation of syncytia.
Co-infection with more than one infecting virus present in the specimen has been reported in a number of studies. The presence of multiple viruses is indicated when more than one well of the 8-well slide has fluorescent cells. The identification of the viruses is based on the individual virus DFA Reagents showing fluorescence. In such a case, it should be reported as “... and ... detected by direct specimen testing.” or “... and ... isolated by cell culture.”
Results from direct specimen testing: The quality of the specimen with respect to the number of epithelial cells in the sample can be assessed by examining the different fields at a magnification of 200X. A satisfactory specimen should have at least 2 columnar epithelial cells per field. A negative result is indicated by the absence of fluorescence in a minimal sampling of 20 columnar epithelial cells. An inadequate sample is indicated by fewer than 20 columnar epithelial cells present in the sample, in which case the test is considered invalid. A new specimen should be obtained and tested or cell culture of the remaining specimen should be initiated.
A satisfactory specimen with no fluorescent cells found should be reported as "Presumptively negative, no Influenza A, Influenza B, Adenovirus, Respiratory Syncytial Virus, Parainfluenza 1, Parainfluenza 2, or Parainfluenza 3 detected by direct specimen testing". However, such negative results should be confirmed using cell culture. In the case culture yields a positive result, it should be reported as "... isolated by cell culture", where ..." is the appropriate virus.
If fluorescent cells are found, continue with the Testing Procedure, staining with the individual virus DFA Reagents. The individual virus DFA Reagent that yields fluorescent cells represents the identification of the respiratory virus. In such a case, it should be reported as "... detected by direct specimen testing", where ..." is the appropriate virus, e.g. Influenza A, Influenza B, Adenovirus, Respiratory Syncytial Virus, Parainfluenza 1, Parainfluenza 2, or Parainfluenza 3.
Results from Culture Isolation / Confirmation: The entire cell spot or monolayer of cells must be examined for virus-infected, fluorescent cells. If no fluorescent cells are found, the results of testing of the specimen should be reported as, "No Influenza A, Influenza B, Adenovirus, Respiratory Syncytial Virus, Parainfluenza 1, Parainfluenza 2, or Parainfluenza 3 isolated by cell culture."
If fluorescent cells are found, continue with the Testing Procedure, staining with the individual virus DFA Reagents. The individual virus DFA Reagent that yields fluorescent cells represents the identification of the respiratory virus. In such a case, it should be reported as "... isolated by cell culture", where ..." is the appropriate
5
{5}
virus, e.g. Influenza A, Influenza B, Respiratory Syncytial Virus, Parainfluenza 1, Parainfluenza 2, Parainfluenza 3, or Adenovirus.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
a. Precision/Reproducibility:
Not applicable
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Not applicable
d. Detection limit:
The Predicate and Subject MAbs were compared by inoculating 96-well cell culture plates with the appropriate virus stock diluted to give ~ 50 viral particles for the entire plate or 1 viral particle per every 2 wells (i.e. 1 TCID50). The plates were incubated at 37°C for 24 hours and then stained with either the Subject Kit or the Predicate D Kit. All plates were stained using the procedure in the product insert. This assay was performed 3 times with each virus and the results indicate no statistical difference from the Predicate kit by a paired t-test.
Each targeted respiratory virus was diluted to a value of 250 virus/mL and serial 2-fold dilutions were then done down to a final value of 0.49 virus/mL. Each dilution of virus was inoculated into 6 shell vials of R-Mix, centrifuged at 700xg for 60 minutes and incubated at 35-37°C for 48 hours. The Subject Kit or the Predicate Kit was used to stain 3 shell vials of each viral dilution according to the product insert. The sensitivity of both fluorescent antibody stains is equivalent, with ~ 0.5 - 1.0 PFU as the minimum viral dilution detected.
In order to determine if mixing the 15 various MAbs in the screening reagent affected their sensitivity, the analytical sensitivity of the subject screen reagent was compared to that of the subject individual reagents by inoculating 96-well cell culture plates with the appropriate virus stock diluted to give ~ 50 viral particles for the entire plate or 1 viral particle per every 2 wells. The plates were incubated at 37°C for 24 hours and then stained with either the subject screening reagent or the subject individual reagents. All plates were stained using the procedure in the product insert. This assay was performed 3 times with each virus and the results indicate that the sensitivity of the screening reagent in the subject kit was not statistically different from that of the subject individual reagents using the paired test.
### Analytical specificity:
The DFA Reagents were tested for cross-reactivity against a wide variety of cells and microorganisms. No cross-reactivity was observed for 64 virus strains (cultured and processed for staining) or for 18 host culture cell types. Eighteen (18) bacterial cultures were stained and examined for cross-reactivity, including Staphylococcus aureus, a protein-A-producing bacterium. Staining of S. aureus appeared as small points of fluorescence while all other
6
{6}
bacterial cultures were negative. [Protein A will nonspecifically bind to the Fc portions of conjugated antibodies. Such binding can be distinguished from viral antigen binding on the basis of morphology, i.e., S. aureus-bound fluorescence appears as small (~1 micron diameter), bright dots.
Stringent conditions for cross-reactivity testing were achieved by using high concentration DFAs and high titers of microorganisms. The DFAs (i.e. directly fluoresceinated monoclonal antibodies) were prepared at 1.5X the concentration that is provided in the kit. Each of the tested viruses was prepared as infected cell monolayers (250 infected cells inoculated into a shell vial culture and incubated for 24 to 48 hours, to yield a $3+$ to $4+$ infection), and processed and stained with the 1.5X DFAs according to the procedure detailed in the product insert. Bacterial strains were cultured, processed as suspensions, then spotted on microscope slides (yielding $>150$ bacteria per 400X microscope field), then stained with the 1.5X DFAs according to the procedure in the product insert. Cell cultures were stained as confluent monolayers.
| | | DFA Reagent (Results Positive (+) or Negative (-) for Reactivity) | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Organism | Strain | Adeno | Flu A | Flu B | Para 1 | Para 2 | Para 3 | RSV |
| Adenovirus | Type 1 | + | - | - | - | - | - | - |
| | Type 3 | + | - | - | - | - | - | - |
| | Type 5 | + | - | - | - | - | - | - |
| | Type 6 | + | - | - | - | - | - | - |
| | Type 7 | + | - | - | - | - | - | - |
| | Type 10 | + | - | - | - | - | - | - |
| | Type 13 | + | - | - | - | - | - | - |
| | Type 14 | + | - | - | - | - | - | - |
| | Type 18 | + | - | - | - | - | - | - |
| | Type 31 | + | - | - | - | - | - | - |
| | Type 40 | + | - | - | - | - | - | - |
| | Type 41 | + | - | - | - | - | - | - |
| Influenza A | Aichi (H3N2) | - | + | - | - | - | - | - |
| | Mal (H1N1) | - | + | - | - | - | - | - |
| | Hong Kong (H3N2) | - | + | - | - | - | - | - |
| | Denver (H1N1) | - | + | - | - | - | - | - |
| | Port Chalmers (H3N2) | - | + | - | - | - | - | - |
| | Victoria (H3N2) | - | + | - | - | - | - | - |
| | New Jersey (H5N1) | - | + | - | - | - | - | - |
| | WS (H1N1) | - | + | - | - | - | - | - |
| | PR (H1N1) | - | + | - | - | - | - | - |
| Influenza B | Hong Kong | - | - | + | - | - | - | - |
| | Maryland | - | - | + | - | - | - | - |
| | Mass | - | - | + | - | - | - | - |
| | Taiwan | - | - | + | - | - | - | - |
| | GL | - | - | + | - | - | - | - |
| | Russia | - | - | + | - | - | - | - |
| RSV | Long | - | - | - | - | - | - | + |
| | Wash | - | - | - | - | - | - | + |
| | 9320 | - | - | - | - | - | - | + |
| Rhinovirus 39 | 209 Picornavirus | - | - | - | - | - | - | - |
| Parainfluenza 1 | C-35 | - | - | - | + | - | - | - |
| Parainfluenza 2 | Greer | - | - | - | - | + | - | - |
| Parainfluenza 3 | C 243 | - | - | - | - | - | + | - |
| Parainfluenza 4a | M-25 | - | - | - | - | - | - | - |
| Parainfluenza 4b | CH19503 | - | - | - | - | - | - | - |
| Metapneumovirus | A1 | - | - | - | - | - | - | - |
| | A2 | - | - | - | - | - | - | - |
| | B3 | - | - | - | - | - | - | - |
{7}
| | B4 | - | - | - | - | - | - |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Coronavirus | OC43 | - | - | - | - | - | - |
| | 229E | - | - | - | - | - | - |
| Herpes simplex virus Type 1 | 1F | - | - | - | - | - | - |
| | MacIntyre | - | - | - | - | - | - |
| Herpes simplex virus Type 2 | MS | - | - | - | - | - | - |
| | Strain G | - | - | - | - | - | - |
| Cytomegalovirus | Towne | - | - | - | - | - | - |
| | Davis | - | - | - | - | - | - |
| | AD169 | - | - | - | - | - | - |
| Varicella-zoster | Webster | - | - | - | - | - | - |
| | Ellen | - | - | - | - | - | - |
| Echovirus | 9 | - | - | - | - | - | - |
| | 11 | - | - | - | - | - | - |
| | 30 | - | - | - | - | - | - |
| | 34 | - | - | - | - | - | - |
| Coxsackievirus | B1 | - | - | - | - | - | - |
| | B2 | - | - | - | - | - | - |
| | B3 | - | - | - | - | - | - |
| | B4 | - | - | - | - | - | - |
| | B5 | - | - | - | - | - | - |
| | B6 | - | - | - | - | - | - |
| Mumps | | - | - | - | - | - | - |
| Rubeola | | - | - | - | - | - | - |
| Rhinovirus | 209 Picornavirus | - | - | - | - | - | - |
| Acholeplasma laidlawii | | - | - | - | - | - | - |
| Bordetella bronchiseptica | | - | - | - | - | - | - |
| Bordetella pertussis | | - | - | - | - | - | - |
| Chlamydia pneumoniae | | - | - | - | - | - | - |
| Clostridium diphtheriae | | - | - | - | - | - | - |
| Haemophillus influenzae type A | | - | - | - | - | - | - |
| Klebsiella pneumoniae | | - | - | - | - | - | - |
| Listeria pneumophila | | - | - | - | - | - | - |
| Moraxella cartarrhalis | | - | - | - | - | - | - |
| Mycobacterium avium | | - | - | - | - | - | - |
| Mycobacterium intracellulare | | - | - | - | - | - | - |
| Mycoplasma hominis type 1 | | - | - | - | - | - | - |
| Mycoplasma orale | | - | - | - | - | - | - |
| Mycoplasma pneumoniae | | - | - | - | - | - | - |
| Mycoplasma salivarium | | - | - | - | - | - | - |
| Pseudomonas aeruginosa | | - | - | - | - | - | - |
| Streptococcus pneumoniae | | - | - | - | - | - | - |
| Streptococcus pyogenes | | - | - | - | - | - | - |
| Ureaplasma urealyticum | | - | - | - | - | - | - |
| Cell cultures: | | | | | | | |
| A549 | | - | - | - | - | - | - |
| BGMK | | - | - | - | - | - | - |
| HEp-2 | | - | - | - | - | - | - |
| LLC-MK2 | | - | - | - | - | - | - |
| MDCK | | - | - | - | - | - | - |
| MRC-5 | | - | - | - | - | - | - |
| MRHF | | - | - | - | - | - | - |
| Mv1Lu | | - | - | - | - | - | - |
| NCI-H292 | | - | - | - | - | - | - |
| pCMK | | - | - | - | - | - | - |
| pRhMK | | - | - | - | - | - | - |
| pRK | | - | - | - | - | - | - |
| RD | | - | - | - | - | - | - |
| RhMK II | | - | - | - | - | - | - |
| R-Mix™ | | - | - | - | - | - | - |
| R-Mix™ Too | | - | - | - | - | - | - |
| Vero | | - | - | - | - | - | - |
| WI-38 | | - | - | - | - | - | - |
{8}
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
The study included 849 original specimens evaluated by this product ("Subject" test) and a currently marketed DFA Screening & Identification Kit ("Predicate" test). All 849 specimens were studied by Direct Specimen (DS) testing with 22 of these specimens having insufficient cell numbers to be evaluated, and one other which could not be evaluated because it exhibited non-specific staining from the Normal Mouse Gamma Globulin DFA Reagent; 520 of the specimens also were studied by Cell Culture (CC) method with one specimen not evaluated because it produced a toxic cell culture monolayer. All but 30 of the specimens were prospectively collected during the 2005-2006 season; those 30 specimens had been archived as Parainfluenza-positive. In addition, a set of 81 clinical isolates were tested by CC methods only. The evaluations were conducted at three laboratory sites. Summary of results are shown in tables below:
| Direct specimen testing – fresh prospectively collected specimens: | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| 326 specimens | Negative | Screen + | Adeno virus | Influenza A | Influenza B | Parainfluenza 1 | Parainfluenza 2 | Parainfluenza 3 | Respiratory Syncytial Virus |
| Predicate Results: | 236 | 90 | 18 | 32 | 18 | 2 | 0 | 2 | 18 |
| Subject Results: | 232 | 94 | 18 | 32 | 19 | 2 | 0 | 5 | 18 |
| Positive Percent Agreement (PPA) | | 95.5% | 100.0% | 100.0% | 100.0% | 100.0% | --- | 100.0% | 100.0% |
| 95% Confidence Interval – PPA | | 89.0-98.2% | 82.4-100% | 89.3-100% | 82.4-100% | 34.2-100% | --- | 34.2-100% | 82.4-100% |
| Negative Percent Agreement (NPA) | 98.3% | | 100.0% | 100.0% | 98.7% | 100.0% | 100.0% | 96.7% | 100.0% |
| 95% Confidence Interval - NPA | 95.7-99.3% | | 95.2-100% | 94.2-100% | 92.9-99.8% | 96.0-100% | 96.1-100% | 90.8-98.9% | 95.2-100% |
| Direct specimen testing – frozen prospectively collected specimens: | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| 474 specimens | Negative | Screen + | Adeno virus | Influenza A | Influenza B | Parainfluenza 1 | Parainfluenza 2 | Parainfluenza 3 | Respiratory Syncytial Virus |
| Predicate Results: | 306 | 168 | 8 | 85 | 19 | 3 | 3 | 9 | 51 |
| Subject Results: | 306 | 168 | 8 | 85 | 19 | 3 | 3 | 9 | 51 |
{9}
| PPA | | 100% | 100% | 100% | 100% | 100% | 100% | 100% |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| 95% CI – PPA | | 97.8-100% | 63.1-100% | 95.7-100% | 82.3-100% | 38.3-100% | 38.3-100% | 70.1-100% |
| NPA | 100% | | 100% | 100% | 100% | 100% | 100% | 100% |
| 95% CI – NPA | 98.8-100% | | 97.7-100% | 95.6-100% | 97.6-100% | 97.8-100% | 97.8-100% | 96.7-100% |
| Cell culture testing – frozen prospectively collected specimens: | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| 490 specimens | Negative | Screen + | Adeno virus | Influenza A | Influenza B | Parainfluenza 1 | Parainfluenza 2 | Parainfluenza 3 | Respiratory Syncytial Virus |
| Predicate Results: | 309 | 181 | 13 | 93 | 23 | 6 | 4 | 9 | 49 |
| Subject Results: | 309 | 181 | 13 | 93 | 23 | 6 | 4 | 9 | 49 |
| PPA | | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% |
| 95% CI – PPA | | 98.0-100% | 73.4-100% | 95.2-100% | 83.1-100% | 55.7-100% | 45.4-100% | 65.5-100% | 91.3-100% |
| NPA | 100% | | 100% | 100% | 100% | 100% | 100% | 100% | 100% |
| 95% CI – NPA | 98.5 – 100% | | 97.3-100% | 95.0-100% | 97.1-100% | 97.4-100% | 97.4-100% | 96.6-100% | 96.6-100% |
Specimens and culture isolates used in these studies came from nasopharyngeal (NP) aspirates, washes, swabs, bronchial alveolar lavages (BAL) and/or tracheal aspirates.
Percent agreement between the Subject and Predicate tests was calculated for prospectively collected specimens. For the DFA Screening Reagent:
- By DS method using fresh specimens, positive percent agreement is 95.5% and negative percent agreement is 98.3% (see Table above), with overall percent agreement 98.8%. By DS method using frozen specimens, both positive percent agreement and negative percent agreement are 100% (see Table above).
- By CC method using frozen specimens, both positive percent agreement and negative percent agreement are 100% (see Table above).
b. Matrix comparison: Not applicable
3. Clinical studies:
a. Clinical Sensitivity: Not applicable.
b. Clinical specificity: Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable): Non-prospective archival specimens: Due to relative low prevalence of Parainfluenza infections in populations of respiratory specimens, few prospectively collected specimens were reactive with the Parainfluenza DFA Reagents. Frozen original specimens previously determined to contain Parainfluenza (types 1, 2, or 3) during the 2006 "respiratory season" were
{10}
obtained from an additional laboratory, and were tested in an internal reference laboratory using the Subject and Predicate Tests by Direct Specimen method (Study 3a-DS; see table below). The same specimens were tested by Cell Culture method (Study 3a-CC, see table below). Original results reported by the laboratory were unknown to the study investigator. Although the study design has a selection bias, this study offers further analytical information on the assay's ability to detect Parainfluenza viruses.
| Study 3a-DS – Direct Specimen Results | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| 26 specimens | Negative | Screen + | Adeno virus | Influenza A | Influenza B | Parainfluenza 1 | Parainfluenza 2 | Parainfluenza 3 | Respiratory Syncytial Virus |
| Predicate Results: | 9 | 17 | 0 | 0 | 0 | 1 | 5 | 11 | 0 |
| Subject Results: | 8 | 18 | 0 | 0 | 0 | 1 | 5 | 12 | 0 |
| PPA | | 100% | --- | --- | --- | 100% | 100% | 100% | --- |
| 95% CI – PPA | | 78.4% - 100% | --- | --- | --- | 16.8% - 100% | 51.1% - 100% | 70.0% - 100% | --- |
| NPA | 88.9% | | 100% | 100% | 100% | 100% | 100% | 85.7% | 100% |
| 95% CI – NPA | 54.3% - >99.9% | | 79.3% - 100% | 79.3% - 100% | 79.3% - 100% | 78.4% - 100% | 73.4% - 100 | 46.7% - 99.5% | 79.3% - 100% |
| Study 3a-CC – Cell Culture Results | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| 29 specimens | Negative | Screen + | Adeno virus | Influenza A | Influenza B | Parainfluenza 1 | Parainfluenza 2 | Parainfluenza 3 | Respiratory Syncytial Virus |
| Predicate Results: | 8 | 21 | 0 | 0 | 0 | 3 | 5 | 13 | 0 |
| Subject Results: | 8 | 21 | 0 | 0 | 0 | 3 | 5 | 13 | 0 |
| PPA | | 100% | --- | --- | --- | 100% | 100% | 100% | --- |
| 95% CI – PPA | | 81.8% - 100% | --- | --- | --- | 38.3% - 100% | 51.1% - 100% | 73.4% - 100% | --- |
| NPA | 100% | | 100% | 100% | 100% | 100% | 100% | 100% | 100% |
| 95% CI – NPA | 62.8% - 100% | | 81.8% - 100% | 81.8% - 100% | 81.8% - 100% | 79.3% - 100% | 77.3% - 100 | 62.8% - 100% | 81.8% - 100% |
Non-prospective archival clinical isolates: A study was conducted using a collection of banked clinical isolates known to contain respiratory viruses that had been frozen from the 2005/2006 respiratory season. These specimens were selected because they were previously shown to contain at least one of the seven virus analytes detected by the Subject Test.
| Study 3b-CC – Cell Culture Results | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| 81 specimens | Negative | Screen + | Adeno virus | Influenza A | Influenza B | Parainfluenza 1 | Parainfluenza 2 | Parainfluenza 3 | Respiratory Syncytial Virus |
| Predicate Results: | 0 | 81 | 11 | 18 | 17 | 4 | 1 | 26 | 5 |
| Subject Results: | 0 | 81 | 11 | 18 | 17 | 4 | 1 | 26 | 5 |
| PPA | | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% |
| 95% CI – PPA | | 94.6% - 100% | 70.0% - 100% | 79.3% - 100% | 78.4% - 100% | 45.4% - 100% | 16.8% - 100% | 84.8% - 100% | 51.1% - 100% |
| NPA | | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% |
| 95% CI – NPA | | 97.3% - 100% | 93.8% - 100% | 93.1% - 100% | 93.2% - 100% | 94.3% - 100% | 94.5% - 100 | 92.2% - 100% | 94.2% - 100% |
{11}
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Respiratory virus infections are often seasonal, with Influenza typically extending from November to April in the northern hemisphere, and Adenovirus infections occurring more often during late winter to early summer. RSV is usually a seasonal (winter and early spring) infection as well, with epidemics lasting up to 5 months, while outbreaks caused by parainfluenza viruses may occur throughout a year.
The clinical studies were comprised of respiratory specimens collected during the winter to early spring months of 2005/2006. Prevalence of the respiratory viruses within the population of specimens that were prospectively collected and tested fresh are noted in the table below:
| Expected Values | Adenovirus | Influenza A | Influenza B | Parainfluenza 1 | Parainfluenza 2 | Parainfluenza 3 | Respiratory Syncytial Virus |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Fresh Specimens (n = 326) | 18 | 32 | 19 | 2 | 0 | 5 | 18 |
| Prevalence | 5.5% | 9.8% | 5.8% | 0.6% | 0 | 1.5% | 5.5% |
The following table summarizes the participant age demographics according to test results for a population of 326 fresh specimens, prospectively collected and evaluated for performance using the predicate assay:
| AgeΦ: | Virus: | Adenovirus | Influenza A | Influenza B | Parainfluenza 1 | Parainfluenza 2 | Parainfluenza 3 | Respiratory Syncytial Virus | Negative |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Totals | | 18 | 32 | 18 | 2 | 0 | 2 | 18 | 236 |
| <1m | | 0 | 0 | 0 | 0 | 0 | 0 | 2 | 1 |
| 1m to 2y | | 8 | 9 | 4 | 1 | 0 | 2 | 8 | 80 |
| 2y to12y | | 8 | 7 | 6 | 0 | 0 | 0 | 1 | 42 |
| 12y to18y | | 1 | 1 | 5 | 0 | 0 | 0 | 0 | 8 |
| 18y to 21y | | 0 | 0 | 1 | 0 | 0 | 0 | 0 | 2 |
| >21y | | 0 | 12 | 1 | 0 | 0 | 0 | 1 | 78 |
| Not reported | | 1 | 3 | 1 | 1 | 0 | 0 | 6 | 25 |
Age: $m =$ months, and $y =$ years
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.